(21 days)
The Raphael Pedicle Screw System is a posterior pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S I vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Raphael Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis).
The Raphael Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma(i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Raphael Hook System can be used in conjunction with Raphael Pedicle Screw System.
The Raphael Pedicle Screw System is a top-loading multiple component, posterior spinal fixation system that consists of pedicle screws, rods, hooks, set screws, connectors, and a transverse (cross) linking mechanism. The Raphael Pedicle Screw System allows surgeons to build a spinal implant construct to stabilize and promote spinal fusion.
This document is a 510(k) premarket notification for the Raphael Pedicle Screw System. It seeks to demonstrate substantial equivalence to a predicate device, the Dio Medical Rexious Spinal Fixation System.
Based on the provided text, there is no study described that proves the device meets specific acceptance criteria in terms of performance derived from non-clinical or clinical testing.
Instead, the submission relies entirely on a claim of substantial equivalence to a previously cleared predicate device.
Here's an analysis based on the information provided and what is explicitly missing in the context of your request:
1. Table of Acceptance Criteria and Reported Device Performance
No specific acceptance criteria are defined in the document for the Raphael Pedicle Screw System related to its performance.
The document states:
"Non-clinical testing was not performed as part of this submission."
"The subject device is identical to the predicate device in terms of indications for use, device dimensions, instrumentation, manufacturing process, cleaning/sterilization process and labeling..."
Therefore, a table of reported device performance for the Raphael Pedicle Screw System cannot be generated from this document, as no performance testing was conducted or reported for this specific submission. The acceptance criteria, by extension, are implicitly met if the device is indeed "identical" to the predicate, whose performance would have been established through prior clearance.
2. Sample Size Used for the Test Set and Data Provenance
Not applicable. No test set was used for this submission because no new performance testing (non-clinical or clinical) was performed. The clearance is based on substantial equivalence to a predicate device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. No test set was used, and therefore no ground truth establishment by experts was required for this submission.
4. Adjudication Method for the Test Set
Not applicable. No test set was used.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a pedicle screw system, not an AI-assisted diagnostic or therapeutic device. No MRMC study was conducted.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a pedicle screw system, not an algorithm or AI device. No standalone performance study was conducted.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)
Not applicable. No new performance data requiring ground truth was generated for this submission. The "ground truth" for clearance in this case is the previously established performance and substantial equivalence of the predicate device.
8. The Sample Size for the Training Set
Not applicable. No training set was used as this is not an AI/ML device.
9. How the Ground Truth for the Training Set was Established
Not applicable. No training set was used.
In summary: The clearance for the Raphael Pedicle Screw System is based on a "Special 510(k)" pathway, where the manufacturer asserts that the device is identical to a previously cleared predicate device (Dio Medical Rexious Spinal Fixation System K113324, K111362, K100765) in terms of indications for use, device dimensions, instrumentation, manufacturing process, cleaning/sterilization process, and labeling. Therefore, no new performance testing or clinical studies were conducted or reported in this specific document.
{0}------------------------------------------------
510(k) Summary Raphael Pedicle Screw System
| SUBMITTED BY | Accel Spine14901 Quorum Dr. Ste. 125Dallas, TX 75254 |
|---|---|
| ESTABLISHMENTREGISTRATION NUMBER | 3009051471 |
| OWNER/OPERATORNUMBER | 10035914 |
| CONTACT PERSON | Aekta PatelGeneral CounselAccel SpinePhone: 214-545-5833Fax: 888-831-4892 |
| SUBMISSION PREPARED BY | Lisa PetersonKaedon Consulting, LLCEmail: Ipeterson@kaedonconsulting.comPhone: 512-507-0746 |
| DATE PREPARED | July 24, 2013 |
| CLASSIFICATION NAME | Spondylolisthesis Spinal Fixation, Spinal Pedicle Fixation |
| DEVICE CLASS | Class II |
| REGULATION NUMBER | 888.3070 (Product Code MNH, KWP, MNI) |
| COMMON NAME | Pedicle Screw Spinal Fixation System |
| PROPRIETARY NAME | Raphael Pedicle Screw System |
| IDENTIFICATION OF PREDICATEDEVICE(S) | Dio Medical Rexious Spinal Fixation System (K113324,K111362, K100765) |
{1}------------------------------------------------
DEVICE DESCRIPTION
The Raphael Pedicle Screw System is a top-loading multiple component, posterior spinal fixation system that consists of pedicle screws, rods, hooks, set screws, connectors, and a transverse (cross) linking mechanism. The Raphael Pedicle Screw System allows surgeons to build a spinal implant construct to stabilize and promote spinal fusion.
INDICATIONS
The Raphael Pedicle Screw System is a posterior pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S I vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Raphael Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis).
The Raphael Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma(i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Raphael Hook System can be used in conjunction with Raphael Pedicle Screw System.
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The purpose of this Special 510(k) is to obtain clearance to market the subject device as the Raphael Pedicle Screw System. The subject device was previously cleared as the Dio Medical Rexious Spinal Fixation System (K113324, K111362, K100765). Documentation demonstrating the legal right to distribute the Raphael Pedicle Screw System is maintained at Accel Spine.
The subject device is identical to the previously cleared Rexious System in terms of indications for use. device dimensions, instrumentation, manufacturing process, cleaning/sterilization process and labeling. The System implant components are supplied non-sterile, are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F136.
DISCUSSION OF NON-CLINICAL TESTING
Non-clinical testing was not performed as part of this submission.
CONCLUSIONS
The subject device is identical to the predicate device in terms of indications for use, device dimensions, instrumentation, manufacturing process, cleaning/sterilization process and labeling, Documentation was provided as part of this Special 510(k) to demonstrate that the Raphael Pedicle Screw System is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure, with three curved lines forming its body and wings. To the left of the bird-like figure, there is a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MID 20093-0002
August 20, 2013
Ms. Aekta Patel General Counsel Accel Spine 14901 Ouorum Drive, Suite 125 Dallas, Texas 75254
Re: K132365
・
Trade/Device Name: Raphael Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI, KWP Dated: July 24, 2013 Received: July 30, 2013
Dear Ms. Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Aekta Patel
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm1 for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
K132365 510(k) Number (if known):
Raphael Pedicle Screw System Device Name:
Indications for Use:
The Raphael Pedicle Screw System is a posterior pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S I vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Raphael Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis).
The Raphael Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma(i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Raphael Hook System can be used in conjunction with Raphael Pedicle Screw System.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ronald P. Jean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K132365
N/A