K Number
K132143

Validate with FDA (Live)

Device Name
EXHALATION VALVE
Date Cleared
2014-06-04

(328 days)

Product Code
Regulation Number
868.5870
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The exhalation valves are used to control Inspiratory pressure and expel the expired air from a patient being ventilated via a single limb breathing system. The exhalation valve and single limb breathing systems are used with adults/pediatrics and prescribed by a physician. A pediatric is defined as 10 kg in weight. The product is not for use in neonates. The device can be used within hospitals and for home care use. It is a single patient use device and can be used for a maximum of 30 days.

Device Description

Double limb breathing systems comprising of an inspiratory and expiratory limb is commonly used for respiratory care. To reduce waste and clutter single limb breathing systems can be used with an exhalation valve to dispel exhaled air from the patient. Some ventilator designs or specifications do not require proximal pressure monitoring; airway pressure is monitored inside the ventilator. There are three variants of the exhalation valve in this submission.

Exhalation valve with proximal pressure port (1924501 &1924504): A single breathing tube connects the patient to the ventilator allowing the patient to receive respiratory care. The exhalation valve is connected at the patient end onto the breathing tube. A pressure monitor line and exhalation valve control line directly connects the valve to the ventilator. Via exhalation valve control line, the ventilator applies a pressure which controls the balloon valve. During the Inspiratory phase the balloon is inflated to close off the exhalation port in the valve body. During the expiratory phase the pressure is released in the valve chamber and the balloon deflates allowing air to be expelled to the surrounding environment via the exhalation port. This prevents the patient rebreathing exhaled gases yet allowing for a single limb breathing system to be used instead of a double limb system. The pressure line monitors the pressure in the valve body. The patient end of the exhalation valve has a swivel connector end.

Exhalation valve without proximal proximal prossure port (1923500): A single breathing tube connects the patient to the ventilator allowing the patient to receive respiratory care. The exhalation valve is connected at the patient end onto the breathing tube. An exhalation valve control line directly connects the valve to the ventilator. Via the exhalation valve control line, the ventilator applies a pressure which controls the balloon valve. During the Inspiratory phase the balloon is inflated to close off the exhalation port in the valve body. During the expiratory phase the pressure is released in the valve chamber and the balloon deflates allowing air to be expelled to the surrounding environment via the exhalation port. This prevents the patient rebreathing exhaled gases vet allowing for a single limb breathing system to be used instead of a double limb system. The patient end of the exhalation valve has a swivel connector end.

Different ventilators require valves which work with different pressure ratios. The MKIII exhalation valve (1923500 and 1924501) has a 1:1.5 pressure ratio which caters for the majority of ventilators. However, there are markets which have applications for ratio of 1:1, and 1:2. The MK3b exhalation valve (1924504) has a 1:2 pressure ratio. The pressure ratio is the pressure difference between the control line pressure and the pressure that will be generated on the patient side of the valve.

AI/ML Overview

This K132143 submission describes the Intersurgical Exhalation Valve (MK3 and MK3b models) and demonstrates its substantial equivalence to a predicate device (Intersurgical 1922500 exhalation valve) through non-clinical testing.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly list "acceptance criteria" in a numerical format that would typically be seen for a device like software. Instead, it presents a comparison of key performance characteristics between the new devices and the predicate, implying that performance comparable to the predicate is the acceptance criteria for substantial equivalence. The "Acceptance Criteria" below are inferred from the predicate's performance or internal requirements.

Characteristic ComparedAcceptance Criteria (Inferred from Predicate/Internal req.)MK3 Exhalation Valve (1923500 non-ported & 1924501 ported) PerformanceMK3b Exhalation Valve (1924504 ported) Performance
Intended UseAdult/pediatric patients using a single limb breathing tubeAdult/pediatric patients using a single limb breathing tubeAdult/pediatric patients using a single limb breathing tube
Duration of UseSingle patient use, up to 24 hours (Predicate)Single patient use, up to 30 daysSingle patient use, up to 30 days
Where usedHospital and homeHospital and homeHospital and home
Flow Resistance at 10L/min≤ 0.6 mbar0.6 mbar0.6 mbar
Flow Resistance at 30L/min≤ 1.0 mbar0.9 mbar0.9 mbar
Pressure Ratio1:2 (Predicate)1:1.51:2
Leakage balloon (ml/min)< 0.5 ml/min< 0.5 ml/min0.0 ml/min
Leakage main body (ml/min)< 3.5 ml/min (and < 5ml/min internal req.)< 0.5 ml/min1.3 ml/min
Conditioned/Storage testingComplete and PassComplete and PassComplete and Pass
TapersPassPassPass
Aging (5 year)N/A (no predicate data)PassPass
30-day testingN/A (24-hour predicate)PassPass
Trigger featureN/A (predicate different design)Trigger by 30msSame as MK3
BiocompatibilityISO 10993ISO 10993ISO 10993
SterilityNon SterileNon SterileNon Sterile
Standards MetISO 5356 (Connectors)ISO 5356 (Connectors)ISO 5356 (Connectors)

