K Number
K133632

Validate with FDA (Live)

Device Name
PERICAN ULTRA
Date Cleared
2014-06-27

(213 days)

Product Code
Regulation Number
868.5150
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The B. Braun Perican Ultra needle is intended for use in injecting local anesthetics and/or analgesics into a patient to provide regional anesthesia and/or to facilitate the placement of a catheter.

Device Description

The B. Braun Perican Ultra needle is a single use, disposable anesthesia needle (18 Ga.) with Tuohy (curved) bevel. The needle is intended for use in injecting local anesthetics and/or analgesics into a patient to provide regional anesthesia and/or to facilitate the placement of a catheter. The proposed needle is composed of a hollow stainless steel cannula with 1 cm depth markings along the length of the cannula. The cannula is also designed with small 'echogenic' etched markings intended to enhance visibility of the needle when using an ultrasound imaging device. The cannula are secured to the winged polycarbonate hub using an epoxy resin and plastic spacer. The Perican Ultra needle is provided with a plastic stylet, which is pre-assembled inside the needle cannula. The hub is designed with a Luer adapter to facilitate attachment of the needle to anesthetic administration devices, such as a syringe.

AI/ML Overview

Here's an analysis of the Perican™ Ultra Needle based on the provided 510(k) Premarket Notification:

The document does not describe a study involving an AI/Machine Learning device. It concerns a medical device, specifically an anesthesia conduction needle. Therefore, many of the requested categories related to AI performance, ground truth, experts, and training/test sets are not applicable.

However, I can extract the acceptance criteria and the methods used to demonstrate the device meets those criteria for this non-AI medical device.

1. Table of Acceptance Criteria and Reported Device Performance

RequirementAcceptance Criteria (Test)Reported Device Performance
Visual Specifications- Needles are free of hooks, burrs, cracks, damage, occlusion, foreign contamination discoloration, excessive roughness.- Inspected for presence of echogenic markings."Results of testing demonstrate that the Perican Ultra needle performs similarly to the predicate device and is substantially equivalent to the predicate devices." (Implies visual specifications were met and echogenic markings were present and acceptable).
Dimensional SpecificationsThe following dimensions are within specification according to the product drawing:- Cannula outer diameter- Cannula inner diameter- Cannula length- Distance to first depth marking- Depth marking width- Tip Gap of Stylet"Results of testing demonstrate that the Perican Ultra needle performs similarly to the predicate device and is substantially equivalent to the predicate devices." (Implies dimensional specifications were met).
Associated Device Testing- Needle permits a 20 Ga. catheter to freely pass through the cannula.- Needle performs adequately with sideport valve assembly (i.e., extension tubing) and threading assist guide attached."Results of testing demonstrate that the Perican Ultra needle performs similarly to the predicate device and is substantially equivalent to the predicate devices." (Implies adequate performance with associated devices).
Ultrasound Visibility- Needle is visible under ultrasound.The device includes "'echogenic' etched markings intended to enhance visibility of the needle when using an ultrasound imaging device." And "Results of testing demonstrate that the Perican Ultra needle performs similarly to the predicate device and is substantially equivalent to the predicate devices." (Implies ultrasound visibility was enhanced and acceptable).
Functional Testing- The Luer of the needle meets the requirements of ISO 594-1 and ISO 594-2.- Needle is free of occlusions and permits flow of fluid.- Needle assembly withstands a minimum pressure without leaking.- Needle withstands bending without breaking.- Needle is sufficiently stiff.- Cannula to hub joint has sufficient tensile strength.- Stylet fit in cannula is sufficient.- Stylet to stylet hub has sufficient tensile strength."Results of functional performance and biocompatibility testing demonstrate that the Perican Ultra needle is substantially equivalent to the predicate devices." (Implies all functional tests were successfully passed).
Sterilization- SAL 10^-6- EO Residual Levels- LAL Bacterial Endotoxin Testing"Results of functional performance and biocompatibility testing demonstrate that the Perican Ultra needle is substantially equivalent to the predicate devices." (Implies all sterilization requirements were met).
BiocompatibilityCytoxicity, Intracutaneous Reactivity, Sensitization, Rabbit Pyrogenicity, Systemic Toxicity, USP <661> Physicochemical Tests for Plastics, Corrosion, Acidity/Alkalinity according to ISO 10993-1:2009."Based on similarities to the predicate device and biocompatibility test results, the Perican Ultra needle materials of construction are considered biocompatible for their clinical application." (Implies all biocompatibility tests were passed successfully).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes for each performance test. It only states that functional performance testing was "completed" and "results of testing demonstrate that the Perican Ultra needle performs similarly to the predicate device." The data provenance (country of origin, retrospective/prospective) is not mentioned, as these are typically bench-top and lab-based engineering tests rather than clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The "ground truth" for this device's performance is established by engineering and laboratory test standards (e.g., ISO standards, dimensional measurements, functional performance metrics), not expert human interpretation of data for diagnostic purposes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are relevant for studies where human interpretation or consensus is required to establish a gold standard. For this device, objective measurements and adherence to established engineering standards serve as the "ground truth."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/Machine Learning device, and therefore no MRMC study, human reader improvement with AI, or effect size is discussed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/Machine Learning device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this medical device is based on:

