K Number
K131584

Validate with FDA (Live)

Date Cleared
2013-08-21

(82 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMDx Flu A/B and RSV for Abbott m2000 assay performed on the Abbott m2000 System is a qualitative in vitro diagnostic test designed for the detection of influenza B, and RSV RNA directly from nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. The assay detects RNA from influenza A. influenza B, and RSV (A and B) using real-time, reverse transcription polymerase chain reaction (rRT-PCR) and fluorogenic target specific hybridization for unique sequences in the viral target genomes. The IMDx Flu A/B and RSV for Abbott m2000 assay is intended for use as an aid in diagnosing influenza A and/or RSV infection.

Negative results for influenza B, or RSV do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment, or patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered when diagnosing respiratory viral infection.

Performance characteristics for influenza A were established during the 2011-2013 influenza seasons when Influenza A/2009 H1N1 and Influenza A/H3 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary,

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The IMDx Flu A/B and RSV for Abbott m2000 assay uses nucleic acid extraction and purification technology, performed on the Abbott m2000sp. combined with rRT-PCR. performed with the Abbott m2000rt, to generate and detect amplified products from influenza A, influenza B, and RSV RNA that is isolated from clinical specimens.

The assay targets the influenza A matrix (M) gene, influenza B non-structural protein (NS1) gene, and RSV A and RSV B fusion (F) gene. The presence of a viral RNA target sequence is indicated by the fluorescent signal generated through the use of fluorescently labeled oligonucleotide probes on the Abbott m2000rt instrument. The probes do not generate a signal unless they are specifically bound to the amplification cycle at which fluorescent signal is detected by the Abbott m2000rt is inversely proportional to the viral RNA target concentration present in the original sample.

An RNA bacteriophage species unrelated to influenza B, or RSV is introduced into each specimen during sample preparation to serve as a process control bacteriophage is lysed simultaneously with influenza A, influenza B and RSV B in the sample, and amplified in the same reaction as the viral RNA targets using rRT-PCR. The process control serves to demonstrate that the entire assay process has proceeded correctly for each sample.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the IMDx Flu A/B and RSV for Abbott m2000 assay:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for diagnostic devices often involve achieving a certain level of sensitivity and specificity against a defined ground truth. Based on the provided summary, the FDA cleared the device, implying that the observed performance met the agency's criteria. While explicit "acceptance criteria" are not listed as target percentages, the "Clinical Agreement Summary" tables present the device's performance metrics.

Target OrganismPerformance MetricAcceptance Criteria (Implied by Clearance)Reported Device Performance (95% CI)
Influenza ASensitivityMet FDA clearance standards97.6% (94% - 99%)
SpecificityMet FDA clearance standards95.7% (94% - 97%)
Influenza BSensitivityMet FDA clearance standards97.1% (90% - 99%)
SpecificityMet FDA clearance standards97.2% (96% - 98%)
RSVSensitivityMet FDA clearance standards97.2% (92% - 99%)
SpecificityMet FDA clearance standards93.0% (91% - 95%)

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: A total of 932 nasopharyngeal swab (NPS) samples were tested and analyzed for performance across two influenza seasons (2011-2012 and 2012-2013).
  • Data Provenance: The samples were prospectively collected from four (2011-2012 season) and three (2012-2013 season) geographically diverse test sites within the United States. These were nasopharyngeal swabs collected for routine influenza testing. The study period was February-April 2012 and March-April 2013.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that the performance was assessed "compared to viral culture followed by direct fluorescent antibody (DFA) detection." This implies laboratory techniques were used as the reference standard, rather than expert clinical consensus or interpretation of images by experts.

4. Adjudication Method for the Test Set

The document mentions that "Discordant samples were tested using the Nanosphere Verigene® Respiratory Virus Plus Nucleic Acid Test (RV+) on the Verigene® System and subsequent results are documented in the footnote." This indicates that discordant results (where the IMDx assay result did not match the initial ground truth) were adjudicated using a "third-party" FDA-cleared molecular assay (the predicate device).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not conducted. This device is an in vitro diagnostic test for detecting viral RNA, not an imaging device or a tool that directly assists human readers in interpreting clinical information.

6. Standalone (Algorithm Only) Performance Study

  • Yes, a standalone performance study was done. The entire study described under "Clinical Agreement" and "Analytical Performance" evaluates the algorithm's performance in detecting viral RNA without human interpretation or intervention in the diagnostic process beyond running the assay. The results presented are exclusively for the device's analytical detection capabilities.

