K Number
K132978

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-06-13

(263 days)

Product Code
Regulation Number
864.7280
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is a qualitative in vitro diagnostic device intended for use in the detection and genotyping of a single point mutation (G1691A, referred to as the Factor V Leiden mutation or FVL) of the Factor V gene, located on Chromosome 1q23, and a single point mutation (G20210A) of the prothrombin gene (referred to as Factor II or FII), located on Chromosome 11p11-q12, from genomic DNA isolated from EDTA anti- coagulated human whole blood samples. The test is to be performed on the IMPACT Dx System and is indicated for use as an aid in the diagnosis of patients with suspected thrombophilia.

Device Description

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is a qualitative, multiplexed genetic testing device for parallel detection and genotyping of the point mutations G1691A of the Factor V gene and G20210A of the Factor II gene from genomic DNA isolated from EDTA anti- coagulated human whole blood samples. The test is to be performed on the IMPACT Dx System.

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is performed using the IMPACT Dx System, which includes the IMPACT Dx NANO and the IMPACT Dx MA, a matrix-assisted laser desorption / ionization time-of-flight (MALDI-TOF) mass spectrometer. The test involves Factor V and Factor II region-specific polymerase chain reaction (PCR) amplification of genomic DNA purified from human whole blood in a multiplexed reaction, followed by allele-specific single base primer extension reactions. The reaction products are desalted, dispensed onto a SpectroCHIP® Array using the IMPACT Dx NANO, and the genotyping products are resolved on the basis of mass using the IMPACT Dx MA.

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test provides reagents for multiplex PCR, deoxynucleotide triphosphate dephosphorylation, and single base extension. The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is comprised of the following components:

  • IMPACT Dx Factor V Leiden and Factor II Primer Set
  • IMPACT Dx PCR Reagent Set
  • IMPACT Dx Extend Reagent Set

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test utilizes a biochemistry process (Sequenom Biochemistry) that involves target-specific PCR amplification and single-base extension reactions with the subsequent analysis of the reaction products of the target nucleic acids by matrixassisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS),

The IMPACT Dx System (System) is a platform for highly accurate and sensitive genomic analysis and is designed for use with FDA cleared or approved assays citing its use. The IMPACT Dx System is comprised of the following instruments, software and consumables;

  • IMPACT Dx NANO (NANO)
  • IMPACT Dx MA (MA)
  • TYPER Dx Software (TYPER Dx)
  • System Consumables
  • SpectroCHIP® Arrays (Chip)
  • Clean Resin
  • 3-Point Calibrant (Calibrant)

The System is intended to be used by trained operators in a professional laboratory to perform the following key tasks:

  • De-salt (using Clean Resin) amplified nucleic acid samples, upon completion of polymerase chain reaction (PCR) and single-base extension reactions following the instructions provided in the Sequenom test-specific package insert;
  • Transfer (using the IMPACT Dx NANO) de-salted nucleic acid samples from a microtiter plate onto a disposable 96-pad sample Chip;
  • Obtain mass spectra (using the IMPACT Dx MA) from samples and 3-Point Calibrant on a Chip; and
  • Analyze (using the TYPER Dx software) the mass spectra of the samples for genotyping results.

The IMPACT Dx System accomplishes genomic analysis and genotyping testing by coupling a biochemistry process (Sequenom biochemistry) that involves target-specific PCR amplification and single-base extension reactions with the subsequent analysis of the reaction products of the target nucleic acids by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). This biochemistry process is homogeneous and does not require purification of the PCR products or the extension products and thus is very amenable to high-throughput genotyping testing.

The IMPACT Dx NANO is a self-contained, enclosed instrument that uses computer-controlled robotics to transfer nanoliter volumes of analyte from a 96-well microtiter plate onto a Chip, which is subsequently processed by means of MALDI-TOF MS analysis on the IMPACT Dx MA. This instrument includes an integrated computer pre-loaded with the Nanodispenser software and provides a simple touch-screen interface for users.

