K Number
K073247

Validate with FDA (Live)

Device Name
US/SS/GS SYSTEM
Date Cleared
2008-03-07

(109 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The US/SS/GS System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The US/SS/GS Fixture System is for one and two stage surgical procedures. It is not for immediate load.

Device Description

The US/SS/GS System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches.

AI/ML Overview

This 510(k) submission (K073247) for the OSSTEM Implant Co., Ltd. US/SS/GS System is for an endosseous dental implant. This document is a summary of the 510(k) submission and primarily focuses on establishing substantial equivalence to previously cleared predicate devices. It does not contain the details of a study with specific acceptance criteria and performance data in the format requested.

Therefore, I cannot populate the table and answer the specific questions about acceptance criteria, device performance, sample sizes, expert involvement, and ground truth methodologies, as this information is not present within the provided text.

The document states:

  • "The US/SS/GS System has been subjected to safety, performance, and product validations prior to release."
  • "Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations."

However, it does not provide the details of these validations or tests, nor does it specify numerical acceptance criteria or the reported performance against them.

In a 510(k) submission for a device like a dental implant, substantial equivalence is often demonstrated through:

  • Comparison of materials, design, indications for use, and technological characteristics to predicate devices.
  • Results from mechanical testing (e.g., fatigue strength, static strength) to ensure the device meets recognized standards.
  • Biocompatibility testing.
  • Dimensional and tolerance analysis.

The provided text only asserts that these tests and validations were done and that the device is substantially equivalent to predicate devices (K062030, K062051, K063861). It does not provide the numerical data, study design, or expert review details that would allow me to fill in the requested information.

{0}------------------------------------------------

K073247

Image /page/0/Picture/1 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo is on the left, with the company name and address on the right. The address is #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea. The phone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website is www.osstem.com.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: November 7 , 2007

  1. Company and Correspondent making the submission:
- Submitter's Name :OSSTEM Implant Co., Ltd.
- Address :#507-8 Geoje3-Dong Yeonje-GuBusan, 611-804, Republic of Korea
- Contact :Mr. JongHyuk Seo
2. Device :

DZE

Trade or (Proprietary) Name : US/SS/GS System Common or usual name : Dental Implant Classification Name : Endosseous Dental Implant 21CFR872.3640 Class II

3. Predicate Device :

The US System, Osstem Implant Co., Ltd, K062030 The SS System, Osstem Implant Co., Ltd, K062051 The GS System, Osstem Implant Co., Ltd, K063861

    1. Description :
      The US/SS/GS System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches.

The US/SS/GS System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.

The US/SS/GS System is substantially equivalent in design, function and intended use to the US System(K062030), SS System(K062051) and GS System(K063861) of Osstem Implant Co., Ltd.

QS-QI-505-3(Rev.0)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo with a circular design on top and the word "osstem" below it. The circular design consists of two curved lines that form an incomplete circle, with the top line being thicker than the bottom line. The word "osstem" is written in a sans-serif font, with all letters in lowercase.

OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com

5. Indication for use :

The US/SS/GS System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The US/SS/GS Fixture System is for one and two stage surgical procedures. It is not for immediate load.

6. Review :

The US/SS/GS System has similar material, indication for use, design and technological characteristics as the predicate device.

The US/SS/GS System has been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.

7. Conclusion :

Based on the information provided in this premarket notification Osstem concludes that the US/SS/GS System is safe and effective and substantially equivalent to the predicate device as described herein

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OSSTEM Implant Company, Limited C/O Mr. MinJoo Kim Manager OSSTEM Incorporated One Ben Fairless Drive Fairless Hills, Pennsylvania 19030

Re: K073247

Trade/Device Name: US/SS/GS System Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: December 18, 2007 Received: December 19, 2007

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

MAR - 7 2008

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Kim

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Suste Y. Michie OMD.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the word "osstem" in lowercase letters below a circular graphic. The graphic appears to be a stylized representation of a circle or ring, possibly with some internal design elements. The text is in a simple, sans-serif font and is positioned directly beneath the graphic.

510(k) Number K _

Device Name: US/SS/GS System

Indication for use : The US/SS/GS System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The US/SS/GS System is for single and two stage surgical procedures. It is not for immediate load.

Prescription Use X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runne

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

KO73247 510(k) Number:

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.