K Number
K092705

Validate with FDA (Live)

Date Cleared
2010-05-28

(267 days)

Product Code
Regulation Number
862.2570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Abbott m2000 system is intended for in vitro diagnostic use in performing FDA cleared and approved nucleic acid testing in clinical laboratories. It comprises the Abbott m2000sp and the Abbott m2000rt instruments. The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing. The Abbott m2000rt is an automated system for performing fluorescence-based PCR to provide quantitative and qualitative detection of nucleic acid sequences.

Device Description

The Abbott m2000 System is an instrument platform that automates steps to perform nucleic acid amplification assays from sample processing through amplification, detection, and data reduction. The Abbott m2000 System comprises the m2000sp and m2000rt instruments, which are operated with separate System Control Center (SCC) workstations. Each instrument contains an independent software application; one for the m2000sp and a second for the m2000rt. The m2000sp instrument is a floor standing, automated sample preparation system. The three main components of the m2000sp are the: Instrument, Cabinet, System Control Center (SCC). The m2000rt instrument is a real-time PCR thermal cycler/reader instrument system. The two main components of the m2000rt are the: Instrument, System Control Center (SCC). The Abbott m2000 System software processes sample preparation and amplification/detection protocols based on pre-determined, assay-specific parameters that are contained in individual assay application specification files that are installed on the SCC. The Abbott m2000sp reads and processes bar coded primary sample tubes and processes up to 96 specimens, controls, and calibrators in batch mode. The m2000 System is capable of processing samples from various matrices, depending on the specific assay application, including plasma, serum, endocervical swabs, urethral swabs, vaginal swabs, and urine. At the completion of the automated sample preparation protocol, the operator seals and manually transfers the PCR plate to the Abbott m2000rt for nucleic acid detection. Bar code and m2000sp data is transferred to the m2000rt electronically.

AI/ML Overview

The Abbott m2000 System is an automated instrument platform intended for in vitro diagnostic use in performing FDA-cleared and approved nucleic acid testing in clinical laboratories. This includes the Abbott m2000sp (automated sample preparation) and m2000rt (automated fluorescence-based PCR for quantitative and qualitative detection of nucleic acid sequences). The system was evaluated in conjunction with the Abbott RealTime CT/NG assay.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

While explicit "acceptance criteria" are not presented as a direct table in the provided document, the study aims to demonstrate substantial equivalence to predicate devices and establish performance characteristics. The key performance metrics reported are sensitivity and specificity for Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG).

MetricAcceptance Criteria (Implicit - based on predicate device performance and clinical relevance)Reported Device Performance (Abbott RealTime CT/NG assay)
Overall CT SensitivityHigh sensitivity, comparable to predicate NAATs95.2%
Overall CT SpecificityHigh specificity, comparable to predicate NAATs99.3%
Overall NG SensitivityHigh sensitivity, comparable to predicate NAATs97.5%
Overall NG SpecificityHigh specificity, comparable to predicate NAATs99.7%
CT LOD (95% Probability)Detectable at very low copy numbers21 copies/assay (95% CI 18 - 28)
nvCT LOD (95% Probability)Detectable at very low copy numbers29 copies/assay (95% Cl 24 - 41)
NG LOD (95% Probability)Detectable at very low copy numbers149 copies/assay (95% CI 130 - 176)
Carryover RateMinimal to no cross-contamination0.91% (5 false positive, 1 equivocal out of 656 negative samples)

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set (Clinical Study):
    • 3,832 male and female, asymptomatic and symptomatic subjects.
    • 3,832 subjects provided urine, endocervical swabs, self-collected vaginal swabs, and clinician-collected vaginal swabs (females) or urine and urethral swabs (males).
    • A total of 6,555 CT results and 6,569 NG results were used in the analysis after excluding subjects for whom patient infected status could not be determined (4 for CT, 7 for NG).
  • Data Provenance: Prospective, multi-center clinical study conducted in the United States. Specimens were collected from 16 geographically diverse sites, including physician private practices, public and private STD clinics, and a hospital emergency room.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not explicitly state the number or qualifications of experts who established the ground truth in the traditional sense (e.g., radiologists interpreting images). Instead, the ground truth was established through a reference standard composed of multiple commercially available nucleic acid amplification tests (NAATs) and culture for NG.