7. Type of Ground Truth Used

  • Clinical Agreement: The primary ground truth for the clinical agreement study was viral culture followed by direct fluorescent antibody (DFA) detection. For discordant samples, a FDA cleared molecular assay (Nanosphere Verigene® Respiratory Virus Plus Nucleic Acid Test) was used for adjudication.
  • Analytical Performance: For analytical studies (reproducibility, analytical sensitivity, reactivity, specificity, interference), the ground truth generally refers to precisely prepared "spiked" samples with known concentrations of viral strains or challenging substances.

8. Sample Size for the Training Set

The document does not provide information regarding the sample size for a training set. This is typical for FDA 510(k) summaries for IVD assays. The development and training of such assays often involve proprietary internal validation data not typically released in a 510(k) summary, which focuses on clinical validation for market clearance.

9. How the Ground Truth for the Training Set Was Established

As with the training set sample size, the document does not provide information on how the ground truth for any potential training set was established. This information is typically considered proprietary to the manufacturer's development process and not part of the public 510(k) summary.

{0}------------------------------------------------

510(k) SUMMARY

K131584

Date of Summary:August 14, 2013
Product NameIMDx Flu A/B and RSV for Abbott m2000
SponsorIntelligent Medical Devices, Inc.19 Blackstone StreetCambridge, MA 02139
CorrespondentMDC Associates, LLCFran White, Regulatory Consultant180 Cabot StreetBeverly, MA 01915

AUG 2 1 2013

Trade or Proprietary Name:IMDx Flu A/B and RSV for Abbott m2000
Common or Usual Name:Respiratory Virus Panel Nucleic Acid Assay System
Product Code:OCC, OOI
Regulation Section:21CFR866.3980
Device Class:Class II

Microbiology (83)

Intended Use

Panel:

The IMDx Flu A/B and RSV for Abbott m2000 assay performed on the Abbott m2000 System is a qualitative in vitro diagnostic test designed for the detection of influenza B, and RSV RNA directly from nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. The assay detects RNA from influenza A. influenza B, and RSV (A and B) using real-time, reverse transcription polymerase chain reaction (rRT-PCR) and fluorogenic target specific hybridization for unique sequences in the viral target genomes. The IMDx Flu A/B and RSV for Abbott m2000 assay is intended for use as an aid in diagnosing influenza A and/or RSV infection.

Negative results for influenza B, or RSV do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment, or patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered when diagnosing respiratory viral infection.

Performance characteristics for influenza A were established during the 2011-2013 influenza seasons when Influenza A/2009 H1N1 and Influenza A/H3 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary,

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

{1}------------------------------------------------

Device Description

The IMDx Flu A/B and RSV for Abbott m2000 assay uses nucleic acid extraction and purification technology, performed on the Abbott m2000sp. combined with rRT-PCR. performed with the Abbott m2000rt, to generate and detect amplified products from influenza A, influenza B, and RSV RNA that is isolated from clinical specimens.

The assay targets the influenza A matrix (M) gene, influenza B non-structural protein (NS1) gene, and RSV A and RSV B fusion (F) gene. The presence of a viral RNA target sequence is indicated by the fluorescent signal generated through the use of fluorescently labeled oligonucleotide probes on the Abbott m2000rt instrument. The probes do not generate a signal unless they are specifically bound to the amplification cycle at which fluorescent signal is detected by the Abbott m2000rt is inversely proportional to the viral RNA target concentration present in the original sample.

An RNA bacteriophage species unrelated to influenza B, or RSV is introduced into each specimen during sample preparation to serve as a process control bacteriophage is lysed simultaneously with influenza A, influenza B and RSV B in the sample, and amplified in the same reaction as the viral RNA targets using rRT-PCR. The process control serves to demonstrate that the entire assay process has proceeded correctly for each sample.

Substantial Equivalency

The IMDx Flu A/B and RSV for Abbott m2000 assay is substantially equivalent to the Verigene® Respiratory Virus Plus Nucleic Acid Test on the Verigene® System (RV+) (K103209). The tables below compare the characteristics of the IMDx Flu A/B and RSV for Abbott m2000 Assay (New Device) and the Nanosphere Verigene Assay (Predicate Device).