The IMPACT Dx MA is a bench top mass spectrometer that processes analyte-loaded Chips by means of MALDI-TOF MS analysis. This instrument includes an integrated computer pre-loaded with the TYPER Dx software, a monitor, and a firewall for secure communication with the IMPACT Dx NANO. The main function of the IMPACT Dx MA is to acquire mass spectra from analytes that have been transferred onto a Chip, which has a chemical matrix on each pad. The mass spectra are captured and further analyzed by the TYPER Dx software.

The TYPER Dx software (TYPER Dx) manages the processing of Sequenom genotyping tests. It is deployed on the computer embedded within the IMPACT Dx MA.

The TYPER Dx software provides the following key functions:

  • Allows users to create and manage panel runs;
  • Monitors analyte transfer activities on the IMPACT Dx NANO;
  • Controls user-initiated, automated mass spectrum acquisition runs on the IMPACT Dx MA;
  • Analyzes the mass spectra acquired by the IMPACT Dx MA and makes genotype calls per a test-specific algorithm;
  • Enables users to view and export results; and
  • Allows an administrator to manage users to ensure secure access to the IMPACT Dx MA and panel run data.
AI/ML Overview

Here's a summary of the acceptance criteria and study detailed in the provided document:

Acceptance Criteria and Device Performance

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test aims for high agreement with a reference method for genotyping Factor II and Factor V genes.

Acceptance Criteria (Implicit)Reported Device Performance (After Repeat Testing)
High overall percent agreement with reference method for FII99.4% (95% CI: 98.6 - 99.8%) for Factor II
High overall percent agreement with reference method for FVL99.3% (95% CI: 98.5 - 99.7%) for Factor V
High positive percent agreement/clinical sensitivity for FII100.0% (95% CI: 96.3 - 100.0%) for Factor II (Heterozygous and Homozygous combined)
High negative percent agreement/clinical specificity for FII99.3% (95% CI: 98.5 - 99.8%) for Factor II (Wild Type)
High positive percent agreement/clinical sensitivity for FVL98.7% (95% CI: 95.3 - 99.8%) for Factor V (Heterozygous and Homozygous combined)
High negative percent agreement/clinical specificity for FVL99.4% (95% CI: 98.6 - 99.8%) for Factor V (Wild Type)
Low rate of No Calls (prior to repeat testing)Factor II: 16 out of 859 samples (1.86%) (Reduced to 5 after repeat testing)
Factor V: 18 out of 860 samples (2.09%) (Reduced to 6 after repeat testing)
Zero Mis-callsZero mis-calls reported for both Factor II and Factor V (both before and after repeat testing)
Analytical Sensitivity (Limit of Detection)0.67 ng per reaction (for wild type, heterozygous, and homozygous mutant alleles for both loci)
Reproducibility (Inter-site, Inter-operator)All operators, except one, from all 3 sites produced 100% agreement between the genotypes after retesting no-calls. One operator had 2 samples yielding repeated no calls; no mis-calls.
No carry-over contaminationNo cross contamination or carry-over was observed
No interference from tested substancesNone of the substances tested adversely impacted the performance of the IMPACT Dx test.

Study Details

2. Sample Size and Data Provenance

  • Test Set Sample Size:

    • Clinical Data (Method Comparison Study): 860 clinical samples
    • Analytical Specificity: 13 clinical genomic DNA from patient samples
    • Analytical Sensitivity: Not explicitly stated, but clinical genomic DNA from patient samples were used with a series dilution.
    • Carry-over Contamination: A panel of 3 clinical genomic DNA samples and a no template control sample.
    • Interferences: Leukocyte-depleted whole blood specimens spiked with cell lines.
    • Reproducibility: 12 human genomic DNA samples.
  • Data Provenance: The document states "clinical genomic DNA obtained from patient samples" and "fresh EDTA anti-coagulated whole blood samples collected from normal human subjects." This implies the data is derived from human clinical samples, likely from retrospective collections for the method comparison and analytical studies, and potentially prospective collection for the fresh blood study. No specific country of origin is mentioned, but the submission is to the FDA (USA).

3. Number of Experts and Qualifications for Ground Truth (Test Set)

Relevant for the method comparison study:

  • Number of Experts: Not explicitly stated as "experts." The ground truth was established by a bi-directional DNA sequencing reference method. While sequencing itself is a technical process, the interpretation of sequencing results could involve expert review. However, the document doesn't detail this.
  • Qualifications of Experts: Not provided, as the ground truth relied on a reference method (bi-directional sequencing) rather than human expert consensus on image interpretation, for example.