  • For CT or NG infection in females: A minimum of two positive results (at least one from each reference NAAT).
  • For CT in females (urine/swab discrepancy): Positive urine and negative endocervical swab from both reference assays resulted in categorization as infected for urine but not for swab specimens.
  • For CT or NG infection in males: A minimum of two positive results were reported.
  • For NG infection (regardless of NAAT) if NG culture positive: If the reference NG culture assay result was positive, the subject was categorized as infected regardless of NAAT results.
  • For non-infection in females: At least one of the reference NAATs reported negative results for all sample types AND if the NG culture assay result was negative.
  • For non-infection in males: A total of at least two negative results reported by the reference NAATs AND if the NG culture assay result was negative.

Therefore, the "ground truth" was established by a consensus of FDA-cleared diagnostic assays rather than individual expert interpretation.

4. Adjudication Method for the Test Set

The adjudication method for determining "patient infected status" (ground truth) was a composite reference standard based on the results of two commercially available NAATs and culture for NG. It essentially used a "2 out of 2" or "2 out of 3" (for NG cases with culture) positive rule for infection, and a "negative across multiple tests" rule for non-infection. In cases of internal discrepancy, specific rules were applied (e.g., for female CT urine vs. swab).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not mentioned. This device is an automated in vitro diagnostic system, not an AI-powered diagnostic imaging or interpretation tool designed to assist human readers. Its performance is evaluated as a standalone diagnostic.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the clinical study primarily evaluated the standalone performance of the Abbott RealTime CT/NG assay on the Abbott m2000 System. The reported sensitivity and specificity values are for the assay and system independent of human interpretation beyond typical lab procedures and result reporting.

7. The Type of Ground Truth Used

The ground truth used was a composite reference standard based on the results of:

  • Two commercially available nucleic acid amplification tests (NAATs) for CT and NG.
  • Culture for NG.

This is a frequently used method for establishing ground truth for infectious disease diagnostics when a single "gold standard" may not always be definitive. It combines multiple reliable diagnostic methods to increase confidence in the true infection status.

8. The Sample Size for the Training Set

The document does not explicitly specify a "training set" size for the Abbott RealTime CT/NG assay and m2000 system. For in vitro diagnostic devices like this, the development process (which would involve data akin to training) is typically extensive and proprietary, focusing on analytical performance and robustness. The clinical study described would be analogous to an independent validation or test set.

9. How the Ground Truth for the Training Set Was Established

Since a dedicated "training set" with ground truth establishment in the AI/machine learning sense is not detailed, this question applies less directly. However, the development of the assay would inherently involve:

  • Analytical studies: Determining Limit of Detection (LOD), analytical sensitivity (serovars, isolates), cross-reactivity, and interference using well-characterized samples (e.g., quantified pure cultures, spiked samples). The ground truth for these analytical studies is based on known concentrations of targets and the presence/absence of interfering substances or cross-reactants. These studies are detailed in Section 9.0 "Summary of Non-Clinical Testing."

{0}------------------------------------------------

l092705

MAY 2 8 2010

Abbott® m2000™ System 510(k) Summary

Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw006

Admin. Sec. 3 510(k) Summary Page 1 of 12

.

{1}------------------------------------------------

Table of Contents

Page
1.0Submitted By:3
2.0Contact Person3
3.0Date of Preparation3
4.0Name of Device4
5.0Predicate Devices4
6.0Device Description4
7.0Intended Use of the Device6
8.0Technological Characteristics of the Device as Compared to the Predicate(s)6
9.0Summary of Non-Clinical Testing7
10.0Summary of Clinical Testing10
11.0Conclusions for Non-clinical and Clinical Testing12

.

