K Number
K132777

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-09-26

(21 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Device Description

The NanoCross catheter is an over-the-wire (OTW) coaxial lumen percutaneous transluminal angioplasty (PTA) balloon catheter compatible with 0.014"guidewires with a distally mounted semi-compliant inflatable balloon and an atraumatic tapered tip. The distal portion of the catheter has a lubricious coating. The manifold includes a lumen marked "THRU". This is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum diameter of 0.014". The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the dilatation balloon with a mixture of contrast medium and saline solution. The balloon has two radiopaque markers for positioning the balloon relative to the stenosis. The radiopaque marker bands indicate the dilating or working section of the balloon.

The NanoCross Elite catheter is available in balloon sizes ranging from 1.5 mm to 4 mm in diameter, and from 20 mm to 210 mm in length; reference labeling for introducer sheath compatibility.

AI/ML Overview

The provided text describes the 510(k) summary for the NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a study measuring clinical performance or AI algorithm accuracy.

Therefore, many of the requested categories for AI/clinical studies are not applicable to this type of regulatory submission. The information below will detail what is available in the provided text.


1. A table of acceptance criteria and the reported device performance

The document lists "bench testing" and "biocompatibility testing" as the studies performed to support substantial equivalence. The acceptance criteria for these tests are implicitly that the device performs "similarly to the predicates" and "met all acceptance criteria." Specific quantitative acceptance criteria or detailed device performance metrics are not provided in this summary.

Acceptance Criteria (General)Reported Device Performance
Met acceptance criteria for all bench tests"The NanoCross Elite catheter met all acceptance criteria for the bench testing with results similar to the predicates."
Biocompatibility per ISO 10993-1 for short duration contact with blood (<24 hours)"The device was tested for biocompatibility per ISO 10993-1... The testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, complement activation, thromboresistance, partial thromboplastin time, and platelet/leukocyte count." (Implies meeting acceptance criteria, though not explicitly stated for individual tests)

List of specific tests performed (acceptance criteria not detailed):

Performed for the NanoCross™ Elite:

  • Balloon Dimensional Verification
  • Catheter Dimensional Verification
  • Balloon Rated Burst Pressure (Unconstrained and In Stent)
  • Balloon Fatigue (Unconstrained and In Stent)
  • Balloon Compliance
  • Inflation/Deflation Time
  • Catheter Bond Strength
  • Kink
  • Device Tracking
  • Insertion Force
  • Balloon Pull-back Force
  • Torque Strength
  • Radiopacity
  • Presence of Coating
  • Re-Insertion Force
  • Pushability

Leveraged from predicate devices (implying the NanoCross Elite is expected to perform similarly):

  • Balloon Dimensional Verification
  • Balloon Rated Burst Pressure (Unconstrained and In Stent)
  • Balloon Fatigue (Unconstrained and In Stent)
  • Balloon Compliance
  • Coating Durability
  • Particle Generation

The following items are NOT APPLICABLE or NOT PROVIDED in this 510(k) summary, as it is a submission for a medical device (PTA balloon catheter) based on substantial equivalence, not an AI algorithm or clinical trial with specific performance metrics such as accuracy, sensitivity, or specificity against a ground truth.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable / Not Provided. This document describes bench and biocompatibility testing, not a clinical study with a test set of data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable / Not Provided. No ground truth for a clinical test set was established.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable / Not Provided. No adjudication for a clinical test set was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not Applicable. This is not an AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable / Not Provided. The "ground truth" for the non-clinical tests would be the established engineering and ISO standards for device performance, but not clinical ground truth as specified in the question.

8. The sample size for the training set

  • Not Applicable. This is not an AI device.

9. How the ground truth for the training set was established

  • Not Applicable. This is not an AI device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for "ev3", which is described as "Your endovascular company." The logo consists of the letters "ev3" in a stylized font, with three circles above the letters. The text "Your endovascular company." is written in a smaller font to the right of the letters.