4. Adjudication Method (Test Set)

Not applicable in the context of this document. The study compares the device's results directly against a bi-directional DNA sequencing reference method. Discrepancies would likely be investigated through re-testing or re-sequencing, as implied by the "Number of FII/FVL Calls After Repeat Testing" in the tables, but formal "adjudication" by multiple human readers is not described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a MRMC comparative effectiveness study was not explicitly done or reported in this document. The study focuses on the analytical and clinical performance of the device itself, comparing it to a reference sequencing method, not on human reader performance with or without AI assistance.

6. Standalone Performance Study

Yes, a standalone (algorithm only without human-in-the-loop performance) study was performed. The entire context of the "Non-clinical Bench Data" and "Clinical Data" sections describes the performance of the IMPACT Dx Factor V Leiden and Factor II Genotyping Test as a diagnostic device, which processes samples and generates genotype calls automatically or semi-automatically via the IMPACT Dx System's TYPER Dx software. The results (e.g., percent agreement with sequencing) reflect the algorithm's performance.

7. Type of Ground Truth Used

The ground truth used for the clinical data (method comparison study) was bi-directional DNA sequencing. This is considered a highly accurate and established reference method for genotyping.

8. Sample Size for the Training Set

The document does not explicitly mention a separate "training set" or its sample size. The reported studies primarily describe testing the device's performance (analytical and clinical validation) against known samples or a reference method. For in-vitro diagnostic devices, particularly those based on established molecular biology principles, the development might involve iterative testing and refinement, but a formal "training set" in the machine learning sense is not typically documented as a distinct phase with a specified sample size in regulatory submissions like this unless the device employs novel AI/ML components requiring such.

9. How Ground Truth for the Training Set Was Established

As a distinct "training set" is not explicitly mentioned, the method for establishing its ground truth is also not described. If development involved internal studies, it's highly probable that ground truth would have been established using methods similar to the validation, such as DNA sequencing or other orthogonal molecular techniques.

{0}------------------------------------------------

ട്ട് sequenom. _ ............................................................................................................................................................

JUN 1 3 2014

510(K) SUMMARY


This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is: K132978

  • A. Submitter Sequenom, Inc. 3595 John Hopkins Court San Diego, California 92121
    Contact

Robin Weiner Senior Vice President, Quality and Regulatory. Affairs Sequenom, Inc. Telephone: (858) 202-9044 Telefax: (858) 202-9020 Email: rweiner@sequenom.com

Date Prepared June 12, 2014 .

B. Device Names and Regulatory Information

ASSAY

Common or Usual Name:Factor V Leiden and Factor II genotyping test
Proprietary NameIMPACT Dx™ Factor V Leiden and Factor II Genotyping Test
Classification NameFactor V Leiden DNA mutation detection systems (21 CFR
864.7280)
ClassificationClass II
Product CodePHJ: System, mass spectrometry, multiplex genotyping, hereditarythrombophilia related mutations
PanelHematology (81)

Pan

C. Predicate Devices

ASSAY :

Roche Factor V Leiden Kit (K033607) Roche Factor II (Prothrombin) G20210A Kit (K033612)

INSTRUMENT

Roche LightCycler® (K033734) AutoGenomics INFINITI™ System (K060564)

IMPACT Dx Factor V Leiden and Factor II Genotyping Test

Confidential Communication Page 1 of 9

{1}------------------------------------------------

s sequenom.

D. Device Description

IMPACT Dx Factor V Leiden and Factor II Genotyping Test Overview

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is a qualitative, multiplexed genetic testing device for parallel detection and genotyping of the point mutations G1691A of the Factor V gene and G20210A of the Factor II gene from genomic DNA isolated from EDTA anti- coagulated human whole blood samples. The test is to be performed on the IMPACT Dx System.