Similarity to Predicate DeviceCharacteristicIMDx Flu A/B and RSV for Abbott m2000Verigene® Respiratory Virus Plus Nucleic Acid Test on the Verigene® System (RV+)
510(k)K131584K103209
Regulation866.3980866.3980
Product CodeOCC, OOIOCC; NSU
Device ClassClass IIClass II
Intended useThe IMDx Flu A/B and RSV for Abbottm2000 assay performed on the Abbott m2000System is a qualitative in vitro diagnostic testdesigned for the detection of influenza A,influenza B, and RSV RNA directly fromnasopharyngeal swabs from patients with signsand symptoms of respiratory infection. Theassay detects RNA from influenza A, influenzaB, and RSV (A and B) using real-time, reversetranscription polymerase chain reaction (rRT-PCR) and fluorogenic target specifichybridization for unique sequences in the viraltarget genomes. The IMDx Flu A/B and RSVfor Abbott m2000 assay is intended for use asan aid in diagnosing influenza A and/orinfluenza B and/or RSV infection.The Verigene® Respiratory Virus Plus NucleicAcid Test (RV+) on the Verigene® System is aqualitative nucleic acid multiplex test intended tosimultaneously detect and identify multiplerespiratory virus nucleic acids in nasopharyngeal(NP) swab specimens from individuals withsigns and symptoms of respiratory tractinfection. The following virus types andsubtypes are identified using the RV+: influenzaA, influenza A subtype H1, influenza A subtypeH3, 2009 H1N1, influenza B, RespiratorySyncytial Virus (RSV) subtype A, and RSVsubtype B. The test is not intended to detectinfluenza C virus. Detecting and identifyingspecific viral nucleic acids from individualsexhibiting signs and symptoms of respiratoryinfection aids in the diagnosis of respiratory viral
CharacteristicIMDx Flu A/B and RSV for Abbott m2000Verigene® Respiratory Virus Plus NucleicAcid Test on the Verigene® System (RV+)
Negative results for influenza A, influenza B,or RSV do not preclude influenza virus orRSV infection and should not be used as thesole basis for diagnosis, treatment, or patientmanagement decisions. Conversely, positiveresults do not rule-out bacterial infection or co-infection with other viruses. The agent(s)detected may not be the definite cause ofdisease. The use of additional laboratorytesting and clinical presentation must beconsidered when diagnosing respiratory viralinfection.infection, if used in conjunction with otherclinical and laboratory findings.Negative results for influenza A, influenza B, orRSV do not preclude influenza virus or RSVinfection and should not be used as the sole basisfor diagnosis, treatment, or patient managementdecisions. Conversely, positive results do notrule-out bacterial infection or co-infection withother viruses. The agent detected may not be thedefinite cause of disease. The use of additionallaboratory testing and clinical presentation mustbe considered in order to obtain the finaldiagnosis of respiratory viral infection.
Performance characteristics for influenza Awere established during the 2011-2012 and2012-2013 influenza seasons when influenzaA/2009 H1N1 and influenza A/H3 were thepredominant influenza A viruses in circulation.When other influenza A viruses are emerging,performance characteristics may vary.Performance characteristics for influenza AVirus were established when influenza A/H3,A/H1, and 2009 H1N1 were the predominantinfluenza A viruses circulating. Thesecharacteristics may vary when other influenza Aviruses are emerging.
If infection with a novel influenza A virus issuspected based on current clinical andepidemiological screening criteriarecommended by public health authorities,specimens should be collected withappropriate infection control precautions fornovel virulent influenza viruses and sent tostate or local health department for testing.Viral culture should not be attempted in thesecases unless a BSL 3+ facility is available toreceive and culture specimens.If infection with a novel influenza A virus issuspected based on current clinical andepidemiological screening criteria recommendedby public health authorities, specimens should becollected with appropriate infection controlprecautions used specifically for novel virulentinfluenza viruses and sent to state or local healthdepartment for testing. Viral culture should notbe attempted in these cases unless a BSL 3+facility is available to receive and culturespecimens.
Sample typeNasopharyngeal swabsNasopharyngeal swabs
SamplePreparationAutomated extraction of nucleic acidsAutomated extraction of nucleic acids
Test PrincipleReal-time, reverse transcription polymerasechain reaction (rRT-PCR) DNA amplificationReal-time, reverse transcription polymerasechain reaction (RT-PCR) DNA amplification
TargetsDetectedinfluenza Ainfluenza BRSV A/Binfluenza Ainfluenza BRSV ARSV B
ControlsPositive ControlNegative ControlProcess ControlPositive ControlNegative ControlInhibition ControlInternal Control
InstrumentationAbbott® m2000TM System (K092705)Verigene ®System (K070804)

Similarity to Predicate Device

{2}------------------------------------------------

.