510(k) Summary

NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter

510(k) SummaryThis 510(k) summary information is submitted in accordancewith the requirements of 21 CFR §807.92.
Applicantev3 Inc.
Submitterev3 Inc.3033 Campus DrivePlymouth, MN 55441-2651Tel: 763-398-7000Fax: 763-591-3248
SEP 2 6 2013
Contact PersonLaura J. Lind
Date PreparedJuly 22, 2013
Device Trade NameNanoCross™ Elite0.014" Over-the-WirePTA Balloon Dilatation Catheter
Device Common NamePTA Dilatation Catheter
Classification NameCatheter, Angioplasty, Peripheral, Transluminal(21 CFR §870.1250, Product Code LIT)
Classification PanelCardiovascular
Predicate DevicesNanoCross™ .014" OTW PTA Dilatation Catheter (K082854,K090849), RapidCross™ PTA Rapid Exchange BalloonDilatation Catheter (K130911).
Intended useThe NanoCross™ Elite 0.014" Over-the-Wire PTA BalloonDilatation Catheter is intended to dilate stenoses in the iliac,femoral, ilio-femoral, popliteal, infra-popliteal, and renalarteries, and for the treatment of obstructive lesions of native orsynthetic arteriovenous dialysis fistulae. This device is alsoindicated for stent post-dilatation in the peripheral vasculature.
Device DescriptionThe NanoCross catheter is an over-the-wire (OTW) coaxiallumen percutaneous transluminal angioplasty (PTA) ballooncatheter compatible with 0.014"guidewires with a distallymounted semi-compliant inflatable balloon and an atraumatictapered tip. The distal portion of the catheter has a lubriciouscoating. The manifold includes a lumen marked "THRU". Thisis the central lumen of the catheter, which terminates at the distal
• Balloon DimensionalVerification• Catheter DimensionalVerification
• Balloon Rated Burst Pressure(Unconstrained and In Stent)• Balloon Fatigue (Unconstrainedand In Stent)
• Balloon Compliance• Inflation/Deflation Time
• Catheter Bond Strength• Kink
• Device Tracking• Insertion Force
• Balloon Pull-back Force• Torque Strength
• Radiopacity• Presence of Coating
• Re-Insertion Force• Pushability

ev3 Corporate World Headquarters / Peripheral Vascular . 3033 Carryus Drive. Plynouth: MN 55441, USA . Ph +1 763 398 7000 eva Neurovascular - 9775 Tolodo Way, Irving, CA 92618 - Ph 11 949 837 3700 ev3 Europe / International Headquarters . 103/108 г.ю LB Boete, 75008 Paris: France . Ph +33 156 86 59 : 10

{1}------------------------------------------------

tip. This lumen is used to pass the catheter over a guidewire with a maximum diameter of 0.014". The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the dilatation balloon with a mixture of contrast medium and saline solution. The balloon has two radiopaque markers for positioning the balloon relative to the stenosis. The radiopaque marker bands indicate the dilating or working section of the balloon.

The NanoCross Elite catheter is available in balloon sizes ranging from 1.5 mm to 4 mm in diameter, and from 20 mm to 210 mm in length; reference labeling for introducer sheath compatibility.

Performance data

Bench testing and biocompatibility testing were performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use.

Using internal Risk Analysis procedures, the following tests were performed for one or more model sizes:

Using the same Risk Analysis procedures, the following tests were leveraged from predicate devices for one or more model sizes:

• Balloon DimensionalVerification• Balloon Rated Burst Pressure(Unconstrained and In Stent)
• Balloon Fatigue(Unconstrained and In Stent)• Balloon Compliance
• Coating Durability• Particle Generation

The device was tested for biocompatibility per ISO 10993-1 for short duration contact with blood (<24 hours). The testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, complement activation, thromboresistance, partial thromboplastin time, and platelet/leukocyte count.

The NanoCross Elite catheter met all acceptance criteria for the

{2}------------------------------------------------

bench testing with results similar to the predicates. Based on the bench test results, comparison to legally marketed predicates, and non-clinical test results, the NanoCross Elite catheter is determined to perform as safely and effectively as the predicates for its intended use. The NanoCross Elite catheter has the following similarities to Summary of the predicate devices: Substantial Similar fundamental scientific technology Equivalence . Same intended use . . Similar operating principle Identical balloon rated burst pressures � . Similar balloon nominal pressure . Similar balloon diameters Similar balloon lengths . o Similar catheter lengths o A lubricious coating . Same cleanroom and shared manufacturing lines Same sterility assurance level and sterilization method . All devices are compatible with 0.014" wires and 4F sheaths. All devices have similar construction and principles of operation. All devices are used by the physician in a similar manner typical of PTA balloon catheters. The NanoCross Elite catheter and the predicates have the same intended use - all devices are intended to treat peripheral arteries. All devices are intended to treat the same target population. The manner in accessing and treating lesions is similar for the devices. The indications for use are identical to the RapidCross™ Catheter. Based on the intended use, technological characteristics, and Conclusion results from safety and performance testing, the NanoCross Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter is considered substantially equivalent to the legally marketed predicate devices NanoCross" .014" OTW PTA Balloon Dilatation Catheter (K082854, K090849) and RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter (K130911).

{3}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 26, 2013

ev3, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K132777

Trade/Device Name: NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter

Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: September 4, 2013 Received: September 5, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Todd D. Courtney -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K132777

Page 1_of_1

Indications for Use Statement

K132777 510(k) Number (if known):

Device Name: NanoCross™ Elite 0.014` Over-the-Wire PTA Balloon Dilatation Catheter

Indications for Use:

The NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter is intended to dilate stenoses in the iliac. femoral. ilio-femoral. popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

Image /page/5/Picture/12 description: The image shows the FDA logo on the left side. On the right side of the logo, the text "Todd D. Courtney - S" is present. The text is aligned vertically, with "Todd D." on the top, "Courtney -" in the middle, and "S" on the bottom.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).