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is performed using the IMPACT Dx System, which includes the IMPACT Dx NANO and the IMPACT Dx MA, a matrix-assisted laser desorption / ionization time-of-flight (MALDI-TOF) mass spectrometer. The test involves Factor V and Factor II region-specific polymerase chain reaction (PCR) amplification of genomic DNA purified from human whole blood in a multiplexed reaction, followed by allele-specific single base primer extension reactions. The reaction products are desalted, dispensed onto a SpectroCHIP® Array using the IMPACT Dx NANO, and the genotyping products are resolved on the basis of mass using the IMPACT Dx MA.

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test provides reagents for multiplex PCR, deoxynucleotide triphosphate dephosphorylation, and single base extension. The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is comprised of the following components:

  • . IMPACT Dx Factor V Leiden and Factor II Primer Set
  • IMPACT Dx PCR Reagent Set .
  • . IMPACT Dx Extend Reagent Set

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test utilizes a biochemistry process (Sequenom Biochemistry) that involves target-specific PCR amplification and single-base extension reactions with the subsequent analysis of the reaction products of the target nucleic acids by matrixassisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS),

IMPACT Dx System Overview

The IMPACT Dx System (System) is a platform for highly accurate and sensitive genomic analysis and is designed for use with FDA cleared or approved assays citing its use. The IMPACT Dx System is comprised of the following instruments, software and consumables;

  • . IMPACT Dx NANO (NANO)
  • IMPACT Dx MA (MA) .
  • TYPER Dx Software (TYPER Dx) ●
  • System Consumables .
    • SpectroCHIP® Arrays (Chip) o
    • Clean Resin o
    • 3-Point Calibrant (Calibrant) o

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication Page 2 of 9

510(k) Summary

{2}------------------------------------------------

sequenom. __

The System is intended to be used by trained operators in a professional laboratory to perform the following key tasks:

  • De-salt (using Clean Resin) amplified nucleic acid samples, upon completion of polymerase . chain reaction (PCR) and single-base extension reactions following the instructions provided in the Sequenom test-specific package insert;
  • Transfer (using the IMPACT Dx NANO) de-salted nucleic acid samples from a microtiter plate . onto a disposable 96-pad sample Chip;
  • . Obtain mass spectra (using the IMPACT Dx MA) from samples and 3-Point Calibrant on a Chip; and
  • Analyze (using the TYPER Dx software) the mass spectra of the samples for genotyping results. .

The IMPACT Dx System accomplishes genomic analysis and genotyping testing by coupling a biochemistry process (Sequenom biochemistry) that involves target-specific PCR amplification and single-base extension reactions with the subsequent analysis of the reaction products of the target nucleic acids by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). This biochemistry process is homogeneous and does not require purification of the PCR products or the extension products and thus is very amenable to high-throughput genotyping testing.

IMPACT Dx NANO

1

The IMPACT Dx NANO is a self-contained, enclosed instrument that uses computer-controlled robotics to transfer nanoliter volumes of analyte from a 96-well microtiter plate onto a Chip, which is subsequently processed by means of MALDI-TOF MS analysis on the IMPACT Dx MA. This instrument includes an integrated computer pre-loaded with the Nanodispenser software and provides a simple touch-screen interface for users.

IMPACT Dx MA

The IMPACT Dx MA is a bench top mass spectrometer that processes analyte-loaded Chips by means of MALDI-TOF MS analysis. This instrument includes an integrated computer pre-loaded with the TYPER Dx software, a monitor, and a firewall for secure communication with the IMPACT Dx NANO. The main function of the IMPACT Dx MA is to acquire mass spectra from analytes that have been transferred onto a Chip, which has a chemical matrix on each pad. The mass spectra are captured and further analyzed by the TYPER Dx software.

TYPER Dx Software

The TYPER Dx software (TYPER Dx) manages the processing of Sequenom genotyping tests. It is deployed on the computer embedded within the IMPACT Dx MA.

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication Page 3 of 9

510(k) Summary

June 12, 2014

{3}------------------------------------------------

sequenom.