.

·

.

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006

1

Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 2 of 12

{2}------------------------------------------------

A summary of the Abbott® m2000" System as required by Section 807.92(c) is provided below.

  • 1.0 Submitted By:
    Abbott Molecular Inc. 1300 E. Touhy Ave. Des Plaines, IL 60018 Phone: (224) 361-7000 eFAX (847) 775-6777

2.0

Contact Person

Paula E. Martin Senior Manager Regulatory and Clinical Affairs Telephone: (224) 361-7333 eFAX (847) 775-6777 Email: paula.martin@abbott.com

Irene Powers Director of Regulatory Affairs Abbott Molecular Inc. Telephone: (847) 938-6930 eFAX (847) 775-6777 Email: irene.powers@abbott.com

3.0

Date of Preparation

August 2009

Abbott m20(){} System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 3 of 12

{3}------------------------------------------------

4.0Name of Device
Trade Name:Abbott® m2000™ System
Components:Abbott® m2000sp™Model Numbers:9K14-01 (G-Series)9K14-02 (E-Series)
Abbott® m2000rt™9K15-01
Common Name:Automated sample preparation and batch analyzer system fornucleic acid amplification and detection.
Components:
The m2000sp is an automated sample preparation system.The m2000rt is an automated batch analyzer for nucleic acidamplification and detection.
Classification Nameand Regulation:The m2000sp is classified as a “Clinical sample concentrator”as described in 21CFR §862.2310 (Class I, Procode JJH).The m2000rt is classified as “Instrumentation for clinicalmultiplex test systems” as described in 21CFR §862.2570(Class II, Procode OOI).
5.0Predicate Devices

Roche COBAS® TaqMan® Analyzer (K012966)

Device Description 6.0

The Abbott m2000 System is an instrument platform that automates steps to perform nucleic acid amplification assays from sample processing through amplification, detection, and data reduction.

The Abbott m2000 System comprises the m2000sp and m2000rt instruments, which are operated with separate System Control Center (SCC) workstations. Each instrument contains an independent software application; one for the m2000sp and a second for the m2000rt.

The m2000sp instrument is a floor standing, automated sample preparation system. The three main components of the m2000sp are the:

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006

Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 4 of 12

{4}------------------------------------------------

  • . Instrument
  • . Cabinet
  • System Control Center (SCC) .

The m 2000rt instrument is a real-time PCR thermal cycler/reader instrument system. The two main components of the m2000rt are the:

  • . Instrument
  • . System Control Center (SCC)

The Abbott m2000 System software processes sample preparation and amplification/detection protocols based on pre-determined, assay-specific parameters that are contained in individual assay application specification files that are installed on the SCC. The Abbott m2000sp reads and processes bar coded primary sample tubes and processes up to 96 specimens, controls, and calibrators in batch mode. The m2000 System is capable of processing samples from various matrices, depending on the specific assay application, including plasma, serum, endocervical swabs, urethral swabs, vaginal swabs, and urine. At the completion of the automated sample preparation protocol, the operator seals and manually transfers the PCR plate to the Abbott in2000rt for nucleic acid detection. Bar code and m2000sp data is transferred to the m2000rt electronically.

{5}------------------------------------------------

Intended Use of the Device 7.0

The Abbott m2000 system is intended for in vitro diagnostic use in performing FDA cleared and approved nucleic acid testing in clinical laboratories. It comprises the Abbott m2000sp and the Abbott m2000rt instruments.

The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.

The Abbott m2000rt is an automated system for performing fluorescencebased PCR to provide quantitative and qualitative detection of nucleic acid sequences.

Technological Characteristics of the Device as Compared to the 8.0 Predicate(s)

The Abbott m2000 System is substantially equivalent to the currently marketed predicate device Roche COBAS TaqMan Analyzer (510(k) K012966.

