K Number
K134033

Validate with FDA (Live)

Date Cleared
2014-03-31

(90 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Miromatrix Biological Mesh is intended to be implanted to reinforce soft tissue.

Device Description

The Miromatrix Biological Mesh is an implantable, animal-sourced, acellular surgical mesh that is derived from porcine liver tissue. The liver tissue undergoes perfusion decellularization and the resulting mesh is comprised primarily of collagen type I. The device is intended to function as a surgical mesh for soft tissue repair while providing a scaffold for tissue incorporation. The Miromatrix Biological Mesh is terminally sterilized in its packaging and is hydrated, moist and flexible when its packaging is opened. The mesh is available in sizes ranging from 1 cm x 2 cm to 20 cm x 30 cm, and may be trimmed or cut as required before being sutured to the surgical site.

AI/ML Overview

Acceptance Criteria and Study for Miromatrix Biological Mesh

This document describes the acceptance criteria and the study performed to demonstrate the substantial equivalence of the Miromatrix Biological Mesh.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Miromatrix Biological Mesh were established by comparing its performance to a predicate device (K980431, SurgiSIS Surgical Mesh) based on recommendations in the "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh," dated March 2, 1999." The device met these criteria by demonstrating comparable safety and mechanical properties.

Test CategoryAcceptance Criteria / Performance MeasuresReported Device Performance
BiocompatibilityDemonstrate comparable safety profile to the predicate device. Tests included: In Vitro Cytotoxicity, Skin Sensitization (Maximization Method), Intracutaneous Reactivity, Acute Systemic Toxicity, In Vitro Bacterial Reverse Mutation (AMES), In Vitro Chromosome Aberration, In Vitro Mammalian Cell Gene Mutation, Intramuscular Implantation, In Vitro Hemolysis, Pyrogenicity, Sub-Chronic Systemic Toxicity.Biocompatibility testing showed a comparable safety profile to the predicate device.
Bench/Laboratory TestingDemonstrate mechanical properties substantially equivalent to the predicate device for reinforcing soft tissue. Tests included: Tensile Strength, Suture Retention Strength, Burst Force, Tear Resistance, DNA Residuals, Collagen Analysis, Viral Inactivation, Endotoxin, Expiration Dating.Bench testing demonstrated that the mechanical properties of the mesh are substantially equivalent for reinforcing soft tissue. The specific results for each parameter (Tensile Strength, Suture Retention Strength, Burst Force, Tear Resistance, etc.) are not detailed but are stated as meeting the equivalence criteria.
Animal TestingDemonstrate tissue incorporation into the mesh.A 30-Day Rat Implantation Study demonstrated tissue incorporation into the mesh.

Summary of Substantial Equivalence: Based on the indications for use, technological characteristics, and performance test results, the Miromatrix Biological Mesh was determined to be substantially equivalent to the predicate device.

2. Sample Size and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the exact sample sizes for each test (e.g., number of samples for tensile strength, number of animals in the rat study).
  • Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. Given the nature of a 510(k) submission for a new device, the studies are inherently prospective.

3. Number, Qualifications, and Adjudication Methods of Experts

These sections are not applicable as the provided document describes the regulatory submission for a medical device (surgical mesh), not an AI/algorithm-based diagnostic or assistive system that typically involves expert review of images or data. The evaluation relied on standard physical, chemical, and biological testing, not expert interpretation of results in a clinical context.

4. Adjudication Method

Not applicable for the reasons stated above.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. The Miromatrix Biological Mesh is a physical implantable device, not an AI or algorithm that assists human readers.

6. Standalone (Algorithm Only) Performance Study

Not applicable. The Miromatrix Biological Mesh is a physical implantable device; it does not involve algorithms or standalone performance in the context of AI.

7. Type of Ground Truth Used

The "ground truth" for this device's evaluation was based on:

  • Biocompatibility Standards: Established guidelines and methods for assessing the biological response to materials, including specific tests for cytotoxicity, systemic toxicity, sensitization, etc.
  • Mechanical Property Standards: Industry-recognized physical tests to measure properties like tensile strength, suture retention, burst force, and tear resistance.
  • Histological Evaluation from Animal Studies: Direct observation and analysis of tissue response and incorporation within the rat implantation study.

Essentially, the ground truth was established by recognized scientific and engineering principles and experimental results, rather than clinical outcomes or expert consensus on interpretations.

