K Number
K132956

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-06-05

(258 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The U fit Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The U fit Dental Implant System is for single and two stage surgical procedures. It is intended for delayed loading.

Device Description

The U fit Implant System is a dental implant system made of Titanium-6A1-4V ELI alloy intended to be surgically placed in the bone of the upper or lower jaw arches for lading affer a conventional healing period. The implants may be used to replace one or more missing teeth. The systems are similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The surface of these systems have been treated with RBM (Resorbable Blasted Media). The Fixture diameters are 3.8, 3.9, 4.0, 4.1, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 mm and lengths are 7. 8.5, 10. 11.5, 13, 14.5 mm in this system. The contained various abutments and accessories in the system are cover screw, abutment screw, healing abutment, solid abutment, transfer abutment, milling abutment, Temporary Abutment, protect cap, solid impression coping, solid lab analog, impression coping/pick-up, impression coping/transfer, and fixture lab analog. The material of abutments and accessories are Ti-6A1-4V ELI(ASTM F-136), POM-C and stainless steel. The abutment diameters are 4.0, 4.1, 4.4, 4.5, 4.8, 5.0, 5.5, 5.8, 6.0, 6.3, 6.5, 6.8, 7.0, 7.3, 7.8 and Heights are 4.0, 5.5, 7.0, 9.0. The implants are supplied sterile and the abutments and accessories are provided non-sterile. The abutments and accessories should be sterilized before use.

AI/ML Overview

The provided text is a 510(k) summary for the U fit Dental Implant System. It describes the device, its intended use, and compares it to predicate devices. However, it does not include information about specific acceptance criteria or a "study that proves the device meets the acceptance criteria" in the format typically associated with clinical performance evaluations for AI/ML devices.

Instead, the submission focuses on demonstrating substantial equivalence to predicate devices through comparisons of materials, intended use, and non-clinical performance testing. The "acceptance criteria" here implicitly refer to meeting the standards outlined in the FDA's guidance for root-form endosseous dental implants and abutments, and satisfying general safety and effectiveness requirements by demonstrating similarity to legally marketed devices.

Here's a breakdown based on the information available:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria in a table format for performance metrics relevant to an AI/ML device (e.g., sensitivity, specificity, AUC). Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence.

Aspect of Acceptance/ComparisonPredicate Device (OSSETEM US/SS/GS System & Dentium IMPLANTIUM ABUTMENTS) PerformanceSubject Device (U fit Dental Implant System) PerformanceComparison Outcome
Intended UseIndicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. For single and two stage surgical procedures. Intended for delayed loading.Identical.Substantially Equivalent
Material (Fixture)Ti-6Al-4V ELI ASTM F136Ti-6Al-4V ELI ASTM F136Substantially Equivalent
Material (Abutments/Accessories)Ti-6Al-4V ELI Gr.23, CP Ti Gr.4, ASTM F136, POM-CTi-6Al-4V ELI Gr.23, CP Ti Gr.4, ASTM F136, POM-C, Stainless SteelSubstantially Equivalent (Minor variations in specific grades, but still generally equivalent materials)
Fixture TypeSubmergedSubmergedSubstantially Equivalent
Surface TreatmentRBMRBMSubstantially Equivalent
BiocompatibilityComplies with ISO 10993 series, ASTM F67, F750-87, USP<88>, <151>, ISO 7405Complies with ISO 10993 series (3, 5, 6, 10, 11), ASTM F67, F750-87, USP<88>, <151>, ISO 7405Substantially Equivalent
Sterilization ValidationComplies with ISO 11737-1 & -2 (Gamma), ISO 17665-1 & -2 (Steam)Complies with ISO 11737-1 & -2 (Gamma), ISO 17665-1 & -2 (Steam). Note: Subject device implants are supplied sterile (gamma), abutments and accessories non-sterile (user steam sterilize). Predicate US/SS/GS Fixture Gamma sterilized; some predicate abutments Gamma, some user Steam.Substantially Equivalent (Sterilization methods are validated and acceptable)
Chemical & SEM Analysis (RBM)No residual post-RBM treatment.No residual post-RBM treatment.Substantially Equivalent
Compliance with GuidanceDesigned, manufactured, and tested in compliance with "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments."Designed, manufactured, and tested in compliance with "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments."Substantially Equivalent
DifferencesSlight differences in fixture designs and diameters compared to the predicateNoted slight differences in fixture designs and diametersDemonstrated to be safe and effective as the predicate and do not raise different questions of safety and effectiveness than the predicate.

2. Sample Size Used for the Test Set and Data Provenance:

This information is not applicable as the submission describes a medical device (dental implant system) made of physical materials, not an AI/ML software device that would use a "test set" of data in the common sense. The "testing" referred to is biocompatibility, sterilization validation, and material property testing, which are typically performed on samples of the manufactured device components rather than a "data set."

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not applicable. The assessment for this type of device does not involve expert readers establishing ground truth from images or data. Ground truth is established through laboratory testing following recognized standards.

