K Number
K130914

Validate with FDA (Live)

Date Cleared
2013-06-21

(80 days)

Product Code
Regulation Number
866.3365
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with the FilmArray Instrument. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrates the presence of organisms as determined by Gram stain.

The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, commonly encountered Staphylococci (including specific differentiation of Staphylococcus aureus), commonly encountered Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, commonly encountered Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis.

The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blaKPC) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.

FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia.

Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some Staphylococci, Enterococci, Streptococci, and Enterobacteriaceae that are not specifically identified by the FilmArray BCID Panel assays.

Device Description

The FilmArray Blood Culture Identification (BCID) Panel is a multiplex nucleic acid test designed to be used with the FilmArray Instrument. The FilmArray BCID pouch contains freeze-dried reagents to perform nucleic acid purification and nested, multiplex PCR with DNA melt analysis. The FilmArray Blood Culture Identification (BCID) Panel simultaneously tests a single positive blood culture sample to provide results for 24 different organisms and organism groups that cause bloodstream infections and three genetic markers that are known to confer antimicrobial resistance (see Table 1).

A test is initiated by loading Hydration Solution and a positive blood culture sample mixed with the provided Sample Buffer into the FilmArray BCID pouch. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The FilmArray Instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister. it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical lysis and standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is exceuted in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed PCR reaction which includes all primers of the outer primer sets. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green Plus-, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 200 stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2" stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism and antimicrobial resistance gene on the panel.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the FilmArray Blood Culture Identification (BCID) Panel, based on the provided text:

FilmArray Blood Culture Identification (BCID) Panel

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria in a dedicated section. However, the "Clinical Performance" and "Reproducibility" sections present performance metrics (sensitivity, specificity, positive percent agreement, negative percent agreement, and percent agreement with expected results) that act as the de facto acceptance criteria. I'll present the overall clinical performance from Tables 6, 7, 8, 9, and 10 and the overall reproducibility from Tables 59 and 60.

Overall Clinical Performance (Aggregated from Prospective and Seeded Arms)

Analyte GroupPerformance MetricValue95% Confidence Interval
Gram-Positive Organisms (Overall)
EnterococcusSensitivity/PPA97.7%93.4-99.5%
Specificity/NPA99.8%99.5-99.9%
Listeria monocytogenesSensitivity/PPA100%90.3-100%
Specificity/NPA100%99.8-100%
StaphylococcusSensitivity/PPA96.5%95.0-97.7%
Specificity/NPA99.1%98.5-99.6%
Staphylococcus aureusSensitivity/PPA98.4%96.1-99.6%
Specificity/NPA99.8%99.5-99.9%
StreptococcusSensitivity/PPA97.5%94.3-99.2%
Specificity/NPA99.8%99.4-99.9%
Streptococcus agalactiaeSensitivity/PPA100%90.3-100%
Specificity/NPA100%99.8-100%
Streptococcus pneumoniaeSensitivity/PPA97.3%85.8-99.9%
Specificity/NPA99.9%99.7-100%
Streptococcus pyogenesSensitivity/PPA100%90.7-100%
Specificity/NPA99.9%99.7-100%
Gram-Negative Organisms (Overall)
Acinetobacter baumanniiSensitivity/PPA100%93.0-100%
Specificity/NPA99.8%99.5-99.9%
EnterobacteriaceaeSensitivity/PPA98.4%96.9-99.3%
Specificity/NPA99.8%99.4-99.9%
Enterobacter cloacae complexSensitivity/PPA97.4%86.5-99.9%
Specificity/NPA99.9%99.6-100%
Escherichia coliSensitivity/PPA98%94.4-99.6%
Specificity/NPA99.8%99.4-99.9%
Klebsiella oxytocaSensitivity/PPA92.2%82.7-97.4%
Specificity/NPA99.9%99.7-100%
Klebsiella pneumoniaeSensitivity/PPA97.1%91.9-99.4%
Specificity/NPA99.6%99.2-99.8%
ProteusSensitivity/PPA100%91.0-100%
Specificity/NPA100%99.8-100%
Serratia marcescensSensitivity/PPA98.7%93.0-100%
Specificity/NPA99.9%99.7-100%
Haemophilus influenzaeSensitivity/PPA100%91.8-100%
Specificity/NPA100%99.8-100%
Neisseria meningitidisSensitivity/PPA100%90.3-100%
Specificity/NPA100%99.8-100%
Pseudomonas aeruginosaSensitivity/PPA98.1%89.7-100%
Specificity/NPA99.9%99.7-100%
Yeast (Overall)
Candida albicansSensitivity/PPA100%94.4-100%
Specificity/NPA99.8%99.5-99.9%
Candida glabrataSensitivity/PPA100%92.7-100%
Specificity/NPA99.9%99.7-100%
Candida kruseiSensitivity/PPA100%90.5-100%
Specificity/NPA100%99.8-100%
Candida parapsilosisSensitivity/PPA96.7%88.7-99.6%
Specificity/NPA99.9%99.7-100%
Candida tropicalisSensitivity/PPA100%91.0-100%
Specificity/NPA100%99.8-100%
Antimicrobial Resistance Genes (PCR/Sequencing Direct from Blood Culture)
mecA (All Staphylococcus)Sensitivity/PPA98.4%96.8-99.3%
Specificity/NPA98.3%96.0-99.4%
vanA/B (Enterococcus)Sensitivity/PPA100%94.4-100%
Specificity/NPA100%94.6-100%
KPC (Enterobacteriaceae & others)Sensitivity/PPA100%91.0-100%
Specificity/NPA100%99.3-100%
Antimicrobial Resistance Genes (PCR/Sequencing of Cultured Isolates)
mecA (All Staphylococcus)Positive Percent Agreement98.9%97.5-99.6%
Negative Percent Agreement87.9%83.9-91.3%
vanA/B (Enterococcus)Positive Percent Agreement100%94.0-100%
Negative Percent Agreement94.4%86.2-98.4%
KPC (Enterobacteriaceae & others)Positive Percent Agreement100%91.0-100%
Negative Percent Agreement100%99.3-100%

Overall Reproducibility (All Sites, Across All Analytes Tested)

Performance MetricValue95% Confidence Interval
Organism Assays (% Agreement)~100%(Min: 98.0%, Max: 100%)
Antimicrobial Resistance Assays (% Agreement)~100%(Min: 98.0%, Max: 100%)

2. Sample Sizes and Data Provenance for the Test Set:

  • Total Test Set Sample Size: 2207 blood cultures
    • Prospective Arm: 1568 specimens
      • 67 specimens excluded (e.g., >8 hours past positivity, incomplete reference data, previous specimen from same subject).
      • Tested either fresh (821 specimens) or frozen (747 specimens).
    • Seeded Arm: 639 specimens
      • 77 cultures excluded (e.g., >8 hours past positivity, not positive by automated blood culture, contaminated).
      • Tested either fresh (419 specimens) or frozen (220 specimens).
  • Data Provenance: The study was a "two-armed clinical study" conducted at eight U.S. clinical sites. This indicates the data is prospective for the clinical arm and retrospective/contrived for the seeded arm (as isolates were seeded into blood culture bottles).

3. Number of Experts and Qualifications for Ground Truth - Test Set:

The document does not explicitly state the number of experts used to establish the ground truth for the test set, nor their specific qualifications (e.g., "radiologist with 10 years of experience").

Instead, the ground truth for organism detection was established using "Standard manual and automated microbiological/biochemical identification methods." For Acinetobacter baumannii, this was supplemented with "16S PCR with bi-directional sequencing."

For antimicrobial resistance genes, the ground truth involved "PCR with bi-directional sequencing for specific resistance gene - direct from blood culture" or "PCR with bi-directional sequencing for specific resistance gene from - appropriate cultured isolates." Additionally, "Standard manual and automated phenotypic antimicrobial susceptibility testing of appropriate cultured isolates" was performed for informational purposes.

While these are standard laboratory methods typically performed by trained microbiologists or laboratory technicians, the number and specific qualifications of the experts overseeing or performing these reference methods are not detailed.

4. Adjudication Method for the Test Set:

The document does not explicitly describe an adjudication method (like 2+1 or 3+1) for discrepant results in the test set.

However, the detailed tables of results often include footnotes explaining discrepancies. For instance, for Acinetobacter baumannii, isolates misidentified by phenotypic methods were subjected to 16S PCR and bi-directional sequencing for definitive characterization. Similarly, for other organisms and resistance genes, bidirectional sequencing was used to investigate false positive and false negative results. This implies an investigative process to reconcile differences, rather than a formal, predefined expert adjudication panel for every case.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was NOT done for this device. The study evaluated the standalone performance of the FilmArray BCID Panel against reference laboratory methods, not its impact on human reader performance or diagnostic accuracy with AI assistance.

6. Standalone Performance Study:

Yes, a standalone study was performed. The entire "Clinical Performance" section (Tables 6-10) reports the performance of the FilmArray BCID Panel as a standalone diagnostic device. The results (sensitivity, specificity, PPA, NPA) are presented for the algorithm/device only, without human-in-the-loop interaction, compared against established laboratory reference methods. The "Reproducibility" section also evaluates the device's technical consistency in a standalone capacity.

7. Type of Ground Truth Used:

The ground truth primarily used was:

  • Expert Consensus/Reference Laboratory Methods:
    • Standard manual and automated microbiological/biochemical identification methods for all organism detections.
    • 16S PCR with bi-directional sequencing for A. baumannii speciation.
    • PCR with bi-directional sequencing for antimicrobial resistance gene detection (from blood culture or cultured isolates).
  • Pathology: Not explicitly mentioned as a primary ground truth for organism identification in this context, which focuses on molecular/microbial identification.
  • Outcomes Data: Not used as ground truth for diagnostic accuracy in this study.

8. Sample Size for the Training Set:

The document does not explicitly specify a sample size for a "training set." The studies primarily focus on clinical validation (prospective and seeded arms) and analytical validation (inclusivity, exclusivity, reproducibility). While the device's algorithm undoubtedly relies on an underlying model, information regarding its specific training data and size is not provided in this regulatory submission summary. The inclusivity studies involve testing a "diverse collection of 303 isolates," which could be considered part of the data informing the assay's design and optimization, but it's not explicitly labeled as a "training set" in the context of a machine learning model.

9. How the Ground Truth for the Training Set Was Established:

As the document does not explicitly identify a "training set" in the context of a machine learning model with corresponding ground truth establishment, this information is not provided. The inclusivity and exclusivity studies, which evaluate the assay's breadth and specificity, likely contribute to the "training" or development of the assay's detection capabilities. The ground truth for these analytical studies was established by known characteristics of the isolates (e.g., ATCC strains, clinical isolates with confirmed identity), and confirmed through standard microbiological identification and sequencing methods.

{0}------------------------------------------------

jC130914

510(k) Summary BioFire Diagnostics, Inc.

JUN 2 1 2013

FilmArray Blood Culture Identification (BCID) Panel Kit

Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

BioFire Diagnostics, Inc. 390 Wakara Way Salt Lake City, UT 84108

Telephone: 801-736-6354 Facsimile: 801-588-0507

Contact: Beth Lingenfelter, ext. 407

ﻟﻠﺴ

Date Submitted: March 30, 2013

Device Name and Classification:

Trade Name: FilmArray BCID Panel

Regulation Number: 21 CFR 866.3365

Classification Name: Multiplex devices that use DNA hybridization to detect bacteria and their resistance markers

Predicate Device:

K122514 - Nanosphere Verigene® Gram-Positive Blood Culture Nucleic Acid Test (BC-BG)

Intended Use:

The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with the FilmArray Instrument. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrates the presence of organisms as determined by Gram stain.

The following gram-positive bacteria, gram-negative bacteria, and veast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, commonly

{1}------------------------------------------------

encountered Staphvlococci (including specific differentiation of Staphylococcus aureus). commonly encountered Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, commonly encountered Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex. Escherichia coli, Klebsiella oxytoca. Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis.

The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blaker) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.

FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia.

Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some Staphylococci, Enterococci, Streptococci, and Enterobacteriaceae that are not specifically identified by the FilmArray BCID Panel assays.

Device Description:

The FilmArray Blood Culture Identification (BCID) Panel is a multiplex nucleic acid test designed to be used with the FilmArray Instrument. The FilmArray BCID pouch contains freeze-dried reagents to perform nucleic acid purification and nested, multiplex PCR with DNA melt analysis. The FilmArray Blood Culture Identification (BCID) Panel simultaneously tests a single positive blood culture sample to provide results for 24 different organisms and organism groups that cause bloodstream infections and three genetic markers that are known to confer antimicrobial resistance (see Table 1).

Gram-Positive BacteriaGram-Negative BacteriaYeast
EnterococcusAcinetobacter baumanniiCandida albicans
Listeria monocytogenesEnterobacteriaceaeCandida glabrata
StaphylococcusEnterobacter cloacae complexCandida krusei
Staphylococcus aureusEscherichia coliCandida parapsilosis
StreptococcusKlebsiella oxytocaCandida tropicalis
Streptococcus agalactiaeKlebsiella pneumoniaeAntimicrobial resistance genes
Streptococcus pneumoniaeProteusmecA – methicillin resistance

Table 1, FilmArray RCID Panel Test Results

{2}------------------------------------------------

Gram-Positive BacteriaGram-Negative BacteriaYeast
Streptococcus pyogenesSerratia marcescensvanA/B - vancomycin resistance
Haemophilus influenzaeblaKPC - carbapenem resistance
Neisseria meningitidis (encapsulated)
Pseudomonas aeruginosa

A test is initiated by loading Hydration Solution and a positive blood culture sample mixed with the provided Sample Buffer into the FilmArray BCID pouch. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The FilmArray Instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister. it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical lysis and standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is exceuted in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed PCR reaction which includes all primers of the outer primer sets. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green Plus-, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 200 stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2" stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism and antimicrobial resistance gene on the panel.

Substantial Equivalence:

.

The Nanosphere Verigene® Gram-Positive Blood Culture Nucleic Acid Test is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause

{3}------------------------------------------------

bloodstream infection. Table 2 outlines the similarities between the two systems and Table 3 outlines the differences.

ElementFilmArray BCID PanelNanosphere Verigene® Gram-PositiveBlood Culture Nucleic Acid Test
OrganismsDetectedEnterococci, Staphylococci (includingspecific differentiation of Staphylococcus aureus ), Streptococci (with specificdifferentiation of Streptococcus agalactiae, Streptococcus pneumoniae ,and Streptococcus pyogenes ) andresistance markers mecA, vanA , and vanB .SameSee below for differences
AnalyteDNASame
TechnologicalPrinciplesMultiplex nucleic acidSameSee below for differences
SampleProcessing andPurificationAutomated by instrumentSame
ControlsTwo controls are included in each reagentpouch to control for sample processing andboth stages of PCR and melt analysis.Internal procedural/instrument qualitycontrols; Internal Negative Control,Sample processing control, externalpositive and negative assay controls.
UserComplexityModerate/LowSame
Table 2. Similarities Between the FilmArray BCID Panel and the Nanosphere Verigene® Gram-
Positive Blood Culture Nucleic Acid Test.
Table 3. Differences Between FilmArray Respiratory Panel Test System and the Nanosphere
Verigene® Gram-Positive Blood Culture Nucleic Acid Test.
ElementFilmArray BCID PanelNanosphere Verigene® Gram-PositiveBlood Culture Nucleic Acid Test
SpecimenTypesPositive blood culture samples containinggram-positive, gram-negative bacteria,and/or yeast.Positive blood culture bottles(unspecified) which contain gram-positive bacteria.
OrganismsDetectedDetection of additional targets: Listeriamonocytogenes, Acinetobacterbaumannii, Enterobacteriaceae(including specific differentiation ofEnterobacter cloacae complex species,Escherichia coli, Klebsiella oxytoca,Klebsiella pneumoniae, Proteus, andSerratia marcescens), Haemophilusinfluenzae, Neisseria meningitidis,Pseudomonas aeruginosa, Candidaalbicans, Candida glabrata, Candidakrusei, Candida parapsilosis, Candidatropicalis, and resistance marker blaKPCTests only for gram positive bacteria.Tests for Listeria spp. rather thanListeria monocytogenes. Includes testingfor additional Staphylococcus spp.:Staphylococcus epidermidis,Staphylococcus lugdunensis, as well astesting for specific Enterococcus spp.:Enterococcus faecalis,Enterococcus faecium. Includes testingfor an additional Streptococcus spp.:Streptococcus anginosus group. Doesnot include testing for blaKPC

{4}------------------------------------------------

ElementFilmArray BCID PanelNanosphere Verigene® Gram-PositiveBlood Culture Nucleic Acid Test
TechnologicalPrinciplesNested multiplex PCR followed by highresolution melting analysis to confirmidentity of amplified product.Qualitative, multiplexed test for thedetection of specific nucleic acid targetsin a microarray format using capture andmediator oligonucleotides for goldnanoparticle probe-based endpointdetection.
InstrumentationFilmArray InstrumentVerigene Reader and Processor SP
Time to resultLess than 1 hour2.5 hours
TestInterpretationAutomated test interpretation and reportgeneration. User cannot access raw data.Diagnostic Software/DecisionAlgorithm.

Summary of Performance Data

Clinical Performance

The clinical performance of the FilmArray BCID Panel was established during a two armed clinical study which was conducted at eight U.S. clinical sites over an eight month time period. The study included a prospective residual blood culture arm and a seeded blood culture arm. In the prospective arm, 1635 prospectively-collected residual blood culture samples (pediatric and adult) were initially included in the study. Sixty-seven (67) specimens were excluded from the study. The most common reasons for exclusion were that the specimens were >8 hours past positivity, incomplete reference/comparator data were provided, or the specimen was from a subject who had a previous specimen included in the study. In the seeded culture arm, analytes proven to be of low prevalence in the prospective arm were evaluated by seeding previously characterized isolates into blood culture bottles and incubating until positivity. A total of 716 seeded cultures were initiated for the study. Seventy-seven (77) cultures were excluded from the study. The most common reasons for exclusion were that the specimens were >8 hours past positivity, the seeded culture was not called positive by the automated blood culture system. or the culture was contaminated or inconsistent with the intended seed organism. The final specimen set consisted of 2207 blood cultures (1568 prospective and 639 seeded). All cultures were grown in Becton Dickinson BACTECTM Plus Aerobic/F Medium. Table 4 provides a summary of demographic information for the 1568 specimens included in the prospective arm of the study.

