K Number
K133961

Validate with FDA (Live)

Date Cleared
2014-06-26

(184 days)

Product Code
Regulation Number
870.1100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

Device Description

The subject devices are multi-parameter patient monitors, specifically the SureSigns VS3 and SureSigns VS4. Modifications to the VS4 include the addition of CO2, SpHb, Respiratory Rate RRa, and Masimo SpO2 measurements. Both VS3 and VS4 have the QuickNBP mode added.

AI/ML Overview

This is a 510(k) summary for Philips SureSigns VS3 and VS4 vital signs monitors, describing modifications to add additional measurement capabilities. The provided text, however, does not contain information about specific acceptance criteria or a detailed study proving the device meets said acceptance criteria with numerical performance data. It broadly states that "Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject device. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the subject device and test results showed substantial equivalence."

Therefore, I cannot fulfill all parts of your request with the provided information.

However, based on the information available, here's what can be extracted:

  • Acceptance Criteria and Reported Device Performance: This information is not explicitly provided in a table or numerical format. The document states that "Pass/Fail criteria were based on the specifications cleared for the subject device and test results showed substantial equivalence," implying that the devices met pre-defined specifications. However, the exact criteria and corresponding performance metrics are not detailed.

  • Sample Size for Test Set and Data Provenance: This information is not explicitly stated in the provided text.

  • Number of Experts and Qualifications: This information is not mentioned in the provided text.

  • Adjudication Method: This information is not mentioned in the provided text.

  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: There is no indication that an MRMC study was done. The device is a vital signs monitor, which typically involves direct measurement rather than interpretation by multiple human readers in the way an AI-assisted diagnostic tool might.

  • Standalone (Algorithm Only) Performance Study: The document mentions that the new features are achieved by using OEM modules that are "FDA cleared under Kxxxxxx." This implies that the performance of these modules as standalone components was already established in their respective clearances. For instance, for CO2 measurement, the Oridion microMediCO2 OEM module was cleared under K094012; for SpHb, RRa, and Masimo SpO2, the Masimo Rainbow SET Radical 7R CO-Oximeter was cleared under K100428; and for Temporal Temperature, the Exergen TemporalScanner Thermometer was cleared under K011291. The QuickNBP mode is described as based on the "same algorithm that provides the regular NBP measurements" in the existing devices. Therefore, while not explicitly called a "standalone study," the reliance on previously cleared, established technologies suggests that their standalone performance has been demonstrated.

  • Type of Ground Truth Used: Not explicitly stated for the overall device's performance. However, for the OEM modules incorporated, their original clearances would have involved appropriate ground truth methods for each physiological parameter (e.g., direct measurement for temperature, arterial blood gas analysis for SpO2 calibration, etc.).

  • Sample Size for Training Set: This information is not applicable as the document describes hardware modifications incorporating existing, cleared OEM modules and leveraging existing algorithms. It does not mention the development or training of new algorithms that would require a distinct training set.

  • How Ground Truth for Training Set Was Established: Not applicable given the nature of the device modifications.

In summary, the provided text primarily focuses on the substantial equivalence argument for modifications to existing vital signs monitors by integrating previously cleared OEM modules and leveraging existing algorithms. It does not contain the detailed performance study information with specific acceptance criteria, sample sizes, expert involvement, or adjudication methods that your request entails for a newly developed AI/diagnostic device.

{0}------------------------------------------------

K133961 JUN 2 6 2014

510K Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).

    1. The submitter and contact person of this pre-market notification is:
      Greg Li Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States

Tel: 978-659-4227 Fax: 978-659-7323 Email: Greg.li@philips.com

    1. Date this summary is prepared: This summary is prepared on December 19, 2013.
    1. The names of the subject devices are as following: SureSigns VS3 SureSigns VS4
    1. The trade names of the devices are SureSigns VS3 and SureSigns VS4.
  • The common usual name for both the VS3 and the VS4 is multi-parameter patient 5. monitor
Device PanelClassificationProCodeDescriptionApplicablesubject devices
Cardiovascular870.1100, IIDSJAlarm, Blood PressureVS3, VS4
870.1110, IIDSKComputer, Blood PressureVS3, VS4
870.1130, IIDXNSystem, Measurement, BloodPressure, Non-InvasiveVS3, VS4
870.2700, IIDQAOximeterVS3, VS4
870.2900, IIDSACable, transducer andelectrode, patient connectorVS3, VS4
General Hospital andPersonal Use880.2910, IIFLLThermometer, Electronic,ClinicalVS3, VS4
Anesthesiology &Respiratory Therapy868.1400, IICCKAnalyzer, GasVS4
    1. The Classification names are as follows:
      |
    1. The modified devices are substantially equivalent to previously cleared devices from Philips and suppliers:
      SureSigns VS3 Vital Signs Monitor cleared under K120132 SureSigns VS4 Vital Signs Monitor cleared under K120132 SureSigns VM8 Patient Monitor cleared under K123900

{1}------------------------------------------------

Masimo Rainbow SET Radical 7R CO-Oximeter cleared under K100428 Exergen TemporalScanner Thermometer cleared under K011291 Oridion Capnography Inc for Capnostream 20 with microMediCO2 module cleared under K094012