4. Adjudication Method:

This information is not applicable. There is no adjudication method described as there are no expert interpretations to reconcile.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This information is not applicable. MRMC studies are used for evaluating diagnostic software where human readers' performance is assessed with and without AI assistance. This device is a physical implant, not a diagnostic software.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

This information is not applicable. This is not an algorithm or software device.

7. Type of Ground Truth Used:

The "ground truth" for verifying the safety and effectiveness of a dental implant system of this type is established through:

  • Material Specifications: Adherence to ASTM standards (e.g., ASTM F136 for Ti-6Al-4V ELI).
  • Biocompatibility Testing: Results from in vitro and in vivo tests (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation, hemocompatibility) according to ISO 10993.
  • Sterilization Validation: Demonstrated sterility assurance level (SAL) according to ISO 11737 and ISO 17665.
  • Surface Characterization: Chemical analysis and SEM imaging to confirm the RBM treatment and lack of residuals.
  • Mechanical Testing (Implied): Although not explicitly detailed in the summary, dental implants typically undergo mechanical strength and fatigue testing to ensure they can withstand physiological loads. This is usually part of meeting the FDA's "Class II Special Controls Guidance Document."

8. Sample Size for the Training Set:

This information is not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable.

{0}------------------------------------------------

K132956

510(k) Submission - U fit Dental Implant System

T Strong, Inc.

510(K) Summary

Submitter

T Strong, Inc. Han Chang Woo 181-4 Uijeon-ri, Jinyeong-eup, Gimhae-si. Gyeongnam 621-801, Republic of Korea Phone: +82-55-342-1771 Fax: +82-55-342-1775

Official Correspondent Kodent, Inc. Susan Park 325 N. Puente st. Unit B Brea, CA 92821 Email: kodentinc@gmail.com Phone: 714-525-0114 Fax: 714-525-01116

Date prepared: 9/17/2013

Device Information

Trade Name: U fit Dental Implant System Common Name: Endosseous dental implant Classification Name: Implant, Endosseous, Root-Form Product code: DZE Subsequent Product Code: NHA Device Class: Class II Regulation number: 21 CFR 872.3640

General Description

The U fit Implant System is a dental implant system made of Titanium-6A1-4V ELI alloy intended to be surgically placed in the bone of the upper or lower jaw arches for lading affer a conventional healing period.

The implants may be used to replace one or more missing teeth. The systems are similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The surface of these systems have been treated with RBM (Resorbable Blasted Media).

The Fixture diameters are 3.8, 3.9, 4.0, 4.1, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 mm and lengths are 7. 8.5, 10. 11.5, 13, 14.5 mm in this system.

The contained various abutments and accessories in the system are cover screw, abutment screw, healing abutment, solid abutment, transfer abutment, milling abutment, Temporary Abutment, protect cap, solid impression coping, solid lab analog, impression coping/pick-up, impression coping/transfer, and fixture lab analog.

The material of abutments and accessories are Ti-6A1-4V ELI(ASTM F-136), POM-C and stainless steel. The abutment diameters are 4.0, 4.1, 4.4, 4.5, 4.8, 5.0, 5.5, 5.8, 6.0, 6.3, 6.5, 6.8, 7.0, 7.3, 7.8 and Heights are 4.0, 5.5, 7.0, 9.0.

The implants are supplied sterile and the abutments and accessories are provided non-sterile. The abutments and accessories should be sterilized before use.

Indications for Use

The U fit Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including: cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed

JUN 0 5 2014

{1}------------------------------------------------

bridgework. The U fit Dental Implant System is for single and two stage surgical procedures. It is intended for delayed loading.

Predicate Device Information

The subject device is substantially equivalent to the following predicate device:

  • . US/SS/GS System manufactured by OSSETEM Co., Ltd. (K073247)
  • . Implantium Abutment manufactured by Dentium Co., Ltd. (K052823)

Comparison Chart

1) Fixtures

Subject devicePredicate device
Device nameU fit Dental Implant SystemUS/SS/GS System
510(k) numberNAK073247
ManufacturerT Strong, Inc.OSSETEM Co., Ltd.
Intended useIdentical to the predicateThe US/SS/GS System is indicated foruse in partially or fully edentulousmandibles and maxillae, in support ofsingle or multiple-unit restorationsincluding; cemented retained, screwretained, or overdenture restorations,and terminal or intermediate abutmentsupport for fixed bridgework. The U fitDental Implant System is for single andtwo stage surgical procedures. It isintended for delayed loading.
MaterialTi-6Al-4V ELIASTM F136Ti-6Al-4V ELIASTM F136
DesignImage: Four dental implantsImage: One dental implant
Fixture TypeSubmergedSubmerged
Implant diameter3.8, 3.9, 4.0, 4.1, 4.5,5.0, 5.5, 6.0, 6.5, 7.0 mm3.3 - 6.0 mm
Implant length7, 8.5, 10, 11.5, 13, 14.5 mm7 - 18 mm
ComponentsVarious abutments and accessoriesVarious abutments and components
Surface treatmentRBMRBM
Gamma sterilizedYesYes
Product CodeDZEDZE

{2}------------------------------------------------

510(k) Submission - U fit Dental Implant System

T Strong, Inc.

