K Number
K132671

Validate with FDA (Live)

Date Cleared
2013-12-20

(115 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor's offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.

Device Description

The Neptune 2 Waste Management System is a fluid waste management system with smoke evacuation. It collects surgical waste fluid within a closed suction system, then disposes of it through a docking station. The system is comprised of a rover and a docker (also referred to as a docking station).

The Neptune 2 Rover (Model Number: 0702-001-000) is a mobile, AC-powered surgical fluid waste collection device. It is intended to collect waste fluids and evacuate smoke during surgical procedures.

The Neptune 2 Docking Station (Model Number: 0702-014-000) is a stationary device for automated rover cleaning and waste offloading. It provides a fixed connection to the hospital's water supply, sewer, and electricity.

AI/ML Overview

The provided text describes the Stryker® Neptune 2 Waste Management System, a fluid waste management system with smoke evacuation, submitted for 510(k) clearance. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving a new device meets specific acceptance criteria through a comparative study with a ground truth established by experts.

Therefore, many of the requested elements (e.g., specific acceptance criteria, sample sizes for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth) are not explicitly detailed in this 510(k) summary. The summary primarily relies on bench testing and comparison to predicate devices to establish safety and effectiveness.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not present a formal table of explicit, quantitative acceptance criteria that the device was tested against and then reports its performance against those criteria in a "study" as one might expect for a new diagnostic or AI-driven insights device. Instead, the "Testing" section states:

"The Neptune 2 Waste Management System meets the specification and performance characteristics as identified in Stryker's internal design control procedures. Bench testing was performed to demonstrate the performance of the device. Testing was completed to verify and validate the device as outlined in the device description. Testing included electrically powered suction safety, electrical safety, vacuum cycling, software, smoke evacuation and filtration. Human factors analysis and usability testing was performed on the user interface, labeling, and training materials."

The "Table 1: Device Comparison Matrix" provides comparative specifications against predicate devices, implying that performance within a similar range or meeting similar functional attributes to the predicates would be considered acceptable. Below is a table summarizing key performance characteristics (rather than explicit acceptance criteria with pass/fail thresholds) of the Neptune 2 and its predicates, as presented in the document.

Feature / Performance CharacteristicStryker® Neptune 2 (Reported Performance)Predicate Device 1 (Stryker® Neptune Waste Management System)Predicate Device 2 (Dornoch Transposal Ultra System)Predicate Device 3 (Valleylab OptiMumm Smoke Evacuator System)
General
ClassificationIIIIIIII
Product CodeJCX, FYDFYD, JCXJCXFYD
Rover Empty Weight300 lbs230 lbs195 lbs or 225 lbs (depending on model)230 lbs (from Neptune 1, similar category)
Electrical Isolation TypeClass I, Type CF Applied PartClass I, Type CF Applied PartInformation not availableClass I, Type CF Applied Part (from Neptune 1, similar category)
Enclosure ProtectionIPXO Ordinary EquipmentIPXO Ordinary EquipmentInformation not availableIPXO Ordinary Equipment (from Neptune 1, similar category)
120V Single-phase Electric Power Requirements12 A16 A4.5 A3 A
Number of Uses Before DockingLimited by fluid capacityLimited by fluid capacityLimited to 2 or 4 proceduresLimited by fluid capacity (from Neptune 1, similar category)
User DisplayLCD screenLCD screenLCD screenDigital display (smoke evacuator) / LCD screen (from Neptune 1)
Suction / Fluid Waste Management
Canister Vacuum Level RegulationElectronic regulatorMechanical regulatorElectronic regulatorElectronic regulator (from Neptune 1, similar category)
Rover Canister Capacity24 L20 L33 L or 52 L (depending on model)24 L (from Neptune 1, similar category)
Canister Graduation LabelYesYesYesYes
Electronic Canister Overfill ProtectionYesYesYesYes
Mechanical Canister Overfill ProtectionFloat-style shut-off valve in roverFloat-style shut-off valve in roverHydrophobic filter located in disposable canister lidFloat-style shut-off valve in rover (from Neptune 1, similar category)
Volume Display Accuracy±150 mL (20L), ±50 mL (4L)±175 mL±150 mL±150 mL (from Neptune 1, similar category)
Volume Display Resolution50 mL (20L), 20 mL (4L)50 mL10 mL50 mL (from Neptune 1, similar category)
Max Suction Range (100% power)67 mm-Hg61 mm-Hg99 mm-Hg61 mm-Hg (from Neptune 1, similar category)
Min Suction Range (20% power)34 mm-Hg0 mm-Hg12 mm-Hg0 mm-Hg (from Neptune 1, similar category)
Smoke Evacuation
Filtration EfficiencyULPAHEPAULPAHEPA
Maximum Flow Rate24 CFM33 CFM20 CFM33 CFM (from Neptune 1, similar category)
Minimum Flow Rate21 CFM0 CFM3 CFM0 CFM (from Neptune 1, similar category)
Variable Speed Settings9Infinite (continuous rotary dial)5Infinite (from Neptune 1, similar category)
Pre-filterMulti-use disposableSingle-use disposableSingle-use disposableSingle-use disposable (from Neptune 1, similar category)
Filter Life IndicatorYesYesYesYes

