K Number
K131661

Validate with FDA (Live)

Date Cleared
2013-11-05

(151 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IMM™ Infusion Syringe Set is intended for delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation.

Device Description

IMM 110 Infusion Syringe Set is single use product which contains disposable parts of one spike infusion set for connecting to fluid reservoir and fluid delivery, and a control syringe for delivery fluid supply from a solution bag to access device for purpose of delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation. The distal end of the set is a dual valve with 6% taper design for connecting to other access device such as needles before use. The IMM Infusion Syringe Set is provided sterile in fluid path, and to be disposed after single patient use.

AI/ML Overview

The provided text describes a 510(k) submission for the IMM™ Infusion Syringe Set. This type of regulatory submission is for medical devices and focuses on demonstrating "substantial equivalence" to a predicate device already legally marketed, rather than proving novel clinical effectiveness through a complex study design like those typically seen for new drugs or high-risk devices.

Therefore, the study presented here is primarily a bench-top performance evaluation to show that the device meets established engineering and safety standards, and functions equivalently to existing devices. It is not a clinical study involving human patients or complex AI algorithms.

Here's a breakdown of the requested information based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

The document states that the device meets the specific requirements established in voluntary standards ISO8536-4 and ISO7886-1. Unfortunately, the specific acceptance criteria (e.g., minimum tensile strength, flow rate tolerances) within these standards, and the exact measured performance of the device against each criterion, are not detailed within this summary. The summary only provides a general statement of compliance.

However, based on the "Element of Comparison" table, some performance characteristics are implicitly mentioned.

Acceptance Criterion (Implicit from Predicate Comparison & Standards)Reported Device Performance
Tubing/fitting bonding strength: Tensile strength>15 N
Flow regulator Clamping efficacy: From zero to maximum in gravity useMet (Similar to predicate)
Sterility: Single useMet (Single use)
Biocompatibility: Materials tested per ISO10993Suitable for intended use
Overall Functional and Performance: Per ISO8536-4 and ISO7886-1Meets requirements

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the functional and performance tests.
The data provenance is not explicitly stated as "country of origin," but the manufacturing company is located in Miaoli, Taiwan. The studies appear to be retrospective bench-top tests conducted internally to support the 510(k) submission.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable in the context of this device and study. The "ground truth" here is compliance with engineering standards and material properties, not clinical diagnoses that would require expert human interpretation. The tests are physical and chemical evaluations.


4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for resolving discrepancies in clinical image interpretation or diagnostic outcomes. For bench-top engineering tests, the outcome is usually a quantitative measurement or a pass/fail against a standard, not subject to subjective adjudication.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (especially those involving image interpretation by human readers, potentially with AI assistance) to assess improvements in reader performance. The IMM™ Infusion Syringe Set is a manual fluid delivery device, not a diagnostic tool where human reader performance is a primary outcome.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

No, a standalone (algorithm only) performance study was not done. This concept is only relevant for AI/ML-based algorithms. The IMM™ Infusion Syringe Set is a physical medical device, not a software algorithm.


7. The Type of Ground Truth Used

The "ground truth" for this device's performance evaluation primarily relies on:

  • Engineering Standards: Compliance with international standards such as ISO8536-4 (Infusion equipment for medical use -- Part 4: Infusion sets for single use, gravity-fed) and ISO7886-1 (Sterile hypodermic syringes for single use -- Part 1: Syringes for manual use). These standards define mechanical properties, dimensional tolerances, flow rates, sterility, etc.
  • Material Compatibility/Safety Standards: Compliance with ISO10993 (Biological evaluation of medical devices).
  • Predicate Device Equivalence: The performance is also benchmarked for equivalence against the predicate devices (K063851 and K792227) in key areas like construction, transparency, bonding strength, flow regulation, syringe size, interface, material, and sterility.

8. The Sample Size for the Training Set

This information is not applicable. "Training set" refers to data used to train AI/ML algorithms. The IMM™ Infusion Syringe Set is a physical medical device, not an AI model.


9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no AI/ML algorithm or training set involved.

{0}------------------------------------------------

Innovative Medical Manufacturing Company

107, Lane 181, Sec.1, Yong Jane Rd., Chunan, Miaoli 350, Taiwan, ROC TEL:037-620236 FAX:037-620239 E-mail:service@immc.com.tw Unified Code:16980249

510(k) Summary of Safety and Effectiveness

1. General Information

Submitted by:Innovative Medical Manufacturing Company107, Lane 181, Sec. 1., Yong Jane RoadChunan, Miaoli 350, Taiwan (ROC)
Contact person:Lucy Huang, Regulatory SpecialistTel: +886-37-620236Fax: +886-37-620239
Date of preparation:10/14/2013
NOV 05 2013