The TYPER Dx software provides the following key functions:

  • Allows users to create and manage panel runs; �
  • Monitors analyte transfer activities on the IMPACT Dx NANO; ●
  • Controls user-initiated, automated mass spectrum acquisition runs on the IMPACT Dx MA; .
  • Analyzes the mass spectra acquired by the IMPACT Dx MA and makes genotype calls per a test-. specific algorithm;
  • Enables users to view and export results; and .
  • Allows an administrator to manage users to ensure secure access to the IMPACT Dx MA and panel run data.

E. Intended Use

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is a qualitative in vitro diagnostic device intended for use in the detection and genotyping of a single point mutation (G1691A, referred to as the Factor V Leiden mutation or FVL) of the Factor V gene, located on Chromosome 1q23, and a single point mutation (G20210A) of the prothrombin gene (referred to as Factor II or FII), located on Chromosome 11p11-q12, from genomic DNA isolated from EDTA anti-coagulated human whole blood samples. The test is to be performed on the IMPACT Dx System and is indicated for use as an aid in the diagnosis of patients with suspected thrombophilia.

Comparison with Predicate Devices F.

ASSAY

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is substantially equivalent to the legally marketed predicate devices. These devices have the same intended use and indications for use and similar technological characteristics and principles of operation. They are all PCR-based tests for the genotyping of the Factor V and Factor II genes, with essentially the same performance characteristics. They use different signal detection methodologies and run on different instrumentation platforms; however, all are based on similar technological principles. Table F.1 below presents the similarities and differences between the assay and the predicate devices.

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication Page 4 of 9 .

510(k) Summary

June 12, 2014

{4}------------------------------------------------

ള്ള ടലവശ്വാനം.

DevicePredicate
IMPACT Dx Factor VLeiden and Factor IIGenotyping TestRoche Factor V Leiden KitRoche Factor II(Prothrombin) G20210AKit
510(k) NumberK132978K033607K033612
Type of TestGenotyping TestGenotyping TestGenotyping Test
MeasurandFactor II and Factor VFactor VFactor II
Target of DetectionSingle-nucleotidepolymorphismSingle-nucleotidepolymorphismSingle-nucleotidepolymorphism
Intended UserHealth Care ProfessionalHealth Care ProfessionalHealth Care Professional
Intended UseQualitative in vitrodiagnostic genotyping testfor the detection of Factor IIand Factor V alleles fromEDTA anti-coagulatedhuman whole blood samplesQualitative in vitrodiagnostic genotyping testfor the detection Factor Vonly in EDTA anti-coagulated whole bloodsamplesQualitative in vitrodiagnostic genotyping testfor the detection of Factor IIonly in EDTA anti-coagulated whole bloodsamples
Indications for UseAid in the diagnosis ofpatients with suspectedthrombophiliaAid in the diagnosis ofpatients with suspectedthrombophiliaAid in the diagnosis ofpatients with suspectedthrombophilia
Specimen TypePurified DNA from humanblood samplesPurified DNA from humanblood samplesPurified DNA from humanblood samples
Technological DetectionPrinciplesGenotyping test forsimultaneous detection(multiplex system) of PCR-amplified DNA sequencesGenotyping test for PCR-amplified DNA sequencesGenotyping test for PCR-amplified DNA sequences
Sample PreparationDNA extraction andpurification performed off-lineDNA extraction andpurification performed off-lineDNA extraction andpurification performed off-line
Oligonucleotide probes andprimersSpecific for Factor V Leiden(G1691A) and Factor IIG20210ASpecific for Factor V Leiden(G1691A)Specific for Factor II(prothrombin G20210A)
Detection ChemistrySNP discrimination byallele-specific singlenucleotide extension coupledwith MALDI-TOF massspectrometryFluorogenic detection ofPCR-amplification productsby melting curve analysisFluorogenic detection ofPCR-amplification productsby melting curve analysis
Analytical Sensitivity200 allele copies (0.67 nginput DNA/reaction)FV: 50 allele copies /reactionFII: 50 allele copies /reaction
InstrumentIMPACT Dx SystemRoche LightCyclerRoche LightCycler
ControlsInternal control per sampleplus external positive andnegative controls requiredper runExternal positive andnegative controls requiredper runExternal positive andnegative controls requiredper run
Reference MethodBi-directional DNAsequencingDNA sequencingDNA sequencing
DevicePredicate
SequenomIMPACT Dx™ SystemRocheLightCycler®AutoGenomicsINFINITI™ System
510(k) NumberK132978K033734K060564
Intended UseSimultaneous, qualitativedetection of multipleanalytes in a PCR amplifiedgenomic DNA sampleutilizing MALDI-TOF MS.A fully automatedamplification and detectionsystem for nucleic acidsusing fluorescencedetection.Designed to measurefluorescence signals oflabeled DNA targethybridized to BioFilmChipmicroarrays
Intended UserClinical laboratoryClinical laboratoryClinical laboratory
Specimen TypePurified nucleic acidsPurified nucleic acidsPurified nucleic acids
Specimen PreparationPerformed off-linePerformed off-linePerformed off-line
Test PrincipleMALDI-TOF detection(multiplex system) of PCR-amplified DNA sequencesFluorogenic detection ofPCR-amplified DNAfragments by melting curveanalysisFluorogenic detection(multiplex system) of PCR-amplified DNA fragments
Detection ProcedureMass spectrometric analysisof target-specificsequences.Optical detection ofstimulated fluorescenceOptical detection ofstimulated fluorescence
Detection ChemistrySingle-base extensionreactions with thesubsequent analysis usingMALDI-TOF MSPaired hybridization probesusing fluorescenceresonance energy transferDirect fluorescence