{3}------------------------------------------------

Differences from Predicate Device
CharacteristicIMDx Flu A/B and RSV for Abbott m2000Verigene® Respiratory Virus Plus Nucleic
Acid Test on the Verigene® System (RV+)
ThroughputBatchSingle use cassette
Viral Sub-NoYes
Typing

These differences do not affect substantial equivalency of the IMDx Flu A/B and RSV for Abbott m2000 and Verigene® Respiratory Virus Plus Nucleic Acid Test. Both assays detect influenza B and RSV viral nucleic acids from nasopharyngeal swabs and the assays have comparable intended uses. The differences noted above do not change the intended use and do not raise different questions of safety and effectiveness.

Performance Characteristics

Clinical Agreement

The performance of the IMDx Flu A/B and RSV for Abbott m2000 assay was assessed compared to viral culture followed by direct fluorescent antibody (DFA) detection during the course of two (2) influenza seasons (2011-2012 and 2012-2013). For the 2011-2012 season (collected February-April 2012), four (4) geographically diverse test sites within the United States prospectively collected influenza A/B and RSV samples enrolled for this study were nasopharyngeal swabs collected for routine influenza testing. A total of four hundred and ninety seven (497) specimens were included in the final data set and analyzed for product performance.

For the 2012-2013 season (collected March-April 2013), three (3) geographically diverse test sites within the United States prospectively collected influenza A/B and RSV samples enrolled for this study were nasopharyngeal swabs collected for routine influenza testing. A total of four hundred and thirty-five (435) specimens were included in the final data set and analyzed for product performance.

Patient Population
Age and Gender Distribution
AgeFemaleMale
2011-20122012-20132011-20122012-2013
≤5 years84 (30.1%)71 (31.8%)76 (34.7%)101 (47.6%)
6 - 21 years38 (13.6%)29 (13.0%)45 (20.5%)28 (13.2%)
22 - 59 years96 (34.4%)76 (34.1%)63 (28.9%)44 (20.7%)
≥ 60 years61 (21.9%)47 (21.1%)34 (15.5%)39 (18.4%)
Season Totals279223218212
Overall Totals502430

Between the two influenza seasons, a total of 932 NPS samples were tested and analyzed for performance. Results are shown below. Discordant samples were tested using the Nanosphere Verigene® Respiratory Virus Plus Nucleic Acid Test (RV+) on the Verigene® System and subsequent results are documented in the footnote.

{4}------------------------------------------------

Influenza A Clinical Agreement Summary

All SitesInfluenza AViral Culture
PositiveNegativeTotal
IMDx Flu A/Band RSV forAbbott m 2000AssayPositive164331197Sensitivity97.6% (94% - 99%) 95% CI
Negative42731735Specificity95.7% (94% - 97%) 95% CI
Total168764932

125/33 samples were confirmed as influenza A Positive using an FDA cleared molecular assay.

24/4samples were confirmed as influenza A Negatives using an FDA cleared nucleic acid test.

Influenza B Clinical Agreement Summary

All SitesViral Culture
Influenza BPositiveNegativeTotal
IMDx Flu A/BPositive6724391Sensitivity97.1% (90% - 99%) 95% CI
and RSV forAbbott m2000AssayNegative2839841Specificity97.2% (96% - 98%) 95%CI
Total69863932

320/24 samples were confirmed as influenza B Positive using an FDA cleared molecular assay.

RSV Clinical Agreement Summary

All SitesRSVViral Culture
PositiveNegativeTotal
IMDx Flu A/Band RSV forAbbott m2000AssayPositive104584162Sensitivity97.2% (92% - 99%) 95% CI
Negative35767770Specificity93.0% (91% - 95%) 95%CI
Total69825932

439/58 samples were confirmed as RSV Positive using an FDA cleared molecular assay.

51/3 samples were confirmed as RSV Positive using an FDA cleared molecular assay.

Analytical Performance

Reproducibility

A sixteen-member panel was used for the reproducibility study. The panel was made with 2 influenza A strains (one HINI and one H3N2), one influenza B strain, one RSV-A strain. Each strain. Each strain was spiked in M4RT viral transport medium at three different approximate target levels: Moderate Positive (~2-3X LoD), Low Positive (1X LoD) and High Negative (0.2 - 0.8X LoD). Panel members were formulated with only one target present (influenza A, influenza B, RSV-A, RSV-B). A negative sample (transport medium only) was also included.