    1. The modifications are as follows:
    • a) Add CO2 measurement to VS4. This is achieved by using the Oridion MicroMedi OEM module. This same module is used in Philips SureSigns VM8 patient monitor for the same measurement. The SureSigns VM8 was FDA cleared under K123900. Oridion microMediCO2 OEM module includes IP1 feature for its CO2 measurement. VM8 doesn't include the IPI feature. The predicate device for VS4 IPI feature is the Oridion microMediCO2 OEM module with. The Oridion microMedi CO2 OEM module was cleared under K094012.
    • b) Add SpHb measurement to VS4. This measurement and the following new added Respiratory (c) and Masimo SpO2 (d) are achieved using the OEM Masimo Rainbow SET module (MX board). The same module is used in the Masimo Rainbow SET Radical 7R CO-Oximeter for the same measurements. And the Masimo Rainbow SET Radical 7R CO-Oximeter is FDA cleared under K100428.
    • c) Add Respiratory Rate RRa measurement to VS4. Details refer to above item (b). In addition. FDA 510(k) K120984 cleared the RRa sensor, RAS-125c. for use on pediatric patients with body weight>10Kg. So RRa will be supported on both adult and pediatric patients.
    • d) Add Masimo SpO2 measurement with Pulse rate to VS4. Details refer to above item (b).
    • e) Add Temporal Temperature measurement to VS4. This is achieved by using the OEM Exergen TemporalScanner Thermometer. This Exergen TemporalScanner Thermometer is FDA cleared under K011291. The OEM thermometer has a RS232 cable added to connect the thermometer to the VS4 monitor.
    • Add QuickNBP mode to both VS3 and VS4. QuickNBP measurement mode f) utilizes a single oscillometric pulse for the determination of the pressure level at a particular pressure step. This compares to the regular NBP measurement mode that utilizes multiple oscillometric pulses to determine the pressure level at a particular pressure step. QuickNBP is based on the same algorithm that provides the regular NBP measurements in SureSigns VS3, VS4 monitors. Same as measure method in STAT mode. The NBP specifications of VS3 and VS4 are not changed due to the addition of the QuickNBP mode.
    1. The subject device VS3 has the same Intended Use and Indications for Use as the legally marketed predicate device, currently marketed VS3:

Indications for Use:

The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of

{2}------------------------------------------------

patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

Patient Types
ParameterAdultPediatricNeonatal
NBP
SpO2
Temperature

The indications for Use of subiect VS4 has following parameters added to the legally marketed SureSigns VS4: CO2, RRa and SpHb

Indications for Use:

The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

ParameterPatient Types
AdultPediatricNeonatal
NBP
SpO2
Temperature
CO2
RRa
SpHb
    1. The subiect devices have the same fundamental technological characteristics as the legally marketed predicate devices. The subject devices use the same algorithms for the measurements as the predicate devices.
    1. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject device. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the subject device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VS3 Vital Signs monitor and the Philips SureSigns VS4 Vital Signs monitor meet all safety and reliability requirements and performance claims and support a determination of substantial equivalence.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002

Philips Medical Systems Greg Li 3000 Minuteman Rd Andover, MA 01810 US

Re: K133961

Trade/Device Name: SureSigns VS3, SureSigns VS4 Regulation Number: 21 CFR 870.1100 Regulation Name: Multi-Parameter Patient Monitor Regulatory Class: Class II Product Code: DSJ, DSK, DXN, DQA, DSA, FLL, CCK Dated: May 28, 2014 Received: May 29, 2014

Dear Greg Li,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability of warranties. We remind you, however, that device labeling must be truthful and not misleading.

June 26, 2014

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

Image /page/3/Picture/10 description: The image shows a logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" around the perimeter. Inside the circle is a stylized symbol that resembles a bird or abstract human figure, composed of three curved lines.

{4}------------------------------------------------

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDeyices/Safety/Reportallyoblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510 (k) Number (if known): K133961

Device Name:

SureSigns VS3 (reference numbers: 863071, 863072, 863073, 863074) SureSigns VS4 (reference number: 863283)

Indications for Use for SureSigns VS3 Vital Signs Monitor

The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

ParameterPatient Types
AdultPediatricNeonatal
NBP
SpO2
Temperature

Indications for Use for SureSigns VS4 Vital Signs Monitor

The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

Patient Types
ParameterAdultPediatricNeonatal
NBPVVV
SpO2VVV
TemperatureVVV
CO2VVV
RRaVVV
SpHbVV

Page 1 of 2

Philips SureSigns VS3 and VS4 Traditional 510K

010_Indications for Use Statement

{6}------------------------------------------------

K133961

AND/OR over-the-counter Use: YES NO Prescription Use: (Part 21 CFFR 801 Subpart D) (21 CFR 807 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 2 of 2

Philips SureSigns VS3 and VS4 Traditional 510K

010_Indications for Use Statement

§ 870.1100 Blood pressure alarm.

(a)
Identification. A blood pressure alarm is a device that accepts the signal from a blood pressure transducer amplifier, processes the signal, and emits an alarm when the blood pressure falls outside a pre-set upper or lower limit.(b)
Classification. Class II (performance standards).