  • Adherence to international standards (ISO 9626:1991/Amd. 1:2001, ISO 7864:1993, ISO 594-1:1986, ISO 594-2:1998, ISO 10993-1:2009), which define acceptable performance characteristics.
  • Objective measurements of physical and functional properties (e.g., cannula dimensions, Luer taper fit, tensile strength, pressure resistance).
  • Visual inspection criteria.

8. The sample size for the training set

Not applicable. This is not an AI/Machine Learning device, so there is no training set in the context of machine learning.

9. How the ground truth for the training set was established

Not applicable. There is no training set in the context of machine learning for this device.

{0}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification Perican™ Ultra Needle June 27, 2014

JUN 2 7 2014

ડી 510(k) SUMMARY

SUBMITTER:B. Braun Medical Inc.901 Marcon BoulevardAllentown, PA 18109-9341610-266-0500
Contact: Lisa Giaquinto, Specialist, Regulatory AffairsPhone: (610) 596-2354
Fax: (610) 266-4962
E-mail: lisa.giaquinto@bbraun.com
DEVICE NAME:Perican™ Ultra Needle
COMMON OR USUAL NAME DEVICEAnesthesia Conduction Needle(An anesthesia conduction needle is a device used to inject localanesthetics into a patient to provide regional anesthesia.)
CLASSIFICATION:Class II, Product Code BSP, 868.5150
PREDICATE DEVICES:Contiplex® FX Continuous Nerve Block Set, BraunMedical Inc., K113059, Class II, CAZ, 868.5140.Tuohy Sono, Pajunk GmbH, K113207, Class II, BSP,868.5150

DESCRIPTION:

The B. Braun Perican Ultra needle is a single use, disposable anesthesia needle (18 Ga.) with Tuohy (curved) bevel. The needle is intended for use in injecting local anesthetics and/or analgesics into a patient to provide regional anesthesia and/or to facilitate the placement of a catheter. The proposed needle is composed of a hollow stainless steel cannula with 1 cm depth markings along the length of the cannula. The cannula is also designed with small 'echogenic' etched markings intended to enhance visibility of the needle when using an ultrasound imaging device. The cannula are secured to the winged polycarbonate hub using an epoxy resin and plastic spacer. The Perican Ultra needle is provided with a plastic stylet, which is pre-assembled inside the needle cannula. The hub is designed with a Luer adapter to facilitate attachment of the needle to anesthetic administration devices, such as a syringe.

K133632 Page 1 of 5

{1}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification Perican™ Ultra Needle June 27, 2014

INDICATIONS FOR USE:

The B. Braun Perican Ultra needle is intended for use in injecting local anesthetics and/or analgesics into a patient to provide regional anesthesia and/or to facilitate the placement of a catheter.

SUBSTANTIAL EQUIVALENCE:

Technological Characteristics

Predicate Device - Contiplex® FX (K113059)

The Perican Ultra needle is indicated for the same use (injecting local anesthetics and/or analgesics into a patient to provide regional anesthesia and/or to facilitate the placement of a catheter) and is similar in design to the Contiplex FX (17 Ga.) needle included as part of the Contiplex FX Continuous Nerve Block Set, cleared in 510(k) # K113059. The Perican Ultra needle is designed with the same stainless steel cannula, polycarbonate needle hub, spacer, adhesive, and guard materials as the Contiplex FX needles. The Perican Ultra needle differs from the predicate device in that the Perican Ultra needle is an 18 Ga. needle as opposed to 17 Ga., contains a stylet, and includes 'echogenic' etched markings to enhance visibility of the cannula under ultrasound, while the Contiplex FX needle does not include these features.

Predicate Device – Tuohy Sono (K113207)

The Perican Ultra needle is intended for the same use and is similar in design to the Tuohy Sono needles marketed by Pajunk GmbH. The Tuohy Sono needles are intended for the transient delivery of anesthetics to provide regional anesthesia and/or analgesia and/or to facilitate placement of a catheter. The needles may be used during all anaesthetic and analgesic procedures according to the physician's indication The cannulas consist of stainless steel tubing bonded to a polycarbonate hub using an epoxy glue. Like the Perican Ultra needles, the Tuohy Sono needles were cleared in a range of diameters (including 18 Ga.), are provided with a stylet, and include 'echogenic' markings to enhance visibility of the cannula under ultrasound.

