K Number
K132697

Validate with FDA (Live)

Date Cleared
2014-06-20

(295 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Encompass® 10/12 Hip Stem is intended for use in total hip replacement surgery. Total hip arthroplasty is intended to provide increased patient mobility and to decrease pain by replacing the damaged hip joint in patients having sufficiently sound bone to support the implants.

The device is intended for use in total hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:

  1. Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis
  2. Previously failed surgery
  3. Proximal femoral neck fractures or dislocation
  4. Idiopathic avascular necrosis of the femoral head
  5. Non-union of proximal femoral neck fractures
  6. Treatment of fractures that are unmanageable using other forms of therapy
  7. Benign or malignant bone tumors, congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis
Device Description

The Encompass® 10/12 Hip Stem is a one-piece, straight femoral stem, designed for single, uncemented use. Device fixation is achieved via press-fit in the medullary canal, which maximizes contact between the stem and bone. The stem is manufactured from wrought titanium alloy Ti-6Al-4V EU per ASTM F136. The proximal portion of the stem is subsequently plasma-sprayed with Commercially Pure Titanium per ASTM F1580. The stem has a neck with a 10/12 trunnion taper for modular attachment to femoral heads. To accommodate various patient anatomies, the stem is offered with and without collars and in a variety of sizes, including the following ranges: lengths (115-143mm), horizontal offsets (32-46mm), and vertical offsets (29-36mm), with a resection angle of 130°, and neck angle of 132°.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device PerformanceComments
Mechanical PerformanceProximal Fatigue (ISO 7206-6:2013)Implied to meet ISO 7206-6:2013 standards for hip stem fatigue performance."Performance testing has been conducted for the subject device Encompass® 10/12 Hip Stem in proximal fatigue in accordance with ISO 7206-6:2013"The document states testing was conducted in accordance with the standard, implying satisfactory results were achieved to demonstrate substantial equivalence, though specific numerical outcomes are not provided.
Mechanical PerformanceDistal Fatigue (ISO 7206-4:2010)Implied to meet ISO 7206-4:2010 standards for hip stem fatigue performance."and distal fatigue in accordance with ISO 7206-4:2010."Similar to proximal fatigue, the device underwent testing according to the standard.
Range of MotionRange of Motion Analysis (ISO 21535:2007(E))Implied to meet ISO 21535:2007(E) standards for range of motion analysis in hip implants."Range of motion analysis was performed per ISO 21535:2007(E)."Testing was performed against this standard, suggesting compliance.
Material PropertiesPlasma Spray Coating (Mechanical Properties & Microstructure)Implied to demonstrate suitable mechanical properties and microstructure for the plasma spray coating."The plasma spray coating underwent testing for mechanical properties and microstructure analysis."Testing was conducted, implying acceptable results for the coating.
Substantial EquivalenceOverall Comparison to PredicatesSimilarity in intended use, indications for use, overall design, materials, manufacturing methods, packaging, mechanical performance, and sterilization.The FDA letter explicitly states the device is "substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices."The primary "acceptance criterion" for this 510(k) submission is substantial equivalence to predicate devices, which the FDA affirmed.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable. The document does not report any clinical studies or human test subjects. The performance testing was non-clinical (mechanical testing of the device itself).
  • Data Provenance: Not applicable. All testing mentioned is non-clinical laboratory testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not applicable. As no clinical studies were performed, there was no human test set requiring expert adjudication for ground truth. The "ground truth" for non-clinical performance testing is based on established engineering standards (ISO).

4. Adjudication Method for the Test Set

  • Not applicable. No clinical studies were performed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a hip stem prosthesis, not an AI or imaging-based diagnostic tool that would typically involve human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a medical implant, not an algorithm, so the concept of "standalone performance" in the context of AI does not apply.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the non-clinical performance studies, the "ground truth" is established by adherence to internationally recognized engineering and medical device standards (e.g., ISO 7206-6, ISO 7206-4, ISO 21535). These standards define acceptable performance parameters for hip stem prostheses.

8. The Sample Size for the Training Set

  • Not applicable. As no clinical studies or AI/algorithm development are reported, there is no "training set." The device is a physical implant.

9. How the Ground Truth for the Training Set was Established

  • Not applicable. There is no training set.