Table 4. Demographic Summary for Prospective Arm of FilmArray BCID Clinical Evaluation
--------------------------------------------------------------------------------------------
Prospective Study Specimens
Total Specimens1568
SexNumber of Specimens
Male917 (58%)
Female651 (42%)
Age GroupNumber of Specimens
≤ 1 year57 (4%)

{5}------------------------------------------------

Prospective Study Specimens
Total Specimens1568
1 - 17 years92 (6%)
18 - 44 years281 (18%)
45 - 64 years583 (37%)
65 - 84 years442 (28%)
≥ 85 years113 (7%)

Positive blood cultures (prospective and seeded) were tested with the FilmArray BCID Panel. The performance of FilmArray BCID was evaluated by comparing the FilmArray BCID test result for each panel member with the appropriate comparator/reference methods shown in Table 5.

Table 5. Reference/Comparator Methods used to Assess FilmArray BCID Performance
---------------------------------------------------------------------------------------
Test ResultReference/Comparator Method(s)
All organism detectionsexcept Acinetobacter baumanniiStandard manual and automated microbiological/biochemical identificationmethods a
Acinetobacter baumannii detectionStandard manual and automated microbiological/biochemical identificationmethodsPlus16S PCR with bi-directional sequencing of all A. calcoaceticus-baumanniicomplex isolates for characterization as A. baumannii or non-A. baumannii a
Antimicrobial resistance gene detectionsin specimens in which an associated organism was detected( mecA from Staphylococcus ; vanA/B from Enterococcus ,KPC from Enterobacteriaceae , Acinetobacter baumannii ,and Pseudomonas aeruginosa )Method 1: PCR with bi-directional sequencing for specific resistance gene- direct from blood culture bMethod 2: PCR with bi-directional sequencing for specific resistance gene from- appropriate cultured isolates bInformational: Standard manual and automated phenotypic antimicrobialsusceptibility testing of appropriate cultured isolates (methicillin resistance,vancomycin resistance, and carbapenem resistance (and/or carbapenemaseproduction) according to current CLSI criteria) c

" Performance of FilmArray BCID develing all organisms was compared to standard manual and automated microbiological/biochemical identification methods. Additionally isolatified as being members of the A. calcoaceticus-baunamii vomplex were subjected to 16S PCR and bi-directional sequencing to categorize the is being A. baunamii for final comparison to the FilmArray BCID A. baunamii-specific results required a sequencing result of adequate quality to match sequences of A. baunamii (or negative result if sequences match non-A. baunannii organisms) deposited in the National Center for Biotechnology Information (NCB) GenBank database (www.nebi.nlm.nlh.gov), with an acceptable E-value. This was required due to the inability of phenotypic identification nethous to adequately discriminate between members of the A. calcoaceticus-baumannii complex.

" Performance of FilmArray BCID detecting antimicrobial resistance genes (nec.4, van KPC) was compared to gene-specific PCR tests with bi-directional sequencing. The assays were designed to amplify different sequences than those targeted by FilmArray BCID. Positive results required a sequencing result of adequate of the expect gene deposited in the National Center for Biotechnology Information (NCBI) GenBank database (www.ncbi.nlm.nih.goy), with an acceptable E-value,

6 Performance of FilmArray BCID as compared to phenotypic antimicrobial susceptibility testing was performed for informational purposes. The phenotypic methods were performed in accordance with current CLSI criteria.

A total of 2207 blood culture specimens (1568 prospective and 639 seeded) were evaluated in the FilmArray BCID clinical evaluation. Specimens were tested by FilmArray BCID either fresh or from frozen aliquots. A total of 1240 specimens were tested fresh (821 prospective and 419 seeded) and 967 specimens were tested frozen (747 prospective and 220 seeded). Clinical sensitivity or positive percent agreement (PPA) was calculated as 100% x (TP/TP + FN). True positive (TP) indicates that both FilmArray BCID and the reference/comparator method had a positive result for a specific

{6}------------------------------------------------

analyte, and false negative (FN) indicates that the FilmArray BCID result was negative while the reference/comparator method was positive. Clinical specificity or negative percent agreement (NPA) was calculated as 100% x (TN/TN + FP). True negative (TN) indicates that both FilmArray BCID and the reference/comparator method had a negative result for a specific analyte, and false positive (FP) indicates that the FilmArray BCID result was positive while the reference/comparator method was negative. The exact binomial two-sided 95% confidence interval was calculated. The results are summarized in Tables 6 - 10.

Gram-Positive BacteriaSensitivity/PPA*Specificity/NPA*
TP/TP + FN%95% CITN/TN + FP%95% CI
EnterococcusProspective Fresh55/5510093.5-100762/76699.598.7-99.9
Prospective Frozen43/4693.582.1-98.6701/70110099.5-100
Seeded Fresh12/1210073.5-100407/40710099.1-100
Seeded Frozen17/1710080.5-100203/20310098.2-100
Overall127/13097.793.4-99.52073/207799.899.5-99.9
ListeriamonocytogenesProspective Fresh0/0--821/82110099.6-100
Prospective Frozen0/0--747/74710099.5-100
Seeded Fresh23/2310085.2-100396/39610099.1-100
Seeded Frozen13/1310075.3-100207/20710098.2-100
Overall36/3610090.3-1002171/217110099.8-100
StaphylococcusProspective Fresh405/41896.994.7-98.3401/40399.598.2-99.9
Prospective Frozen364/37996.093.6-97.8359/36897.695.4-98.9
Seeded Fresh0/0--418/41999.898.7-100
Seeded Frozen1/11002.5-100219/21910098.3-100
Overall770/79896.595.0-97.71397/140999.198.5-99.6
StaphylococcusaureusProspective Fresh133/13697.893.7-99.5685/68510099.5-100
Prospective Frozen120/12199.295.5-100622/62699.498.4-99.8
Seeded Fresh0/0--419/41910099.1-100
Seeded Frozen0/0--220/22010098.3-100
Overall253/25798.496.1-99.61946/195099.899.5-99.9
StreptococcusProspective Fresh73/7794.887.2-98.6740/74499.598.6-99.9
Prospective Frozen63/6498.491.6-100683/68310099.5-100
Seeded Fresh18/1810081.5-100401/40110099.1-100
Seeded Frozen44/4410092.0-100175/17699.496.9-100
Overall198/20397.594.3-99.21999/200499.899.4-99.9
Streptococcusagalactiae(Group B)Prospective Fresh8/810063.1-100813/81310099.5-100
Prospective Frozen10/1010069.2-100737/73710099.5-100
Seeded Fresh3/310029.2-100416/41610099.1-100
Seeded Frozen15/1510078.2-100205/20510098.2-100
Overall36/3610090.3-1002171/217110099.8-100
StreptococcuspneumoniaeProspective Fresh15/1510078.2-100805/80699.999.3-100
Prospective Frozen10/1010069.2-100737/73710099.5-100
Seeded Fresh4/580.028.4-99.5413/41499.898.7-100
Seeded Frozen7/710059.0-100213/21310098.3-100
Overall36/3797.385.8-99.92168/217099.999.7-100
Streptococcuspyogenes(Group A)Prospective Fresh5/510047.8-100815/81699.999.3-100
Prospective Frozen2/210015.8-100745/74510099.5-100
Seeded Fresh9/910066.4-100410/41010099.1-100
Seeded Frozen22/2210084.6-100198/19810098.2-100
Overall38/3810090.7-1002168/216999.999.7-100
Table 6. FilmArray BCID Clinical Performance Summary - Gram-Positive Organism Results
(Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification

ª Sensitivity and Specificity celer to performate with the prospective speciment (PPA) and Negative Percent Agreement (NPA) refer to performance with the seeded specimens.

b 3/4 false positive Enterococcus specimens contained Staphylococcus; the false positive results may be due to cross-reactivity.

5 Isolaes from 16/28 false negalive Staphylococus species species S, peties S, petterkofer by bi-directional sequencing. Bidirectional sequencing confirmed the presence of Staphylococcus in 10/12 false positive specimens; 2 were S. epidermidis, and 1 was S. haemolyticus.

4 Bidirectional sequencing identified 2 isolates from S. aureus false negative specimens as S. epidernidis; they were not S. aweus. Bidirectional sequencing confirmed the presence of S. aurens in 1/4 false positive and one false negative S. arreas were in sequentially-tested specimens and may be due to sample mix-up.

& Bidirectional sequencing confirmed the presence of S. mitis in 1/5 false positive Streptococcus specimens

{7}------------------------------------------------

Table 7. FilmArray BCID Clinical Performance Summary - Gram-Negative Organism Results (Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification plus 16S Sequencing for Speciation for A. baumannii)

Gram-Negative BacteriaSensitivity/PPA1Specificity/NPA1
TP/TP + FN%95% CITP/TP + FP%95% CI
AcinetobacterbaumanniiProspective Fresh7/710059.0-100813/81499.999.3-100
Prospective Frozen7/710059.0-100739/74099.999.2-100
Seeded Fresh20/2010083.2-100397/39999.598.2-99.9
Seeded Frozen17/1710080.5-100202/20399.597.3-100
Overall51/5110093.0-1002151/215699.899.5-99.9
EnterobacteriaceaeProspective Fresh153/15698.194.5-99.6665/66510099.4-100
Prospective Frozen150/15497.493.5-99.3589/59399.398.3-99.8
Seeded Fresh93/9310096.1-100326/32610098.9-100
Seeded Frozen94/9598.994.3-100125/12510097.1-100
Overall490/498c98.496.9-99.31705/1709c99.899.4-99.9
Enterobacter cloacaecomplexProspective Fresh10/1190.958.7-99.8809/81099.999.3-100
Prospective Frozen11/1110071.5-100734/73699.799.0-100
Seeded Fresh8/810063.1-100411/41110099.1-100
Seeded Frozen9/910066.4-100211/21110098.3-100
Overall38/3997.486.5-99.92165/216899.999.6-100
Escherichia coliProspective Fresh77/7997.591.2-99.7742/74210099.5-100
Prospective Frozen68/6998.692.2-100674/67899.498.5-99.8
Seeded Fresh4/410039.8-100414/41599.898.7-100
Seeded Frozen1/11002.5-100219/21910098.3-100
Overall150/153d9894.4-99.62049/2054d99.899.4-99.9
Klebsiella oxytocaProspective Fresh4/410039.8-100817/81710099.5-100
Prospective Frozen1/2501.3-98.7744/74599.999.3-100
Seeded Fresh32/3688.973.9-96.9383/38310099.0-100
Seeded Frozen22/2210084.6-100198/19810098.2-100
Overall59/64e92.282.7-97.42142/214399.999.7-100
KlebsiellapneumoniaeProspective Fresh33/3497.184.7-99.9786/78799.999.3-100
Prospective Frozen35/3794.681.8-99.3705/71099.398.4-99.8
Seeded Fresh13/1310075.3-100403/40699.397.9-99.8
Seeded Frozen21/2110083.9-100199/19910098.2-100
Overall102/105f97.1≤91.9-99.42093/2102f99.699.2-99.8
ProteusProspective Fresh11/1110071.5-100810/81010099.5-100
Prospective Frozen11/1110071.5-100736/73610099.5-100
Seeded Fresh2/210015.8-100417/41710099.1-100
Seeded Frozen15/1510078.2-100205/20510098.2-100
Overall39/3910091.0-1002168/216810099.8-100
Serratia marcescensProspective Fresh14/1410076.8-100807/80710099.5-100
Prospective Frozen8/810063.1-100739/73910099.5-100
Seeded Fresh28/2810087.7-100390/39199.798.6-100
Seeded Frozen26/2796.381.0-99.9193/19310098.1-100
Overall76/77g98.793.0-1002129/2130g99.999.7-100
HaemophilusinfluenzaeProspective Fresh5/510047.8-100816/81610099.5-100
Prospective Frozen3/310029.2-100744/74410099.5-100
Seeded Fresh29/2910088.1-100390/39010099.1-100
Seeded Frozen6/610054.1-100214/21410098.3-100
Overall43/4310091.8-1002164/216410099.8-100
NeisseriameningitidisProspective Fresh1/11002.5-100820/82010099.6-100
Prospective Frozen0/0-747/74710099.5-100
Seeded Fresh30/3010088.4-100389/38910099.1-100
Seeded Frozen5/510047.8-100215/21510098.3-100
Overall36/3610090.3-1002171/217110099.8-100
PseudomonasaeruginosaProspective Fresh19/1910082.4-100802/80210099.5-100
Prospective Frozen32/339784.2-99.9713/71499.999.2-100
Seeded Fresh0/0-419/41910099.1-100
Seeded Frozen0/0-220/22010098.3-100
Overall51/52h98.189.7-1002154/215599.999.7-100
YeastSensitivity/PPA*Specificity/NPA*
TP/TP + FN%95% CITN/TN + FP%95% CI
Candida albicansProspective Fresh12/1210073.5-100808/80999.999.3-100
Prospective Frozen4/410039.8-100740/74399.698.8-99.9
Seeded Fresh47/4710092.5-100372/37210099.0-100
Seeded Frozen1/11002.5-100219/21910098.3-100
Overall64/6410094.4-1002139/214399.899.5-99.9
Candida glabrataProspective Fresh6/610054.1-100813/81599.899.1-100
Prospective Frozen6/610054.1-100741/74110099.5-100
Seeded Fresh32/3210089.1-100387/38710099.1-100
Seeded Frozen5/510047.8-100215/21510098.3-100
Overall49/4910092.7-1002156/215899.999.7-100
Candida kruseiProspective Fresh2/210015.8-100819/81910099.6-100
Prospective Frozen2/210015.8-100745/74510099.5-100
Seeded Fresh28/2810087.7-100391/39110099.1-100
Seeded Frozen5/510047.8-100215/21510098.3-100
Overall37/3710090.5-1002170/217010099.8-100
Candida parapsilosisProspective Fresh3/310029.2-100818/81810099.6-100
Prospective Frozen4/410039.8-100742/74399.999.3-100
Seeded Fresh47/4995.986.0-99.5370/37010099.0-100
Seeded Frozen5/510047.8-100214/21599.597.4-100
Overall59/6196.788.7-99.62144/214699.999.7-100
Candida tropicalisProspective Fresh0/0-821/82110099.6-100
Prospective Frozen3/310029.2-100744/74410099.5-100
Seeded Fresh31/3110088.8-100388/38810099.1-100
Seeded Frozen5/510047.8-100215/21510098.3-100
Overall39/3910091.0-1002168/216810099.8-100

4 Sensitivity and Specificity refer to performance with the prospective speciment (PPA) and Negative Percent Agreement (NPA) refer to performance with the seeded specimens

b Bidirectional sequencing identified isolates from 4 false positive species 3); this species appears to cross-rece with the A. baunanii assay. These four ised identified as A. baunamii by phenotypic methods. 6 other isolates originally identified as

{8}------------------------------------------------

A. baunamii by phenotypic methods were identified by bidirectional sequencials (genomospecies 13; 4 isolates), A bereziniae, and A. radioresistens; these 6 isolates did not cross-react with the A. baumannii assay.

5 One false positive and one false negatively in sequentially-tested specimens and may be due to sample mix-up. One isslate from another false negative speciment as E. coll by phenotypic methods, was identified as Pasteurella, and not E. coli, by bidirectional sequencing.

4 One false positive and one false negative E. coli were in sequentially-tested specimens and may be due to sample mix-up.

  • Bidirectional sequencing identified 4/5 isolates from false negative K. axyoca species, Rapollella ornithinolvica, and not K. oxyloca. The misidentification is a known limitation of phenotypic testing methods for this species.

The isolate from one false negative K. preunen was identified as the closely related organism, Routella and not K. preunoniae, 69 false positive K. pneumoniae results appear to be due to cross-reactivity with Entercobacter and Roulella ornithinolytica (misidentified as K. oxytoca by phenotypic methods).

8 Bidirectional sequencing identified the isolative S. marcescens specimen as being in the S. proteonucallans(grimesii group and not S. marcescens. The one false S. marcescens result uppears to be due to cross-reactivity with Raoulella ornithinolyica (misidentified as K. oxytoca by phenotypic methods).

" Bidirectional sequencing identified the isolate P. aeraginosa specimen as the closely related species Bendomonas stutzeri and not P. aeruginosu.

Table 8. FilmArray BCID Clinical Performance Summary - Yeast Organism Results
(Comparator Method: Standard Manual/Automoted Microbiological/Rioghemical Identification)

" Scusitivity and Specificity refer to performance with the prospective speciment (PPA) and Negative Percent Agreement (NPA) refer to performance with the seeded specimens

b Bidirectional sequencing identified the isolative C. parapsilosis specimens as being the closely related species C. metapsilosis. This misidentification is a known limitation of phenotypic identification methods,

As the antimicrobial resistance gene results are not reported in the absence of a presumptively associated organism, performance was calculated only for samples in which FilmArray BCID detected an appropriate organism. Performance was calculated separately against the two comparator methods; PCR/sequencing direct from the blood culture specimens and PCR/sequencing from organisms isolated from the blood cultures. When comparing to PCR/sequencing from bacterial isolates, performance was only calculated for specimens in which FilmArray BCID detected an appropriate organism and from which an appropriate organism isolate was obtained (i.e., antimicrobial resistance gene results could be obtained for both methods). The NPA for mecA and vanA/B are lower when comparing to PCR/sequencing from bacterial isolates than to

{9}------------------------------------------------

PCR/sequencing direct from blood culture primarily due to the reference methods not isolating a resistant clone of an applicable organism. This may be due to heterogeneous resistance within a population of cultured organisms or co-culturing of multiple indistinguishable applicable organisms with different resistance profiles (e.g., culturing a resistant Staphylococcus along with a sensitive Staphylococcus).