2) Abutments

,所

Subject DevicePredicate Device
Device NameU fit Dental Implant SystemIMPLANTIUM ABUTMENTS
510k #NAK052823
ManufacturerT Strong, Inc.Dentium, Inc.
Product CodeNHANHA
Cover Screw
Design
Intended UseTo provide sealing effect for fixtureTo provide sealing effect for fixture
MaterialTi-6Al-4V ELI Gr.23Ti-6Al-4V ELI Gr.23
SterilizationSteam Sterilization by userSteam Sterilization by user
Abutment Screw
Design
Intended UseTo connect abutment to fixtureTo connect the abutment to the fixture
MaterialTi-6Al-4V ELI, ASTM F136Ti-6Al-4V ELI, ASTM F136
SterilizationSteam Sterilization by userSteam Sterilization by user
Healing Abutment
Design
Intended UseTo help the soft tissue of gum naturallyformedTo help the soft tissue of gum naturallyformed
MaterialTi-6Al-4V ELI Gr.23Ti-6Al-4V Eli, Gr. 23
SterilizationSteam Sterilization by userGamma Sterilization
Solid Abutment
Design
Intended UseCement retained restorationCement retained restoration
MaterialTi-6Al-4V ELI Gr.23CP Ti Gr.4
SterilizationSteam Sterilization by userSteam Sterilization by user
Transfer Abutment
DesignImage: Transfer Abutment Design 1Image: Transfer Abutment Design 2
Intended UseCement retained restorationCement retained restoration
MaterialTi-6Al-4V ELI Gr.23CP Ti Gr.4
SterilizationSteam Sterilization by userSteam Sterilization by user
Milling Abutment
DesignImage: Milling Abutment Design 1Image: Milling Abutment Design 2
Intended UseCement retained restorationCement retained restoration
MaterialTi-6Al-4V ELI Gr.23CP Ti Gr3. ASTM F67
SterilizationSteam Sterilization by userSteam Sterilization by user
Temporary Abutment
DesignImage: Temporary Abutment Design 1Image: Temporary Abutment Design 2
Intended UseTo manufacture temporary prosthesesTo manufacture temporary prostheses
MaterialTi-6Al-4V ELI Gr.23CP Ti Gr.4
SterilizationSteam Sterilization by userSteam Sterilization by user

{3}------------------------------------------------

Comparison Analysis

The U fit Dental Implant System has a substantially equivalent intended use as the identified predicate. The U fit Dental Implant System is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments, and they are all constructed of titanium.

The U fit Dental Implant System is substantially equivalent in materials, indications and intended use, packaging, labeling, and performance to the predicate device currently marketed in the U.S.

The only differences between the subject device and the predicate are slight differences in fixture designs and diameters.

Any differences in technology characteristics are accompanied by information that demonstrated the device is safe and effective as the predicate and do not raise different questions of safety and effectiveness than the predicate.

{4}------------------------------------------------

Non-Clinical Data

The manufacturing and development process for the U fit Dental Implant System is compliance with ISO 13485(Medical devices - Quality management systems - Requirements for regulatory purposes), ISO 14971(Medical devices - Application of risk management to medical devices), ISO 10993-1(Biological evaluation of medical devices), EN 980(Symbols for use in the labelling of medical devices), EN 1041(Information supplied by manufacturer of medical devices). Biocompatibility testing has been performed in accordance with ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO10993-10, ISO 10993-11, ASTM F67, ASTM F750-87, USP<88>, USP<151>, ISO 7405. Sterilization validating testing has been performed in accordance with ISO 11737-1 & ISO 11737-2 for gamma sterilization and ISO 17665-1 and ISO 17665-2 for steam sterilization. Analysis of chemical and SEM images have been performed to verify that there is no residual after RBM treatment on the Fixture. Non-clinical testing consisted of testing in accordance with the FDA guidance "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The results of the non-clinical testing demonstrate that the U fit Dental Implant System is substantially equivalent to the predicate devices.

Conclusion

The U fit Dental Implant System constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. The risks of using the device as recommended pose no greater risks than other implant system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, U fit Dental Implant System and its predicate devices are substantially equivalent.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

June 5, 2014

T Strong, Incorporated C/O Ms. Susan Park Kodent, Incorporated 325 N. Puente Street, Unit B Brea, California 92821

Re: K132956

Trade/Device Name: U Fit Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: April 24, 2014 Received: April 24, 2014

Dear Ms. Park:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Ms. Park

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (2) CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21

CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S: Runner -S

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

and the commend of the comments of the country of

Enclosure

{7}------------------------------------------------

Indication for Use

K132956 510(K) Number (if known):

Device Name: U fit Dental Implant System

The U fit Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The U fit Dental Implant System is for single and two stage surgical procedures. It is intended for delayed loading.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1

Sheena A. Green -S 2014.06.05 07:48:23-04'00

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.