K Number
K132843

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-01-08

(120 days)

Product Code
Regulation Number
866.3365
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-negative bacteria as determined by gram stain.

BC-GN detects and identifies the following:

Bacterial Genera and SpeciesResistance Markers
Acinetobacter spp.CTX-M (blaCTX-M)
Citrobacter spp.KPC (blaKPC)
Enterobacter spp.NDM (blaNDM)
Proteus spp.VIM (blaVIM)
Escherichia coli1IMP (blaIMP)
Klebsiella pneumoniaeOXA (blaOXA)
Klebsiella oxytoca
Pseudomonas aeruginosa

BC-GN will not distinguish Escherichia coli from Shigella spp. (S. dysenteriae, S. flexneri, S. boydii, and S. sonnei)

BC-GN is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however. is not used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by BC-GN, to detect mixed infections that may not be detected by BC-GN, for association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.

Device Description

The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN) is a molecular assay which relies on detection of specific nucleic acid targets in a microarray format. For each of the bacterial nucleic acid sequences detected by the BC-GN test, unique Capture and Mediator oligonucleotides are utilized, with gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides are covalently bound to the microarray substrate and hybridize to a specific portion of the nucleic acid targets. The Mediator oligonucleotides have regions which bind to a different portion of the same nucleic acid targets and also have sequences which allow binding of gold nanoparticle probes. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency and provide accurate detection of target capture.

The BC-GN test is performed on the Verigene System, a sample-to-result, fully automated. bench-top molecular diagnostics workstation consisting of two components: the Verigene Reader and the Verigene Processor SP. For the BC-GN test, the Verigene System allows automated nucleic acid extraction from positive bacteria-containing blood culture specimens and target detection of bacteria-specific DNA. The BC-GN test utilizes single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.

The Reader is the Verigene System's central control unit and user interface, and, with a touch-screen control panel and barcode scanner, guides the user through test processing. imaging, and test result generation. The Verigene Processor SP executes the test procedure. automating the steps of (1) Sample Preparation- cell lysis and magnetic bead-based bacterial DNA isolation from blood culture samples, and (2) Hybridization-- detection and identification of bacterial-specific DNA in a microarray format by using gold nanoparticle probe-based technology. Once the specimen is loaded by the operator, all other fluid transfer steps are performed by an automated pipette that transfers reagents between wells of the trays and loads the specimen into the Test Cartridge for hybridization. Single-use disposable test consumables and a self-contained Verigenc Test Cartridge are utilized for each sample tested with the BC-GN test.

To obtain the test results after processing is complete. the user removes the Test Cartridge from the Processor SP, and inserts the substrate holder into the Reader for analysis. Light scatter from the capture spots is imaged by the Reader and intensities from the microarray spots are used to make a determination regarding the presence (Detected) or absence (Not Detected) of a bacterial nucleic acid sequence/analyte. This determination is made by means of software-based decision algorithm resident in the Reader.

AI/ML Overview

The Nanosphere Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN) is a qualitative multiplexed in vitro diagnostic test designed for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers directly from positive blood culture media.

Here's an analysis of its acceptance criteria and the supporting studies:

1. Table of Acceptance Criteria (Performance Goals) and Reported Device Performance (Method Comparison Study)

The document does not explicitly state "acceptance criteria" as a set of predefined thresholds for performance metrics. However, the "Method Comparison" study presents the device's performance against reference methods, which implicitly serve as the comparison points for its effectiveness. The reported performance metrics are Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA), along with their 95% Confidence Intervals.

Organism / Resistance MarkerImplied Acceptance Criterion (Performance Goal; generally high agreement >90-95% for diagnostic tests)Reported Device Performance (PPA (95% CI))Reported Device Performance (NPA (95% CI))
Acinetobacter spp.High PPA and NPA98.2% (90.5-100)99.9% (99.6-100)
Citrobacter spp.High PPA and NPA100% (92.8-100)99.9% (99.6-100)
Enterobacter spp.High PPA and NPA97.6% (93.0-99.5)99.4% (98.8-99.7)
Proteus spp.High PPA and NPA100% (93.8-100)99.9% (99.6-100)
Escherichia coliHigh PPA and NPA99.8% (98.9-100)99.4% (98.7-99.8)
Klebsiella pneumoniaeHigh PPA and NPA93.1% (89.7-95.7)100% (99.7-100)
Klebsiella oxytocaHigh PPA and NPA92.2% (82.7-97.4)99.6% (99.0-99.8)
Pseudomonas aeruginosaHigh PPA and NPA97.6% (93.3-99.5)100% (99.7-100)
CTX-M (Resistance Marker)High PPA and NPA98.7% (95.4-99.8)99.9% (99.5-100)
OXA (Resistance Marker)High PPA and NPA95.3% (86.9-99.0)99.9% (99.5-100)
KPC (Resistance Marker)High PPA and NPA100% (93.1-100)100% (99.7-100)
VIM (Resistance Marker)High PPA and NPA100% (91.4-100)100% (99.7-100)
NDM (Resistance Marker)High PPA and NPA100% (91.4-100)100% (99.7-100)
IMP (Resistance Marker)High PPA and NPA100% (92.6-100)100% (99.7-100)