Sensitivity /PPASpecificity /NPA
Antimicrobial Resistance GenesTP/TP + FN%95% CITN/TN + FP%95% CI
mecA - Methicillin Resistance Gene
mecAAll StaphylococcusDetectedProspective Fresh253/25798.4%96.1-99.6%147/15098.0%94.3-99.6%
Prospective Frozen233/23798.3%95.7-99.5%134/13698.5%94.8-99.8%
Seeded Fresh1/1100%n/a0/0
Seeded Frozen1/1100%n/a0/0
Overall488/49698.4%96.8-99.3%281/28698.3%96.0-99.4%
mecAStaphylococcusDetected;S. aureus DetectedProspective Fresh67/6997.1%89.9-99.6%64/64100%94.4-100%
Prospective Frozen70/70100%94.9-100%54/54100%93.4-100%
Seeded Fresh0/00/0
Seeded Frozen0/00/0
Overall137/13998.6%94.9-99.8%118/118100%96.9-100%
mecAStaphylococcusDetected;S. aureusNot DetectedProspective Fresh186/18898.9%96.2-99.9%83/8696.5%90.1-99.3%
Prospective Frozen163/16797.6%94.0-99.3%80/8297.6%91.5-99.7%
Seeded Fresh1/1100%n/a0/0
Seeded Frozen1/1100%n/a0/0
Overall351/35798.3%96.4-99.4%163/16897.0%93.2-99.0%
vanA/B - Vancomycin Resistance Genes
vanA/BEnterococcusDetectedProspective Fresh23/23100%85.2-100%36/36100%90.3-100%
Prospective Frozen13/13100%75.3-100%30/30100%88.4-100%
Seeded Fresh12/12100%73.5-100%0/0
Seeded Frozen16/16100%79.4-100%1/1100%n/a
Overall64/64100%94.4-100%67/67100%94.6-100%
KPC - Carbapenem Resistance Gene (Carbapenemase)
KPCEnterobacteriaceaeand/orA. baumannii and/orP. aeruginosaDetectedProspective Fresh3/3100%29.2-100%177/177100%97.9-100%
Prospective Frozen3/3100%29.2-100%187/187100%98.0-100%
Seeded Fresh10/10100%69.2-100%105/105100%96.5-100%
Seeded Frozen23/23100%85.2-100%89/89100%95.9-100%
Overall39/39100%91.0-100%558/558100%99.3-100%
KPCEnterobacteriaceaeDetectedProspective Fresh3/3100%29.2-100%150/150100%97.6-100%
Prospective Frozen3/3100%29.2-100%151/151100%97.6-100%
Seeded Fresh10/10100%69.2-100%83/83100%95.7-100%
Seeded Frozen23/23100%85.2-100%71/71100%94.9-100%
Overall39/39100%91.0-100%455/455100%99.2-100%
KPCEnterobacteriaceaeNot Detected;A. baumannii and/orP. aeruginosaProspective Fresh0/027/27100%87.4-100%
Prospective Frozen0/036/36100%90.3-100%
Seeded Fresh0/022/22100%84.6-100%
Seeded Frozen0/018/18100%81.5-100%
Overall0/0103/103100%96.5-100%

Table 9. FilmArray BCID Clinical Performance Summary -- Antimicrobial Resistance Genes (Comparator Method: PCR/Sequencing Direct from Blood Culture).

4 Sensitivity and Specificity refer to performance with the prospective speciment (PPA) and Negative Percent Agreement (NPA) refer to performance with the seeded specimens.

Table 10. FilmArray BCID Clinical Performance Summary - Antimicrobial Resistance Genes (Comparator Method: PCR/Sequencing of Cultured Isolates)

ANTIMICROBIAL RESISTANCE GENESPositive Percent AgreementNegative Percent Agreement *
TP/TP + FN%95% CITN/TN + FP%95% CI
mecA - Methicillin Resistance Gene
mecAAll StaphylococcusDetectedProspective Fresh234/23699.2%97.0-99.9%149/17286.7%80.6-91.3%
Prospective Frozen219/22298.6%96.1-99.7%135/15189.4%83.4-93.8%
Seeded Fresh0/0--0/0--
Seeded Frozen1/1100%n/a0/0--

{10}------------------------------------------------

ANTIMICROBIAL RESISTANCE GENESPositive Percent AgreementNegative Percent Agreement *
TP/TP + FN%95% CITN/TN + FP%95% CI
mecAStaphylococcusDetected;S. aureus DetectedOverall454/45998.9%97.5-99.6%284/32387.9%83.9-91.3%
Prospective Fresh64/6598.5%91.7-100%65/6895.6%87.6-99.1%
Prospective Frozen66/66100%94.6-100%54/5893.1%83.3-98.1%
Seeded Fresh0/0--0/0--
Seeded Frozen0/0--0/0--
mecAStaphylococcusDetected;S. aureusNot DetectedOverall130/13199.2%95.8-100%119/12694.4%88.9-97.7%
Prospective Fresh170/17199.4%96.8-100%84/10480.8%71.9-87.8%
Prospective Frozen153/15698.1%94.5-99.6%81/9387.1%78.6-93.2%
Seeded Fresh0/0--0/0--
Seeded Frozen1/1100%n/a0/0--
Overall324/32898.8%96.9-99.7%165/19783.8%77.9-88.6%
vanA/B - Vancomycin Resistance Genes
vanA/BEnterococcusDetectedProspective Fresh20/20100%83.2-100%36/3992.3%79.1-98.4%
Prospective Frozen12/12100%73.5-100%30/3196.8%83.3-99.9%
Seeded Fresh12/12100%73.5-100%0/0--
Seeded Frozen16/16100%79.4-100%1/1100%n/a
Overall60/60100%94.0-100%67/7194.4%86.2-98.4%
KPC - Carbapenem Resistance Gene (Carbapenemase)
KPCEnterobacteriaceaeand/orA. baumannii and/orP. aeruginosaDetectedProspective Fresh3/3100%29.2-100%177/177100%97.9-100%
Prospective Frozen3/3100%29.2-100%187/187100%98.1-100%
Seeded Fresh10/10100%69.2-100%105/105100%96.5-100%
Seeded Frozen23/23100%85.2-100%89/89100%95.9-100%
Overall39/39100%91.0-100%558/558100%99.3-100%
KPCEnterobacteriaceaeDetectedProspective Fresh3/3100%29.2-100%151/151100%97.6-100%
Prospective Frozen3/3100%29.2-100%152/152100%97.6-100%
Seeded Fresh10/10100%69.2-100%83/83100%95.7-100%
Seeded Frozen23/23100%85.2-100%71/71100%94.9-100%
Overall39/39100%91.0-100%457/457100%99.2-100%
KPCEnterobacteriaceaeNot Detected:A. baumannii and/orP. aeruginosaDetectedProspective Fresh0/0--26/26100%86.8-100%
Prospective Frozen0/0--35/35100%90.0-100%
Seeded Fresh0/0--22/22100%84.6-100%
Seeded Frozen0/0--18/18100%81.5-100%
Overall0/0--101/101100%96.4-100%

"Isolates for 12 Staphylococi, 4 Enterobacteriaceae/A. bannamii/P. aeruginosa did not grow from the subcultured blood culture and could therefore not be tested with the PCR/bi-directional sequencing method. These blood cultures were considered negative for the antimicabial resistance genes by comparator method, and FilmArray performance has been calculated as True Negative (when FilmArray is negative for the analyte) or False Positive for the analyte) for the analyte) for each of these isolates.

Performance of FilmArray BCID as compared to phenotypic antimicrobial susceptibility testing (AST) results was calculated for informational purposes. Results stratified by AST method are presented in Tables 11-13. Some PPA are lower when comparing results from bacterial isolates than to PCR/sequencing direct from blood culture because phenotypic AST testing is capable of detecting antimicrobial resistance due to mechanisms other than acquisition of mecA, vanA/B, or KPC.

Table 11. mec4 Performance - Comparison to Phenotypic Antimicrobial Susceptibility Testing (AST) Methods

Note: AST results were not provided for several isolates.

PHENOTYPIC METHODSPositive Percent AgreementTP/TP + FN% (95%CI)Negative Percent AgreementTN/TN + FP% (95%CI)
ProspectiveCefoxitin Disc Diffusion22/22100%15/15100%

{11}------------------------------------------------

All StaphylococcusChromogenic Agar42/4691.3%25/3278.1%
Automated AntimicrobialSusceptibility Testing366/38096.3%226/26286.3%
All Methods430/44896.0%(93.7 - 97.6%)266/30986.1%(81.7 - 89.7%)
ProspectiveStaphylococcus,S. aureus DetectedChromogenic Agar10/1190.9%8/8100%
Automated AntimicrobialSusceptibility Testing117/11998.3%108/11296.4%
All Methods127/13097.7%(93.4 - 99.5%)116/12096.7%(91.7 - 99.1%)
SeededStaphylococcusAutomated AntimicrobialSusceptibility Testing1/1100%0/0-

Table 12. vanA/B Performance - Comparison to Phenotypic Vancomycin AST Methods

PHENOTYPIC METHODSPositive Percent AgreementNegative Percent Agreement
TP/TP + FN% (95%CI)TN/TN + FP% (95%CI)
ProspectiveEnterococcusVancomycin Screen Agar3/3100%5/5100%
Vancomycin Disc Diffusion0/10.0%--
Automated AntimicrobialSusceptibility Testing29/3096.7%55/5894.8%
All Methods32/34a94.1%(80.3 - 99.3%)60/6395.2%(86.7 - 99.0%)
SeededEnterococcusVancomycin Disc Diffusion14/14100%1/1100%
Vancomycin Screen Agar14/14100%--
All Methods28/28100%(87.7 - 100%)1/1100% (n/a)
Combined Prospectiveand SeededEnterococcusAll Methods60/62a96.8%(88.8 - 99.6%)61/6495.3%(86.9 - 99.0%)
  • Two isolates (one E. galinarum and one E. Jaecalis) that were vancomycin resistant by phenotypic AST the van AB genes by bi-directional sequence analysis.

Table 13. KPC Performance - Comparison to Phenotypic Carbapenem AST Methods Note: AST results were not provided for several isolates.

Note: Acinetobacter baunannii and Pseudonosa are commonly resistant to carbapencess due to mechanisms other than acquisition of the KPC gene (blakec). These bacteria very rarely carry the KPC gene.

PHENOTYPIC METHODSPositive Percent AgreementTP/TP + FN% (95%CI)Negative Percent AgreementTN/TN+ FP% (95%CI)
ProspectiveA. baumanniiAutomated AntimicrobialSusceptibility Testing0/100%4/4100%
SeededA. baumanniiMeropenem Disc Diffusion0/300%9/9100%
A. baumannii - All Methods0/400% (n/a)13/13100%(75.3-100%)
ProspectiveP. aeruginosaAutomated AntimicrobialSusceptibility Testing0/100%32/32100%
Meropenem Disc Diffusion--6/6100%
Meropenem/Ertapenem DiscDiffusion0/10%2/2100%
P. aeruginosa - All Methods0/110% (n/a)40/40100%(91.2-100%)
ProspectiveK. pneumoniaeAutomated AntimicrobialSusceptibility Testing6/6100%64/64100%
SeededK. pneumoniaeMeropenem Disc Diffusion19/19100%1/1100%
Modified Hodge Test11/11100%1/1100%

{12}------------------------------------------------

PHENOTYPIC METHODS(Meropenem)Positive Percent AgreementNegative Percent Agreement
TP/TP + FN% (95%CI)TN/TN + FP% (95%CI)
K. pneumoniae - All Methods36/36100%(90.3-100%)66/66100%(94.6-100%)
ProspectiveE. cloacaeAutomated AntimicrobialSusceptibility Testing-22/22100%
Automated AntimicrobialSusceptibility Testing-3/3100%
SeededE. cloacaeMeropenem Disc Diffusion0/10%
Modified Hodge Test(Meropenem)2/2100%11/11100%
E. cloacae - All Methods2/3 *66.7%(9.4-99.2%)36/36100%(90.3-100%)
ProspectiveE. coliAutomated AntimicrobialSusceptibility Testing-144/144100%
SeededE. coliModified Hodge Test(Meropenem)1/1100%4/4100%
E. coli - All Methods1/1100% (n/a)148/148100%(97.5-100%)
ProspectiveP. mirabilisAutomated AntimicrobialSusceptibility Testing-21/21100%
SeededMeropenem Disc Diffusion-4/4100%
P. mirabilisModified Hodge Test(Meropenem)0/10%11/11100%
P. mirabilis - All Methods0/1 *0% (n/a)36/36100%(90.3-100%)
ProspectiveAll OtherEnterobacteriaceaeAutomated AntimicrobialSusceptibility Testing-43/43100%
SeededAll OtherEnterobacteriaceaeAutomated AntimicrobialSusceptibility Testing-42/42100%
Meropenem Disc Diffusion-13/13100%
Modified Hodge Test(Meropenem)-61/61100%
All Other Enterobacteriaceae - All Methods-159/159100%(97.7-100%)
  • Two isolates (one E. cloacce and one P. mirabilis) that were carbapenem resistant by phenotypic AST testing were negative for the KPC gene by bi-directional sequence analysis.

Table 14. Stratification of Enterococcus Clinical Performance by Species

(Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification)

Enterococcus speciesPositive Agreement
ProspectiveSeeded
E. avium2/2 (100%)-
E. casseliflavus1/2 (50%)1/1 (100%)
E. durans1/1 (100%)-
E. faecalis55/56 (98.2%)8/8 (100%)
E. faecalis + E. faecium1/1 (100%)-
E. faecium36/37 (97.3%)9/9 (100%)
E. gallinarum2/2 (100%)1/1 (100%)
Enterococcus sp. (not speciated)-10/10 (100%)
Overall Enterococcus98/101 (97.0%)95%CI = 91.6-99.4%29/29 (100%)95%CI = 88.1-100%

Table 15. Stratification of Stuphylococcus Clinical Performance by Species

(Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification)

Staphylococcus speciesPositive Agreement
ProspectiveSeeded
S. aureus256/257 (99.6%)-
S. auricularis0/1 (0%)-
S. capitis15/17 (88.2%)-

{13}------------------------------------------------

Positive Agreement
Staphylococcus speciesProspectiveSeeded
S. capitis + S. epidermidis1/1 (100%)-
S. capitis + S. hominis1/1 (100%)-
S. capitis + S. lugdunensis1/1 (100%)-
S. carnosus0/1 (0%)-
S. cohnii1/1 (100%)-
S. cohnii + S. hominis1/1 (100%)-
S. epidermidis200/201 (99.5%)1/1 (100%)
S. epidermidis + S. hominis4/4 (100%)-
S. epidermidis + Staphylococcus sp. (not speciated)2/2 (100%)-
S. haemolyticus19/19 (100%)-
S. haemolyticus + S. hominis1/1 (100%)-
S. hominis65/65 (100%)-
S. hominis + Staphylococcus sp. (not speciated)1/1 (100%)-
S. intermedius2/2 (100%)-
S. intermedius + Staphylococcus sp. (not speciated)1/1 (100%)-
S. lentus1/1 (100%)-
S. lugdunensis5/5 (100%)-
S. saprophyticus2/2 (100%)-
S. sciuri0/1 (0%)-
S. simulans3/3 (100%)-
S. warneri4/5 (80%)-
Staphylococcus sp. (not speciated)³180/200 (90%)-
Overall Staphylococcus769/797 (96.5%)95%CI= 95.0-97.7%1/1 (100%)95%CI= n/a

4 Of the 20 unspeciated staphylococci not detected by FilmArray BCID, 16 were identified as S. pettenkoferi, 2 as S. epidermidis, I as S. capitis, and I as S. caprae by 16S sequence analysis. The 180 unspeciated Staphylococci that were detected by FilmArray BCID were not sequenced.

Streptococcus speciesPositive Agreement
ProspectiveSeeded
Group A (Pyogenic)
S. pyogenes7/7 (100%)31/31 (100%)
Group B (Pyogenic)
S. agalactiae18/18 (100%)18/18 (100%)
Group C/G (Pyogenic)
S. canis1/1 (100%)-
S. equi/S. dysgalactiae1/1 (100%)-
Streptococcus group C2/2 (100%)-
Streptococcus group G2/2 (100%)-
Group D (Bovis Group)
S. bovis3/3 (100%)-
S. equinus1/1 (100%)-
Group F (Anginosus Group)
S. anginosus4/4 (100%)-
S. anginosus group1/1 (100%)-
S. intermedius3/3 (100%)-
S. constellatus2/2 (100%)-
Mitis Group
S. gordonii1/1 (100%)-
S. mitis8/9 (88.9%)-
S. mitis + viridans streptococci1/1 (100%)-
S. mitis/S. oralis2/2 (100%)-
S. mitis/S. oralis + viridans streptococci1/1 (100%)-
S. oralis5/5 (100%)-
S. parasanguinis1/1 (100%)-
S. parasanguinis + viridans streptococci1/1 (100%)-
S. pneumoniae25/25 (100%)12/12 (100%)
S. sanguinis2/2 (100%)-
Table 16. Stratification of Streptococcus Clinical Performance by Species
(Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification)

{14}------------------------------------------------

Streptococcus speciesPositive Agreement
ProspectiveSeeded
Salivarius Group
S. salivarius1/2 (50%)-
S. salivarius + S. sanguinis group1/1 (100%)-
Other
S. vestibularis1/1 (100%)-
Viridans streptococci(not further speciated)40/43 (93.0%)1/1 (100%)
Streptococcus sp. (not speciated)1/1 (100%)-
Overall Streptococcus136/141 (96.5)95%CI=91.9-98.8%62/62 (100%)95%CI=94.2-100%

Table 17. Stratification of Enterobacteriaceae Clinical Performance by Genus/Species. (Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification)

Enterobacteriaceae genus/speciesProspectiveSeeded
Citrobacter freundii2/2 (100%)-
Citrobacter freundii + Escherichia coli1/1 (100%)-
Citrobacter koseri1/2 (50%)-
Enterobacter aerogenes5/5 (100%)2/2 (100%)
Enterobacter aerogenes + Klebsiella oxytoca1/1 (100%)-
Enterobacter cloacae19/19 (100%)17/17 (100%)
Enterobacter cloacae complex3/3 (100%)-
Enterobacter gergoviae1/1 (100%)-
Enterobacter sakasaki1/1 (100%)-
Enterobacter sp.1/1 (100%)-
Escherichia coli141/144 (98%)5/5 (100%)
Escherichia coli + Klebsiella pneumoniae2/2 (100%)-
Escherichia coli + Providencia stuartii "1/1 (100%)-
Escherichia hermannii1/1 (100%)-
Klebsiella oxytoca5/5 (100%)58/58 (100%)
Klebsiella pneumoniae67/68 (99%)34/34 (100%)
Klebsiella pneumoniae + Pantoea agglomerans1/1 (100%)-
Leclercia adacarboxylata1/1 (100%)-
Morganella morganiib + Proteus mirabilis1/1 (100%)-
Pantoea agglomerans1/1 (100%)-
Pantoea sp.0/2 (0%)-
Proteus mirabilis21/21 (100%)15/15 (100%)
Proteus vulgaris1/1 (100%)2/2 (100%)
Salmonella group B1/1 (100%)-
Salmonella group C1/1 (100%)-
Salmonella sp.1/1 (100%)-
Salmonella typhi1/1 (100%)-
Serratia marcescens22/22 (100%)54/55 (98%)
Overall Enterobacteriaceae303/310 (97.7%)95% CI = 95.4-99.1%187/188 (99.5%)95%CI = 97.1-100%

ª FilmArray BCID does not detect Providenicia stuartii; the positive Enterobacteriaceae result is likely due to the presence of Escherichia coli in the blood culture.

b FilmArray BCID does not detect Morganii; the positive Enterobacteriaceae result is likely due to the presence of Proteus mirabilis in the blood culture.