Each panel member was tested in replicates of three, twice a days, for a total of 10 experiment runs. Testing was conducted at three site, the runs were performed by two operators, with each operator performing one run each day.

To conduct the site-to-site reproducibility study, panel members (moderate positive, high negative, and negative) for each virus were randomized and sample identities were blinded to the user. Each panel member was tested in replicates of three, twice a day for five days, for a total of 10 runs. Testing was conducted at three sites. At cach site, the runs were performed by two operator performing one run each day. The entire study was conducted using the same instrument system (Abbott m2000rt) at each site, and one reagent lot of

{5}------------------------------------------------

the IMDx Flu A/B and RSV for Abbott m2000 kits. The aggregate percent CV values across all sites were ≤ 3.7% for all targets, and for all concentrations tested.

Site 1Site 2Site 3All 3 Sites
Specific PanelMemberLevel% Agreement(Agreementwith expectedresult)Avg. CN(%CV)% Agreement(Agreementwith expectedresult)Avg. CN(%CV)% Agreement(Agreementwith expectedresult)Avg. CN(%CV)% Agreement(95% CI)Avg. CN(%CV)
ModeratePositive100.00(30/30)33.52(2.46)100.00(30/30)34.28(2.17)93.33(28/30)33.76(2.58)97.78(94.73 - 100.00)33.85(2.28)
influenza A(H1N1)Low Positive100.00(30/30)34.95(1.67)100.00(30/30)35.83(2.41)96.67(29/30)35.16(3.21)97.78(94.73 - 100.00)35.31(2.35)
HighNegative36.67(11/30)36.92(4.67)33.33(10/30)37.88(1.99)33.33(10/30)37.43(3.31)34.44(24.63 - 44.26)37.38(3.04)
ModeratePositive100.00(30/30)33.26(1.77)100.00(30/30)33.65(1.37)100.00(30/30)33.31(1.59)100.00(100.00 - 100.00)33.41(1.53)
Influenza A(H3N2)Low Positive100.00(30/30)34.61(1.29)100.00(30/30)34.85(1.22)100.00(30/30)35.04(1.96)100.00(100.00 - 100.00)34.83(1.50)
HighNegative6.67(2/30)36.53(1.41)16.67(5/30)37.31(1.81)20.00(6/30)37.28(1.88)14.44(7.18-21.71)37.04(1.72)
ModeratePositive100.00(30/30)32.25(1.37)96.67(29/30)32.18(1.33)100.00(30/30)32.22(0.79)98.89(96.72 - 100.00)32.22(1.13)
Influenza BLow Positive96.67(29/30)33.91(0.81)96.67(29/30)33.89(1.21)73.33(22/30)33.89(1.00)88.89(82.40 - 95.38)33.90(1.01)
HighNegative30.00(9/30)34.53(1.07)36.67(11/30)34.74(0.90)53.33(16/30)34.50(1.07)40.00(29.88 - 50.12)34.53(0.93)
ModeratePositive100.00(30/30)32.62(3.52)100.00(30/30)32.10(4.82)93.33(28/30)32.47(2.60)97.78(94.73 - 100.00)32.39(3.64)
RSV-ALow Positive96.67(29/30)33.63(2.15)96.67(29/30)33.57(2.47)100.00(30/30)33.15(3.21)97.78(94.73 - 100.00)33.45(2.52)
HighNegative6.67(2/30)34.93(2.80)6.67(2/30)35.03(1.90)23.33(7/30)34.50(3.07)12.22(5.46 - 18.99)34.82(2.48)
ModeratePositive96.67(29/30)32.68(1.90)100.00(30/30)32.39(2.14)90.00(27/30)31.78(2.54)95.56(91.30 - 99.81)32.29(2.18)
RSV-BLow Positive86.67(26/30)35.20(1.91)86.67(26/30)35.21(1.50)70.00(21/30)34.63(2.14)81.11(73.02 - 89.20)35.02(1.83)
HighNegative70.00(21/30)36.78(0.95)86.67(26/30)36.55(1.20)80.00(24/30)36.06(1.76)78.89(70.46 - 87.32)36.23(0.83)
NegativeNegative100.00(30/30)-1.00(0.00)100.00(30/30)-1.00(0.00)100.00(30/30)-1.00(0.00)100.00(100.00 - 100.00)-1.00(0.00)

Analytical Sensitivity (Limit of Detection)

The Analytical Sensitivity, or Limit of Detection (LoD), is defined as the concentration at which ≥ 95% of all replicates tested positive. The LoD for the IMDx Flu A/B and RSV for Abbott m2000 assay was determined using two strains of influenza A, two strains of influenza B, one strain of RSV-A, and one strain of RSV-B. Volumes of each dilution tested were in accordance with sample processing instructions provided in this package insert, and are representative of the amount of material collected by swab sampling methods. The results are summarized in the table below.