Note: The document does not explicitly state numerical acceptance criteria. The "implied acceptance criterion" section is a general interpretation based on typical regulatory expectations for diagnostic accuracy.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: A total of 1412 specimens were analyzed in the method comparison study.
    • 604 prospectively-collected fresh specimens
    • 272 prospectively-collected frozen specimens
    • 239 selected frozen specimens
    • 297 simulated frozen specimens
  • Data Provenance: The study was conducted at thirteen (13) investigational sites. The document does not specify the countries of origin for these sites or the data itself, but such clinical trials for FDA submissions are typically conducted in the US or under comparable regulatory frameworks. The inclusion of "prospectively-collected fresh specimens" and "prospectively-collected frozen specimens" indicates prospective data collection, while "selected frozen specimens" and "simulated frozen specimens" indicate retrospective or artificially prepared samples. This suggests a mixed approach to sample collection.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts used to establish ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience"). Instead, it describes the methods used to establish ground truth:

  • For bacterial organisms: "standard culture-based automated phenotypic bacterial identification reference methods." This implies that the ground truth was established by clinical microbiology laboratories following established protocols, likely interpreted by qualified clinical microbiologists or medical laboratory scientists.
  • For resistance markers: "the combination of PCR amplification and bidirectional sequencing confirmation." This indicates molecular biology techniques, which would also be performed and interpreted by appropriately trained laboratory personnel.

4. Adjudication Method for the Test Set

The document does not mention an explicit "adjudication method" involving multiple human readers for the test set results. The ground truth was established using standard reference laboratory methods, not through an adjudication process of human interpretations of the device's output. The device's results were directly compared to these reference methods.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not reported. This device is an in vitro diagnostic test (an algorithm-only device without human-in-the-loop performance being evaluated in the clinical study presented) rather than an imaging or interpretive AI device where human reader performance is typically assessed and compared to AI-assisted performance. The study focuses on the agreement of the device's output with standard reference methods. Therefore, an effect size of human readers improving with AI vs. without AI assistance is not applicable here.

6. Standalone (Algorithm-Only Without Human-in-the-Loop Performance) Study

Yes, the method comparison study directly assesses the standalone performance of the Verigene BC-GN test (algorithm only). The device's results are automatically generated by the Verigene System (Reader and Processor SP) and then compared to the ground truth established by reference laboratory methods, without human intervention in the interpretation of the device's output that would then be compared to human interpretation.

7. Type of Ground Truth Used

The ground truth used was:

  • Culture-based automated phenotypic bacterial identification reference methods for bacterial organisms.
  • PCR amplification and bidirectional sequencing confirmation for resistance markers.

8. Sample Size for the Training Set

The document does not explicitly specify a "training set" in the context of machine learning model development. This device appears to be a molecular diagnostic assay using a microarray and a software-based decision algorithm, rather than a system heavily reliant on a continuously learning or adaptable AI model that would typically have a distinct training phase with a specific dataset.

However, analytical studies involved extensive testing that could be considered analogous to developing and refining the device's performance characteristics:

  • Analytical Sensitivity (LOD): Tested 12 bacterial strains.
  • Analytical Reactivity (Inclusivity): Tested 195 strains of 44 different organisms (including 79 with resistance markers).
  • Analytical Specificity (Exclusivity): Tested 172 "non-BC-GN panel" organisms.
  • Competitive Inhibition / Mixed Growth: Multiple studies involving combinations of bacterial organisms.
  • Precision/Repeatability: 18-member panel tested for 864 replicates.
  • Reproducibility: 18-member panel tested for 1620 replicates across 3 external sites.

These studies contribute to the design and validation of the test's targets, probes, and decision algorithm, but a "training set" in the context of statistical machine learning for inferential models is not presented.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, a distinct "training set" with ground truth in the typical machine learning sense is not explicitly described. For the analytical studies (e.g., LOD, inclusivity, exclusivity), the ground truth for bacterial identification and resistance marker presence would have been established using well-characterized laboratory strains and standard microbiological and molecular identification techniques. These involve:

  • Pure cultures: For LOD and inclusivity, known bacterial strains are used.
  • Conventional identification methods: Such as cell morphology, growth characteristics, biochemical tests, and potentially 16S rRNA sequencing for difficult or novel strains.
  • Molecular techniques for resistance markers: Such as PCR and sequencing to confirm the presence and identity of specific resistance genes.

These methods are the gold standards in microbiology for characterizing bacteria and their genetic elements, forming the basis of the device's design and analytical performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "Nanosphere" with a stylized logo to the left. The logo is a circle divided into two halves, one filled with a solid pattern and the other with horizontal lines. The word "Nanosphere" is written in a simple, sans-serif font, with each letter outlined in black.