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document explicitly states "Bench testing was performed...Testing was completed to verify and validate the device." However, specific sample sizes (e.g., number of test articles, number of test runs) for each component of the bench testing (electrical safety, vacuum cycling, smoke evacuation, filtration, software, human factors) are not detailed.
  • Data Provenance: The testing was conducted internally by Stryker. No information is provided regarding the country of origin of the data or whether it was retrospective or prospective, as this is a device for fluid and smoke management, not an AI or diagnostic device that typically uses patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable and not provided in the 510(k) summary. This type of detail is typically associated with studies for diagnostic devices or AI algorithms where expert consensus establishes a "ground truth" against which device performance is measured (e.g., presence or absence of a disease). For a waste management system, the "ground truth" is typically defined by engineering specifications and functional testing (e.g., vacuum pressure achieved, fluid volume collected, smoke particles filtered).

4. Adjudication Method for the Test Set

This is not applicable and not provided in the 510(k) summary for the same reasons as #3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. MRMC studies are typically used to assess the performance of diagnostic devices or AI algorithms when interpreted by multiple human readers, often comparing performance with and without AI assistance. This device is a surgical waste management system and does not involve human interpretation of cases.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

No, a standalone algorithm-only study was not done because this is a physical medical device. It does not employ an "algorithm" in the sense of a software component making diagnostic or analytical decisions. Its performance is related to its mechanical and electrical functions.

7. The Type of Ground Truth Used

The "ground truth" for this device would be defined by engineering specifications, regulatory standards (e.g., electrical safety, biocompatibility), and functional performance targets (e.g., flow rates, vacuum levels, filtration efficiency, fluid capacity). The device was tested to verify it met these internal specifications and regulatory requirements. The biocompatibility testing specifically referenced ISO-10993 standards.

8. The Sample Size for the Training Set

This is not applicable and not provided. "Training set" refers to data used to train machine learning algorithms. The Neptune 2 is a physical medical device, not an AI/ML algorithm.

9. How the Ground Truth for the Training Set Was Established

This is not applicable as there is no training set for an AI/ML algorithm for this device.

{0}------------------------------------------------

KV32671

DEC 2 0 2013

stryker

Instruments

510(k) Summary

510(k) Owner:Stryker Instruments4100 E. Milham AvenueKalamazoo, MI 49001(p) 269-389-3549(f) 269-389-5412
Contact Person:Michelle Jump
Registration No.:1811755
Trade Name:Stryker® Neptune 2 Waste Management System
Common Name:Portable suction and smoke evacuation device
Classification Name and Regulation NumberPrimary Product Code
-----------------------------------------------------------------
Product CodeDeviceRegulation NumberClass
JCXapparatus, suction, ward use, portable, ac-powered21 CFR 872.4780II
Secondary Product Code
--------------------------
Product CodeDeviceRegulation NumberClass
FYDapparatus, exhaust, surgical21 CFR 878.5070II
Predicate Devices:Stryker® Neptune Waste Management System (K012991)Dornoch Transposal Ultra System (K081047).Valleylab OptiMumm Smoke Evacuator System (K980915)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