Table 18. Stratification of Proteus Clinical Performance by Species.
(Comparator Method: Standard Manual/Automated Microbiological/Biochemical Identification)
Proteus speciesPositive Agreement
ProspectiveSeeded
Proteus mirabilis22/22 (100%)15/15 (100%)
Proteus vulgaris-2/2 (100%)

{15}------------------------------------------------

Comments of Children Comments of Children Comments of Children Comments of------1 2 000 1 0000 1 1 2 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 11 8 1 8 8 8 8 8
Overall Proteus22/22 (100%)95%CI= 84 6-100%17/17 (100%)020/ 1= 80-5-100%
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.

FilmArray BCID reported a total of 81 prospective specimens with discernible multiple organism detections (5.2% of all prospective specimens; 81/1568). The majority of multiple detections (74/81; 91.3%) contained two discernible organisms, while 6.2% (5/81) contained three discernible organisms, and 2.5% (2/81) contained four discernible organisms. The most prevalent multiple detection was Enterococcus with Staphylococcus (S. aureus not detected) (1.3% of all specimens; 20/1568). Out of the 81 polymicrobial specimens, 29 contained one or more analytes that had not been detected with the reference/comparator methods, i.e., discrepant result.

Table 19. Discernible Multiple Detection Combinations as Determined by FilmArray BCID

Distinct Multiple Detection Combinationsas Determined by FilmArray BCIDTotal SpecimensDiscrepantSpecimensDiscrepant Result(s)(Organism NotDetected byReference Method)
Organism 1 ResultsOrganism 2 ResultsOrganism 3 ResultsOrganism 4 Results
Enterobacter cloacaecomplex,EnterobacteriaceaeEscherichia coli,EnterobacteriaceaeKlebsiella oxytoca,EnterobacteriaceaeKlebsiellapneumoniae,Enterobacteriaceae11E. cloacae, E. coli, K. oxytoca
Candida albicansCandida glabrataStaphylococcusStreptococcus11C. albicans
Candida albicansCandida parapsilosisEnterococcus11C. parapsilosis
EnterococcusPseudomonas aeruginosaStaphylococcus aureus, Staphylococcus10
EnterococcusProteus, EnterobacteriaceaeStaphylococcus10
EnterococcusStaphylococcusStreptococcus11Streptococcus
Candida albicansStaphylococcusStreptococcus10
StaphylococcusStreptococcus agalactiae, Streptococcus10
Proteus, EnterobacteriaceaeStaphylococcus aureus, Staphylococcus11Staphylococcus, S. aureus
Staphylococcus aureus, StaphylococcusStreptococcus agalactiae, Streptococcus10
Staphylococcus aureus, StaphylococcusStreptococcus pneumoniae, Streptococcus11Streptococcus, S. pneumoniae
Escherichia coli, EnterobacteriaceaeStaphylococcus aureus, Staphylococcus30
EnterococcusStaphylococcus aureus, Staphylococcus31Staphylococcus, S. aureus
Candida albicansStaphylococcus aureus, Staphylococcus11C. albicans
Acinetobacter baumanniiStaphylococcus aureus, Staphylococcus10
Staphylococcus aureus, StaphylococcusPseudomonas aeruginosa11P. aeruginosa
Staphylococcus aureus, StaphylococcusStreptococcus40
Organism 1 ResultsOrganism 2 ResultsOrganism 3 ResultsOrganism 4 ResultsTotal SpecimensDiscrepant SpecimensDiscrepant Result(s) (Organism Not Detected by Reference Method)
Staphylococcus11Enterobacteriaceae, E. coli
EnterococcusEscherichia coli, Enterobacteriaceae11Enterobacteriaceae, E. coli
Acinetobacter baumanniiKlebsiella pneumoniae, Enterobacteriaceae21A. baumannii
Enterobacter cloacae complexKlebsiella pneumoniae, Enterobacteriaceae11E. cloacae complex
Klebsiella pneumoniae, EnterobacteriaceaeEnterococcus31K. pneumoniae, Enteric
Klebsiella pneumoniae, EnterobacteriaceaeEscherichia coli, Enterobacteriaceae53E. coli, K. pneumoniae (2)
Candida glabrataProteus, Enterobacteriaceae11C. glabrata
Proteus, EnterobacteriaceaeEnterococcus11
EnterococcusStaphylococcus206Staphylococcus (3), Enterococcus (3)
StaphylococcusPseudomonas aeruginosa11Staphylococcus
Escherichia coli, EnterobacteriaceaeStreptococcus21Streptococcus
Klebsiella pneumoniae, EnterobacteriaceaeStreptococcus10
StaphylococcusStreptococcus70
Candida albicansEnterococcus20
Candida kruseiEnterococcus10
Candida glabrataEnterococcus10
EnterococcusStaphylococcus10
Candida albicansCandida glabrata11C. glabrata
Candida albicansEnterococcus11C. albicans
EnterobacteriaceaeEnterococcus10
Acinetobacter baumanniiPseudomonas aeruginosa20
EnterobacteriaceaePseudomonas aeruginosa10
EnterobacteriaceaeStaphylococcus11Staphylococcus
Total Specimens with Multiple Detections8129

{16}------------------------------------------------

Table 20. Additional Specimens with Multiple Isolates Identified by Reference/Comparator Methods Note: Organisms shaded gray are not targeted by FilmArray BCID (i.e., off-panel organisms). This list docs not include multiple detection combinations already represented in the previous table of FilmArray BCID multiple detections.

{17}------------------------------------------------

Distinct Multiple Detections by Reference/Comparator methods
Isolate 1Isolate 2Isolate 3Isolate 4Total SpecimensDiscrepant SpecimensDiscrepant Result(s)(Targeted OrganismsNot Detected byFilmArray BCID)
Aeromonas sobriaPantoea agglomeransPantoea agglomeransPseudomonasaeruginosa10
Enterococcus faecalisFlavobacteriumspeciesKlebsiella pneumoniaeStaphylococcus species11Staphylococcus
Klebsiella pneumoniaeStaphylococcus speciesStaphylococcus speciesViridans streptococci11Staphylococcus,Streptococcus
Neisseria speciesViridans streptococciViridans streptococciViridans streptococci10
Acinetobacter lwoffiiCorynebacteriumspeciesStaphylococcusepidermidis10
Coryneform bacteriumspeciesStaphylococcus aureusStreptococcus oralis10
EnterococcuscasseliflavusEscherichia coliStaphylococcus aureus11Enterococcus
Klebsiella pneumoniaeKlebsiella pneumoniaeStreptococcusmitis/oralis10
Pantoea speciesStaphylococcusintermediusStaphylococcus species11Enterobacteriaceae
Staphylococcus aureusStaphylococcushaemolyticusStreptococcusparasanguis10
Staphylococcus capitisStaphylococcusepidermidisStaphylococcuslugdunensis10
Streptococcusmitis/oralisViridans streptococciViridans streptococci10
Viridans streptococciViridans streptococciViridans streptococci10
Abiotrophia defectivaStaphylococcus species11Staphylococcus
Acinetobacterbaumannii (seq. = A.nosocomialis/calcoaceticus)Acinetobacterbaumannii (seq. = A.nosocomialis/calcoaceticus)10
Acinetobacter lwoffiiKlebsiella pneumoniae11Streptococcus
Acinetobacter lwoffiiViridans streptococci11Staphylococcus
Acinetobacter lwoffiiStaphylococcus species11K. pneumoniae ,Enterobacteriaceae
Aerococcus viridansKlebsiella pneumoniae11Staphylococcus
Aerococcus speciesStaphylococcusepidermidis11Staphylococcus
Bacillus pumilusPseudomonasfluorescens/putida10
BrevundimonasdiminutaWeeksella virosa10
Candida parapsilosisKocuria kristinae10
Citrobacter freundiiEscherichia coli10
Citrobacter koseriEnterococcus faecium10
CorynebacteriumjeikeiumCorynebacteriumspecies10
CorynebacteriumspeciesCorynebacteriumspecies10
CorynebacteriumspeciesEnterococcus faecalis10
CorynebacteriumspeciesMicrococcus species10
CorynebacteriumspeciesStaphylococcus aureus20
CorynebacteriumspeciesStaphylococcushaemolyticus20
CorynebacteriumspeciesStaphylococcushominis20
CorynebacteriumspeciesStaphylococcus species31Staphylococcus
Total SpecimensDiscrepant SpecimensDiscrepant Result(s) (Targeted Organisms Not Detected by FilmArray BCID)
Isolate 1Isolate 2Isolate 3Isolate 4
Enterobacter aerogenesKlebsiella oxytoca10
Enterococcus faecalisEnterococcus faecium10
Enterococcus faecalisStenotrophomonas maltophilia10
Enterococcus faecalisViridans streptococci11Enterococcus
Enterococcus faeciumEnterococcus faecium10
Escherichia coliEscherichia coli30
Escherichia coliPasteurella multocida11E. coli , Enterobacteriaceae
Escherichia coliProvidencia stuartii10
Escherichia coliStenotrophomonas maltophilia10
Haemophilus influenzaeMoraxella catarrhalis10
Klebsiella pneumoniaePantoea agglomerans11K. pneumoniae
Lactobacillus acidophilusStreptococcus species10
Micrococcus speciesStaphylococcus epidermidis10
Morganella morganiiProteus mirabilis10
Neisseria speciesStaphylococcus hominis10
Rhodococcus speciesStaphylococcus warneri11Staphylococcus
Staphylococcus aureusStaphylococcus aureus20
Staphylococcus aureusStaphylococcus cuprae10
Staphylococcus aureusStaphylococcus species20
Staphylococcus aureusStreptococcus salivarius11Streptococcus
Staphylococcus capitisStaphylococcus capitis10
Staphylococcus capitisStaphylococcus epidermidis10
Staphylococcus capitisStaphylococcus hominis10
Staphylococcus capitisStreptococcus pneumoniae11Staphylococcus
Staphylococcus cohniiStaphylococcus hominis10
Staphylococcus epidermidisStaphylococcus hominis40
Staphylococcus epidermidisStaphylococcus species20
Staphylococcus haemolyticusStaphylococcus hominis10
Staphylococcus hominisStaphylococcus hominis10
Staphylococcus hominisStaphylococcus species10
Staphylococcus speciesStaphylococcus species30
Staphylococcus speciesStenotrophomonas maltophilia10
Streptococcus parasanguinisViridans streptococci10
Streptococcus salivariusStreptococcus sanguis group10
Viridans streptococciStreptococcus mitis10
Distinct Multiple Detections by Reference/Comparator methodsTotal SpecimensDiscrepant SpecimensDiscrepant Result(s) (Targeted Organisms Not Detected by FilmArray BCID)
Isolate 1Isolate 2Isolate 3Isolate 4
Viridans streptococciViridans streptococci30
Total8616

: : : :

{18}------------------------------------------------

·

{19}------------------------------------------------

The reference method detected 201 off-panel organism isolates (i.e., those not targeted by FilmArray BCID) from the 1568 prospective cultures. The majority of these isolates belong to groups of organisms commonly considered to be blood culture contaminants (49 Corynebacterium/Diphtheroids, 33 Bacillus sp., and 27 Micrococcus sp., among others). Occurrence of off-panel organisms in the prospective arm of the clinical evaluation is presented in Table 21.

Off-Panel OrganismNumberIdentifiedOff-Panel OrganismNumberIdentified
Abiotrophia sp. or Granulicatella sp. (formerlynutritionally-deficient Streptococci)7Flavobacterium species1
Achromobacter xylosoxidans1Fusarium species1
Acinetobacter sp. (not A. baumannii)23Kocuria kristinae1
Actinomyces odontolyticus2Lactobacillus acidophilus1
Actinomyces species1Lactobacillus species2
Aerococcus species1Micrococcus luteus1
Aerococcus viridans2Micrococcus luteus/lylae1
Aeromonas sobria1Micrococcus species25
Bacillus cereus19Moraxella catarrhalis1
Bacillus pumilus1Moraxella osloensis1
Bacillus species13Moraxella species1
Brevibacterium species1Mycobacterium fortuitum complex1
Brevibacterium ensei1Mycobacterium species1
Brevundimonas diminuta1Neisseria species2
Brevundimonas vesicularis1Paenibacillus species1
Burkholderia cepacia complex2Pasteurella multocida2
Candida kefyr1Pasteurella species1
Capnocytophaga species1Propionibacterium species1
Chryseobacterium meningosepticum(Elizabethkingia/Flavobacterium)1Pseudomonas fluorescens/putida2
Chryseobacterium indologenes1Pseudomonas species3
Chryseomonas luteola1Rhizobium radiobacter2
Corynebacterium jeikeium1Rothia (Stomatococcus) mucilaginosa4
Corynebacterium mucifaciens1Sphingomonas mucosissima1
Corynebacterium species/Diphtheroids47Stenotrophomonas maltophilia10
Cryptococcus neoformans2Weeksella virosa1

Table 21. Occurrence of Off-Panel Organisms as Determined by Reference/Comparator Methods

{20}------------------------------------------------

Selected Analytic Studies

Growth and Detection

A study was performed to establish the range of expected organism concentrations in blood cultures that would be tested with the FilmArray BCID Panel from the time of positivity up to eight hours after positivity. All organism growth and testing was performed using seeded blood culture bottles (BACTECTM Plus Aerobic/F Medium incubated in the BACTEC™ 9050 continuously monitoring blood culture instrument). Each microorganism was mixed with human whole blood and seeded directly into blood culture bottles for growth. At the time of positivity (and/or eight hours after positivity), the blood culture was removed from the instrument for plate enumeration (determination of CFU/mL) and FilmArray BCID testing. Three independent positive cultures (bottles) were evaluated for each organism at each time point and FilmArray testing was performed in triplicate for each bottle.

Table 22 summarizes the concentration of organism (CFU/mL) determined for a representative panel of 30 isolates. The number and percent of correct positive BCID Panel test results is provided for each isolate and overall (% Detected). A correct result means that both the correct organism and antimicrobial resistance gene (where applicable) were detected in the sample.

At Positivity8 Hours After Positivity
Species/Isolate(s) TestedPer Bottle (CFU/mL)Mean (CFU/mL)# Detected/Total (% Detected)Per Bottle (CFU/mL)Mean (CFU/mL)# Detected/Total (% Detected)
Gram-Positive Bacteria
Enterococcus faecalis [vanB+]JMI 3684.60E+081.80E+082.62E+083.01E+089/9(100%)7.25E+088.90E+081.07E+098.95E+089/9(100%)
Enterococcus faecium [vanA+]JMI 4751.47E+081.53E+081.59E+081.53E+089/9(100%)2.23E+081.64E+081.55E+081.81E+089/9(100%)
Enterococcus hiraeATCC 491351.26E+082.76E+083.25E+082.42E+089/9(100%)8.00E+086.60E+087.20E+087.27E+089/9(100%)
Listeria monocytogenesCDC F2380 (ATCC 43256)4.50E+081.22E+091.18E+099.50E+089/9(100%)1.76E+092.31E+091.67E+091.91E+099/9(100%)
Staphylococcus aureusATCC 116321.48E+082.00E+072.56E+076.45E+079/9(100%)8.75E+089.80E+081.21E+086.59E+089/9(100%)
Staphylococcus aureus[MRSA/mecA]ATCC BAA-17471.41E+075.65E+066.05E+068.60E+069/9(100%)5.70E+073.85E+079.75E+076.43E+079/9(100%)
Staphylococcus epidermidisATCC 122281.38E+089.85E+071.16E+081.18E+089/9(100%)2.12E+083.95E+081.56E+097.22E+089/9(100%)
Staphylococcus epidermidis[MRSE/mecA]3.60E+073.75E+077.65E+079/9(100%)1.35E+096.80E+081.44E+099/9(100%)

Table 22. Summary of Organism Concentration (CFU/mL) in Positive Blood Cultures and Correct Detection of Organisms in Positive Blood Cultures by the FilmArray BCID Panel

{21}------------------------------------------------

At Positivity8 Hours After Positivity
Species/Isolate(s) TestedPer Bottle (CFU/mL)Mean (CFU/mL)# Detected/Total (% Detected)Per Bottle (CFU/mL)Mean (CFU/mL)# Detected/Total (% Detected)
ATCC 298871.56E+084.50E+082.29E+093.15E+08
Streptococcus agalactiaeATCC 138131.22E+089.15E+084.96E+089/9(100%)5.80E+084.30E+084.42E+089/9(100%)
Streptococcus mitisATCC 159141.57E+081.51E+096.90E+087.86E+089/9(100%)1.50E+092.03E+092.91E+092.15E+099/9(100%)
Streptococcus pneumoniaeATCC BAA-2553.45E+082.67E+081.31E+096.41E+089/9(100%)1.03E+096.00E+081.37E+091.00E+099/9(100%)
Streptococcus pyogenesATCC 196152.53E+082.44E+083.80E+082.92E+089/9(100%)2.38E+085.70E+088.90E+085.66E+089/9(100%)
Gram-Negative Bacteria
Acinetobacter baumanniiATCC 99552.17E+081.44E+082.45E+082.02E+089/9(100%)4.85E+083.85E+084.35E+084.35E+089/9(100%)
Enterobacter cloacaeATCC 130474.20E+083.95E+081.50E+083.22E+089/9(100%)2.23E+091.46E+092.19E+091.96E+099/9(100%)
Escherichia coliATCC 438889.80E+076.10E+071.93E+081.17E+089/9(100%)1.17E+091.39E+097.70E+078.79E+089/9(100%)
Klebsiella oxytocaATCC 131827.40E+086.85E+083.85E+086.03E+089/9(100%)3.05E+091.86E+091.20E+092.04E+099/9(100%)
Klebsiella oxytoca [+KPC]JMI 78186.15E+079.15E+073.05E+076.12E+079/9(100%)1.96E+092.00E+091.13E+091.70E+099/9(100%)
Klebsiella pneumoniaeATCC 138834.35E+082.10E+089.15E+085.20E+089/9(100%)1.60E+091.65E+091.58E+091.61E+099/9(100%)
Klebsiella pneumoniae [+KPC]JMI 7661.21E+082.50E+082.05E+081.92E+089/9(100%)1.14E+099.10E+087.70E+089.40E+089/9(100%)
Proteus mirabilisATCC 299063.25E+071.04E+089.10E+077.58E+079/9(100%)1.04E+099.80E+087.30E+089.17E+089/9(100%)
Serratia marcescensATCC 271378.35E+081.46E+094.90E+089.28E+089/9(100%)1.053+091.37E+091.02E+091.15E+099/9(100%)
Serratia marcescens [+KPC]JMI 6974.90E+083.90E+081.02E+083.27E+089/9(100%)2.19E+091.40E+092.42E+081.28E+099/9(100%)
Haemophilus influenzae (type b)ATCC 102112.80E+083.60E+082.23E+082.88E+089/9(100%)3.25E+093.35E+092.74E+093.11E+099/9(100%)
Neisseria meningitidisATCC 437442.07E+083.90E+081.55E+082.51E+089/9(100%)6.65E+087.65E+087.85E+087.38E+089/9(100%)
Pseudomonas aeruginosaATCC 278531.34E+081.76E+089.75E+071.36E+089/9(100%)1.35E+091.39E+081.76E+091.08E+099/9(100%)
Yeast
Candida albicans9.05E+039/98.80E+049/9
ATCC 102318.00E+043.12E+04(100%)1.03E+059.70E+04(100%)
4.65E+031.00E+05
Candida glabrata1.26E+061.45E+069/91.47E+072.01E+079/9
ATCC 155451.11E+06(100%)2.65E+07(100%)

BioFire Diagnostics, Inc. 510(k)
FilmArray BCID Panel

.