StrainLoD
Influenza A/Swine/NY/02/09 (H1N1)$3.9 x 10^0$ TCID50/mL
Influenza A/Brisbane/10/07 (H3N2)$1.51 x 10^1$ TCID50/mL
Influenza B/Florida/04/2006$2.82 x 10^2$ TCID50/mL
RSV A (RSVA Type A)$4.17 x 10^0$ TCID50/mL
RSV B CH93-(18)-18$1.65 x 10^0$ TCID50/mL
StrainConcentration Tested
A/California/7/2009 (H1N1)1.07 x 102 CEID50/mL
A/New Caledonia/20/99 (H1N1)2.62 x 101 TCID50/mL
A/Solomon Islands/3/2006 (H1N1)6.19 x 100 TCID50/mL
A/PR/8/34 (H1N1)1.56 x 100 TCID50/mL
A/WS/33 (H1N1)4.88x 10-1 CEID50/mL
A/Brisbane/59/07 (H1N1)2.49 x 101 TCID50/mL
A/Swine/Canada/6294/09 (H1N1)6.57 x 100 TCID50/mL
A/NJ/8/76 (H1N1)5.91 x 100 TCID50/mL
A/NWS/33 (H1N1)4.05 x 100 CEID50/mL
A/FM/1/47 (H1N1)1.20 x 101 CEID50/mL
A/Mal/302/54 (H1N1)1.86 x 100 CEID50/mL
A/Denver/1/57 (H1N1)7.48 x 101 CEID50/mL
A/Virginia/ATCC2/2009 (H1N1)5.91 x 100 CEID50/mL
A/Wisconsin/67/05 (H3N2)4.85 x 101 TCID50/mL
A/MRC2 (H3N2)1.56 x 102 CEID50/mL
A/Aichi/2/26 (H3N2)2.85 x 102 CEID50/mL
A/Victoria/3/75 (H3N2)8.87 x 100 CEID50/mL
A/Port Chalmers/1/73 (H3N2)3.64 x 102 TCID50/mL
A/Perth/16/09 (H3N2)3.79 x 100 TCID50/mL
A/Hong Kong/8/68 (H3N2)5.10 x 100 TCID50/mL
A/Rhode Island/01/2010 (H3N2)1.50 x 103 TCID50/mL
A/New York/55/2004 (H3N2)2.58 x 102 TCID50/mL
A/Uruguay/716/2007 (H3N2)9.79 x 102 TCID50/mL
A/Florida/2/2006 (H3N2)2.28 x 102 TCID50/mL
A/Victoria/361/2011 (H3N2)3.84 x 102 CEID50/mL
A/Indiana/10/2011 (H3N2v)4.61 x 102 TCID50/mL
A/Texas/71/2007 (H3N2v)1.56 x 100 TCID50/mL
A/Indiana/08/2011 (H3N2v)2.60 x 100 TCID50/mL
B/Mass/3/661.78 x 101 CEID50/mL
B/Allen/451.00 x 104 CEID50/mL
B/Lee/401.00 x 104 CEID50/mL
B/Maryland/1/591.00 x 103 CEID50/mL
B/Florida/07/041.82 x 101 TCID50/mL
B/Florida/02/20063.43 x 100 TCID50/mL
B/Hong Kong/5/721.19 x 101 CEID50/mL
B/RUSSIA/699.99 x 100 CEID50/mL
B/TAIWAN/2/62 (93-02)1.00 x 103 CEID50/mL
B/GL/1739/541.00 x 106 CEID50/mL
B/Brisbane/60/082.01 x 10-1 TCID50/mL
B/Wisconsin/01/20103.20 x 100 CEID50/mL
B/Santiago/4360/20071.61 x 100 CEID50/mL
B/Texas/39/20065.91 x 101 CEID50/mL
B/Ohio/01/20051.19 x 102 CEID50/mL
RSV/A27.42 x 100 TCID50/mL
RSVA/Long7.08 x 103 TCID50/mL
RSVA 1998/12-212.10 x 101 TCID50/mL
RSVA 1998/3-26.97 x 100 TCID50/mL
RSVA 2001/2-204.09 x 100 TCID50/mL
RSVA 2001/3-129.97 x 100 TCID50/mL
RSVB/93208.39 x 100 TCID50/mL
RSVB/WV/14617/851.79 x 100 TCID50/mL

Analytical Reactivity

A panel of 55 strains (31 influenza A, 15 influenza B, and 9 RSV) was tested for their ability to be detected by the IMDx Flu A/B and RSV for Abbott m2000 assay. Each strain was diluted in viral transport media and tested in triplicate. All strains tested were positive for their representative assay targets.