{2}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification Perican™ Ultra Needle June 27, 2014

Performance Testing

The following performance standards have been utilized in the evaluation of the Perican Ultra needle:

ISO 9626:1991/Amd. 1: 2001, "Stainless steel needle tubing for the manufacture of medical devices. "

ISO 7864:1993, "Sterile hypodermic needles for single use."

ISO 594-1:1986, "Conical Fittings with a 6 % Luer taper for syringes, needles and certain other medical equipment -- Part 1: General Requirements. "

ISO 594-2:1998, "Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 2: Lock fittings. "

Functional performance testing (summarized in Table 1) was completed with the proposed Perican Ultra needle to demonstrate that the needle performs as intended. Results of testing demonstrate that the Perican Ultra needle performs similarly to the predicate device and is substantially equivalent to the predicate devices.

RequirementTest
Visual Specifications-Needles are free of hooks, burrs, cracks, damage, occlusion, foreign contamination discoloration, excessive roughness.-Inspected for presence of echogenic markings
Dimensional SpecificationsThe following dimensions are within specification according to the product drawing:-Cannula outer diameter-Cannula inner diameter-Cannula length-Distance to first depth marking-Depth marking width- Tip Gap of Stylet
Associate Device Testing-Needle permits a 20 Ga. catheter to freely pass through the cannula-Needle performs adequately with sideport valve assembly (i.e. extension tubing) and threading assist guide attached
Ultrasound Visibility-Needle is visible under ultrasound

Table 1: Performance Test Requirements

{3}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification
Perican™ Ultra Needle June 27, 2014

RequirementTest
Functional Testing-The Luer of the needle meets the requirements of ISO 594-1and ISO 594-2-Needle is free of occlusions and permits flow of fluid-Needle assembly withstands a minimum pressure withoutleaking-Needle withstands bending without breaking-Needle is sufficiently stiff-Cannula to hub joint has sufficient tensile strength-Stylet fit in cannula is sufficient-Stylet to stylet hub has sufficient tensile strength
Sterilization-SAL 10-6-EO Residual Levels-LAL Bacterial Endotoxin Testing

Biocompatibility

The materials of construction of the Perican Ultra needle are either 1) non-patient contacting materials, 2) used in a predicate device with the same type and duration of patient contact, and/or 3) are patient contacting and have been evaluated for biocompatibility through testing, which meets and/or exceeds the tests recommended for consideration in ISO 10993-1: 2009, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." The minimum tests completed with the Perican Ultra patientcontacting materials of construction are provided in Table 2: Biocompatibility Test Requirements.

RequirementTest
BiocompatibilityCytoxicity, Intracutaneous Reactivity, Sensitization, RabbitPyrogenicity, Systemic Toxicity, USP <661>Physicochemical Tests for Plastics, Corrosion,Acidity/Alkalinity
Table 2: Biocompatibility Test Requirements

Based on similarities to the predicate device and biocompatibility test results, the Perican Ultra needle materials of construction are considered biocompatible for their clinical application.

{4}------------------------------------------------

K133632

Page 5 of 5

(

B. Braun Medical Inc. D: Braun Moorked Into:
510(k) Premarket Notification
Perican™ Ultra Needle
June 27, 2014

Conclusion

Results of functional performance and biocompatibility testing demonstrate that the Perican Ultra needle is substantially equivalent to the predicate devices. イ

and the comments of the comments of the comments of the comments of the comments of the

{5}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2014

B. Braun Medical, Inc. Lisa Giaquinto, MS, RAC Regulatory Affairs Specialist 901 Marcon Blvd. Allentown, PA 18109

Rc: K133632

Trade/Device Name: Perican™ Ultra Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: May 27, 2014 Received: May 29, 2014

Dear Ms. Giaquinto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications , for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register:

{6}------------------------------------------------

Page 2 - Ms. Giaquinto

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

hup://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Teiashri a RID/ODE/CDRI /

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Sincerely yours.

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K133632

Device Name Perican Ultra Needle

Indications for Use (Describe)

The B. Braun Perican Ultra needle is injecting local anesthetics and/or analgesics into a patient to provide regional anesthesia and/or analgesia and/or to facilitate the placement of a catheter.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

CONTRACTOR CONTRACTOR FOR FOR FOR FOR FOR FOR FOR CONCY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/7/Picture/12 description: The image shows the text "Todd D" on the top left corner. Below it, the date "2014.06.26" is written. The text "-04'00'" is on the right side of the image.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

5

§ 868.5150 Anesthesia conduction needle.

(a)
Identification. An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.(b)
Classification. Class II (performance standards).