{22}------------------------------------------------

Species/Isolate(s) TestedAt Positivity8 Hours After Positivity
Per Bottle(CFU/mL)Mean(CFU/mL)# Detected/Total(% Detected)Per Bottle(CFU/mL)Mean(CFU/mL)# Detected/Total(% Detected)
Candida kruseiATCC 908781.97E+064.82E+069/9(100%)1.91E+073.16E+079/9(100%)
5.65E+062.68E+07
2.47E+063.55E+07
6.35E+063.25E+07
Candida parapsilosisATCC 908752.56E+063.12E+069/9(100%)6.70E+075.35E+079/9(100%)
3.60E+063.80E+07
3.20E+065.55E+07
Candida tropicalisATCC 660291.50E+069.70E+059/9(100%)1.10E+071.36E+079/9(100%)
7.45E+052.04E+07
6.65E+059.45E+06
Overall Correct Detectiona(Organism and AntimicrobialResistance Genes)At Positivity:270/270(100%)8 Hours After Positivity:270/270(100%)

" In addition to the correct results (Sreptococus, Sreptococus agadactiae, Haenophilus influenzae, Neisseria neningitids, and Candida krase) were observed in a single run (1/540; 0.2%), The correct results were obtained when retested.

Inclusivity

Analytical reactivity (inclusivity) of the BCID Panel was evaluated in a study that assessed the ability of the BCID Panel to detect a diverse collection of 303 isolates of genetically, phenotypically, and geographically diverse bacteria and yeast, and to properly indicate the presence of four different antimicrobial resistance genes.

Each isolate was initially tested in blood culture matrix at a concentration consistent with the levels of organism enumerated from blood cultures at the time of positivity (see Growth and Detection section above). If the expected result was obtained at the injual test level, no further testing was performed. If an isolate was not detected initially, additional testing was performed at 10-100 fold higher concentrations. If detected at the higher concentration(s), the species/isolate is indicated as detected with reduced sensitivity and the concentration of organism that was detected is indicated. If not detected at the highest concentration, the isolate is listed as not detected by the FilmArray BCID Panel. Results are provided below for each FilmArray BCID Panel test result.

When possible, in silico analysis of sequence data was used to make predictions of assay reactivity for less common species that may be detected by the FilmArray BCID Panel but were not tested.

Gram-Positive Bacteria

Enterococcus

Enterococcus Detected[~1x108 CFU/mL]Enterococcus Detected withReduced Sensitivity[~1x109 CFU/mL]EnterococcusNot Detecteda
Enterococcus aviumATCC 49463Enterococcus saccharolyticusATCC 43076Enterococcus pseudoaviumATCC 49372
Enterococcus casseliflavusATCC 700668Enterococcus disparATCC 51266Enterococcus raffinosusATCC 49427
Enterococcus cecorumATCC 43198

Table 23. Enterococcus Inclusivity Results

{23}------------------------------------------------

Enterococcus Detected[~1x108 CFU/mL]Enterococcus Detected withReduced Sensitivity[~1x109 CFU/mL]EnterococcusNot Detecteda
EnterococcusduransATCC 11576
ATCC 49532
ATCC 49533
EnterococcusfaecalisJMI 12536
ATCC 51299
ATCC 700802
JMI 368
EnterococcusfaeciumATCC 27270
ATCC 35667
ATCC BAA-2127
JMI 536
ATCC 700221
JMI 475
EnterococcusflavescensATCC 49996
EnterococcusgallinarumATCC 49608
Enterococcus hiraeATCC 8043
EnterococcusmalodoratusATCC 43197
EnterococcusmundtiiATCC 43187

a Not detected at the highest test concentrations ~1x10"-1x10" CFU/mL.

Listeria monocytogenes

Table 24. Listeria monocytogenes Inclusivity Results

Listeria monocytogenes Detectedᵃ
SpeciesSerotypeIsolate ID
Listeria monocytogenes1/2aFSL-C1-056b
Listeria monocytogenes1/2aFSL-J2-020b
Listeria monocytogenes1/2bFSL-J2-064b
Listeria monocytogenes1/2bHUM-2009042206c
Listeria monocytogenes4bATCC 43256
Listeria monocytogenes4bATCC 13932

a Estimated concentration in a positive blood culture is ~5x10° CFU/mL.

b Isolates obtained from Cornell University.

& Isolates obtained from the Colorado Department of Public Health (CDPH).

Staphylococcus (including Staphylococcus aureus)

Table 25. Stuphylococcus aureus Inclusivity Results

Staphylococcus/Staphylococcus aureus Detecteda
SpeciesIsolate IDStrain InformationPFGE Type
Methicillin-sensitive S. aureus (MSSA)
Staphylococcus/Staphylococcus aureus Detecteda
SpeciesIsolate IDStrain InformationPFGE Type
Staphylococcus aureusATCC BAA-174996:308USA 900
Staphylococcus aureusATCC BAA-1759N7129USA 900
Staphylococcus aureusATCC BAA-1765102-04USA 1200
Staphylococcus aureusbATCC 12600NCTC 8532 Type strainUnknown
Staphylococcus aureusbATCC 11632S13Unknown
Staphylococcus aureusATCC BAA-2419Mass/2010Unknown
Staphylococcus aureusATCC BAA-2420Mass/2010Unknown
Staphylococcus aureusATCC BAA-2421Mass/2010Unknown
Staphylococcus aureus1060728n/aUnknown
Staphylococcus aureusAnt1n/aUnknown
Staphylococcus aureusLem8n/aUnknown
Staphylococcus aureusMAL8134n/aUnknown
Staphylococcus aureusMAQn/aUnknown
Staphylococcus aureusPer2n/aUnknown
Staphylococcus aureusRARn/aUnknown
Staphylococcus aureusS313n/aUnknown
Staphylococcus aureusSal3n/aUnknown
Staphylococcus aureusVer2n/aUnknown
Staphylococcus aureus ssp. aureusbATCC 10832Wood 46Unknown
Staphylococcus aureus ssp. aureusbATCC 14154RoseUnknown
Staphylococcus aureus ssp. aureusATCC 25923Seattle/1945Unknown
Borderline Oxacillin-resistant S. aureus (BORSA)
Staphylococcus aureusSUN1dn/aUnknown
Staphylococcus aureusSUN2dn/aUnknown
Staphylococcus aureusSUN3dn/aUnknown
Staphylococcus aureusSUN4dn/aUnknown
Staphylococcus aureusSUN5dn/aUnknown
Staphylococcus aureusSUN6dn/aUnknown
Methicillin-resistant S. aureus (MRSA)
Staphylococcus aureus ssp. aureusATCC BAA-38E2125 DenmarkUnknown
Staphylococcus aureus ssp. aureusATCC 43300F-182 KansasUnknown
Staphylococcus aureus ssp. aureusATCC 700698Mu3 Japan/1996Unknown
Staphylococcus aureus ssp. aureusATCC BAA-1720MRSA252 UKUnknown
Staphylococcus aureus ssp. aureusATCC BAA-39HUSA304 Hungary/1993Unknown
Staphylococcus aureusNARSA NRS705NY-12 New York/2005USA 100
Staphylococcus aureusNARSA NRS701MN-082 Minn/2006USA 200
Staphylococcus aureus ssp. aureusATCC BAA-1717TCH1516 TexasUSA 300
Staphylococcus aureusNARSA NRS703MN-095 Minn/2006USA 300
Staphylococcus aureusNARSA NRS683GA-298 Georgia/2005USA 300
Staphylococcus aureusNARSA NRS662CO-34 Colorado/2005USA 300
Staphylococcus aureusNARSA NRS707NY-155 New York/2005USA 300
Staphylococcus aureusATCC BAA-1707MW2 N. Dakota/1998USA 400
Staphylococcus aureusNARSA NRS691GA-62 Georgia/2005USA 500
Staphylococcus aureusNARSA NRS648CA-347 California/2005USA 600
Staphylococcus aureusNARSA NRS689GA-442 Georgia/2006USA 700
Staphylococcus aureus ssp. aureusATCC BAA-42HDE288 Portugal/1996USA 800
Staphylococcus aureusNARSA NRS668CO-72 Colorado/2005USA 800
Staphylococcus aureusATCC BAA-174794:1013 Vermont/1993USA 1000
Staphylococcus aureusNARSA NRS676CT-19 Conn/2005USA 1000
Staphylococcus aureusNARSA NRS745CA-629 California/2006USA 1000
Staphylococcus aureusATCC BAA-17647031 AlaskaUSA 1100
Staphylococcus aureusATCC BAA-1691HFH-30137 Michigan/2003Not 100-1100
Staphylococcus aureusATCC BAA-1700HFH-33798 Illinois/2004Not 100-1100
Staphylococcus aureusATCC BAA-2312M10/0061 Ireland/2010Unknown
Staphylococcus aureusATCC BAA-2313M10/0148 Ireland/2010CC130

0

{24}------------------------------------------------

a Detected at the initial test concentration of 5x106CFU/mL.

{25}------------------------------------------------

6 Initial test concentration was 5x105 CFU/mL.

& Isolates obtained from University of Rennes, France,

4 Isolates obtained from Sunnybrook Research Institute, affiliated with the University of Toronto,

€ Tested as a seeded blood culture at the time of positivity.

Staphylococcus Detected[~5x10 6 CFU/mL]Staphylococcus Detected withReduced Sensitivity[~5x10 7 CFU/mL]StaphylococcusNot Detected b
Staphylococcus lutraeATCC 700373Coagulase-positive staphylococci (non-S.aureus)Staphylococcus intermediusATCC 29663
Staphylococcus pseudointermediusATCC 49444
Staphylococcus schleiferi subsp. coagulansATCC 49545
Coagulase-negative staphylococci (CoNS)
Staphylococcus capraeATCC 51548Staphylococcus capitis subsp. capitisATCC 27842Staphylococcus auricularisClinical isolate d
Staphylococcus cohniiATCC 29972Staphylococcus pasteuriATCC 51127Staphylococcus carnosusATCC 51365
ATCC 12228Staphylococcus saprophyticusATCC 15305Staphylococcus lentus eATCC 700403
ATCC 29886Staphylococcus simulansClinical isolates fStaphylococcus pettenkoferi5 clinical isolates
Staphylococcus epidermidisATCC 55133Staphylococcus warneriATCC 25614Staphylococcus schleiferi subsp. schleiferiATCC 43808
ATCC 29887Staphylococcus sciuriATCC 29060
ATCC 51625
ATCC 35984
Staphylococcus equorumATCC 43958
Staphylococcus haemolyticusATCC 29968
Staphylococcus hominis ssp. hominisATCC 25615
Staphylococcus lugdunensisATCC 43809
Staphylococcus xylosusATCC 29966

Table 26, Results of Staphylocaccus (non-S. aureus) Inclusivity Testing 2

4 All 54 S. aureus isolates (Table 25 above) received Staphylococcus Detected results.

6 Not detected when tested at a concentration of ≥5x10. CFU/mL.

5 Isolates identified as Staphylococcus internedius by automated identification systems were detected in two clinical specimens.

4 Staphylococcus auricularis was not tested in analytic studies, but was not detected in a clinical blood culture.

4 An isolate identified as Staphylococcus lentus by an automated identification system was detected in one clinical specimen.

I Staphylococcus simulars was not tested in analytic studies, but was detected in three clinical blood cultures at unknown concentration.

Based on inclusivity testing results for Staphylococci and in silico analysis of available sequences, the following predictions of reactivity are provided for many less common CoNS species that were not tested.

Please note that performance of the FilmArray BCID Panel for these organisms has not been established.

{26}------------------------------------------------

Detection PredictedaDetection Predicted withReduced SensitivitybDetection Not Predictedc
Staphylococcus gallinarumStaphylococcus microtiStaphylococcus arlettae
Staphylococcus kloosiiStaphylococcus simiaeStaphylococcus chromogenes
Staphylococcus succinusStaphylococcus condimenti
Staphylococcus fleurettii
Staphylococcus piscifermentans
Staphylococcus pulvereri
Staphylococcus rostri
Staphylococcus saccharolyticus
Staphylococcus vitulinus

Table 27. In Silico Predictions of Stuphylococcus Reactivity

4 Predicted result of Staphylococcus Detected when present in a blood culture sample at a concentration of ≥5x10 °CF'U/mL.

b Predicted result of Staphylococcus Detected when present in a blood culture sample at a concentration of ≥ 5x10 CFUmL. & Predicted result of Staphylococcus Not Detected at relevant concentrations.

Streptococcus (including S. agalactiae, S. pneumoniae, and S. pyogenes)

Streptococcus Detected*
SpeciesIsolate IDStrain Information
Streptococcus pyogenesATCC 19615
Streptococcus pyogenesPCMC 20100107C102
Streptococcus pyogenesATCC 49399Group A (Pyogenic group)
Streptococcus pyogenesATCC 12344
Streptococcus pyogenesATCC 12384
Streptococcus agalactiaeATCC 13813Type strain - Serotype la/c
Streptococcus agalactiaePCMC 20100107CI03Untyped clinical isolate
Streptococcus agalactiaeATCC 12403Type IIIGroup B (Pyogenic group)
Streptococcus agalactiaeATCC BAA-611Serotype V
Streptococcus agalactiaeNCTC 8017Unknown serotype
Streptococcus dysgalactiae ssp. equisimilisATCC 12388Group C/G (Pyogenic group)
Streptococcus bovisATCC 33317Group D (Bovis group)
Streptococcus equinisATCC 9812
Streptococcus mutansATCC 25175Group E (Mutans group)
Streptococcus anginosusATCC 33397
Streptococcus intermediusATCC 27335Group F (Anginosus group)
Streptococcus constellatusATCC 27513
Streptococcus gordoniiATCC 10558Mitis group
Streptococcus parasanguinisATCC 31412
Streptococcus sanguinisATCC 10556
Streptococcus mitisATCC 15914
Streptococcus oralisATCC 10557
Streptococcus pseudopneumoniaeATCC BAA-960
Streptococcus pneumoniaeATCC BAA-255Strain R6 (no capsule)
Streptococcus pneumoniaeATCC 700672Serotype 14

. Table 28. Streptococcus Inclusivity Results

{27}------------------------------------------------

Streptococcus Detecteda
SpeciesIsolate IDStrain Information
Streptococcus pneumoniaeSerotype 4
Streptococcus pneumoniaeATCC 700673
Streptococcus pneumoniaeSerotype 19A
Streptococcus pneumoniaeATCC BAA-341
Streptococcus pneumoniaeSerotype 5
Streptococcus salivariusATCC 13419Salivarius group
Streptococcus gallolyticusATCC BAA-2069Uncertain grouping

4 Detected at the initial test concentration of ~1x10* CFU/mL.

Table 29. Streptococcus agalactiae Inclusivity Results

Streptococcus/Streptococcus agalactiae (Group B) Detecteda
SpeciesIsolate IDStrain Information
Streptococcus agalactiaeATCC 13813Type strain - Serotype 1a/c
Streptococcus agalactiaePCMC 20100107CI03Untyped clinical isolate
Streptococcus agalactiaeATCC 12403Type IIIGroup B (Pyogenic group)
Streptococcus agalactiaeATCC BAA-611Serotype V
Streptococcus agalactiaeNCTC 8017Unknown serotype

" Detected at the initial test concentration of ~1x108 CFU/mL.

Table 30. Streptococcus pneumoniae Inclusivity Results

Streptococcus/Streptococcus pneumoniae Detecteda,b
SpeciesIsolate IDStrain Information
Streptococcus pneumoniaeATCC BAA-255Strain R6 (no capsule)
Streptococcus pneumoniaeATCC 700672Serotype 14
Streptococcus pneumoniaeATCC BAA-334Serotype 4Mitis group
Streptococcus pneumoniaeATCC 700673Serotype 19A
Streptococcus pneumoniaeATCC BAA-341Serotype 5

ª Detected at the initial test concentration of ~1x10*CFU/mL.

b Based on sequence analysis, the BCID Panel may not detect S. pneumoniae serotypes I I A and 19, or may detect these serotypes with reduced sensitivity compared to other serotypes.

Streptococcus/Streptococcus pyogenes (Group A) Detecteda
SpeciesIsolate IDStrain Information
Streptococcus pyogenesATCC 19615
Streptococcus pyogenesPCMC 20100107CI02
Streptococcus pyogenesATCC 49399Group A (Pyogenic group)
Streptococcus pyogenesATCC 12344
Streptococcus pyogenesATCC 12384

Table 31. Streptococcus nvogenes Inclusivity Results

4 Detected at the initial test concentration of ~1x108 CFU/mL.

Based on results of inclusivity testing and in silico analysis of available sequences, the following predictions of reactivity are provided for less common Streptococcus species that were not tested. As shown in Table 32 below, the analysis predicts that many species

{28}------------------------------------------------

will be detected at concentrations expected in positive blood cultures (10°-10° CFU/mL), and others (particularly Mutans group species) will likely not be detected due to sequence mismatches with the assay primers.

Please note that performance of the FilmArray BCID Panel for these organisms has not been established.

Detection PredictedaDetection Predicted with ReducedSensitivitybDetection Not Predicted
Streptococcus australisStreptococcus parauberisStreptococcus cricetic
Streptococcus equiStreptococcus downeic
Streptococcus ictaluriStreptococcus macacaec
Streptococcus infantisStreptococcus porcinus
Streptococcus infantariusStreptococcus urialis
Streptococcus pasteurianus
Streptococcus perois
Streptococcus suis
Streptococcus thermophilus
Streptococcus vestibularis

Table 32. In Silico Predictions of Streptococcus Reactivity

4 Predicted result of Streptococcus Detected when present in a blood culture sample at a concentration of -1x10 CFU/mL.

8 Predicted result of Streptococcus Detected when present in a blood culture sample at a concentration of ≥1x10" CFU/mL,

° Mutans group streptococci.