4100 E. Milham Avenue
Kalamazoo. MI 49001
t: 269 323 7700 1: 269 389 5412
www.sstryker.com

{1}------------------------------------------------

DeviceDescription:The Neptune 2 Waste Management System is a fluid waste management systemwith smoke evacuation. It collects surgical waste fluid within a closed suctionsystem, then disposes of it through a docking station. The system is comprised of arover and a docker (also referred to as a docking station).The Neptune 2 Rover (Model Number: 0702-001-000) is a mobile, AC-poweredsurgical fluid waste collection device. It is intended to collect waste fluids andevacuate smoke during surgical procedures.The Neptune 2 Docking Station (Model Number: 0702-014-000) is a stationarydevice for automated rover cleaning and waste offloading. It provides a fixedconnection to the hospital's water supply, sewer, and electricity.The following Stryker accessories are for use with the Neptune 2 WasteManagement System:
Model Number Product Name 0700-001-026 Neptune Docking Detergent 0702-020-000 Neptune 2 4-Port Manifold 0702-020-001 Neptune 2 Specimen Collection 4-Port 0702-025-000 Neptune 2 Single-Port Manifold 0702-034-000 Fluid Suction HEPA Filter 0702-040-000 Smoke Evacuator ULPA Filter 0702-045-023 Smoke (Evacuator) Tubing, 3/8 inch x 10 feet 0700-026-000 Smoke (Evacuator) Tubing, 7/8 inch x 10 feet 0702-045-027 PenAdapt®
Indicationsfor Use:The Neptune 2 Waste Management System is intended to be used in the operatingroom, pathology, surgical centers, and doctor's offices to collect and dispose ofsurgical fluid waste as well as collect smoke generated from electrocautery or laserdevices.
ContraindicationsThe Neptune 2 Waste Management System is contraindicated against:• Connection directly to chest tubes.• Connection to closed wound drainage systems.
TestingThe Neptune 2 Waste Management System meets the specification andperformance characteristics as identified in Stryker's internal design controlprocedures. Bench testing was performed to demonstrate the performance of thedevice. Testing was completed to verify and validate the device as outlined in thedevice description. Testing included electrically powered suction safety, electricalsafety, vacuum cycling, software, smoke evacuation and filtration. Human factorsanalysis and usability testing was performed on the user interface, labeling, andtraining materials.
BiocompatibilityBiocompatibility testing of the Neptune 2 Waste Management System confirmed thatthe device meets the applicable requirements of the FDA Blue Book MemorandumG95-1 entitled Use of International Standards ISO-10993 Biological Evaluation of

·

{2}------------------------------------------------

Medical Devices Part -1: Evaluation and Testing and are biocompatible.
SubstantialEquivalence(SE) Rationale:The Neptune 2 Waste Management System, when compared to its predicates, has an equivalent intended use, protocols for use, mode of operation, technological characteristics and performance specifications.(See attached Table 1: Device Comparison Matrix)
Safety andEffectiveness:Based upon the comparison to the predicate devices, the Stryker Neptune 2 Waste Management System is substantially equivalent to the predicate devices.
Submitted by:Michelle JumpAssociate Regulatory Affairs Analyst
Signature
Date Submitted:December 20, 2013

{3}------------------------------------------------

4100 E. Milham Avenue
Kalamazoo, Mi 49001
1: 269 323 7700 f: 269 389 5412
www.siryker.com

stryker

Instruments

Stryker Neptune 2

Covidien (Valleylab)

Dornoch Transposal

Stryker Neptune 1

able 1: Device Comparison Matri

Model NumberClearanceClassificationRegulationProduct CodeIndications for Use
Rover: 0700-010-00510kII21 CFR 878.5070FYDThe Neptune Waste Management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.
Duo: DU100Quad: QD100K012991II21 CFR 878.4780JCXThe Dornoch Transposal Ultra System is self-powered suction / vacuum pump intended to collect and dispose of liquid waste within Hospital Operating Rooms, Pathology Labs, Surgical Outpatient Centers, and Doctor's Offices.
OptiMumm: 945 102 097K081047II21 CFR 878.5070FYDThe Indications for Use of the Valleylab OptiMumm Smoke Evacuator system are for the removal of smoke and incidental fluids produced during electrosurgery and/or laser surgery. The removal of smoke from the surgical site improves visibility and reduces potential health hazards associated with surgical smoke.
Rover: 0702-010-000K980915II21 CFR 878.478021 CFR 878.5070JCXFYDThe Neptune Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor's offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.
PendingII