510(k) Summarv

The Summary for this 510(k) submission is submitted in accordance with the requirements of SMDA 1900 and CFR 807.92

510(k) Number:

Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN) K132843:

Summary Preparation Date:

January 6, 2014

Submitted by:

Nanosphere, Inc. 4088 Commercial Avenue Northbrook, IL 60062 Phone: 847-400-9000 Fax: 847-400-9176

Contact:

Mark A. Del Vecchio Vice President, Regulatory Affairs

Proprietary Names:

For the instrument: Verigene® System For the assay: Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN)

Common Names:

For the instrument:

Bench-top molecular diagnostics workstation

For the assay:

Gram-negative blood culture assay Acinetobacter spp. blood culture assay Citrobacter spp. blood culture assay Enterobacter spp. blood culture assay Proteus spp. blood culture assay Klebsiella pneumoniae blood culture assay Klebsiella oxytoca blood culture assay Pseudomonas aeruginosa blood culture assay Escherichia coli blood culture assay CTX-M resistance marker blood culture assay OXA resistance marker blood culture assay KPC resistance marker blood culture assay VIM resistance marker blood culture assay NDM resistance marker blood culture assay IMP resistance marker blook culture assay

{1}------------------------------------------------

Regulatory Information:

Regulation section:

866.3365 Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures

Clussification:

Class II

Panel:

Microbiology (83)

Product Code(s):

PEN Gram-Negative Bacteria and Associated Resistance Markers

Other codes used by predicate devices:

Gram-Positive Bacteria and their Resistance Markers PAM

PEO Fungal Organisms, Nucleic Acid-Based Assay

OOI Real Time Nucleic Acid Amplification System

Predicate Devices:

FilmArray Blood Culture Identification (BCID) Panel (K130914, Biofire Diagnostics) Verigene Gram-Positive Blood Culture (BC-GP) Nucleic Acid Test (K122514 / K113450)

Intended Use:

The Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in virro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-negative bacteria as determined by gram stain.

BC-GN detects and identifies the following:

Bacterial Genera and SpeciesResistance Markers
Acinetobacter spp.CTX-M (blaCTX-M)
Citrobacter spp.KPC (blaKPC)
Enterobacter spp.NDM (blaNDM)
Proteus spp.VIM (blaVIM)
Escherichia coli1IMP (blaIMP)
Klebsiella pneumoniaeOXA (blaOXA)
Klebsiella oxytoca
Pseudomonas aeruginosa

BC-GN will not distinguish Escherichia coli from Shigella spp. (S. dysenteriae, S. flexneri, S. boydii, and S. somet)

BC-GN is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however. is not used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by BC-GN, to detect mixed infections that may not be detected by BC-GN, for association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.

{2}------------------------------------------------

Technological Characteristics:

The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN) is a molecular assay which relies on detection of specific nucleic acid targets in a microarray format. For each of the bacterial nucleic acid sequences detected by the BC-GN test, unique Capture and Mediator oligonucleotides are utilized, with gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides are covalently bound to the microarray substrate and hybridize to a specific portion of the nucleic acid targets. The Mediator oligonucleotides have regions which bind to a different portion of the same nucleic acid targets and also have sequences which allow binding of gold nanoparticle probes. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency and provide accurate detection of target capture.

The BC-GN test is performed on the Verigene System, a sample-to-result, fully automated. bench-top molecular diagnostics workstation consisting of two components: the Verigene Reader and the Verigene Processor SP. For the BC-GN test, the Verigene System allows automated nucleic acid extraction from positive bacteria-containing blood culture specimens and target detection of bacteria-specific DNA. The BC-GN test utilizes single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.

The Reader is the Verigene System's central control unit and user interface, and, with a touch-screen control panel and barcode scanner, guides the user through test processing. imaging, and test result generation. The Verigene Processor SP executes the test procedure. automating the steps of (1) Sample Preparation- cell lysis and magnetic bead-based bacterial DNA isolation from blood culture samples, and (2) Hybridization-- detection and identification of bacterial-specific DNA in a microarray format by using gold nanoparticle probe-based technology. Once the specimen is loaded by the operator, all other fluid transfer steps are performed by an automated pipette that transfers reagents between wells of the trays and loads the specimen into the Test Cartridge for hybridization. Single-use disposable test consumables and a self-contained Verigenc Test Cartridge are utilized for each sample tested with the BC-GN test.

To obtain the test results after processing is complete. the user removes the Test Cartridge from the Processor SP, and inserts the substrate holder into the Reader for analysis. Light scatter from the capture spots is imaged by the Reader and intensities from the microarray spots are used to make a determination regarding the presence (Detected) or absence (Not Detected) of a bacterial nucleic acid sequence/analyte. This determination is made by means of software-based decision algorithm resident in the Reader.

Page 3 of 15

{3}------------------------------------------------

Performance Data - Analytical Testing

Analytical Sensitivity / Limit of Detection (LOD)

The following analytical sensitivity (LOD) has been established for the detection of each of the eight (8) BC-GN test bacterial targets and six (6) resistance markers, established by testing a total of 12 bacterial strains, most of which contained the listed resistance markers.