{6}------------------------------------------------

.

:

{7}------------------------------------------------

StrainConcentration Tested
A/duck/Pennsylvania/10218/84(H5N2)*23 pg/μL
A/HongKong/33982/2009 (H9N2)*57 pg/μL

Denotes testing on purified genomic RNA

Analytical Specificity: Cross Reactivity

A panel of 36 organisms from respiratory pathogens or human microbiota, and purified human DNA were tested at approximately 1x106 CFU/mL for bacteria, 1x105 TCIDs/mL for viruses, and 1.0x104 copies/mL for human DNA using the IMDx Flu A/B and RSV for Abbott m2000 assay. No cross reactivity was observed for the IMDx Flu A/B and RSV for Abbott m2000 assay.

Microbial Interference

The same panel of 36 organisms from respiratory pathogens or human microbiota, and purified human DNA at concentrations of approximately 1x10 CFU/mL for bacteria, 1x10 TCID-30/mL for viruses, and 1.0x10 copies/mL for human DNA were added to sample tubes containing one of six target organisms; two influenza B and two RSV strains in viral transport medium. The two influenza A strains used for this study were influenza A/Swine/NY/02/2009 (H1N1) and influenza A/Brisbane/10/2007 (H3N2). The two influenza B strains used for this study were influenza B/Florida/04/2006 (V/87) and influenza B/Malaysia/2506/2004 (YA/88). The two RSV strains used for this study were RSV A Type A and RSV B CH93-18(18). Assay analytes were present in the samples at concentrations corresponding to 2-3X LoD. Each of the six strains was tested against the 36 member panel in triplicate using the IMDx Flu A/B and RSV for Abbott m2000 assay. No evidence of interference was observed.

Organism
Adenovirus type I
Adenovirus type 7A
Bordetella pertussis
Candida albicans
Coronavirus
Corynebacterium ulcerans
Coxsackievirus
Cytomegalovirus (CMV)
Epstein-Barr Virus (EBV)
Escherichia coli
Haemophilus influenza
Human Herpes Virus 6 (HHV6), Z29 strain
Human Herpes Virus 7 (HHV7), SB Strain
Human genomic DNA*
Klebsiella pneumoniae
Lactobacillus acidophilus Z048
Legionella pneumoniae
Moraxella catarrhalis
Mycoplasma hominis
Mycoplasma pneumoniae
Neisseria meningitidis
Neisseria gonorrhoeae
Parainfluenza virus 1
Parainfluenza virus 2
Parainfluenza virus 3

{8}------------------------------------------------

Organism
Pseudomonas aeruginosa
Staphylococcus aureus MRSA
Staphylococcus aureus MSSA
Staphylococcus epidermidis MRSE
Streptococcus pneumoniae
Streptococcus salivarius
Measles Virus
Mumps
Metapneumovirus 3 type B1
Metapneumovirus 9 type A1
Rhinovirus
Influenza A/Swine/NY/02/09**
Influenza A/Brisbane/10/07**
Influenza B/Florida/04/06**
Influenza B/Malaysia/2506/04**
RSVA Type A strain**
RSVB CH93-18(18)**

Human Genomic DNA ≥ 1.0 X10
** * Tested in competitive interference study

** Tested in competitive interference study.

Competitive Interference

A competitive interference, or co-infection, study was performed to test whether a high titer of one virus would interfere with the detection of a second target virus that was present at low titer. High titered samples were formulated at concentrations of 2 x 10 TCID30mL, and low titered targets were formulated at a concentration of approximately 2-3X LoD for that strain. One strain each of influenza B, and RSV was used in this study. Each virus type was tested at low titers in conjunction with a high titer of each of the other two virus types, and samples were tested in triplicate. There was no observed interference with co-infection samples from the competitive interference study; high concentration targets did not interfere with the detection of a second target that was near LoD.