2. Device Information

Common Name:Syringe Kit
Proprietary Name:IMMTM Infusion Syringe Set
ClassificationPiston Syringe; Intravascular Administration Set
Name:
Product code:FMF(880.5860)FPA (880.5440)

3. Predicate Device

Trade Name510(k) NumberDecision Date
Vascular Auto-Fill Syringe KitK0638512007/1/26
Multi-Ad Dispensing systemK7922271979/12/20

4. Device Description

IMM 110 Infusion Syringe Set is single use product which contains disposable parts of one spike infusion set for connecting to fluid reservoir and fluid delivery, and a control syringe for delivery fluid supply from a solution bag to access device for purpose of delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation. The distal end of the set is a dual valve with 6% taper design for connecting to other access device such as needles before use. The IMM Infusion Syringe Set is provided sterile in fluid path, and to be disposed after single patient use.

{1}------------------------------------------------

Innovative Medical Manufacturing Company

107, Lane 181, Sec.1, Yong Jane Rd., Chunan, Miaoli 350, Taiwan, ROC TEL:037-620236 FAX:037-620239 E-mail:service@immc.com.tw Unified Code:16980249

5. Indication for Use

IMM™ Infusion Syringe Set is intended for delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation.

6. Technological Characteristics

The IMM™ Infusion Syringe Set is single use device, and can be delivered in unit sterile package or in bulk non-sterile shipment for further packaging. It allows easy operation and replacement before use.

7. Substantial equivalence

The IMM™ Infusion Syringe Set has the same indication for use and similar characters with the predicated device. The technological characteristics of the IMM device are as same as the predicate device.

Element ofComparisonSubject DeviceK131661VascularK063851B. BraunK792227
Intended UseFor deliveringcontrolled volumeof fluids fromreservoirs oflabeled route ofadministration offluids by manualoperationSameSimilar
ConstructionSpike inlet, PVCtubing, dual checkvalve, syringeSimilar, exceptadditionalextensionSpike inlet, PVCtubing, dual checkvalve, syringe
TransparencyTubing of fluid lineis transparencySameSame
Tubing/fittingbondingstrengthTensile strength>15 NSimilarSimilar
Flow regulatorClampingefficacyFrom zero tomaximum ingravity useSameSame
Syringe12ml12ml10ml
InterfaceLuer fittingLuer fittingLuer fitting
MaterialPP, PVC, PC, ABSPP, PVC, PC, ABSSimilar
SterilitySingle useSameSame
SterilizationmethodEtOEtOEtO

{2}------------------------------------------------

Innovative Medical Manufacturing Company

107, Lane 181, Sec.1, Yong Jane Rd., Chunan, Miaoli 350, Taiwan, ROC TEL: 037-620236 FAX:037-620239 E-mail:service@immc.com.tw Unified Code:16980249

8. Performance Summary

The non-clinical functional and performance tests demonstrated that IMM™ Infusion Syringe Kit meets the specific requirements established in voluntary standards: ISO8536-4 and ISO7886-1.

9. Material

The Infusion syringe set is composed of materials that have been tested in accordance with ISO10993 and have been determined to be suitable for the intended use of this product.

10. Safety and Effectiveness

All verification and validation test data indicated that the IMM Infusion syringe set can perform as intended use and is substantially equivalent to the predicate device.

5-3

{3}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 5,2013

Innovative Medical Manufacturing Company Ms. Lucy Huang Regulation Specialist No .. 107, Lane 181, Sect 1., Yong Jane Road Chunan, MIAOLI CHINA 350

Re: K131661

Trade/Device Name: IMM™ Infusion Syringe Set Regulation Number: 21 CFR 880.5860, 21 CFR 880.5840 Regulation Name: Piston Syringe, Intravascular Administration Set Regulatory Class: II Product Code: FMF, FPA Dated: September 13, 2013 Received: September 25, 2013

Dear Ms.Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Huang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/7 description: The image shows the text "Mary Surfammer -S". The text is in a bold, sans-serif font. The word "Mary" is in a larger font size than the rest of the text. The word "Surfammer" is stylized with a wave-like design. The "-S" is in a smaller font size than the rest of the text.

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K131661 510(k) Number (if known):

Device Name: IMM™ Infusion Syringe Set

Indications for Use:

IMM™ Infusion Syringe Set is intended for delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation.

Prescription Use Over-The-Counter Use (Part 21 CFR 801 Subpart AND/OR (21 CFR 801 Subpart C) D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of __ 1_

Image /page/5/Picture/9 description: The image shows the logo of the Food and Drug Administration (FDA). The logo is in black and white and features the letters "FDA" in a stylized, geometric font. The letters are outlined with multiple lines, giving them a three-dimensional appearance. The "A" in FDA is shaped like a triangle.

Richard C. Chapman 2013.11.04 16:10:11 -05'00'

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).