Table F.1 ・ Substantial Equivalence – Comparison to Predicate Devices

:

: - -------------------------------------------------------------------------------------------------------------------------------------------------------------------------


.

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication Page 5 of 9

·

{5}------------------------------------------------

sequenom.

INSTRUMENT

The IMPACT Dx System is substantially equivalent to the currently legally marketed predicate devices. These devices have the same intended use and similar technological characteristics and principles of operation. They are PCR-based genotyping instruments using detection of single nucleotide polymorphisms and have essentially the same performance characteristics. Table F.2 below presents the similarities and differences between the IMPACT Dx System and predicate devices.

Table F.2
Substantial Equivalence -- Comparison with Predicate Device

G. Non-clinical Bench Data

A series of internal and external analytical studies were conducted which demonstrated that the IMPACT Dx Factor V Leiden and Factor II Genotyping Test has equivalent performance compared to the predicate devices with respect to analytical sensitivity, analytical specificity, reproducibility and potential interfering substances and mutations. In addition, studies were conducted to verify that sample-to-sample contamination did not occur during the test procedure when used with the IMPACT Dx System. Lastly, in a study using fresh EDTA anti-coagulated whole blood samples collected from normal human subjects, DNA was extracted using the recommended DNA extraction method. The IMPACT Dx Factor V Leiden and Factor II Genotyping Test demonstrated 100% concordance with the reference bi-directional sequencing method on genotypes of both Factor II and Factor V genes.

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication Page 6 of 9

June 12, 2014

{6}------------------------------------------------

sequenom.

Analytical Specificity

To determine the specificity of the extension primers were evaluated individually in the extension reactions using 13 clinical genomic DNA from patient samples encompassing all relevant genotypes for Factor VL and Factor II. All clinical samples produced the expected genotype calls, in respect to the extension primer under evaluation. When all extension primers in the IMPACT Dx Factor V Leiden and Factor II Genotyping Test were evaluated together, similarly, all clinical samples produced the expected genotype calls for both FII and FVL.

Interfering Mutations

In studies to evaluate the effect of mutations in close proximity to genotyping at locus 1691 of the FVL gene, as well as locus 20210 on the FII gene, the variants were tested in each of the 4 reactions. No FVL or FII genotype will be reported on the patient specimens when the Factor V forward assay fails to produce a genotype call due to the presence of an interfering 1690 C>T or 1692A>C or on a patient specimen when the Factor II forward assay fails to produce a genotype call due to the presence of either the 20207A>C or 20209C>T interfering mutations. No miscall or unexpected call from any of the samples was observed during the evaluation.