These devices are similar in that they are designed to amplify, detect and report results of DNA or RNA qualitative and/or quantitative assays. The primary similarities and differences between the m2000 System and the predicate device are shown in Table 1 in the attachment (the differences are shown in underlined italics in the table).

Because the Abbott RealTime CT/NG assay 510(k) is being submitted at the same time as this 510(k), it should be noted that the CT/NG assay, which is used on the Abbott m2000rt, is shown to be substantially equivalent to the Gen-Probe Combo-2 Assay (K043224). The Abbott RealTime CT/NG assay is also shown to be substantially equivalent to the Becton Dickinson ProbeTec ET Chlamydia trachomatis /Neisseria gonorrhoeae Amplified DNA Assay (K012351). Please refer to the Abbott RealTime CT/NG assay 510(k) submission for more information.

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 6 of 12

{6}------------------------------------------------

9.0 Summary of Non-Clinical Testing

9.1 Analytical Sensitivity

The Limit of Detection (LOD) claim for the Abbott RealTime CT/NG assay is 320 copies of Chlamydia trachomatis (CT) target DNA and 320 copies of Neisseria gonorrhoeae (NG) target DNA per assay. The assay targets the Chlamydia trachomatis cryptic plasmid (present at approximately 7 to 10 copies per Chlamydia organism) and the multicopy opacity gene of Neisseria gonorrhoeae (repeated up to 11 times per organism). Thus, 320 copies of target DNA is equivalent to approximately 30 to 40 organisms per assay.

The LOD of the Abbott RealTime CT/NG assay is defined as CT and NG target DNA concentration detected with a probability of 95% or greater. The CT and NG DNA concentrations detected with 95% probability were determined by testing dilutions of CT and NG target DNA. Probit analysis of the data determined that the concentration of CT DNA detected with 95% probability was 21 copies/assay (95% CI 18 - 28), the concentration of nvCT DNA detected with 95% probability was 29 copies/assay (95% Cl 24 - 41), and the concentration of NG DNA detected with 95% probability was 149 copies/assay (95% CI 130 - 176).

The claimed assay LOD was confirmed by testing samples that contained 320 copies of CT, nvCT and NG target DNA per assay. The detection rate was 100% (405/405) for CT, 100% (403/403) for nvCT, and 99.5% (403/405) for NG in the assay.

An additional study was conducted to challenge the performance of the Abbott RealTime CT/NG assay in samples containing high target numbers of CT, nvCT, or NG in the presence of low target numbers of the opposite analyte. Samples were prepared to contain 320 CT or nvCT target DNA copies and 1 x 107 NG target DNA copies per assay, or 1 x 107 CT or nvCT target DNA copies and 320 NG target DNA copies per assay. The detection rate of 320 copies of CT or nvCT DNA in the presence of high NG target was 100% (405/405). The detection rate of 320 copies of NG DNA in the presence of high CT or nvCT target was 100% (405/405).

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 7 of 12

{7}------------------------------------------------

The analytical sensitivity of the Abbott RealTime CT/NG assay for detecting Chlamydia trachomatis serovars A through L was determined by testing dilutions of each serovar. Serovars A through K and L1 through L3 were detected at less than 1 Inclusion Forming Units (IFU) per assay. Additionally, nvCT was diluted and was also detected at less than 1 IFU per assay.

The analytical sensitivity of the Abbott RealTime CT/NG assay for detecting 28 different isolates of Neisseria gonorrhoeae was determined by testing dilutions of each isolate. All isolates were detected at less than 1 Colony Forming Unit (CFU)/assay.

9.2 Evaluation of Potential Cross-Reactants

A total of 111 strains of bacteria, viruses, parasites, yeast, and fungi were tested for potential cross reactivity in the Abbott RealTime CT/NG assay (Table 2). These included organisms that are phylogenetically related to CT and NG, and those that can be found in the urogenital tract. Purified DNA or RNA was diluted to a final concentration of 1 x 10' copies/assay. HBV DNA and HCV RNA were added directly into the PCR reaction at approximately 4 x 10 and 6 x 106 copies per reaction, respectively. All results were negative for both CT and NG.