Potentially Interfering Substances

The susceptibility of the IMDx Flu A/B and RSV for Abbott m2000 assay to interference by elevated levels of cndogenous substances or exogenous was evaluated. Six viral strains were evaluated: two influenza A strains: A/Swine NY/02/09 (H/N) and A/Brisbane/10/07 (H3N2); two influenza B strains: B/Florida/07 and B/Malaysia/2506/04; two RSV strains: RSV-A and RSV-B CH93-(18)-18. These test organisms were diluted to a concentration of 2-3X LoD using M4RT viral transport medium. The test panel consisted of substances that may be found in nasopharyngeal swab specimens (listed in the Table below). Testing was conducted using one lot of IMDx Flu A/B and RSV for Abbott m2000 amplification reagents. Substances were diluted in viral transport media to concentrations that would either replicate or exceed the highest concentration expected to be found in a clinical sample. Each of the 6 viral strains was tested in triplicate in the presence of each substance. No interference in the performance of the IMDx Flu A/B and RSV for Abbott m2000 assay was observed.

SubstanceActive Ingredients and Potential Interferents inSubstance
Nasal SpraysOxymetazoline
Phenylephrine
Sodium chloride (with preservatives)
Nasal GelGalphimia glauca
Luffa operculata
Sulfur

{9}------------------------------------------------

SubstanceActive Ingredients and Potential Interferents inSubstance
Fluticasone
Mometasone furoate
Budesonide
Flunisolide
Triamcinolone acetonide
Beclomethasone
NSAIDsAspirin
Ibuprofen
Naproxen
AcetaminophenAcetaminophen
Relenza®Zanamivir
Antibacterial, systemicTobramycin
BenzocaineBenzocaine
Antibiotic nasal ointmentMupirocin
Allergy medicineHistamine hydrochloricum
Mucus (bovine)Mucin protein, type 1-S
Blood (Human)Whole Blood with EDTA

In addition to the substances listed above, the FluMist® influenza vaccine was tested for its ability to interfere with the IMDx Flu A/B and RSV for Abbott m2000 assay. FluMist contains live attenuated reassortants of each of the three A/California/07/2009 (H1N1), Influenza A/Perth/16/2009 strains: Influenza (H3N2), and Influenza B/Brisbane/60/2008 and is administered intranasally. Initial testing of FluMist® influenza vaccine at a concentration of 10%5 to 1015 units gave detected signals for influenza B, but the influenza A and B CN values were higher than expected, indicating that the assay detected the viruses present in the vaccine. When the vaccine was diluted to 1 x 10 ° and then tested using the IMDx Flu A/B and RSV for Abbott m2000 assay, influenza A and B were not detected. When the diluted vaccine was used for interference was observed; all test strains of influenza A, influenza B, and RSV were detected as expected.

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features a stylized caduceus, a symbol often associated with healthcare, with three intertwined strands. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

INTELLIGENT MEDICAL DEVICES, INC. C/O FRAN WHITE, MDC Associates, LLC. 180 CABOT STREET BEVERLY MA 01915

August 21, 2013

Re: K131584

Trade/Device Name: IMDx Flu A/B and RSV for Abbott m2000 Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: II Product Code: OCC, OOI Dated: May 29, 2013 Received: May 31, 2013

Dear Ms. White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements moset forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{11}------------------------------------------------

Page 2-Ms. White

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sally A Hojvat -S

Sally Hojvat, M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indications for Use

510(k) Number (if known): K131584

Device Name: IMDx Flu A/B and RSV for Abbott m2000

Indications for Use:

The IMDx Flu A/B and RSV for Abbott m2000 assay performed on the Abbott m2000 System is a qualitative in vitro diagnostic test designed for the detection of influenza B, and RSV RNA directly from nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. The assay detects RNA from influenza A. influenza B, and RSV (A and B) using real-time, reverse transcription polymerase chain reaction (rRT-PCR) and fluorogenic target specific hybridization for unique sequences in the viral target genomes. The IMDx Flu A/B and RSV for Abbott m2000 assay is intended for use as an aid in diagnosing influenza A and/or RSV infection.

Negative results for influenza B, or RSV do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment or patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered when diagnosing respiratory viral infection.

Performance characteristics for influenza A were established during the 2011-2013 influenza seasons when Influenza A/2009 H1N1 and Influenza A/H3 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices and Radiological Health (OIR)

Tamara V. Feldblyum -S 2013.08.15 15:53:36 -04'00'

Page I of I

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.