2. Sample size used for the test set and the data provenance

The submission describes non-clinical laboratory testing. There is no mention of a "test set" in the context of clinical data or patient samples. The tests were performed on the device prototypes themselves. The data provenance is laboratory testing conducted by Intersurgical Incorporated. There is no mention of country of origin for specific test data, but the company is based in Liverpool, NY, USA. The testing is prospective for the new devices, demonstrating their performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a non-clinical device performance study based on engineering specifications and physical measurements, not a diagnostic study requiring expert interpretation of results or establishment of "ground truth" as it would pertain to medical imaging or diagnostics.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is a non-clinical device performance study with objective measurements, not a study requiring adjudication of expert opinions.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (exhalation valve) and not an AI or diagnostic imaging device that would typically undergo MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a mechanical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this non-clinical study, the "ground truth" is defined by established engineering principles, international standards (e.g., ISO 5356-1, ISO 10993-1), and the performance characteristics of the legally marketed predicate device. For example, resistance to flow is measured against a standard, and biocompatibility is confirmed against ISO 10993. The "performance" of the predicate device serves as a benchmark for substantial equivalence.

8. The sample size for the training set

Not applicable. There is no concept of a "training set" for a physical medical device submission of this nature. The devices themselves are the "samples" produced and tested.

9. How the ground truth for the training set was established

Not applicable. No training set as per item 8.

{0}------------------------------------------------

K132143

JUN 0 4 2014

Section 5 510(k) Summary

510K Submitter Address and Establishment Registration Number:

Registration Number: 1319447 Name: INTERSURGICAL INCORPORATED 417 Electronics Parkwav Address: Liverpool, NY 13088 Date: May 2, 2014 Contact Person: Michael Zalewski - VP RAIQA/CS Phone Number: 315-451-2900 X 202 Fax Number: 315-451-3696

Classification: 21 CFR 868.5870, Classification Name: Non-rebreathing Valve, Classification Product Code: 73 CBP, Device Class: II, 510K Submission: Traditional.

Predicate Device:

The 1923500 & 1924501 MK3 exhalation valve and the 1924504 MK3b exhalation valves are substantially equivalent to the Intersurgical 1922500 exhalation valve (K984481). The predicate device can be used on a single limb breathing systems. The valve sits on a T-piece connected at the patient end of the breathing tube. When the patient inspires, pressure applied via the exhalation line inflates the valve membrane causing it to close off the exhalation port. Then when the patient exhales the pressure on the membrane is released allowing the exhaled air to escape. The predicate valve has a 1:2 pressure ratio similar to the MKIIIb. The device is for single patient use on adults and pediatrics.

The function of all three valves is the same, however the MK3 and MK3b can seal with a low inspiratory drive in order that the ventilator can trigger a spontaneous breath. IPPB machines in the past had no requirement for PEEP control. however ventilators are generally now fitted with controllable PEEP modes

The predicate device is aesthetically different to the MK3 and MK3b, the MK3 valves are now inline, reducing the need for additional connectors.

Description of Device:

Double limb breathing systems comprising of an inspiratory and expiratory limb is commonly used for respiratory care. To reduce waste and clutter single limb breathing systems can be used with an exhalation valve to dispel exhaled air from the patient. Some ventilator designs or specifications do not require proximal pressure monitoring; airway pressure is monitored inside the ventilator. There are three variants of the exhalation valve in this submission.

{1}------------------------------------------------

Description of Device:

Exhalation valve with proximal pressure port (1924501 &1924504)

A single breathing tube connects the patient to the ventilator allowing the patient to receive respiratory care. The exhalation valve is connected at the patient end onto the breathing tube. A pressure monitor line and exhalation valve control line directly connects the valve to the ventilator. Via exhalation valve control line, the ventilator applies a pressure which controls the balloon valve. During the Inspiratory phase the balloon is inflated to close off the exhalation port in the valve body. During the expiratory phase the pressure is released in the valve chamber and the balloon deflates allowing air to be expelled to the surrounding environment via the exhalation port. This prevents the patient rebreathing exhaled gases yet allowing for a single limb breathing system to be used instead of a double limb system. The pressure line monitors the pressure in the valve body. The patient end of the exhalation valve has a swivel connector end.