{4}------------------------------------------------

lest toose. Connection to closed ound drainage systems.The Neptune 2 Waste Management System is contraindicated stated is the states • Connection directly sohest tubes.19 x 23 x 70"swiveling hospitagrade casterRubber bumpersurfaces of the rover vith a soft, lint-free cloth noistened with a non-nstructions. Caution against using solvents, lubricants, lubricants, other chemicals,brasive, hospital lisinfectant prepared ccording to theVipe the externancluding glutaraldehyde c similar chemicalnanufacturer'cleanersNoYes
None. :: , '.5 -. "
None24" x 24" x 55" or 23" x 33" x 33" x 55" depending on modeswiveling casterlastic bumperNoepending on mode2 or 4Yes
None18 x 25 x 51"swiveling hospitgrade castersRubber bumperWipe the surfaces of the Restitio and the criation and the station and the station and the stations and the count of the count of the world coultions codes of cides to on simcleansers.YesTYes
ntraindicationstem ConfiguraSizeCastersBumpersaning instructiredlover is battery-power luring dockingof canisterVolume measuremen vith digital display

{5}------------------------------------------------

Offloading and canistercleaning accessoryNeptune 1Docking Station(0700-005-000)Ultra EvacDocking StationUL-EV100Neptune 2Docking Station(0702-014-000)
Rover empty weight230 lbs195 lbs or 225 lbs(depending on model)300 lbs
Electrical isolation typeClass I, Type CF AppliedPartInformation not availableClass I, Type CF AppliedPart
Enclosure protectionIPXO Ordinary EquipmentInformation not availableIPXO Ordinary Equipment
120V single phaseelectric powerrequirements16 A4.5 A12 A
Number of uses beforedockingLimited by fluid capacityLimited to 2 or 4procedures(depending on model)Limited by fluid capacity
User displayLCD screenLCD screenLCD screen
User interaction to dockPush rover into dockerPush Rover adjacent toDocker and Insertdocker's gas-pump-stylenozzle into roverPush rover into docker
Waste/water connectionto roverDry-break hydrauliccouplingsInformation not availableDry-break hydrauliccouplings
Rover-docker couplingsize.875" body diameter.625" OD1.325" body diameter
Rover-docker couplingorientationHorizontalApproximately 25° fromhorizontalVertical
Water inlet controlSolenoid valveSolenoid valveSolenoid valve
Backflow preventionvalveInternalExternalInternal
Weight90 lbs100 lbs95 lbs
Waste/water connectionto facilityQuick disconnect fittingsInformation not availableQuick disconnect fittings

{6}------------------------------------------------

Neptune docking

etergent (0700-00)
xed rubber pa 0 - 480 mm-H 23″ x 23″ x 16 igital displa
(mm-Hg) nfrared Sprinkle НЕРА 3 A Yes Yes Yes Yes Yes ನ . ---ﺎﺭﻳﺲ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ … .. 、『 ్ ప్ర ﺃ ﺍﻟ : : . . ﻳﻮ . ・ ﻨﻪ : : .............................................................................................................................................................................. 。 : ・・・・
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ . . . . . .. . :: : . ormation not availab tra Enzyme (UL-EZ20
ectrical contac 4 - 700 mm-Hj 0" x 16" x 28 igital display
(mm-Hg) Sprinkler HEPA 15 A Yes Yes Yes Yes No ਨ ters with adjustab
extensions Neptune docking
tergent (0700-00)
4 - 483 mm-H Analog gauge
mm-Hg, in-Hg, in-Hg 5" x 20" x 22 Sprinkle nfrarec HEPA Yes 3 A Yes Yes Yes (દડ I nister cleaning meth ectromagnet for
olding rover to docke cuum limit adjustab multaneous levels cuum control ranj ptional wall suctio Rinses canister with
facility water cker dimension cuum Syster 20V single phas
ectric power
equirements offloads waste to
anitary sewer Canister cleaning
Jetergent uction filtration
fficiency anister vacuun ocker support over-Docker
ommunication ndicator uction orts