Gram-Negative Bacteria Acinetobacter baumannii

Acinetobacter baumannii Detecteda
SpeciesIsolate ID
Acinetobacter baumanniiATCC 9955
Acinetobacter baumanniiATCC BAA-1605
Acinetobacter baumanniiATCC 17961
Acinetobacter baumanniiATCC 19003
Acinetobacter baumanniiATCC BAA-2093
Acinetobacter baumanniiATCC 15308

Table 33. Acinetobacter baumannii Inclusivity Results

4 Detected at the initial test concentration of ~1x108 CFU/mL.

Enterobacteriaceae (including Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens)

Table 34. Results of Enterobacteriaceae Inclusivity Testing
Enterobacteriaceae Detected[~5×107 CFU/mL or 1×108 CFU/mL]Enterobacteriaceae Detectedwith Reduced Sensitivity[~5×108-1×109 CFU/mL]EnterobacteriaceaeNot Detecteda
Cedeceae davisiaeATCC 43023Edwardsiella tardaATCC 15947Morganella morganii subsp. morganiiATCC 25829
Citrobacter freundiiATCC 43864Enterobacter gergoviaeATCC 33028Pantoea (Enterobacter) agglomeransbATCC 27155
Citrobacter koseriATCC 29223Hafnia alveiATCC 51815Providencia (Proteus) acalifaciensATCC 51902
Cronobacter muytjensiiATCC 51329Salmonella bongoriSGSC 3041Providencia (Proteus) rettgeriATCC 9250

Table 34, Results of Euterobacteriacage Inclusivity Testing

{29}------------------------------------------------

Enterobacteriaceae Detected[~5×107 CFU/mL or 1×108 CFU/mL]Enterobacteriaceae Detectedwith Reduced Sensitivity[~5×108-1×109 CFU/mL]EnterobacteriaceaeNot Detectedn
Cronobacter (Enterobacter) sakazakiiATCC 29544Serratia fonticolaATCC 29844Providencia stuartiATCC 33672
Enterobacter aerogenesATCC 13048Serratia odoriferaATCC 33077Rahnella aquatilisATCC 33071
Enterobacter aerogenesATCC 29751Serratia rubidaeaATCC 27593Serratia liquefaciensATCC 27592
Enterobacter asburiaeATCC 35953Tatumella ptyseosATCC 33301
Enterobacter amnigenusATCC 51816Serratia plymuthicaATCC 183
Enterobacter cloacae9 isolatescYersinia enterocoliticaATCC 6025
Enterobacter hormaecheiATCC 49162
Enterobacter kobeiATCC BAA-260d
Enterobacter nimipressuralisATCC 9912d
Escherichia coli5 isolatese
Escherichia fergusoniiATCC 35469
Escherichia hermaniiATCC 33650
Escherichia vulnerisATCC 33821
Klebsiella oxytoca11 isolatesf
Klebsiella pneumoniae10 isolatesg
Klebsiella variicolaATCC BAA-830
Kluyvera ascorbataATCC 33433
Kluyvera (Enterobacter) intermediusATCC 33110
Leclercia adecarboxylataATCC 23216
Proteus species10 isolatesh
Raoultella ornithinolyticaATCC 31898
Raoultella planticolaATCC 31900
Raoultella terrigenaATCC 33257
Salmonella enterica-cholerasiusATCC 10708
Salmonella enterica-heidelbergATCC 8326
Salmonella enterica-paratyphiSGSC 3222
Salmonella enterica-typhimuriumATCC 13311
Serratia marcescens6 isolatesi
Serratia entomophilaATCC 43705
Serratia ficariaATCC 33105
Shigella boydiijATCC 8700
Shigella dysenteriaejPHM-2004008089
Shigella flexnerijATCC 12022
Shigella sonneijATCC 11060
Yokenella regensburgeiATCC 35313

4 Not Detected at the highest test concentration of 1×10°-1×10°CFU/mL.

b Not Detected in this study, but Pantoea agglomerans was detected by the BCID Panel in a clinical blood culture.

° See Enterobacter cloacae complex table.

4 Tested as purified nucleic acid at a concentration of 0.63ug/mL (equivalent 10 ~1.0×10° CFU/mL).

{30}------------------------------------------------

€ See Escherichia coli table.

¹ Sec Klebsiella oxytoca table.

8 See Klebsiella pneumoniae table.

" Sec Proteus table.

1 See Serratia marcescens table, T Tested as a seeded blood culture within 1 hour of positivity.

ested as a seeded blood culture within 1 hour of positivity.

Based on results of inclusivity testing and in silico analysis of available sequences, the following predictions of reactivity are provided for less common Enterobacteriaceae that were not tested.

Please note that performance of the FilmArray BCID Panel for these organisms has not been established.

Table 35. In silico Predictions of Enterobacteriaceae Reactivity
Detection Predicted withReduced SensitivityaDetection Not PredictedUnknown Reactivityb
Brenneria spp.Photorhabdus spp.Buttiauxella spp.
Dickeya spp.Serratia grimesiiEwingella americana
Erwinia spp.Serratia proteamaculansLeminorella spp.
Pectobacterium spp.Xenorhabdus spp.Moellerella spp.
Yersinia spp.

a Predicted result of Enterobacteriaceae Detected when present in a blood culture sample at a concentration of ≥ 1x10 CFU/mL

b Sequence duta not available for in silico reactivity predictions.

Enterobacter cloacae complex

Enterobacter cloacae complex Detected[ $~1\times10^8$ CFU/mL]Enterobacter cloacae complexNot Detecteda
Enterobacter asburiaeATCC 35953Enterobacter nimipressuralisbATCC 9912
Enterobacter cloacae subsp. cloacaeATCC BAA-1143Enterobacter kobeiATCC BAA-260
Enterobacter cloacae subsp. cloacaeATCC 13047
Enterobacter cloacae subsp. cloacaeNCTC 10005
Enterobacter cloacae subsp. cloacaeATCC 49141
Enterobacter cloacaesubsp. dissolvensbATCC 23373
Enterobacter hormaecheiATCC 49162

Table 36. Summary of Enterobacter cloacae complex Inclusivity Results

3 Not Detected at highest test concentration of 1×10th CFU/mL.

Tested as purified nucleic acid at a concentration of 0.63pg/mL (equivalent to ~1×10 CFU/mL). Detected by Enterobacicinaceae assay.

Escherichia coli

Table 37. Escherichia coli Inclusivity Results

Escherichia coli Detecteda
SpeciesIsolate IDStrain Info
Escherichia coliATCC 43888CDC B6914-MS1serotype O157:H7
Escherichia coliATCC 491057482-1-1serotype O15
Escherichia coliATCC 25922FDA-Seattle1946

{31}------------------------------------------------

Escherichia coliATCC 35401H10407serotype O78:H11
Escherichia coliATCC BAA-201Produces ESBL TEM-3

2 Detected at the initial test concentration of 5×107 CFU/mL.

Klebsiella oxytoca

Table 38. Klebsiella oxytoca Inclusivity Results

Klebsiella oxytoca DetectedaKlebsiella oxytoca Not Detectedb
SpeciesIsolate IDStrain InfoSpeciesIsolate IDStrain Info
Klebsiella oxytocaATCC 13182n/aKlebsiella oxytocab,cJMI 10678MY/2011
Klebsiella oxytocaATCC 49131n/a
Klebsiella oxytocaATCC 700324n/a
Klebsiella oxytocaATCC 43086n/a
Klebsiella oxytocaATCC 8724n/a
Klebsiella oxytocaJMI 14611AR/2011
Klebsiella oxytocaJMI12707MA/2011
Klebsiella oxytocaJMI 7818AR/2004
Klebsiella oxytocaJMI 2661NY/2003
Klebsiella oxytocaJMI 2523n/a

4 Detected at the initial test concentration of 5×107 CFU/mL.

b Detected as Enterobacteriaceae at the initial test concentration of 5×10' CFU/mL but Not Detected for Klebsiella anytoca at the highest test concentration of 1×1010 CFU/mL.

Sequence analysis confirmed this isolate as a variant K. oxytoca that will not be detected by the FilmArray BCID Panel Koxytoca assay.

Klebsiella pneumoniae

Table 39. Klebsiella pneumoniae Inclusivity Results

Klebsiella pneumoniae Detecteda
SpeciesIsolate IDStrain Information
Klebsiella pneumoniaeATCC BAA-1706n/a
Klebsiella pneumoniae ssp. pneumoniaeATCC 13883Type strain
Klebsiella pneumoniae ssp. ozaenaeATCC 11296NCTC 5050
Klebsiella pneumoniae ssp. rhinoscleromatisATCC 13884NCTC 5046 Type strain
Klebsiella pneumoniaeATCC 700603n/a
Klebsiella pneumoniaeATCC BAA-1705n/a
Klebsiella pneumoniaeJMI 766n/a
Klebsiella pneumoniaeJMI 328n/a
Klebsiella pneumoniaeJMI 8091n/a
Klebsiella pneumoniaeJMI 438n/a
Klebsiella variicolabATCC BAA-830F2R9/ 2001 Type strain

a Detected at the initial test concentration of 1×10°CFU/mL.

. Identical sequence to K. pneumoniae variant 342. Both K. pneumoniae variant 342 and Klebsiella variicola have been recovered from clinical specimens and will be identified by the BCID Panel and most phenotypic laboratory methods as Klebsiella pneumonice. Proteus

Table 40. Proteus Inclusivity Results

Proteus Detecteda
SpeciesIsolate ID
ATCC 29906
JMI 10793
Proteus mirabilisATCC 25933
ATCC 33583
ATCC 7002
Proteus hauseriATCC 13315

{32}------------------------------------------------

ATCC 700826
Proteus penneriATCC 33519
Proteus vulgarisATCC 33420ATCC 27973

4 Detected at the initial test concentration of 1×107 CFU/mL.

Serratia marcescens

Table 41. Serratia marcescens Inclusivity Results

Serratia marcescens Detectedª
SpeciesIsolate IDStrain Information
Serratia marcescensATCC 13880Type strain
Serratia marcescensATCC 14756n/a
Serratia marcescensATCC 27137n/a
Serratia marcescensATCC 43297n/a
Serratia marcescensJMI 697CT/2009
Serratia marcescensJMI 8089TX/2004

4 Detected at the initial test concentration of 1×10°CFU/mL.

Haemophilus influenzae

Table 42. Huemophilus influenzae Inclusivity Results

Haemophilus influenzae Detectedª
SpeciesIsolate IDStrain Information
Haemophilus influenzaeATCC 33929Non-typeable
Haemophilus influenzaeATCC 51907Non-typeable
Haemophilus influenzae ssp. aegyptusATCC 11116Non-typeable
Haemophilus influenzaeATCC 9006Type a
Haemophilus influenzaeATCC 31512Type b
Haemophilus influenzaeATCC 10211Type b
Haemophilus influenzaeATCC 49699Type c
Haemophilus influenzaeATCC 9008Type d
Haemophilus influenzaeATCC 8142Type e
Haemophilus influenzaeATCC 700223Type f

ª Detected in a positive blood culture tested within 1 hour of positivity. The concentration of H. influenza in a positive blood culture at the time of positivity is estimated to be ~1×10* CFU/mL.

Neisseria meningitidis (encapsulated)

Neisseria meningitidis DetectedaNeisseria meningitidis Not Detectedb
SpeciesIsolate IDSerogroupSpeciesIsolate IDSerogroup
Neisseria meningitidisATCC 43744W135Neisseria meningitidis(unencapsulated)Clinical isolatecNone
Neisseria meningitidisATCC 13077ANeisseria meningitidis(unencapsulated)Clinical isolatecNone
Neisseria meningitidisATCC 13090BNeisseria meningitidis(unencapsulated)Clinical isolatecNone
Neisseria meningitidisATCC 13102CNeisseria meningitidisClinical isolatecNone

Table 43. Neisseria meningitidis Inclusivity Results

{33}------------------------------------------------

Neisseria meningitidis DetectedaNeisseria meningitidis Not Detectedb
SpeciesIsolate IDSerogroupSpeciesIsolate IDSerogroup
(unencapsulated)
Neisseria meningitidisATCC 13113DNeisseria meningitidisClinical isolatedB
Neisseria meningitidisATCC 35561Y

Detected in a seeded blood culture tested within 1 hour of positivity (estimated concentration ~1×10° CFU/mL).

6 Not Detected in a seeded blood culture tested 1-5 hours after positivity.

· Clinical isolates of unencapsulated N. meningitids were tested positive blood cultures to confirm that they would not be detected by the BCID Panel.

4 DNA from a clinical isolate with a variant crA gene was tested at a concentration equivalent to 2.5×10° CFUmL, DNA obtained from University of Lausanne, Institute of Microbiology, Switzerland.

Pseudomonas aeruginosa

Table 44. Pseudomonas aeruginosa Inclusivity Results

Pseudomonas aeruginosa Detectedª
SpeciesIsolate ID
Pseudomonas aeruginosaATCC 27853
Pseudomonas aeruginosaATCC 10145
Pseudomonas aeruginosaATCC 19429
Pseudomonas aeruginosaATCC 25619
Pseudomonas aeruginosaATCC BAA-1744
Pseudomonas aeruginosaATCC 35554

4 Detected at the initial test concentration of 1×10℃FU/mL.

Yeast

Candida albicans

Candida albicans Detecteda
SpeciesIsolate IDStrain Info
Candida albicansATCC 10231Serotype A - 3147
Candida albicansATCC MYA-427A39 [DUMC 136.97]
Candida albicansATCC MYA-2876SC5314
Candida albicansATCC 11651171D
Candida albicansATCC 22972M 97
Candida albicansATCC 90028NCCLS 11

Table 45. Results of Candida albicans Inclusivity Testing

4 Detected at the initial test concentration of 1×10+CFU/mL.

Candida glabrata

Table 46. Results of Candida glabrata Inclusivity Testing

Candida glabrata Detecteda
SpeciesIsolate IDStrain Info
Candida glabrataATCC 15545NRRL YB-4025
Candida glabrataATCC 3255426247-1
Candida glabrataATCC 2001CBS138
Candida glabrataATCC 15126CBS15126
Candida glabrataATCC MYA-2950n/a

{34}------------------------------------------------

4 Detected at the initial test concentration of 1×106 CFU/mL.

Candida krusei

Table 47. Results of Candida krusei Inclusivity Testing
-----------------------------------------------------------
Candida krusei Detecteda
SpeciesIsolate IDStrain Info
Candida kruseiATCC 90878B74
Candida kruseiATCC 20174889-08-008
Candida kruseiATCC 14243n/a
Candida krusei/IssatchenkiaorientalisbATCC 28870CBS 2052
Issatchenkia orientalisbATCC 6258NRRL Y-413

" Detected at the initial test concentration of 1×106 CFU/mL.

b Issatchenkia orientalis and Pichia kudriavzevii are anamorphs of C. krusei,

Candida parapsilosis

Table 48. Results of Candida parapsilosis Inclusivity Testing

Candida parapsilosis DetectedaCandida parapsilosis Detected with Reduced Sensitivityb
SpeciesIsolate IDStrain InfoSpeciesIsolate IDStrain Info
Candida parapsilosisATCC 90875B78Candida parapsilosisATCC 96142MCO462[UTHSC R-648]
Candida parapsilosisATCC 34136ST-89
Candida parapsilosisATCC 96138MCO433
Candida parapsilosisATCC 22019CBS604

Detected at the initial test concentration of 1×10°CFU/mL.

b Detected at a test concentration of 1×107 CFU/mL.

Candida tropicalis

Table 49. Results of Candida tropicalis Inclusivity Testing
Candida tropicalis Detecteda
SpeciesIsolate IDStrain Info
Candida tropicalisATCC 66029AmMS 227
Candida tropicalisATCC 750Type Strain
Candida tropicalisATCC 90874B79
Candida tropicalisATCC MYA-2734508-12.1
Candida tropicalisbATCC 201380API 90 01 105 (Vitek QC)

4 Detected at the initial test concentration of 1×10°CFU/mL.

b Target concentration was 5×105 CFU/mL, final test concentration was 1×106 CFU/mL (2×).

Antimicrobial Resistance Genes

mecA

Table 50. mecA Inclusivity Results

mecA Detecteda,b
SpeciesIsolate IDStrain InformationSCCmec Type
Methicillin-sensitive S. aureus (MSSA) with SCCmec cassette ( mecA positive)
Staphylococcus aureusATCC BAA-2419Mass/2010II

{35}------------------------------------------------

mecA Detecteda,bSCCmecType
SpeciesIsolate IDStrain Information
Staphylococcus aureusATCC BAA-2420Mass/2010II
Staphylococcus aureusATCC BAA-2421Mass/2010II
Methicillin-resistant S. epidermidis (MRSE) (mecA positive)
Staphylococcus epidermidisATCC 29887255-01B
Staphylococcus epidermidis cATCC 51625CCF 15990Unknown
Staphylococcus epidermidisATCC 35984RP62A
Methicillin-resistant S. aureus (MRSA) (mecA positive)
Staphylococcus aureus ssp. aureusATCC BAA-38E2125 DenmarkI
Staphylococcus aureus ssp. aureusATCC 43300F-182 KansasII
Staphylococcus aureus ssp. aureusATCC 700698Mu3 Japan/1996II
Staphylococcus aureus ssp. aureusATCC BAA-1720MRSA252 UKII
Staphylococcus aureusNARSA NRS705NY-12 New York/2005II
Staphylococcus aureusNARSA NRS701MN-082 Minn/2006II
Staphylococcus aureusNARSA NRS648CA-347 California/2005II
Staphylococcus aureus ssp. aureusATCC BAA-39HUSA304 Hungary/1993III 3A&5
Staphylococcus aureusNARSA NRS703MN-095 Minnesota/2006IV
Staphylococcus aureusNARSA NRS683GA-298 Georgia/2005IV
Staphylococcus aureusNARSA NRS662CO-34 Colorado/2005IV
Staphylococcus aureusNARSA NRS707NY-155 New York/2005IV
Staphylococcus aureusATCC BAA-1707MW2 N. Dakota/1998IV
Staphylococcus aureusNARSA NRS691GA-62 Georgia/2005IV
Staphylococcus aureusNARSA NRS689GA-442 Georgia/2006IV
Staphylococcus aureusNARSA NRS668CO-72 Colorado/2005IV
Staphylococcus aureusATCC BAA-174794:1013 Vermont/1993IV
Staphylococcus aureusNARSA NRS676CT-19 Conn/2005IV
Staphylococcus aureusATCC BAA-17647031 AlaskaIV
Staphylococcus aureusATCC BAA-1691HFH-30137 Michigan/2003IV
Staphylococcus aureusATCC BAA-1700HFH-33798 Illinois/2004IV
Staphylococcus aureus ssp. aureusATCC BAA-1717TCH1516 TexasIVa
Staphylococcus aureusNARSA NRS745CA-629 California/2006V
Staphylococcus aureus ssp. aureusATCC BAA-42HDE288 Portugal/1996VI
Methicillin-resistant S. aureus with mecALGA251/mecC variant
Staphylococcus aureusATCC BAA-2312M10/0061 Ireland/2010XI
Staphylococcus aureusATCC BAA-2313M10/0148 Ireland/2010XI

4 Detected at the initial test concentration of 5×10℃CFU/mL.

" Staphylococcus Detected and/or Staphylococcus aureus Detected results also reported, as appropriate.
" Initial test concentration was 5×10° CFU/mL.