Analytical Sensitivity

Clinical genomic DNA obtained from patient samples encompassing all relevant genotypes for Factor V and Factor II genes were used to determine the limits of detection of the IMPACT Dx Factor V Leiden and Factor II Genotyping Test. A series dilution of input DNA levels ranging from 25 ng down to 0.0015 ng was used. The limit of detection was very similar for wild type, heterozygous and homozygous mutant alleles for both loci, and is 0.67 ng per reaction. The recommended input DNA level is 25 ng per reaction.

Carry-over Contamination

A study to evaluate cross contamination and carry-over was conducted with a panel of 3 clinical genomic DNA samples and a no template control sample. No cross contamination or carry-over was observed as all the no template control samples produced the expected no-call results after the runs.

Interferences

Leukocyte-depleted whole blood specimens spiked with cell lines from individuals encompassing all relevant genotypes for both the Factor V and Factor II genes were used in this evaluation. The following substances, at the concentrations listed below, were evaluated for their potential interfering effect on the test results.

  • Hemoglobin, 200 mg/dL .
  • K2-EDTA, 3.4 µM ●
  • . Heparin, 3000 units/L
  • . Cholesterol. 500 mg/dL
  • Bilirubin, 60 mg/dL .
  • Ethanol, 500 mg/dL

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication

510(k) Summary

Page 7 of 9

{7}------------------------------------------------

s sequenom.

These substances are representative of metabolites produced during pathological conditions (endogenous) or compounds introduced during sample preparation (exogenous). All the substances were added to whole blood samples before genomic DNA extraction, except ethanol, which was added after the DNA extraction. None of the substances tested adversely impacted the performance of the IMPACT Dx test.

Reproducibility

The reproducibility of the IMPACT Dx Factor V Leiden and Factor II Genotyping Test was assessed at three external clinical sites. Twelve human genomic DNA samples, including all genotypes (wild type (GG), heterozygous mutant (GA) and homozygous mutant (AA) for both FV and FII genes) were used in this study. Each DNA sample was tested in singlet by each of 3 sites using one discrete IMPACT Dx System at each site. Two operators each conducted 5 runs on 5 non-consecutive days for a total of 10 testing runs per site. The number of correct calls, defined as the number of samples yielding the expected genotypes for both FV and FII genes, no calls and mis-calls were calculated for each operator, and all sites and operators combined. All operators, except one, from all 3 sites produced 100% agreement between the genotypes after all no call results were retested. One operator (Operator 2, Site 1) had 2 samples vielding repeated no calls; there were no mis-calls.

H. Clinical Data

In a method comparison study, 860 clinical samples were tested in the IMPACT Dx Factor V Leiden and Factor II Genotyping Test and compared to bi-directional sequencing reference method. Table H.1 presents the percent agreement between the IMPACT Dx Factor V Leiden and Factor II Genotyping Test and the reference method along with the respective lower confidence boundary (LCB) of the 95% confidence interval (CI) for Factor II. Table H.2 presents similar information for Factor V.

The test demonstrated an overall percentage agreement of 99.4% (Cl: 98.6 - 99.8%) and 99.3% (Cl: 98.5 - 99.7%) for the Factor II and Factor V genes, respectively, with the reference method.

GenotypeBySequencing#SamplesNumber of FII Calls Before Repeat TestingNumber of FII Calls After Repeat Testing
CorrectCalls# NoCalls¹#MissedCalls²%Agreement95%LCB³CorrectCalls# NoCalls¹#MissedCalls²# RepeatSamples⁴%Agreement95%LCB³
Wild Type76274715098.097.0757501599.398.5
Heterozygous78771098.794.178001100.097.1
Homozygous191900100.088.619000100.088.6
All Samples85984316098.197.0854501699.498.6

Table H.1 Comparison of the IMPACT Dx Factor V Leiden and Factor II Genotyping Test and DNA Sequencing Results - Factor II

IMPACT Dx Factor V Leiden and Factor II Genotyping Test
K132978
Confidential Communication
Page 8 of 9

{8}------------------------------------------------

s sequenom.