ction 005

{7}------------------------------------------------

Canister vacuum levelregulationCollection CanistersMechanical regulatorElectronic regulatorElectronic regulator20 hrs. ' •"25 hrs80 hrs
Rover canister capacity20 L33 L or 52 L(depending on model)24 LYesત્વ જિડYes
Canister graduation labelYesYesYesTﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ.Iరా
Electronic canisteroverfill protectionYesYesYes7/8":11- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3/8", 7/8″3/8", 7/8"
Mechanical canisteroverfill protectionFloat-style shut-off valvein roverHydrophobic filterlocated in disposablecanister lid.Float-style shut-off valvein roverYes-YesYes
Volume display accuracy$\pm 175$ mL$\pm 150$ mL$\pm 150$ mL, $\pm 50$ mL(20 L canister, 4 Lcanister)ED్ :: :- 1 -formation not availaEO
Volume displayresolution50 mL10 mL50 mL, 20 mL(20 L canister, 4 Lcanister)SO 10993-1. .ﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ50 1099320 10993-
Lighted canister contentsNoYesNo
Smoke Evacuatorfertica45 degree્રﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤforizonta
Filtration efficiencyHEPAULPAULPA375"x .320" and 1x .37--: : : : : : : : : : : : : :375"
Maximum flow rate33 CFM20 CFM24 CFMSingle patient usingle patient usSingle patient
Minimum flow rate0 CFM3 CFM21 CFM1-port o 4-port3-portﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ1-port of 4-port
Maximum Suction Range(100% power)61 mm-Hg99 mm-Hg67 mm-HgYesYes:-. .Yes
Minimum suction range(20% power)0 mm-Hg12 mm-Hg34 mm-Hg4.7 oz3.1 oz1.6 oz
Pre-filterSingle-use disposableSingle-use disposableMulti-use disposableon-sterilelon-steril::.Non-steril
Filter life indicatorYesYesYes
Variable speed settingsInfinite(continuous rotary dial)59YesYesYes
ectric motor-driveformation not availaectric motor-driv
t

: :

:

{8}------------------------------------------------

{9}------------------------------------------------

TravelMaximum height from floorEnvironmental Conditions
Operation TempOperation HumidityOperation Barometric PressureStorage / Transportation
29 in89 in10 - 40°C30 - 75%700 - 1060 hPa-20 - 40°C
Information not availableInformation not available10 - 40°C30 - 75%700 - 1060 hPa-20 - 40°C
32 in102 inInformation not availableInformation not availableInformation not availableInformation not available
10 - 40°C30 - 75%700 - 1060 hPa-20 - 40°C
Storage / Transportation HumidityStorage / Transportation Barometric Pressure
10 - 75%500 - 1060 hPa
10 - 75%500 - 1060 hPa
Information not availableInformation not available
10 - 75%500 - 1060 hPa

. ﺑ

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized human figure with outstretched arms, representing care and protection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure. The logo is printed in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 20, 2013

Stryker Corporation Michelle Jump, RAC, M.S. Regulatory Affairs Associate Analyst 4100 E. Milham Ave, KALAMAZOO MI 49001

Re: K132671

Trade/Device Name: Neptune 2 Waste Management System Regulation Number: 21 CFR 872.4780 Regulation Name: Apparatus, suction, ward use, portable, ac-powered Regulatory Class: II Product Code: JCX, FYD Dated: November 14, 2013 Received: November 15, 2013

Dear Ms. Jump:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{11}------------------------------------------------

Page 2 - Ms. Jump

.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

Sincerely yours,

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Image /page/11/Picture/7 description: The image contains a signature and some text. The text reads "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID FOR". The signature is illegible.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, . Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

510(K) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: _ Stryker® Neptune 2

Indications for Use

ﺮ ﺍﻟﻤﻌﺎﺩﻟﺔ ﻣﻮﺍﺩ ﺍﻟﻤﻮﺿﻮﻋﺎﺕ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﻠﺘﻤﻴﺔ -

The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor's offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

and/or

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Glaverie -S 2013.12.17 18:13:45 -05'0

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.