"Initial test concentration was 5×10² CFU/mL.

yan A/B

Table 51. vanA/B Inclusivity Results

vanA/B Detecteda,b
SpeciesIsolate IDStrain.Information
Enterococcus faecium [vanA]JMI 536TX/2006
Enterococcus faecium [vanA]ATCC 700221Connecticut
Enterococcus faecium [vanA]JMI 475IN/2003
Enterococcus faecalis [vanA]JMI 12536Mass/2002

{36}------------------------------------------------

vanA/B Detecteda,b
SpeciesIsolate IDStrain Information
Enterococcus faecalis [vanB]ATCC 51299Missouri
Enterococcus faecalis [vanB]ATCC 700802Missouri/1987
Enterococcus faecalis [vanB]JMI 368VA/2003

a Detected at the initial test concentration of 1×10° CFU/mL.

b Enterococcus Detected results also reported.

КРС

Table 52. KPC Inclusivity Results

KPC Detecteda,bSpeciescIsolate IDKPC TypeStrain Information
Enterobacter cloacaeBAA-2341Unknown1101152
Enterobacter hormaecheiBAA-2082Unknownn/a
Escherichia coliBAA-2340Unknown1101362
Klebsiella oxytocaJMI 2523Unknownn/a
Escherichia coliClinical IsolateKPC-2n/a
Enterobacter cloacaeClinical IsolateKPC-2n/a
Klebsiella oxytocaJMI 7818KPC-2AR/2004
Klebsiella pneumoniaeJMI 328KPC-2n/a
Klebsiella pneumoniaeATCC BAA-1705KPC-2Modified Hodge Test Control
Serratia marcescensJMI 697KPC-2CT/2009
Enterobacter cloacaeClinical IsolateKPC-3n/a
Klebsiella oxytocaJMI 2661KPC-3NY/2003
Klebsiella pneumoniaeJMI 766KPC-4n/a
Klebsiella pneumoniaeJMI 8091KPC-4n/a
Klebsiella pneumoniaeJMI 438KPC-4n/a

3 Detected at the initial test concentration of 5×10° CFU/mL for K. oxyloca isolates and 1×10°CFU/mL for K. pneumonice and S marcescens isolates. Detected in a seeded blood culture tested within 1 hour of positivity for Enterobacter spp. and E. coli. 6 Enterobacteriaceae and corresponding species specific Detected results also reported.

^ Other isolates which carry the KPC gene (i.e. Acineobacer baumamii, Pseudomonas aeruginosa, and Entervacere other than those listed above) were not evaluated.

Exclusivity

The potential for cross-reactivity with BCID Panel assays was evaluated by testing high concentrations of organism in contrived or seeded blood culture samples. The test concentration was equal to or greater than the level of organism estimated to be in a blood culture sample 8 hours after positivity (approximately 10°-10" CFU/mL for bacteria and 10'-10° CFU/mL for yeast), or the highest concentration possible based on the organism stock. The selection of organisms focused on species that may be found in positive blood cultures (clinically relevant) and/or those that are closely related to target organisms (nearest neighbors). Organisms were also selected based on antimicrobial resistance phenotypes and the presence or absence of the antimicrobial resistance genes identified by the BCID Panel. The tested organisms were divided into two categories: onpanel organisms and off-panel organisms.

On-panel organisms were tested to verify that they only react with the appropriate assays on the panel. On-panel exclusivity testing included gram-positive bacteria, gram-negative bacteria and yeast, representing 29 genera and 98 individual species.

Off-panel organisms were expected to have negative test results for all of the assays on the FilmArray BCID Panel (or positive organism results but negative results for the

{37}------------------------------------------------

antimicrobial resistance genes detected by the FilmArray BCID Panel). Off-panel testing included gram-positive bacteria, gram-negative bacteria, yeast, viruses, and Mycoplasmataceae.

Results are presented for all organisms that were tested and received the expected FilmArray BCID Panel test result(s) (no cross-reactivity, Tables 53 - 57), followed by a summary of species or isolates with which cross-reactivity was observed (Table 59).

ON PANEL
Enterococcus SpeciesStaphylococcus aureusCoagulase-NegativeStaphylococciStreptococcus Species
E. aviumE. casseliflavus(2 isolates)E. cecorumE. disparE. duransE. faecalis (3 isolates)E. faceium (2 isolates)aE. gallinarium(2 isolates)E. hiraeE. raffinosusListeria monocytogenesL. monocytogenesMSSA (18 isolates)Resistant S. aureus -BORSA (6 isolates)MRSA (mecA)VRSA (mecA, vanA)Coagulase-PositiveStaphylococciS. intermediusS. lutraeS. pseudointermediusS. schleiferi ssp.coagulansS. capitis ssp. capitisS. capraeS. cohniiS. epidermidis(2 isolates)S. haemolyticusS. hominisS. lugdunensisS. pasteuriS. saprophyticusS. schleiferi ssp.schleiferiS. sciuriS. warneriS. xylosusS. agalactiaeS. anginosusS. bovisS. dysgalactiaeS. gallolyticusS. mitisS. mutansS. parasanguinisS. pneumoniaeS. pseudopneumoniaeS. pyogenesS. salivarius
OFF PANEL
Gram-positive CocciGram-positive BacilliListeria SpeciesGram-positiveAnaerobes
GranulicatellaadiacensbGemella morbillorumLactococcus lactisMacrococcuscaseolyticusMicrococcus luteusVagococcus fluvialisActinomycesodontolyticusBacillus cereusCorynebacteriumjeikeiumLactobacillusacidophilusMycobacteriumtuberculosiscRhodococcus equiRothia mucilaginosaL. (murrayi) grayiL. innocuadL. ivanoviissp. londoniensisL. seeligeriL. welshimeriClostridium perfringensPeptostreptococcusanaerobiusPropionibacteriumacnes

Table 53. Non-Cross-Reactive Gram Positive Bacteria

Abbreviations: BORSA=Borderline-oxacillin resistant Staphylococcus aureus; MRSA=Methicillin Resistant Stay/hylococcus arreus; VRSA=Vancomycin Resistant Staphylococcus aureus.

"One isolate was tested at a concentration of Sng/mL Extracted DNA: ~ 1×10€ CFU/mL,

  • A false positive Sreptococcus result was observed in the initial test of this isolate. The expected negative results were observed in multiple subsequent tests. No cross-reactivity between and the BCID Panel Streptococcus assays is predicted by sequence analysis.

C Tested at a concentration of 7.33×106 CFU/mL.

4 In silico analysis predicts that cross-reactivity between the Lmonocytogenes assay and some atypical strains of L. innocwa is possible, however, no cross-reactivity was observed in this testing,

NOTE: While not observed in this study, cross-reactivity between the Enterococcus assay and some Staphylococcus species may occur when the staphylococci are present in a sample at very high concentrations.

{38}------------------------------------------------

Table 54. Non-Cross-Reactive Gram Negative Bacteria
ON PANEL
AcinetobacterbaumanniiEnterobacteriaceae Isolatesa
A. baumannii(2 isolates)Cedeceae davisiaeCitrobacter fruendiEscherichia hermaniiEscherichia vulnerisProvidencia acalifaciensProvidencia rettgeri
HaemophilusinfluenzaeCitrobacter koseriCronobacter muytjensiCronobacter sakazakiiHafnia alveiKlebsiella oxytoca(3 isolates)Providencia stuartiRahnella aquatilisRaoultella terrigena
H. influenzae (type b)Enterobacter amnigenusEnterobacter asburiaeKlebsiella pneumoniae(6 isolates)Raoultella planticolaSalmonella enterica
Neisseria meningitidisEnterobactercancerogenusKluyvera ascorbataKluyvera intermediusSerratia liquefaciensSerratia fonticola
N. meningitidisEnterobacter cloacaeEnterobacterhormaecheiLeclerciaadecarboxylataMorganella morganiiSerratia marcescens(2 isolates)Serratia plymuthica
PseudomonasaeruginosaEnterobacter gergoviaeEscherichia coli(2 isolates)Pantoea agglomeransbProteus mirabilisTatumella ptyseosYersinia enterocolitica
P. aeruginosaProteus penneriProteus vulgarisYokenella regensburgei
OFF PANEL
Acinetobacter SpeciesHaemophilus SpeciesPseudomonas SpeciesGram-negative Bacilli
A. calcoaceticusA. haemolyticusA. johnsoniiA. juniiA. IwoffiiA. radioresistensA. schindleriA. ursingiiA. nosocomialis(genomospecies13TU; 2 isolates )H. parahaemolyticusH. parainfluenzaeH. parasuisH. somnusP. fluorescensP. luteolaP. nitroreducensP. oryzihabitansP. pertucinogenaP. stutzeriAeromonas hydrophilaBrevundimonasdiminutaMoraxella catarrhalis(3 isolates)StenotrophomonasmaltophilaVibrio parahaemolyticus
Neisseria SpeciesN. siccaN. elongateN. perflavaN. mucosaN. lactamicaGram-negativeAnaerobesGram-negativeCoccobacilli
Bacteroides fragilisVeillonella parvulaBordetella pertussisCampylobacter fetusChlamydia trachomatisLegionellapneumophilia

^a Some isolates were not detected by the FilmArray BCID Panel and are discussed in the inclusivity evaluation.

ª Some isolates were not detected by the FilmArray BCID Panel and anchisvity evaluation.

  • In silico analysis indicates that cross-reactivity between the Enterebacter effecti agglomerans may be possible. However, no cross-reactivity was observed in this study.

€ Tested at a concentration of 2.63×10*CFU/mL.

ON PANELOFF PANEL
Candida SpeciesCandida SpeciesNon-Candida Fungi
C. albicansC. dubliniensisC. sojaeAspergillus fumigatus
C. glabrataC. lusitaniaeC. viswanathiiDebaryomyces hansenii
C. kruseiC. metapsilosisC. guilliermondiiKluyveromyces lactis
C. parapsilosisC. multigemmisaSaccharomycescerevisiae
C. tropicalisSchizosaccharomycespombe

Table 55. Non-Cross-Reactive Fungi

{39}------------------------------------------------

4 In silico analysis predicts that cross-reacivity between the Cparapsilosis assay and C. multigemmis is possible, however, no crossreactivity was observed in this testing.

OFF PANEL
Mycoplasmataceae IsolatesViruses
Mycoplasma hominis ( $3.16 \times 10^7$ CFU/mL)Cytomegalovirus ( $1.67 \times 10^4$ TCID $_{50}$ /mL)
Ureaplasma urealyticum ( $1.57 \times 10^6$ CFU/mL)Epstein Barr Virus ( $1.00 \times 10^5$ TCID $_{50}$ /mL)
Herpes Simplex Virus - Type 1 (1:30 dilution of stock)
Varicella Zoster Virus ( $8.17 \times 10^3$ TCID $_{50}$ /mL)

.

Table 56. Non-Cross-Reactive Viruses and Mycoplasmataceae

Table 57. Non-Cross-Reactive Antimicrobial Resistance

ON PANELOFF PANEL®
mec A
Methicillin Resistant Staphylococci (mecA)Borderline Oxacillin Resistant S. aureus(BORSA)
Staphylococcus epidermidis-MRSEmecAmecAmecAlvanAStaphylococcus aureus-BORSA (6 isolates)
Staphylococcus aureus-MRSAMethicillin Sensitive Staphylococci
Staphylococcus aureus-VRSAStaphylococcus aureus-MSSA (18 isolates)Staphylococcus epidermidis-MRSE (1 isolate)Staphylococcus spp. (16 isolates)
van A/B
Vancomycin Resistant Enterococci (vanA/B)Vancomycin Resistant Enterococci (non-vanA/B)
Enterococcus faecalisEnterococcus faeciumvanBvanAEnterococcus casseliflavusEnterococcus casseliflavusEnterococcus gallinariumEnterococcus gallinariumvanCvanCvanCvanC
Vancomycin Sensitive Enterococci
Enterococcus spp. (8 isolates)
KPC
Carbapenem Resistant Enterobacteriaceae (KPC)Carbapenem Resistant Enterobacteriaceae (non-KPC)
Klebsiella oxytocaKlebsiella pneumoniaeKPC-2KPC-4Klebsiella pneumoniaeKlebsiella pneumoniaeUnknownNDM
Serratia marcescensKPC-2Carbapenem Sensitive/Beta-lactam ResistantIsolates
Klebsiella pneumoniaeKlebsiella pneumoniaeEscherichia coliAcinetobacter baumanniiMoraxella catarrhalisMoraxella catarrhalisCarbapenem Sensitive IsolatesAmpCSHVTEM-3/CTX-1blaOXAblaOXABRO-1(bla)/orf3
Enterobacteriaceae (51 isolates)Acinetobacter baumannii (1 isolate)Pseudomonas aeruginosa (2 isolates)

ª Off-panel refers to the antimicrobial resistance gene. Organisms may be positive for organism assay(s).

b Ten of the 18 isolates are known to harbor remnants of SCCmec cassette.

{40}------------------------------------------------

Table 58. Predicted and Observed Cross-Reactivity Between BCID Panel Assays and On-Panel or
Off-Panel Organisms Tested for Exclusivity
BCID Panel Assay/ResultCross-Reactive Organism(s)/Isolate(s)/Gene
Gram-positive Bacteria
EnterococcusSome coagulase-negative staphylococcia
Gram-negative Bacteria
Acinetobacter baumanniiAcinetobacter calcoaceticus-baumannii (ACB) complex species:Acinetobacter calcoaceticus (ssp. anitratus)bAcinetobacter pittii (formerly genomospecies 3)b
Escherichia coli/EnterobacteriaceaeShigella species:Shigella boydiiShigella dysenteriaeShigella flexneriShigella sonneiEscherichia fergusonnii
Klebsiella pneumoniae/EnterobacteriaceaeKlebsiella variicola (or Klebsiella pneumoniae variant 342)Enterobacter aerogenesRaoultella ornithinolyticac
Serratia marcescens/EnterobacteriaceaeSerratia species (Serratia entomophilae, Serratia ficaria, Serratiaodoriferad, Serratia rubidaead)Raoultella ornithinolyticacPseudomonas aeruginosa (ATCC 25619)fPseudomonas putidaf
Haemophilus influenzaeHaemophilus haemolyticusg
Yeast
Candida parapsilosisCandida orthopsilosis (Group III Candida parapsilosis)h
Antimicrobial Resistance Genes
vanA/BvanMi
  • Cross-reactivity was not observed in this study but is predicted by in silico analysis to occur only with some species (i.e. S epidermis, S. capitis and S. haemolyticus) when present in a sample at very high levels. This cross-cacivity was observed infrequently in pre-analytical studies and the clinited occurrence of -0.2% of all prospective patient samples).

b Acinelobacer calcoaceticus-baunanni (ACB) complex species are often mis-identified as A. baumamii by autonuslus microbial identification methods.

SCross-cactivity was not observed when ATCC 31898 was tested in the inclusivity study at a concentration ~1x10° CFU/mL, but cross-reactivity was observed in clinical cultures containing R. ornithinolytica.

d Cross-reactivity was observed only at high organism concentration (≥10° CFU/mL); rare human pathogens.

e Pseudomonas putida is a rare opportunistic pathogen.

4 No cross-reactivity observed with five other Pseudomonas aeruginosa isolates tested at ≥10° CFU/mL.

8 Haemophilus haemolyticus is a commensal organism of the respiratory tract that is rarely isolated from blood culture.

  • Candida orthopsilosis is mis-identified as C. parapsilosis by automated and manual microbial identification methods.

| Vancomycin-resistant Enterococcus faecium isolated in Asia, 2011; vanB resistance phenotype.

Reproducibility

A multicenter reproducibility study was performed to determine between-site and overall reproducibility of the BCID Panel. Reproducibility testing occurred at three test sites using a panel of six simulated blood culture specimens, each spiked with various

{41}------------------------------------------------

combinations of two different organisms (analytes). To best represent the composition of specimens likely to be tested by the BCID Panel, half of the analytes were at a concentration consistent with the level of organism in a blood culture bottle at the time of positivity, and half of the analytes were at a concentration similar to that observed in bottles eight hours after positivity (see Growth and Detection above). Negative results for each assay were obtained from samples that were not spiked with a corresponding organism (analyte not in the sample).

The data incorporate a range of potential variation introduced by seven different operators, three different pouch lots, and ten different FilmArray Instruments. Every specimen was tested on eight different days, for a total of 90 replicates per analyte. A summary of results (percent (%) agreement with the expected result) for each analyte is provided in the following tables.