Table H.2
Comparison of the IMPACT Dx Factor V Leiden and Factor II Genotyping Test and DNA Sequencing
Results – Factor V
GenotypeBySequencing#SamplesNumber of FVL Calls Before Repeat TestingNumber of FVL Calls After Repeat Testing
CorrectCalls# NoCalls1#MissedCalls2%Agreement95%LCB3CorrectCalls# NoCalls1#MissedCalls2# RepeatSamples4%Agreement95%LCB3
Wild Type71069713098.297.1706401399.498.6
Heterozygous1321275096.292.213020598.595.3
Homozygous181800100.088.018000100.088.0
All Samples86084218097.996.7854601899.398.5

When the heterozygous and homozygous mutant genotypes are combined into one "Positive" category and compared to one "Negative" (wild type) category, the test demonstrated a positive percent agreement or clinical sensitivity of 100.0% (95% CI: 96.3 – 100.0%) and a negative percent agreement or clinical specificity of 99.3% (95% Cl: 98.5 - 99.8%) for Factor II. Similarly, for Factor V, positive percent agreement or clinical sensitivity was 98.7% (95% CI: 95.3 - 99.8%), and negative percent agreement or clinical specificity was 99.4% (95% Cl: 98.6 - 99.8%) for Factor V.

I. Conclusion

The intended use of the IMPACT Dx Factor V Leiden and Factor II Genotyping Test is to measure a single point mutation (G1691A, referred to as the Factor V Leiden mutation or FVL) of the Factor V gene, located on Chromosome 1q23, and a single point mutation (G20210A) of the prothrombin gene (referred to as Factor II or F11), located on Chromosome 11p11-q12, from genomic DNA isolated from EDTA anti-coagulated human whole blood samples as an aid in the diagnosis of patients with suspected thrombophilia.

Performance data demonstrate that the IMPACT Dx Factor V Leiden and Factor II Genotyping Test using the IMPACT Dx System is as safe and effective as predicate devices. Thus the IMPACT Dx System and the IMPACT Dx Factor V Leiden and Factor II Genotyping Test are substantially equivalent to the other legally marketed devices, supporting premarket 510(k) clearance,

IMPACT Dx Factor V Leiden and Factor II Genotyping Test K132978 Confidential Communication Page 9 of 9

510(k) Summary

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 13, 2014

Sequenom, Inc. Ms. Robin Weiner Senior Vice President, Quality and Regulatory Affairs 3595 John Hopkins Court San Diego, CA 92121

Re: K132978

Trade/Device Name: IMPACT Dx™ Factor V Leiden and Factor II Genotyping Test on the IMPACT Dx™ System Regulation Number: 21 CFR 864.7280 Regulation Name: Factor V Leiden DNA Mutation Detection Systems Regulatory Class: Class II Product Code: PHJ Dated: May 13, 2014 Received: May 14, 2014

Dear Ms. Weiner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{10}------------------------------------------------

Page 2 - Ms. Robin Weiner

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

' Reena Philip -S

Reena Philip, Ph.D. Director Division of Molecular Genetics and Pathology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K132978

Device Name

IMPACT Dx™ Factor V Leiden and Factor II Genotyping Test on the IMPACT™ Dx System

Indications for Use (Describe)

The IMPACT Dx Factor V Leiden and Factor II Genotyping Test is a qualitative in vitro diagnostic device intended for use in the detection and genotyping of a single point mutation (G1691A, referred to as the Factor V Leiden mutation or FVL) of the Factor V gene, located on Chromosome 1q23, and a single point mutation (G20210A) of the probrombin gene (referred to as Factor II or FII), located on Chromosome 11p11-q12, from genomic DNA isolated from EDTA anticoagulated human whole blood samples. The test is to be performed on the IMPACT Dx System and is indicated for use as an aid in the diagnosis of patients with suspected thrombophilia.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Donna M. Roscoe -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review Instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

§ 864.7280 Factor V Leiden DNA mutation detection systems.

(a)
Identification. Factor V Leiden deoxyribonucleic acid (DNA) mutation detection systems are devices that consist of different reagents and instruments which include polymerase chain reaction (PCR) primers, hybridization matrices, thermal cyclers, imagers, and software packages. The detection of the Factor V Leiden mutation aids in the diagnosis of patients with suspected thrombophilia.(b)
Classification. Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Factor V Leiden DNA Mutation Detection Systems.” (See § 864.1(d) for the availability of this guidance document.)