Exhalation valve without proximal proximal prossure port (1923500)

A single breathing tube connects the patient to the ventilator allowing the patient to receive respiratory care. The exhalation valve is connected at the patient end onto the breathing tube. An exhalation valve control line directly connects the valve to the ventilator. Via the exhalation valve control line, the ventilator applies a pressure which controls the balloon valve. During the Inspiratory phase the balloon is inflated to close off the exhalation port in the valve body. During the expiratory phase the pressure is released in the valve chamber and the balloon deflates allowing air to be expelled to the surrounding environment via the exhalation port. This prevents the patient rebreathing exhaled gases vet allowing for a single limb breathing system to be used instead of a double limb system. The patient end of the exhalation valve has a swivel connector end.

Different ventilators require valves which work with different pressure ratios. The MKIII exhalation valve (1923500 and 1924501) has a 1:1.5 pressure ratio which caters for the majority of ventilators. However, there are markets which have applications for ratio of 1:1, and 1:2. The MK3b exhalation valve (1924504) has a 1:2 pressure ratio. The pressure ratio is the pressure difference between the control line pressure and the pressure that will be generated on the patient side of the valve.

Indications for Use:

The exhalation valves are used to control Inspiratory pressure and expel the expired air from a patient being ventilated via a single limb breathing system. The exhalation valve and single limb breathing systems are used with adults/pediatrics and prescribed by a physician. The device can be used within hospitals and for home care use. It is a single patient use device and can be used for a maximum of 30 days.

{2}------------------------------------------------

Technology Characteristics Summary

The intended use of the Intersurgical Exhalation Valves is comparable to the referenced predicate device. The comparison of the data shows similar values for the key performance characteristics. Proposed devices show similar values for resistance to flow, pressure ratio and leakage when compared to the legally marketed devices.

Non-clinical test results are submitted to confirm product safety and substantial equivalence to predicate device.

CharacteristicComparedMK3 Exhalation Valve-1923500 non-portedand 1924501 portedMK3b ExhalationValve - 1924504portedPredicate ExhalationValve 1922500K984481
Intended Use:Target populationAdult/pediatric patientusing a single limbbreathing tube.Adult/pediatric patientusing a single limbbreathing tube.Adult/pediatric patientusing a single limbbreathing tube.
Indications for useProduct has been designedand validated for 30 dayswhereas predicate productwas not. It is animprovement on theproduct, the performanceremains comparable to thepredicate.The exhalation valvesare used to controlInspiratory pressure andexpel the expired airfrom a patient beingventilated via a singlelimb breathing system.The exhalation valve andsingle limb breathingsystems are used withadults/pediatrics andprescribed by aphysician. The devicecan be used withinhospitals and for homecare use. It is a singlepatient use device andcan be used for amaximum of 30 days.The exhalation valvesare used to controlInspiratory pressure andexpel the expired airfrom a patient beingventilated via a singlelimb breathing system.The exhalation valveand single limbbreathing systems areused withadults/pediatrics andprescribed by aphysician. The devicecan be used withinhospitals and for homecare use. It is a singlepatient use device andcan be used for amaximum of 30 days.A one-way valve thatprevents the escape ofinspiratory gases andrebreathing of ofexpired gases whileproviding a means ofegress for expired gases.Fits all 22 male T-piecesand allows for the use ofPEEP valve as the gasesexit the valve via a22mm port.Recommended change:every 24 hours or morefrequently if visibledeterioration isobserved. Single patientuse.

Device Comparison Table

{3}------------------------------------------------

Device Comparison Table

CharacteristicComparedMK3 Exhalation Valve3500 non-ported1924531 DOI3b.Exhalation11:48Predicate ExhalaValve 1922500
Where usedHospital and homeHospital and homeHospital and home
Product LabelingBreathing circuitwithexhalation valve.Breathing circuit withexhalation valve.Breathingcircuitwithexhalation valve.
Single Use or Reusable?Single Patient UseSingle Patient UseSingle Patient Use
DesignandPerformance:
FlowResistancetoat10L/min in mbar0.60.6 same body as MK30.6
ResistancetoFlowat30L/min in mbar0.90.9 same body as MK31.0
Pressure Ratio1:1.51:21:2
Leakageballooninml/min<0.50.0<0.5
Leakagein mainbodyml/min<0.51.33.5
Conditioned/storagetesting (-20 degrees + 50degrees)Same as mk3bComplete and passComplete and pass
TapersPassPassPass
Aging 5 yearPassPassNo test work conducted
30 day testingPassSame as MK3Pass
TriggerTrigger by 30msSame as MK3N/a different design
Internal Diameter of seatmm17.17 +/- 0.0515.5 +/- 0.04N/a different design
Materials:
BodyHDPEHDPEStyrene Butadiene
CapPolypropylenePolypropylenePolypropylene
ChamberPolycarbonatePolycarbonateN/A
ValveSiliconeSiliconeSilicone
Swivel ColorTrans (Clear) StyreneButadieneGreen StyreneButadieneN/A
Energy Used/Delivered:Air flow through deviceused to deliverinspiratory air andexhale expiratory airto/from patientAir flow through deviceused to deliverinspiratory air andexhale expiratory airto/from patientAir flow through deviceused to deliver inspiratoryair and exhale expiratoryair to/from patient
Compatibility:Designed for use withsingle limb breathingtubeDesigned for use withsingle limb breathingtubeDesigned for use withsingle limb breathing tube
Biocompatibility:ISO 10993ISO 10993
Sterility:Non SterileNon SterileNon Sterile
Applicable Standards:
Standards Met:ISO 5356 (Connectors)ISO 5356 (Connectors)ISO 5356 (Connectors)