{0}------------------------------------------------

JUN 2 0 2014

510(k) Summary

Name of Sponsor:Ortho Development Corporation
12187 South Business Park Drive
Draper, Utah 84020
510(k) Contact:Mike Ensign
Director of Regulatory Affairs and Quality Assurance
Telephone: (801) 553-9991
Facsimile: (801) 553-9993
Email: mensign@odev.com
Date Prepared:June 19, 2014
Proprietary Name:Encompass® 10/12 Hip Stem
Common Name:Hip Stem Prosthesis
Classification:Class II device
21 CFR 888.3358, Hip joint metal/polymer/metal semi-constrained
porous-coated uncemented prosthesis
Device Product Code:LPH, MBL
Predicate Devices:Encompass™ Press-Fit Hip Stem (K053293) Ortho Development
Primaloc™ Cementless Hip System (K953977) Ortho Development
Ovation 10/12 Hip Stem (K131022) Ortho Development
Bi-Metric® Porous Primary Femoral Component (K020580) Biomet

Device Description

The Encompass® 10/12 Hip Stem is a one-piece, straight femoral stem, designed for single, uncemented use. Device fixation is achieved via press-fit in the medullary canal, which maximizes contact between the stem and bone. The stem is manufactured from wrought titanium alloy Ti-6Al-4V EU per ASTM F136. The proximal portion of the stem is subsequently plasma-sprayed with Commercially Pure Titanium per ASTM F1580. The stem has a neck with a 10/12 trunnion taper for modular attachment to femoral heads. To accommodate various patient anatomies, the stem is offered with and without collars and in a variety of sizes, including the following ranges: lengths (115-143mm), horizontal offsets (32-46mm), and vertical offsets (29-36mm), with a resection angle of 130°, and neck angle of 132°.

{1}------------------------------------------------

Intended Use

The Encompass® 10/12 Hip Stem is intended for use in total hip replacement surgery. Total hip arthroplasty is intended to provide increased patient mobility and to decrease pain by replacing the damaged hip joint in patients having sufficiently sound bone to support the implants.

Indications for Use

The device is intended for use in total hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:

    1. Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis
  • Previously failed surgery 2.
    1. Proximal femoral neck fractures or dislocation
    1. Idiopathic avascular necrosis of the femoral head
    1. Non-union of proximal femoral neck fractures
    1. Treatment of fractures that are unmanageable using other forms of therapy
    1. Benign or malignant bone tumors, congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis

Basis of Substantial Equivalence

The subject device Encompass® 10/12 Hip Stem is substantially equivalent to the previously cleared predicate devices based on similarities in intended use, indications for use, overall design, materials, manufacturing methods, packaging, mechanical performance, and sterilization.

Non-Clinical Test Summary

Performance testing has been conducted for the subject device Encompass® 10/12 Hip Stem in proximal fatigue in accordance with ISO 7206-6:2013 and distal fatigue in accordance with ISO 7206-4:2010. Range of motion analysis was performed per ISO 21535:2007(E). The plasma spray coating underwent testing for mechanical properties and microstructure analysis.

Clinical Test Summary

No clinical studies were performed.

Conclusions

Based on the similarities to the predicate devices, and a review of the testing, the devices are substantially equivalent to femoral stem components that were cleared under K053293 and K020580.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized symbol resembling a person with outstretched arms, with three curved lines representing the body and arms. The symbol is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the symbol. The text and symbol are both in black and are set against a white background.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

June 20, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ortho Development Corporation Mr. Tom Haueter Regulatory Affairs Manager 12187 South Business Park Drive Draper, Utah 84020

Rc: K132697

Trade/Device Name: Encompass" 10/12 Hip Stem Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, MBL Dated: May 16, 2014 Received: May 19, 2014

Dear Mr. Haueter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. Tom Haueter

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours,

Mark N. Melkerson -A

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Explration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K132697

Device Name Encompass® 10/12 Hip Stem

Indications for Use (Describe)

This device is intended for use in total hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:

  1. Notably impaired hip joint due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis
    1. Previously failed surgery
    1. Proximal femoral neck fractures or dislocation
    1. Idiopathic avascular necrosis of the femoral head
    1. Non-union of proximal femoral neck fractures
    1. Treatment of fractures that are unmanageable using other forms of therapy
  1. Benign or malignant bone tumors, congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

:

:

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

.

Elizabeth & Frank -S

Division of Orthopedic Devices

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the The burden and for the catedian of withing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.