BCID PanelTest ResultOrganism TestedTest ConcentrationTest SiteResults% AgreementwithExpected Result
EnterococcusEnterococcus faecium [ vanA ]JMI4751.50E+08 CFU/mLSite A30/300/30180/180100%[98.0% - 100%]
Site B30/300/30
Site C30/300/30
All Sites90/900/90
EnterococcusEnterococcus faecalis [ vanB ]JMI 3688.95E+08 CFU/mLSite A30/300/30
Site B30/300/30
Site C30/300/30
All Sites90/900/90
NegativeSite A0/120120/120360/360100%[99.0%-100%]
Site B0/120120/120
Site C0/120120/120
All Sites0/360360/360
ListeriamonocytogenesNegativeSite A0/180180/180540/540100%[99.3%-100%]
Site B0/180180/180
Site C0/180180/180
All Sites0/540540/540
StaphylococcusStaphylococcus aureus[MRSA]ATCC BAA-17478.60E+06 CFU/mLSite A30/300/3090/90100%[96.0%-100%]
Site B30/300/30
Site C30/300/30
All Sites90/900/90
NegativeSite A0/150150/150449/450a99.8%[98.8%-100%]
Site B1/150a149/150
Site C0/150150/150
All Sites1/450449/450
StaphylococcusaureusStaphylococcus aureus[MRSA] ATCC BAA-17478.60E+06 CFU/mLSite A30/300/3090/90100%[96.0% -100%]
Site B30/300/30
Site C30/300/30
All Sites90/900/90
NegativeSite A0/150150/150450/450100%[99.2%-100%]
Site B0/150150/150
Site C0/150150/150
All Sites0/450450/450
StreptococcusStreptococcus pyogenesATCC 196155.70E+08 CFU/mLSite A30/300/3090/90100%[96.0%-100%]
Site B30/300/30
Site C30/300/30
All Sites90/900/90
NegativeSite A0/150150/150450/450100%[99.2% - 100%]
Site B0/150150/150
Site C0/150150/150
All Sites0/450450/450
StreptococcusagalactiaeNegativeSite A0/180180/180540/540100%[99.3%-100%]
Site B0/180180/180
Site C0/180180/180

Table 59. Summary of Reproducibility Results -- Organism Assays

{42}------------------------------------------------

BCID PanelTest ResultOrganism TestedTest ConcentrationTest SiteResults% AgreementwithExpected Result
StreptococcuspneumoniaeNegativeAll Sites0/540540/540
Site A0/180180/180540/540
NegativeSite B0/180180/180100%
Site C0/180180/180[99.3% - 100%]
All Sites0/540540/540
Streptococcus pyogenesATCC 196155.70E+08 CFU/mLSite A30/300/3090/90
StreptococcuspyogenesSite B30/300/30100%
Site C30/300/30[96.0% - 100%]
All Sites90/900/90
NegativeSite A0/150150/150450/450
Site B0/150150/150100%
Site C0/150150/150[99.2% - 100%]
All Sites0/450450/450
Acinetobacter baumanniiATCC 99552.00E+08 CFU/mLSite A30/300/3090/90
AcinetobacterbaumanniiSite B30/300/30100%
Site C30/300/30[96.0% - 100%]
All Sites90/900/90
NegativeSite A0/150150/150450/450
Site B0/150150/150100%
Site C0/150150/150[99.2% - 100%]
All Sites0/450450/450
Klebsiella pneumoniae [KPC]JMI 7669.40E+08 CFU/mLSite A30/300/30
Site B30/300/30
Site C30/300/30180/180
All Sites90/900/90100%
EnterobacteriaceaeProteus mirabilisATCC 299069.20E+08 CFU/mLSite A30/300/30[98.0% - 100%]
Site B30/300/30
Site C30/300/30
All Sites90/900/90
NegativeSite A0/120120/120
Site B0/120120/120360/360
Site C0/120120/120100%
All Sites0/360360/360[99.0% - 100%]
Enterobactercloacae complexNegativeSite A0/180180/180540/540
Site B0/180180/180100%
Site C0/180180/180[99.3%-100%]
All Sites0/540540/540
Escherichia coliNegativeSite A0/180180/180540/540
Site B0/180180/180100%
Site C0/180180/180[99.3% - 100%]
All Sites0/540540/540
Klebsiella oxytocaNegativeSite A0/180180/180540/540
Site B0/180180/180100%
Site C0/180180/180[99.3%-100%]
All Sites0/540540/540
Klebsiella pneumoniae [KPC]JMI 7669.40E+08 CFU/mLSite A30/300/30
KlebsiellapneumoniaeSite B30/300/3090/90
Site C30/300/30100%[96.0% - 100%]
All Sites90/900/90
NegativeSite A0/150150/150450/450
Site B0/150150/150100%
Site C0/150150/150[99.2%-100%]
All Sites0/450450/450
Proteus mirabilisATCC 299069.20E+08 CFU/mLSite A30/300/3090/90
ProteusSite B30/300/30100%
Site C30/300/30[96.0%-100%]
All Sites90/900/90
NegativeSite A0/150150/150450/450
Site B0/150150/150100%
Site CAll Sites0/1500/450150/150450/450[99.2%-100%]
SerratiamarcescensNegativeSite A0/180180/180540/540
Site B0/180180/180100%

:

.

:

:

{43}------------------------------------------------

BCID PanelTest ResultOrganism TestedTest ConcentrationTest SiteResults% AgreementwithExpected Result
Site C0/180180/180[99.3% - 100%]
All Sites0/540540/540
Site A0/180180/180
HaemophilusinfluenzaeNegativeSite B1/180 *179/180539/540 *
Site C0/180180/18098.0%
All Sites1/540539/540[99.0% - 100%]
Site A0/180180/180
NeisseriameningitidisNegativeSite B0/180180/180540/540
Site C0/180180/180100%
All Sites0/540540/540[99.3% - 100%]
Site A30/300/30
Pseudomonas aeruginosaATCC 27853Site B30/300/3090/90
1.40E+08 CFU/mLSite C30/300/30100%
PseudomonasaeruginosaAll Sites90/900/90[96.0% - 100%]
Site A0/150150/150
NegativeSite B0/150150/150450/450
Site C0/150150/150100%
All Sites0/450450/450[99.2% - 100%]
Site A30/300/30
Candida albicansATCC 10231Site B30/300/3090/90
3.10E+04Site C30/300/30100%
Candida albicansAll Sites90/900/90[96.0% - 100%]
NegativeSite A0/150150/150
Site B0/150150/150450/450
Site C0/150150/150100%
All Sites0/450450/450[99.2% - 100%]
Candida glabrataATCC 15545Site A30/300/30
Site B30/300/3090/90
Site C30/300/30100%
All Sites90/900/90[96.0% - 100%]
Candida glabrata2.00E+07NegativeSite A0/150150/150
Site B0/150150/150450/450
Site C0/150150/150100%
All Sites0/450450/450[99.2% - 100%]
Candida kruseiATCC 90878Site A30/300/30
Site B30/300/3090/90
Site C30/300/30100%
All Sites90/900/90[96.0% - 100%]
Candida krusei3.20E+07NegativeSite A0/150150/150
Site B0/150150/150450/450
Site C0/150150/150100%
All Sites0/450450/450[99.2% - 100%]
NegativeSite A0/180180/180
CandidaparapsilosisSite B1/180 *179/180539/540 *
Site C0/180180/18099.8%
All Sites1/540539/540[99.0% - 100%]
Candida tropicalisATCC 66029Site A30/300/30
Site B30/300/3090/90
Site C30/300/30100%
All Sites90/900/90[96.0% - 100%]
Candida tropicalis9.70E+05NegativeSite A0/150150/150
Site B0/150150/150450/450
Site C0/150150/150100%
All Sites0/450450/450[99.2% - 100%]

" A single pouch run at Site B generated four false positive results: Staphylococcus, mecA (see below), Haemophilus influenzae, and Candida parapsilosis.

Table 60. Summary of Reproducibility Results - Antimicrobial Resistance Gene Assays

BCID PanelTest ResultOrganism TestedTest ConcentrationTest SiteResults% Agreementwith ExpectedTest Result
DetectedNotDetectedN/A
vanA/BEnterococcus faecium [vanA]Site A30/300/300/30180/180

{44}------------------------------------------------

Site B30/300/300/30100%
JMI4751.50E+08 CFU/mLSite C30/300/300/30[98.0% - 100%]
All Sites90/900/900/90
Enterococcus faecalis [vanB]JMI 368Site A30/300/300/30
8.95E+08 CFU/mLSite B30/300/300/30
Site C30/300/300/30
All Sites90/900/900/90
Site A0/1200/120120/120
No Associated OrganismSite B0/1200/120120/120360/360
Site C0/1200/120120/120100%
[99.0% - 100%]
All Sites0/3600/360360/360
Staphylococcus aureus[MRSA]ATCC BAA-Site A30/300/300/30
1747Site B30/300/300/3090/90
mecA8.60E+06 CFU/mLSite C30/300/300/30100%
All Sites90/900/900/90[96.0% - 100%]
Site A0/1500/150150/150
No Associated OrganismSite B1/150 "0/150149/150449/450"
Site C0/1500/150150/15099.8%
All Sites1/4500/450449/450[98.8% - 100%]
Klebsiella pneumoniae [KPC]JMI 766Site A30/300/300/30
9.40E+08Site B30/300/300/3090/90
KPCSite C30/300/300/30100%
All Sites90/900/900/90[96.0% - 100%]
Proteus mirabilis ATCCSite A0/9090/900/90
29906 .Site B0/9090/900/90270/270
andSite C0/9090/900/90100%
Pseudomonas aeruginosaATCC 27853All Sites0/270270/2700/270[98.6% - 100%]
Site A0/600/6060/60
Site B0/600/6060/60180/180
No Associated OrganismSite C0/600/6060/60100%
All Sites0/1800/180180/180[98.0% - 100%]

² A single pouch run at Site B generated a false positive mecA result.

,

The reproducibility of Tm for each analyte and positive assay was also evaluated and a summary is provided in the following tables.

.

Organism TestedTest ConcentrationReproducibility of Tm
BCID PanelAssayTest SiteTm MeanTm Std DevTm MinTm MaxRange(max-min)
Gram-Positive Bacteria
EnterococcusEnterococcus faecium [vanA]JMI4751.50E+08 CFU/mLSite A82.50.481.984.02.1
Site B82.6-0.282.383.00.7
Site C82.30.281.982.80.9
All Sites82.50.381.984.02.1
EnterococcusEnterococcus faecalis [vanB]JMI 3688.95E+08 CFU/mLSite A82.00.381.582.40.9
Site B82.20.281.882.81.0
Site C81.60.481.082.41.4
All Sites81.90.481.082.81.8
SaureusStaphylococcus aureus [MRSA]Site A77.10.376.677.81.2
Reproducibility of Tm
BCID PanelAssayOrganism TestedTest ConcentrationTest SiteTm MeanTm Std DevTm MinTm MaxRange(max-min)
ATCC BAA-17478.60E+06 CFU/mLSite B77.30.376.877.81.0
Site C76.90.276.577.51.0
All Sites77.10.376.577.81.3
Streptococcus pyogenesATCC 19615Site A81.90.481.583.62.1
Streptococcus5.70E+08 CFU/mLSite B82.10.181.882.30.5
Site C81.80.281.582.10.6
All Sites81.90.381.583.62.1
Streptococcus pyogenesATCC 19615Site A79.00.478.579.81.3
Spyogenes5.70E+08 CFU/mLSite B79.20.378.779.81.1
Site C78.80.378.579.51.0
All Sites79.00.378.579.81.3
Gram-Negative Bacteria
Site A80.60.480.081.21.2
AbaumanniiAcinetobacter baumanniiATCC 9955Site B80.80.280.481.20.8
2.00E+08 CFU/mLSite C80.30.479.580.91.4
All Sites80.50.479.581.21.7
Site A88.60.388.189.11.0
EntericKlebsiella pneumoniae [KPC]JMI 766Site B88.80.188.689.20.5
9.40E+08 CFU/mLSite C88.30.387.888.81.0
All Sites88.60.387.889.21.4
Site A87.90.387.388.51.2
KpneumoniaeKlebsiella pneumoniae [KPC]JMI 7669.40E+08 CFU/mLSite B88.10.287.888.40.6
Site C87.60.386.788.11.5
All Sites87.80.486.788.51.8
ProteusProteus mirabilisATCC 29906Site A81.20.380.681.81.2
Site B81.40.281.281.90.7
9.20E+08 CFU/mLSite C81.20.280.781.60.9
All Sites81.30.380.681.91.2
Site A87.90.387.388.51.2
PaeruginosaPseudomonas aeruginosaATCC 27853Site B88.20.387.889.51.7
1.40E+08 CFU/mLSite C88.50.288.189.11.0
All Sites88.20.487.389.52.2
Yeast
Site A79.80.379.380.31.0
CalbicansCandida albicansATCC 10231Site B80.10.279.780.50.8
3.10E+04Site C79.50.378.980.21.3
All Sites79.80.478.980.51.7
Site A75.30.374.776.11.3
CglabrataCandida glabrataATCC 15545Site B75.40.374.976.41.5
2.00E+07Site C75.70.275.476.10.7
All Sites75.50.374.776.41.7
Candida kruseiATCC 90878Site A84.50.484.185.21.2
Ckrusei3.20E+07Site B84.70.384.385.31.1
Site C85.00.384.685.81.3
All Sites84.80.484.185.81.8
CtropicalisCandida tropicalisATCC 66029Site A79.10.378.680.11.6
9.70E+05Site B79.20.278.879.60.8
Site C79.50.279.380.00.7
Reproducibility of Tm
BCID PanelAssayOrganism TestedTest ConcentrationTest SiteTm MeanTm Std DevTmMinTmMaxRange(max-min)
All Sites79.30.378.680.11.6

Table 61. Summary of Tm Analysis for Positive Organism Assays

{45}------------------------------------------------

{46}------------------------------------------------

Table 62. Summary of Tm Analysis for Positive Antimicrobial Resistance Gene Assays

Reproducibility of Tm
BCID PanelTest ResultOrganism TestedTest ConcentrationTest SiteTm MeanTm Std DevTmMinTmMaxRange(max-min)
vanA/BEnterococcus faecium [vanA]JMI4751.50E+08 CFU/mLSite A85.70.485.186.71.6
Site B86.00.385.586.51.0
Site C85.60.385.186.31.2
All Sites85.70.485.186.71.6
vanA/BEnterococcus faecalis [vanB]JMI 3688.95E+08 CFU/mLSite A86.00.385.386.61.3
Site B86.30.285.986.91.0
Site C85.70.485.186.61.5
All Sites86.00.485.186.91.8
mecAStaphylococcus aureus [MRSA]ATCC BAA-17478.60E+06 CFU/mLSite A73.60.373.174.41.3
Site B73.70.373.274.21.0
Site C73.40.372.874.11.3
All Sites73.60.372.874.41.7
KPCKlebsiella pneumoniae [KPC]JMI 7669.40E+08Site A86.20.385.586.71.1
Site B86.40.286.186.70.5
Site C85.90.385.286.41.2
All Sites86.10.485.286.71.5

Interference

Substances that could be present in blood culture samples or introduced during sample handling were evaluated for their potential to interfere with assay performance. A potentially interfering substance (see Table 63) was added to a simulated positive acrobic blood culture sample which contained simulated blood culture matrix (human whole blood that had been incubated in a blood culture bottle) and one of six different organism mixes. Each organism mix contained two live pathogens at a concentration equivalent to the level determined to be present when a blood culture bottle is detected as positive by the blood culture instrument. None of the substances tested were found to compete or interfere with the assays in the BCID Panel.

Endogenous SubstancesExogenous SubstancesTechnique-SpecificSubstances
HemoglobinFluconazoleCeftriaxoneBleach
TriglyceridesVancomycinTetracyclineEthanol
BilirubinCiprofloxacinAmoxicillin/Clavulanate
γ-globulinGentamicin sulfateHeparin
HumanImipenemSodium Polyanetholesulfonate (SPS)
Genomic DNA
On-Panel Competing MicroorganismsOff-Panel Competing Microorganisms

Table 63. Potentially Interfering Substances

{47}------------------------------------------------

Staphylococcus epidermidisCorynebacterium jeikeium
Escherichia coliBacillus cereus
Streptococcus mitisMicrococcus luteus
Clostridium perfringens
Propionibacterium acnes
Blood Culture Media/Bottle Types
BACTEC Plus Aerobic/FBacT/ALERT SA Standard AerobicVersaTREK REDOX 1
BACTEC Standard AerobicBacT/ALERT SN Standard AnaerobicVersaTREK REDOX 2
BACTEC Standard AnaerobicBacT/ALERT FA Aerobic FAN
BACTEC Plus Anaerobic/FBacT/ALERT FN Anaerobic FAN
BACTEC Pediatric PlusBacT/ALERT PF Pediatric FAN
BACTEC Lytic/10 Anaerobic/FBacT/Alert FA Plus Aerobic

Note: While not shown to interfere in this evaluation, the BacT/ALERT blood culture bottles that contain charcoal have the potential to generate false positive results presence of nucleic acids from non-viable organisms and are listed as contraindicated for use with the FilmArray BCID Panel.

.

:

{48}------------------------------------------------

Image /page/48/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a symbol that resembles a stylized caduceus, which is a symbol often associated with healthcare and medicine. The symbol consists of a staff with two snakes coiled around it.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 21 2013

BETH LINGENFELTER, M.S. BIOFIRE DIAGNOSTICS, INC. 390 WAKARA WAY · SALT LAKE CITY UT 84108

Re: K130914

Trade/Device Name: FilmArray Blood Culture Identification Panel Regulation Number: 21 CFR 866.3365 Regulation Name: Multiplex Nucleic Acid Assay for Identification of Microorganisms and

Resistance Markers from Positive Blood Cultures

Regulatory Class: II Product Code: PAM, PEO, PEN, OOI Dated: March 30, 2013 Received: April 02, 2013

Dear Ms. Lingenfelter:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosal of the enactment date of the Medical Device Amendments, or to conninered prices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Act (71ct) that do not 10, subject to the general controls provisions of the Act. The I ou may, merciolo, manes of the Act include requirements for annual registration, listing of general controls providion practice, labeling, and prohibitions against misbranding and actives, good manufacturing probles, as only and information related to contract liability addition. Thease noter u, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classimod too above) . Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{49}------------------------------------------------

Page 2-Ms. Lingenfelter

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer 809), prease contact the Division of Sman Manatavarons, 196-7100 or at its Internet address Assistance at its lon-free number (600) 050 2017 07 (0) (2) [2] [2] [2] [2] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] 0 ] http://www.rda.gov/MedicalDcvrccs/Resourcesor o premarket notification" (21CFR Part the regulation entitled, Misoranomy of reference to premains wonest works.
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office http://www.ida.govetrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the You may obtain other general information on your respected with the sistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (800) 038-2041 Of (501)icalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sally A. Hojvat -S

Sally Hojvat, Ph.D., M.Sc Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{50}------------------------------------------------

Indications for Use

K130914 510(k) Number (if known): Device Name: FilmArray Blood Culture Identification (BCID) Panel

The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with the FilmArray Instrument. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrates the presence of organisms as determined by Gram stain.

The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, commonly encountered Staphylococci (including specific differentiation of Staphylococcus aureus), commonly encountered Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, commonly encountered Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated). Pseudomonas aeruginosa, Candida albicans, Candida glabrata. Candida krusei. Candida parapsilosis, and Candida tropicalis.

The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blagg) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.

FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia.

Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some Staphylococci, Enterococci, Streptococci, and Enterobacteriaceae that are not specifically identified by the FilmArray BCID Panel assays.

{51}------------------------------------------------

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE------------------------------------------------------------------------------------------------------------------------------------------ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Image /page/51/Picture/5 description: The image shows the text "John Hobson -S 2013:06.19 16:40:22 -04'00'". The text is in a sans-serif font and is black. The text is slightly blurred. The background of the image is white.

A.L. Wetz
Division Sign-Off

Office of in Vitro Diagnostics and Radiotogical Health

...... . .

510(K)_K (309) T

§ 866.3365 Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

(a)
Identification. A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.(b)
Classification. Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).