TargetResistanceMarker(s)No. ofStrainsTestedLOD (CFU/mL)
Acinetobacter spp.OXA2$4.0x10^5$ to $4.6x10^6$
Citrobacter spp.VIM2$6.9x10^6$ to $1.3x10^7$
Enterobacter spp.KPC2$4.1x10^6$ to $1.1x10^7$
Proteus spp.-2$1.9x10^5$ to $7.7x10^5$
Klebsiella pneumoniaeOXA; CTX-M1$1.2x10^7$
Klebsiella oxytocaCTX-M1$2.0x10^7$
Escherichia coliNDM1$3.7x10^6$
Pseudomonas aeruginosaIMP1$2.3x10^7$

Analytical Reactivity (Inclusivity)

Analytical reactivity was evaluated by triplicate testing of a large number of gram-negative bacterial species, both with and without antibiotic resistance markers, as appropriate. The panel organisms were selected to cover the genetic diversity of each BC-GN target and resistance marker. A total of 195 strains of 44 different organisms were tested including 36 strains of 12 Acinetobacter species, 41 strains of 11 Citrobacter species, 29 strains of 8 Enterobacter species, 16 strains of 5 Proteus species, 8 strains of 4 Shigella species, 10 strains of Klebsiella oxytoca, 25 strains of Klebsiella pneumoniae, 13 strains of Pseudomonas aeruginosa, and 17 strains of Escherichia coli. Of these 195 strains, 79 contained one or more resistance markers associated with 11 different bacterial species including a total of 38 strains containing CTX-M, 17 containing OXA, 12 containing IMP, 10 containing VIM, 10 containing KPC, and 9 containing NDM. The bacterial and resistance targets for all of the 195 strains were correctly detected (including the expected cross-reactivity with the 4 Shigella species) with three exceptions. Two (2) of the 13 Acinetobacter radioresistens strains observed false negative results for the OXA marker for 9 of 24 replicates tested. In both strains however, the bacterial target was correctly identified as "Acinetobacter spp." in all replicates tested. One of the Citrobacter amalonaticus strains observed false negative results Citrobacter spp. for two of nine replicates tested.

{4}------------------------------------------------

Analytical Specificity (Exclusivity)

Analytical specificity was assessed using organisms phylogenetically related to panel organisms detected by the BC-GN test, organisms without gene sequence information. common blood-borne pathogens, as well as organisms potentially present as contaminants in blood culture specimens. The organisms tested were divided into two distinct panels. The first panel consisted of the 195 "BC-GN panel" organisms, which in total comprised the analytical inclusivity study samples. There were no false positive results observed for any of the samples, indicating no cross-reativity of any of the panel members.

The second panel consisted of 172 "non-BC-GN panel" organisms which were not expected to be detected by the BC-GN test, including 88 gram-negative bacteria (including Acinetobacter baumanii containing OXA-51), 71 gram-positive bacteria, six (6) gramnegative cocci bacteria, and seven (7) yeast strains. Of these 172 strains tested, 159 demonstrated no cross-reactivity with the BC-GN test, while thirteen (13) organisms were determined to cross-react with the tests, as listed below. In addition. BC-GN will not distinguish E. coli from Shigella spp. including S. dysenteriae, S. flexneri, S. bovdii, and S. sonnei.

BC-GN Test Target for Which CrossReactivity ObservedCross Reactive Organism
Citrobacter spp.Buttiauxella gaviniaeEnteric group 137
Enterobacter spp.Klebsiella variicolaLeclercia adecarboxylata
Escherichia coliEscherichia albertiiS. dysenteriaeS. flexneriS. boydiiS. sonnei
Klebsiella oxytocaKluyvera ascorbataRaoultella ornithinolyticaRaoultella planticolaCedecea davisae
CTX-MKluyvera georgiana*Leminorella grimontiiEnterococcus raftinosusCandida parapsilosisblaKLUAblaKLUGblaKLUY

*Organism confirmed by bi-directional sequencing to contain CTX-M

{5}------------------------------------------------

Interfering Substances

The potential inhibitory effects of substances that may be present in patient blood specimens and a blood culture bottle additive present in different blood culture media were tested with the BC-GN test at biologically or experimentally relevant concentrations. The design of the study took into consideration that the BC-GN test sample preparation process inherently acts to minimize the potential for an interferent present in the blood culture samples to impact the test. The effects of potential interfering substances were evaluated with one representative strain of each of the eight (8) BC-GN test bacterial targets and six (6) resistance markers in the presence of several endogenous substances. Hemoglobin, triglycerides, conjugated and unconjugated bilirubin, y-globulin and Sodium Polyanethol Sulfonate (SPS) were selected for testing. Control samples containing no interferents were also tested. No interfering effects were observed.

