K Number
K033612

Validate with FDA (Live)

Date Cleared
2003-12-18

(31 days)

Product Code
Regulation Number
864.7280
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Factor II (Prothrombin) G20210A Kit is an in vitro diagnostic test for the detection and genotyping of a single point mutation (G to A at position 20210) of the human Factor II gene, from DNA isolated from human whole peripheral blood. The test is intended to be used on the LightCycler using SW 3.5. The sample preparation must be performed according to the workflow procedure described in the package insert. The Factor II (Prothrombin) G20210A Kit is indicated as an aid to diagnosis in the evaluation of patients with suspected thrombophilia.

Device Description

The Factor II (Prothrombin) G20210A Kit is an in vitro diagnostic test for the detection and genotyping of the Factor II (Prothrombin) G20210A mutation, from DNA isolated from human whole peripheral blood.

AI/ML Overview

Acceptance Criteria and Device Performance Study for K033612

This analysis describes the acceptance criteria and the study used to demonstrate that the Factor II (Prothrombin) G20210A Kit meets these criteria, as detailed in the provided 510(k) summary (K033612).

1. Table of Acceptance Criteria and Reported Device Performance

FeatureAcceptance Criteria (Implied)Reported Device Performance
Substantial EquivalenceHigh agreement with a DNA sequencing method.99.5% agreement with a DNA sequencing method.
PrecisionConsistent and reproducible results.Within-run: Tm1: 0.14 – 0.26 % CV, Tm2: 0.14 – 0.24 % CV, ΔTm: 0.58- 0.90% CV Total: Tm1: 0.19 – 0.33 % CV, Tm2: 0.22 – 0.44 % CV, ΔTm: 1.23- 1.53 % CV Overall median: 0.36%
Analytical Sensitivity (LDL)Ability to detect low concentrations of allele copies.< 50 allele copies per reaction
Reagent StabilityStable performance over time and under specified conditions.Store up to stated expiration date at -15 to -25°C. Protect from light. Freeze immediately after use. Kit reagents may be frozen and thawed up to five times.
InterferencesMinimal interference from common substances in samples.High concentrations of heparin might interfere. No interference from EDTA or citrate anticoagulants.
Precautions and WarningsIdentification of known limitations or potential false results.One rare mutation (A20218G) will lead to false-positive result after performing the melting curve analysis.

Note: The 510(k) summary does not explicitly state numerical acceptance criteria for each test; however, the "Performance Characteristics" section implicitly sets the acceptance criteria by reporting the observed performance, which is deemed sufficient for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 572 samples were tested for the substantial equivalence study.
  • Data Provenance: The document does not specify the country of origin of the data. It also does not explicitly state whether the samples were collected retrospectively or prospectively.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not provide information on the number or qualifications of experts used to establish the ground truth for the test set.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method used for the test set. The ground truth was established by a "DNA sequencing method."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No information on a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is provided in the document. This type of study is more common for imaging devices where human interpretation is a critical part of the diagnostic process. This device is an in vitro diagnostic test for genetic mutation detection, which typically involves automated or semi-automated analysis rather than human reader interpretation in the context of an MRMC study.

6. Standalone Performance Study

Yes, a standalone performance study was done. The "Substantial equivalence" section directly compares the device's performance to a DNA sequencing method (the ground truth), reporting 99.5% agreement. This demonstrates the algorithm's (or kit's) performance without explicit human intervention in the interpretation of the Factor II G20210A Kit results.

7. Type of Ground Truth Used

The type of ground truth used for the substantial equivalence study was DNA sequencing method. This is considered a gold standard for genetic mutation detection.

8. Sample Size for the Training Set

The document does not specify the sample size for a training set. This is common for IVD kits where the primary validation is often based on the performance of the prepared reagents and assay protocol against a gold standard, rather than training a machine learning algorithm.

9. How the Ground Truth for the Training Set Was Established

As no training set is explicitly mentioned for a machine learning algorithm, there is no information on how its ground truth would have been established. For the overall validation of the kit, the DNA sequencing method served as the established ground truth.