Additionally, a total of 32 culture isolates were tested for potential cross reactivity in the Abbott RealTime e assay. These included 27 organisms listed in Table 2, and Neisseria cinerea. Neisseria lactamica, Neisseria sicca, Ca Ski cells containing HPV 16, and Hela cells containing HPV 18. Ca Ski cells containing HPV 16 and Hela cells containing HPV 18 were tested at 106 cells per assay, C. pneumoniae and C. psittaci were tested at 100 EB per assay, HSV-1 and HSV-2 were tested at 106 genomes per assay, and the rest of the organisms were tested at 10° Colony Forming Units (CFU) per assay. All results were negative for both CT and NG.

9.3 Evaluation of Potentially Interfering Substances

The potential for interference in the Abbott RealTime CT/NG assay was assessed with substances that may be found in swab and/or urine specimens. Substances were spiked

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 8 of 12

{8}------------------------------------------------

into a swab and/or urine matrix containing 320 copies of CT and NG target DNA per assay, and into a swab and/or urine matrix without CT or NG DNA.

No interference in the performance of the Abbott RealTime CT/NG assay was observed in the presence of the substances listed in Table 3.

Interference in the performance of the Abbott RealTime CT/NG assay may be observed with the following substances:

  • Talcum powder at concentrations greater than 0.1% in urine specimens. .
  • . Phenazopyridine hydrochloride (the active ingredient in URISTAT) at concentrations greater than 3 mg/mL in urine specimens.
  • . Mucus at concentrations greater than 0.1% for urine specimens and 1% for swab specimens.

9.4 Precision Study

A precision study was performed at three sites, two external and one internal. Each site was provided a fifteen-member panel. Nine panel members targeted different combinations of CT and NG concentrations and six panel members targeted different combinations of nvCT and NG concentrations. Five replicates of each panel member were tested in each run. Thirty runs (10 per site) were performed for a total of 150 replicates of each panel member. The study included three amplification reagent lots. Each site tested two amplification reagent lots. A variance components analysis for a nested model was performed on delta cycle (DC) values, and the results are summarized in Table 4 and Table 5, respectively.

9.5 Carrvover

Potential carryover was determined by performing a study in which high copy CT positive samples were interspersed with negative samples arranged in a checkerboard pattern. The positive samples were CT DNA at a concentration of 10' copies/ml. The carryover rate is defined as the number of CT negative samples that are reported as positive or equivocal over the total number of CT-negative samples tested. Each run included 47 negative samples and 46 positive samples. A total of 14 runs were evaluated

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 9 of 12

{9}------------------------------------------------

using two lots of the RealTime CT/NG amplification reagents on four m2000sp and m2000rt instrument pairs.

A total of 656 valid negative samples were evaluated for potential carryover effect. A total of 5 false positive and 1 equivocal results were observed. The carryover rate was 0.91%.

10.0 Summary of Clinical Testing

Performance characteristics of the Abbott RealTime CT/NG assay were established in a multi-center clinical study conducted in the United States. Specimens were collected from subjects at 16 geographically diverse sites that included physician private practices, public and private STD clinics, and a hospital emergency room. A total of 3,832 male and female, asymptomatic and symptomatic subjects were enrolled. Study subjects were classified as symptomatic if the subject reported STD-related symptoms. Specimens collected from each female subject included urine, endocervical swabs, self-collected vaginal swab, and clinician-collected vaginal swabs. Specimens collected from each male subject included urine and urethral swabs. Specimen testing methods included the Abbott RealTime CT/NG assay, two commercially available nucleic acid amplification tests (NAAT) for CT and NG, and culture for NG. The NAATs and the NG culture were used as reference assays in the clinical study.