Fresh vs. Frozen Samples

The stability of specimens subjected to multiple freeze/thaw cycles was evaluated by conducting a fresh versus frozen specimen study. This study was performed to demonstrate that frozen storage of samples does not affect the accuracy of test results such that frozen clinical blood culture samples could be utilized in the evaluation of the BC-GN test. This study was conducted using positive blood culture specimens which were tested before and after freezing (at or below -70 °C) to allow a comparison. Representative strains of each of the eight (8) BC-GN test bacterial targets and six (6) resistance markers, as well as two additional organisms, Hafnia alvei and Staphylococcus epidermidis which served as negative organism controls, were grown to bottle positivity. Cultured samples were tested at baseline (fresh/unfrozen) and then frozen at or below -70 ℃, thawed and tested, and the freeze/thaw and testing of the specimen repeated to evaluate two freeze-thaw cycles. A comparison of matched sets of specimens demonstrated 100% agreement between the fresh and frozen samples for all the analytes and for all the conditions tested. The results from this study demonstrate that the BC-GN test results are not compromised when samples are exposed to up to two freeze/thaw cycles.

Competitive Inhibition / Mixed Growth

A competitive inhibition study was conducted to evaluate the impact of combinations of gram-negative organisms present in the same blood culture bottle at similar concentrations prior to culturing. Combinations of bacterial organisms representative of all BC-GN test target analytes were co-inoculated into individual blood culture bottles at clinically-relevant starting concentrations and incubated to positivity. The BC-GN test correctly detected the bacteria and resistance marker(s) for four (4) of the eight target organisms, irrespective of combination(A. baumannii/OXA, C. freundii/VIM, K. pneumoniae/OXA and CTX-M, and P. mirabilis) present in co-inoculated blood culture bottles. demonstrating that these organisms are not subject to competitive inhibition from BC-GN test panel members when both are present at concentrations expected in routine clinical practice. For the remaining four (4) organisms, at least one of the expected bacterial targets or resistance markers was not detected. This was due most likely to the slower growth rates of several organisms relative to other co-infected organisms, as faster growing organisms may reach a higher concentration at bottle positivity compared with a slower growing organism. This may lead to slowergrowing organism concentrations at bottle positivity that are below the limit of detection of the test.

{6}------------------------------------------------

A second competitive inhibition study was conducted to further evaluate representative combinations of the four (4) organisms for which at least one false negative result was observed during the first study. This involved retesting nine specific organism combinations. This study was designed to confirm whether false negative results observed with these combinations were due to the slower growth rates of the undetected organisms relative to the detected organisms. Therefore, growth rate was eliminated as an experimental variable by testing mixed culture organisms at "bottle positivity" concentrations and above the LoD for each organism. Of the nine (9) combinations, eight (8) combinations vielded expected calls. For one combination, K. oxylocalCTX-M and E. coli/NDM, the K. oxytoca target was detected at a rate of 78% (7/9); however, 100% detection was observed for the other three targets in this sample (E. coli, NDM, and CTX-M). This demonstrated that except in one instance, growth rate, not competitive inhibition was a contributing factor to the initial observed false negative results.Carry-over / Cross-contamination

Twelve Verigene SP instruments were used to assess the potential for carryover/crosscontamination by alternately running "high positive" samples followed by negative samples. Representative strains of the eight (8) bacterial target organisms and six (6) resistance markers were used to prepare the high positive samples. All of the high positive samples yielded the expected "Detected" results for the intended bacterialmarker and "Not Detected" results for the other analytes. The negative samples gave a "Not Detected" call for all analytes. There was no evidence of carryover/cross-contamination observed.

{7}------------------------------------------------

Cutoff Verification

In order to verify the assay cut-off, the target mean intensity values observed with the BC-GN test were examined for a panel of eighteen (18) contrived specimens, representing all of the BC-GN test analytes and two (2) additional organisms not detected by the BC-GN test, which served as negative organism controls. Taking into consideration replicate testing, a complete set of 1:10 dilutions of all the organisms, and the number of probes on the BC-GN test microarray, 2460 data points were compiled for the threshold evaluation, using logistic fit and ROC statistics.

Validation of Additional Blood Culture Bottle Types

In order to achieve a"universal bottle-type" claim, the performance of the BC-GN test was evaluated for twelve (12) types of blood culture media (in addition to the BACTEC PLUS-Aerobic/F bottle used for the other analytical validation studies) using three different automated blood culture monitoring systems, providing a significant representation of commercially-available blood culture media bottles and monitoring systems (see listing of culture bottles tested below). This study was conducted to demonstrate that BC-GN test target organisms grow to a sufficient concentration in the various bottle types and additionally to demonstrate the stability of bacterial DNA is sufficiently stable over an extended time period of 36 hours once bottle positivity has been reached.