{0}------------------------------------------------

K033612

DEC 1 8 2003510(k) Summary
IntroductionAccording to the requirements of 21 CFR 807.92, the following informationprovides sufficient detail to understand the basis for a determination ofsubstantial equivalence.
Submittername, address,contactRoche Diagnostics Corporation9115 Hague RoadIndianapolis, IN 46250(317) 521 - 2386
Contact Person: Robert A. GreggDate Prepared: November 14, 2003
Device NameProprietary name: Factor II (Prothrombin) G20210A Kit
Common name: Factor II (Prothrombin) G20210A Kit
Classification name: Factor deficiency test
DeviceDescriptionThe Factor II (Prothrombin) G20210A Kit is an in vitro diagnostic test for thedetection and genotyping of the Factor II (Prothrombin) G20210A mutation,from DNA isolated from human whole peripheral blood.
Intended useThe Factor II (Prothrombin) G20210A Kit allows the detection andgenotyping of a single point mutation (G to A at position 20210) of thehuman Factor II gene, from DNA isolated from human whole peripheralblood. The test is performed on the LightCycler Instrument utilizingpolymerase chain reaction (PCR) for the amplification of Factor II DNArecovered from clinical samples and fluorigenic target-specific hybridizationfor the detection and genotyping of the amplified Factor II DNA.
The test is intended to be used on the LightCycler using SW 3.5. The samplepreparation must be performed according to the workflow proceduredescribed in the package insert.

:

{1}------------------------------------------------

Indications forUseThe Factor II (Prothrombin) G20210A Kit is an in vitro diagnostic test forthe detection and genotyping of the Factor II (Prothrombin) G20210Amutation as an aid to diagnosis in the evaluation of patients with suspectedthrombophilia.
SubstantialequivalenceThe Factor II (Prothrombin) G20210A Kit gives equivalent results to a DNAsequencing method. A total of 572 samples were tested and 99.5% agreementwas observed between the methods.
PerformancecharacteristicsThe following table lists performance characteristics of the Factor II(Prothrombin) G20210A Kit
-------------------------------------------------------------------------------------------------------------------------------------

Table 2 - Performance Characteristics

FeatureFactor II (Prothrombin) G20210A Kit performance
PrecisionWithin-runTm1: 0.14 – 0.26 % CV Tm2: 0.14 – 0.24 % CV ΔTm: 0.58- 0.90% CV Total Tm1: 0.19 – 0.33 % CV Tm2: 0.22 – 0.44 % CV ΔTm: 1.23- 1.53 % CV Overall median: 0.36%
Analyticalsensitivity (LDL)< 50 allele copies per reaction
Reagent StabilityStore up to stated expiration date at -15 to -25°C. Protect from light Freeze immediately after use Kit reagents may be frozen and thawed up to five times
InterferencesHigh concentrations of heparin might interfere with the polymerase chain reaction. No interference from EDTA or citrate anticoagulants.
Precautions andWarnings• One rare mutation will lead to false-positive result after performing themelting curve analysis (i.e., the mutation A20218G)

.

: 上一篇:

:

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird-like figure.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 8 2003

Robert Gregg, Ph.D. Director, Regulatory Submissions Roche Diagnostics Corporation 9115 Hague Road Indianapolis, IN 46250

Re: K033612

Trade/Device Name: Factor II (prothrombin) G20210A Kit Regulation Number: 21 CFR 864.7280 Regulation Name: Factor V Leiden DNA mutation detection systems Regulatory Class: Class II Product Code: NPR Dated: November 14, 2003 Received: November 17, 2003

Dear Dr. Gregg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

K033612 510(k) Number (if known): NAA

Device Name:

Factor II (Prothrombin) G20210A Kit

Indications For Use:

The Factor II (Prothrombin) G20210A Kit is an in vitro diagnostic test for the detection and genotyping of a single point mutation (G to A at position 20210) of the human Factor II gene, from DNA isolated from human whole peripheral blood.

The Factor II (Prothrombin) G20210A Kit is indicated as an aid to diagnosis in the evaluation of patients with suspected thrombophilia.

The test is intended to be used on the LightCycler Instrument. The sample preparation must be performed according to a workflow procedure described in the package insert.

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

Tas for
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K033612

§ 864.7280 Factor V Leiden DNA mutation detection systems.

(a)
Identification. Factor V Leiden deoxyribonucleic acid (DNA) mutation detection systems are devices that consist of different reagents and instruments which include polymerase chain reaction (PCR) primers, hybridization matrices, thermal cyclers, imagers, and software packages. The detection of the Factor V Leiden mutation aids in the diagnosis of patients with suspected thrombophilia.(b)
Classification. Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Factor V Leiden DNA Mutation Detection Systems.” (See § 864.1(d) for the availability of this guidance document.)