8. Sample Size for the Training Set

Not applicable. This device is not an AI/machine learning product and therefore does not have a "training set" in the computational sense. The "training" in the context of device development would refer to the iterative design and testing phases, but not a formal training set for an algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reasons above.

{0}------------------------------------------------

ડ. 510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Miromatrix Biological Mesh is provided below.

Device Common Name:Surgical Mesh
Device Proprietary Name:Miromatrix Biological Mesh
Submitter:Miromatrix Medical, Inc.18683 Bearpath TrailEden Prairie, MN 55347www.miromatrix.com
Contact:Stephen RhodesBiologics Consulting Group, Inc.400 N. Washington St., Suite 100Alexandria, VA 22314Phone: 301-579-3700Fax: 703-548-7457Email: srhodes@bcg-usa.com
Alternate Contact:Jeff RossVP Product DevelopmentMiromatrix Medical, Inc.Phone: 763-458-8801Email: jross@miromatrix.com
Date Prepared:December 30, 2013
Classification Regulation:21 CFR 878.3300
Panel:General & Plastic Surgery
Product Code:FTM
Predicate Device:K980431, SurgiSIS Surgical MeshCook Biotech, Inc.

CONFIDENTIAL

{1}------------------------------------------------

Indication for Use:

The Miromatrix Biological Mesh is intended to be implanted to reinforce soft tissue.

Device Description:

The Miromatrix Biological Mesh is an implantable, animal-sourced, acellular surgical mesh that is derived from porcine liver tissue. The liver tissue undergoes perfusion decellularization and the resulting mesh is comprised primarily of collagen type I. The device is intended to function as a surgical mesh for soft tissue repair while providing a scaffold for tissue incorporation. The Miromatrix Biological Mesh is terminally sterilized in its packaging and is hydrated, moist and flexible when its packaging is opened. The mesh is available in sizes ranging from 1 cm x 2 cm to 20 cm x 30 cm, and may be trimmed or cut as required before being sutured to the surgical site.

Comparison to the Predicate:

The Miromatrix Biological Mesh has similar indications for use and technological characteristics as the predicate surgical mesh. Both devices are porcine-derived, acellular meshes that are comprised primarily of collagen type I. The meshes reinforce soft tissue while providing a scaffolding for tissue repair.

Mechanical testing, biocompatibility testing and animal testing were conducted per the recommendations in: Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance – Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh," dated March 2, 1999." The following tests were conducted:

Biocompatibility Testing In Vitro Cytotoxicity Skin Sensitization (Maximization Method) Intracutaneous Reactivity Acute Systemic Toxicity In Vitro Bacterial Reverse Mutation (AMES) In Vitro Chromosome Aberration In Vitro Mammalian Cell Gene Mutation Intramuscular Implantation In Vitro Hemolysis Pyrogenicity Sub-Chronic Systemic Toxicity

Bench/Laboratory Testing Tensile Strength Suture Retention Strength Burst Force Tear Resistance DNA Residuals Collagen Analysis

CONFIDENTIAL

{2}------------------------------------------------

Viral Inactivation Endotoxin Expiration Dating

Animal Testing 30-Day Rat Implantation Study

The biocompatibility testing showed the comparable safety profile of the Miromatrix Biological Mesh and the predicate. Bench testing demonstrated that the mechanical properties of the mesh are substantially equivalent for reinforcing soft tissue. Animal testing demonstrated tissue incorporation into the mesh.

Summary of Substantial Equivalence:

Based on the indications for use, technological characteristics and performance test results, the Miromatrix Biological Mesh is substantially equivalent to the predicate.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

March 31, 2014

Miromatrix Medical, Inc. do Biologics Consulting Group, Inc. Stephen P. Rhodes 400 North Washington Street, Suite 100 Alexandria, Virginia 22314

Re: K134033

Trade/Device Name: Miromatrix Biological Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: December 31, 2013 Received: December 31, 2013

Dear Mr. Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Stephen P. Rhodes

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

  • for Binita Ashar, MD, MBA, FACS Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31. 2013 See PRA Statement on last page.

510(k) Number (if known) K134033

Device Name Miromatrix Biological Mesh

Indications for Use (Describe)

The Miromatix Biological Mesh is intended to be implanted to reinforce soft tissue.

Type of Use (Select one or both, as applicable) 网 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

Please do not write Below This Line - continue on a separate Page if Needed.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Jiyoung Dang -S

FORM FDA 3881 (9/13)

Page 1 of 2

PSC Pac Section Sermon (201) 44)-6789

CONFIDENTIAL

Page 14

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.