Blood Culture Bottle(Manufacturer)Blood Culture System
BD BACTEC(Becton Dickenson)Standard/10 Aerobic/FBACTEC 9050
PLUS - Anaerobic/F
Peds Plus/F
Standard/Anaerobic/F
Lytic/10 Anaerobic/F
BacT/ALERT(BioMerieux)SA Standard AerobicBacT/ALERT 3D
FA FAN® Aerobic
PF Pediatric FAN
SN Standard Anaerobic
FN FAN Anaerobic
VersaTREK(Thermo Scientific)REDOX 1 EZ Draw® / AerobicVersaTREK
REDOX 2 EZ Draw® / Anaerobic

In the study, representative BC-GN test bacterial organisms (all eight bacteria and six resistence markers) were inoculated into each of the twelve different bottle types which were spiked with anti-coagulated human whole blood. The bottles were placed on the appropriate culture system and upon reaching bottle positivity, an aliquot of each sample was tested with the BC-GN test in triplicate. To evaluate specimen stability in the various bottles, one bottle of each organism was then stored for 36 hours at each of three temperature conditions: refrigerated, ambient, and on the blood culture system, after which BC-GN testing was performed. The test results demonstrated that these twelve (12) blood culture bottles are appropriate for use with the BC-GN test and that specimens are stable in those bottles at various temperature storage conditions for up to 36 hours after reaching bottle positivity.

{8}------------------------------------------------

Precision / Repeatability

results from the study are contained in the table below.

Precision was evaluated by testing an 18-member panel, containing eight (8) unique specimens representing each target analyte (both bacterial and resistance markers) detected by the BC-GN test, as well as two negative controls, one consisting of negative blood culture media and the second containing an organism not detected by the BC-GN test (Hafnia alver). The 18-member panel was tested in-house by Nanosphere twice daily by two operators over twelve (12) non-consecutive days for a total of forty-eight replicates per sample. Except for the negative controls, organisms were tested at Bottle Positivity and Bottle Positivity + 8 hours incubation. A total of 864 tests were conducted and the final call rate for the study was 99.9% (863/864). There was one inaccurate call involving a Klebsiella pneumoniae/OXA/CTX-M specimen, whereby BC-GN detected "K. oxytoca", in addition to the correct expected calls, resulting in a call accuracy of 99.9% (862/863) for the study. The

SampleBottle PositivityBottle Positivity + 8hours
Organism/SpecimenResistanceMarker(s)Final CallRateAccuracyFinal CallRateAccuracy
Negative Control - Blood CultureMediaN/A100%(48/48)92.6-100100%(48/48)92.6-100--
Hafnia alveiN/A97.9%(47/48)88.9-100100%(47/47)92.5-100--
Acinetobacter baumaniiOXA100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100
Citrobacter freundiiVIM100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100
Enterobacter cloacueKPC100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100
Escherichia coliNDM100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100
Klebsiella pneumoniaeOXA,CTX-M100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-10097.9%(47/48)88.9-100
Klebsiella oxytocaCTX-M100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100
Proteus mirabilisN/A100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100
Pseudomonas aeruginosaIMP100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100100%(48/48)92.6-100

{9}------------------------------------------------

Reproducibility

Reproducibility was evaluated by testing an 18-member panel (the same sample panel that was tested for the evaluation of precision with one exception), containing eight (8) unique specimens representing each target analyte detected by BC-GN, as well as two negative controls, one consisting of negative blood culture media only and the second containing an organism not detected by BC-GN (Morganella morganii). The 18-member panel was tested at three (3) external sites twice daily in triplicate on five (5) non-consecutive days for a total of 90 replicates per sample. Except for the negative controls, organisms were tested at Bottle Positivity and Bottle Positivity + 8 hours incubation. The final call rate and accuracy for the study was 100% (1620/1620). The results from the study are contained in the table below.

SampleBottle PositivityBottle Positivity + 8 hours
Organism/SpecimenResistanceMarker(s)Final CallRateAccuracyFinal CallRateAccuracy
Negative Control - BloodCulture MediaN/A100%90/90(96.0-100)100%90/90(96.0-100)--
Morganella morganiiN/A100%90/90(96.0-100)100%90/90(96.0-100)--
Acinetobacter baumaniiOXA100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Citrobacter freundiiVIM100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Enterobacter cloacaeKPC100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Escherichia coliNDM100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Klebsiella pneumoniaeOXA.CTX-M100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Klebsiella oxytocaCTX-M100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Proteus mirabilisN/A100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)
Pseudomonas aeruginosaIMP100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)100%90/90(96.0-100)

{10}------------------------------------------------

Method Comparison

The combined results for the method comparison study conducted at thirteen (13) investigational sites are presented below. A total of 1412 specimens were analyzed, 604 of which were prospectively-collected fresh specimens, 272 of which were prospectivelycollected frozen specimens, 239 of which were selected frozen specimens, and 297 of which were simulated frozen specimens. Performance of the BC-GN test related to the detection of the target organisms was determined by comparing the BC-GN test result to results obtained using standard culture-based automated phenotypic bacterial identification reference methods. For the resistance markers, reference method testing included the combination of PCR amplification and bidirectional sequencing confirmation.