{4}------------------------------------------------

Summary of Testing:

Both subject and predicate devices are for use in the hospital and home and are single patient use.

The target population is adult and pediatric patients using a single limb breathing tube. A pediatric population is defined as 10 kg in weight. The product is not for use in neonates.

The resistance to flow through the valve is the same for both predicate and the subiect valve.

The pressure ratio for the MK3b and predicate device are exactly the same, meaning it will provide the same performance on the ventilators.

The subject and predicate valve have the same principal of operation.

Although there are a few performance differences between the subject and the predicate devices the intended use is the same. They are used in the same configuration i.e. on a breathing system, connected to a ventilator, except the subject devices incorporate the connector into the design making it more economical, and saves on deadspace.

The leakage in the balloon and the main body is slightly higher in the predicate device than the subject devices, however this leakage is minimal. Intersurgical's internal requirement is for devices to leak less than 5ml/min, however most standard requirements are much larger than this. There is no standard requirement for exhalation valves. The less leakage the more efficient the device will be.

The trigger feature is an additional function which the predicate device is unable to perform, this is a benefit with the new valves and does not alter the intended use, it just means the subject valves can be used on more ventilators. The new valves are validated for patient trigged spontaneous breaths.

The predicate device was previously validated for 24 hours whereas the subject device have been shown to work without any impairment to the performance for 30 days. The predicate device has not been tested for this duration.

Nonclinical tests submitted to demonstrate substantial equivalence for the Exhalation Valve include Resistance to Flow, Pressure Ratio, Leakage and Tapers when compared to the legally marketed device. All materials used in the Exhalation Valve have been evaluated according to tests outlined in ISO 10993-1 and meet the requirements of Bluebook Memo, General Program Memorandum G95-1 biocompatibility testing for genetotoxicity, implantation, cytotoxicity, sensitization, and irritation. The Exhalation Valve connectors meet the requirements of Anesthetic and respiratory equipment - conical connectors: Part 1: Cones and Sockets ISO 5356-1:2004.

{5}------------------------------------------------

Substantial Equivalence:

Intersurgical Incorporated has demonstrated that the proposed device is as safe and as effective as the predicate device. It is considered to be substantially equivalent to the currently marketed predicate device which has been previously reviewed for market clearance by the FDA.

K132143

Premarket Notification [510(k)] Number

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

June 4, 2014

Intersurgical. Inc. Michael Zalewski Vice President, RA/QA/CS 417 Electronics Parkway Liverpool, NY 13088-6098

Re: K132143

Trade/Device Name: Exhalation valve Regulation Number: 21 CFR 868.5870 Regulation Name: Non-rebreathing Valve Regulatory Class: II Product Code: CBP Dated: May 05, 2014 Received: May 06, 2014

Dear Mr. Zalewski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Mr. Michael Zalewski

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

MarySBunner-S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Slatement on last page.

510(k) Number (if known) K132143

Device Name

Indications for Use (Describe)

Product # 1923500 - MK3 Exhalation Valve without a proximal pressure port Product # 1924501 - MK3 Exhalation Valve with a proximal pressure port Product # 1924504 - MK3b Exhalation Valve with a proximal pressure port

Product # 1924904 - RRS-30 Exhalation Valve with a proximal pressure port.

indications For Use: The exhalation valves are used to control Inspiratory pressure and expel the expired air from a patient being ventilated via a single limb breathing system. The exhalation valve and single limb breathing systems are used with adults/pediatios and prescribed by a physician. A pediation is defined as 10 kg in weight. The product is not for use in neonates. The device can be used within hospitals and for home care use. It is a single patient use device and ean be used for a maximum of 30 days.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Anya C. Harry -S
2014.06.04
05:38:03 -04'00'

§ 868.5870 Nonrebreathing valve.

(a)
Identification. A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.(b)
Classification. Class II (performance standards).