OrganismPercent Agreement
Positive(95% CI)Negative(95% CI)
Acinetobacter spp.98.2%55/56(90.5-100)99.9%1355/1356(99.6-100)
Citrobacter spp.100%49/49(92.8-100)99.9%1362/1363(99.6-100)
Enterobacter spp.97.6%120/123(93.0-99.5)99.4%1281/1289(98.8-99.7)
Proteus spp.100%58/58(93.8-100)99.9%1353/1354(99.6-100)
Escherichia coli99.8%517/518(98.9-100)99.4%889/894(98.7-99.8)
Klebsiella pneumoniae93.1%285/306(89.7-95.7)100%1106/1106(99.7-100)
Klebsiella oxytoca92.2%59/64(82.7-97.4)99.6%1342/1348(99.0-99.8)
Pseudomonas aeruginosa97.6%124/127(93.3-99.5)100%1285/1285(99.7-100)

Summary of Clinical Test Performance-Bacterial Results (n=1412)

{11}------------------------------------------------

Percent Agreement
Resistance Marker TargetPositive(95% CI)Negative(95% CI)
CTX-M98.7%151/153(95.4-99.8)99.9%1112/1113(99.5-100)
OXA95.3%61/64(86.9-99.0)99.9%1201/1202(99.5-100)
KPC100%51/51(93.1-100)100%1215/1215(99.7-100)
VIM100%41/41(91.4-100)100%1225/1225(99.7-100)
NDM100%41/41(91.4-100)100%1225/1225(99.7-100)
IMP.100%48/48(92.6-100)100%1218/1218(99.7-100)

Substantial Equivalence

As demonstrated by the information stated below, the Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN test) has been shown to be as safe and effective as the BioFire FilmArray Blood Culture Identification (BCID) Panel Kit (K130914) and the Nanosphere Verigene Gram-Positive Blood Culture (BC-GP) Nucleic Acid Test (K122514). The BC-GN test has similar intended use and indications, technological characteristics, and performance characteristics. The minor differences between the BC-GN test and its predicate device raises no new issues of safety or effectiveness. Clinical and analytical performance data demonstrate that the BC-GN test is as safe and effective as the predicate device. Thus. the BC-GN test is substantially equivalent.

{12}------------------------------------------------

DevicePredicate #1
ItemsVerigene BC-GN TestBioFire FilmArray Blood CultureIdentification Panel
510(k)#K132843K130914
Regulation866.3365Same
Product CodePENPEN
Device ClassIISame
Intended UseQualitative in vitro diagnostic test fordetection and identification ofmicroorganism nucleic acidsSame
Indications for UseTo aid in the diagnosis of bacterialbloodstream infectionsSame
Warnings andPrecautionsSub-culturing of positive blood cultures isnecessary to recover organisms forsusceptibility testing, identification oforganisms not detected by the test,differentiation of mixed growth, associationof antimicrobial resistance marker genes toa specific organism, or for epidemiologicaltyping.Same
Contraindication(s)Not to be used to monitor bloodstreaminfections/treatment for bacterial infection.Same
Organisms andResistance MarkersDetectedProteus spp.Escherichia coliKlebsiella pneumoniaeKlebsiella oxytocaPseudomonas aeruginosaKPCSame(see below for differences)
Sample Processing andPurificationOn-board, automated DNA extractionSame
Technology PrinciplesMultiplex nucleic acid-basedSame(see below for differences)
Interpretation ofResultsDiagnostic Software/Decision AlgorithmSame
Sensitivity(Positive % Agreement)Acinetobacter spp.: 98.2% (90.5-100%)Enterobacter spp.: 97.5% (93.0-99.5%)Escherichia coli: 99.6% (98.9-100%)Klebsiella oxytoca: 92.2% (82.7-97.4%)Klebsiella pneumoniae: 93.1% (89.7-95.7%)Proteus spp.: 100% (93.8-100%)KPC: 100% (93.1-100%)A. baumannii: 100% (93.0-100%)Enterobacteriaceae: 98.4% (96.9-99.3%)Escherichia coli: 98.0% (94.4-99.6%)K. oxytoca: 92.2% (82.7-97.4%)K. pneumoniae: 97.1% (91.9-99.4%)Proteus spp.: 100% (91.0-100%)KPC: 100% (91.0-100%)
Specificity(Negative % Agreement)Acinetobacter spp.: 100% (99.6-100%)Enterobacter spp.: 99.6% (98.8-99.7%)Escherichia coli: 99.4% (98.7-99.8%)Klebsiella oxytoca: 99.5% (99.0-99.8%)Klebsiella pneumoniae: 100% (99.7-100%)Proteus spp.: 100% (99.6-100%)KPC: 100% (99.7-100%)A. baumannii: 99.8% (99.5-99.9%)Enterobacteriaceae: 99.8% (99.4-99.9%)Escherichia coli: 99.8% (99.4-99.9%)K. oxytoca: 99.9% (99.7-100%)K. pneumoniae: 99.6% (99.2-99.8%)Proteus spp.: 100% (99.8-100%)KPC: 100% (99.2-100%)
End UsersOperators with no clinical lab experience toexperienced clinical laboratory technologists.Same
ItemsDevicePredicate #1
Verigene BC-GN TestBioFire FilmArray Blood CultureIdentification Panel
510(k)#K132843K130914
Organisms andResistance MarkersDetectedAcinetobacter spp.Citrobacter spp.Enterobacter spp.CTX-MNDMVIMIMPOXAAcinetobacter baumanniiEnterobacteriaceaeHaemophilus influenzaNeisseria meningitidis (encapsulated)Serratia marcescens
Test FormatDisposable single-use, multi-chamberedfluidic test cartridgeDisposable single-use, pouch containingfreeze-dried reagents
ControlsTwo Internal Processing Controls (wholeorganism complete assay control andsingle-stranded DNA Hybridizationcontrol)Two controls are included in each reagentpouch to control for sample processingand both stages of PCR and melt analysis.
DNA AmplificationmethodNoneNested, multiplex PCR with DNA meltanalysis.
Time to Result2 hoursLess than one hour
Detection MethodGold/Ag nanoparticle probe detection ofbacterial-specific DNA on complementaryoligo- microarrayAmplicon detection by melt curveanalysis of fluorescence signal from DNAbinding dye.
InstrumentationVerigene Reader and Processor SPFilmArray Instrument

Similarities between the Verigene BC-GN Test and Predicate Device #1

  • Data for predicate obtained from BioFire FilmArray Blood Culture Identification Panel (K130914) Decision Summary.

·

.

{13}------------------------------------------------

Differences between the Verigene BC-GN Test and Predicate Device #1

.

100 come the comments of

·

{14}------------------------------------------------

ItemsDevicePredicate #2
Verigene BC-GN TestVerigene BC-GP Test
510(k)K132843K122514
Regulation866.3365Same
Product CodePENPAM
Device ClassClass IISame
Intended UseQualitative multiplex in vitro diagnostic test for thedetection and identification of microorganism nucleicacidsSame
Indications for UseTo aid in the diagnosis of infectious diseasesSame
Time to Result~2.5 hrs~2.0 hrs
Test CartridgeDisposable single-use, multi-chambered fluidiccartridge.Same
Sample PrepAutomated onboard extraction of nucleic acidsperformed on the Processor SP using silica coatedmagnetic beads and chaotropic salts.Same
Quality controlInternal procedural/instrument quality controls;internal negative control, sample processing control;external positive and negative assay controlsSame
Detection MethodGold/Ag nanoparticle probe detection of bacterial-specific DNA on complementary oligo- microarraySame
Method Steps/ComplexitySingle pipetting step specimen, prior to NA extractionacid isolation, hybridization, and direct detectionSame
Interpretation ofResultsDiagnostic Software/ Decision AlgorithmSame
SoftwareCustom embedded software application running underembedded operating systemSame
ReaderOptical intensities are analyzed after target-specifichybridization of probesSame
End UsersOperators with no clinical lab experience toexperienced clinical laboratory technologistsSame

Similarities between the Verigene BC-GN Test and Predicate Device #2 ·

.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, clutching a staff with a snake wrapped around it.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

NANOSPHERE, INC MARK DEL VECCHIO 4088 COMMERCIAL AVENUE NORTHBROOK IL 60062

January 8, 2014

Re: K132843

Trade/Device Name: Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN) Regulation Number: 21 CFR 866.3365

Regulation Name: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures

Regulatory Class: II Product Code: PEN, NSU Dated: December 20, 2013 Received: December 27, 2013

Dear Mr. Del Vecchio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{16}------------------------------------------------

Page 2—Mr. Del Vecchio

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809). please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm:

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincercly yours.

Uwe Scherf -S for

Sally Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{17}------------------------------------------------

Indications for Use

510(k) Number (if known): K132843

Device Name: Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN)

Indications For Use:

The Verigene® Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-toresult Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood cutture system and which contain gram-negative bacteria as determined by gram stain.

BC-GN detects and identifies the following:

Bacterial Genera and SpeciesResistance Markers
Acinetobacter spp.CTX-M (blaCTX-M)
Citrobacter spp.KPC (blaKPC)
Enterobacter spp.NDM (blaNDM)
Proteus spp.VIM (blaVIM)
Escherichia coli1IMP (blaIMP)
Klebsiella pneumoniaeOXA (blaOXA)
Klebsiella oxytoca
Pseudomonas aeruginosa

BC-GN will not distinguish Escherichia coli from Shigella spp. (S. dysenteriae, S. flexner, S. boydii, and S. sonnel)

BC-GN is indicated for use in coniunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by BC-GN, to detect mixed infections that may not be detected by BC-GN, for association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.

Prescription Use X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Image /page/17/Picture/16 description: The image shows the name John Hobson -S, followed by the date and time 2014.01.07 10:37:53. The last line shows the time zone offset -05'00'. There is a logo in the background that is difficult to read.

Page 1 of

§ 866.3365 Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

(a)
Identification. A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.(b)
Classification. Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).