For females, self-collected vaginal swab and urine specimens were collected first, followed by endocervical swab for culture. Remaining swab specimen collection was randomized to minimize bias. For males, urethral swab for culture was collected first. Remaining swab specimen collection was randomized to minimize bias. Urine specimen was collected after the swab specimens.

For each subject, a patient infected status was determined based on the combined results from the reference assays. A female subject was categorized as infected for CT or NG if a minimum of two positive results (at least one from each reference NAAT) were reported. For CT, female subjects with positive results on urine and negative results on endocervical swab specimens from both reference assays were categorized as infected for

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 10 of 12

{10}------------------------------------------------

urine and not infected for swab specimens. A male subject was categorized as infected for CT or NG if a minimum of two positive results were reported. If the reference NG culture assay result was positive, the subject was categorized as infected regardless of NAAT results.

A female subject was categorized as not infected with CT or NG if at least one of the reference NAATs reported negative results for all sample types and if the NG culture assay result was negative. A male subject was categorized as not infected with CT or NG if a total of at least two negative results were reported by the reference NAATs and if the NG culture assay result was negative.

If patient infected status could not be determined due to missing and/or indeterminate results from the reference assays, the subject was excluded from the analysis. Patient infected status could not be determined for 4 subjects for CT and 7 subjects for NG.

Tables 6 through 24 summarize the clinical trial data.

Abbott RealTime CT/NG test results were compared to the patient infected status for calculation of assay sensitivity and specificity. A total of 6,555 CT and 6,569 NG results were used in the analysis. The results were analyzed by gender, sample type, and the presence of symptoms. The overall sensitivity and specificity for CT was 95.2% and 99.3%, respectively. The overall sensitivity and specificity for NG was 97.5% and 99.7%, respectively. Sensitivity and specificity for CT for female subjects and male subjects are presented in Tables 6 and 7, respectively. Sensitivity and specificity for NG for female subjects and male subjects are presented in Tables 8 and 9, respectively.

A comparison of patient infected status, individual test results from the reference assays and Abbott RealTime CT/NG assay was performed. CT results for infected and noninfected female subjects are presented in Tables 10 and 11, and for infected and noninfected male subjects in Tables 12 and 13. NG results for infected and non-infected female subjects are presented in Tables 14 and 15, and for infected and non-infected male subjects in Tables 16 and 17.

Abbolt m2000 System 510(k) May 2010 Admin See 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 11 of 12

{11}------------------------------------------------

The prevalence of CT and NG in this study was dependent on several factors including age, gender, clinic type, and the method of testing. The prevalence per collection site determined by the Abbott RealTime CT/NG assay for endocervical swab specimens is presented in Table 18, for clinician-collected and self-collected vaginal swab specimens is presented in Table 19; for female urine specimens in Table 20; and for male urethral swab and male urine specimens in Tables 21 and 22, respectively.

The Positive and Negative Predictive Values (PPV and NPV) were calculated using hypothetical prevalence rates and the Abbott RealTime CT/NG assay sensitivity and specificity determined from the clinical study. The overall sensitivity and specificity for CT was 95.2% and 99.3%, respectively. The overall sensitivity and specificity for NG was 97.5% and 99.7%, respectively. Estimates of the PPV and NPV for the Abbott RealTime CT/NG assay are presented in Table 23 for CT and Table 24 for NG. "

11.0 Conclusions for Non-clinical and Clinical Testing

Based on the results of the studies summarized above, the Abbott m2000 System, in conjunction with the Abbott RealTime CT/NG assay, has been shown to be substantially equivalent.

Abbott m2000 System 510(k) May 2010 Admin Sec 3_510(k) Summary_mw006 Admin Sec. 3.0 Reason for Submission 510(k) Summary Page 12 of 12

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Abbott Molecular, Inc .. c/o Paula Martin Senior Manager, Regulatory and Clinical Affairs 1300 E. Touhy Ave. Des Plaines, IL 60018

MAY 2 8 2010

K092705 Re:

Trade/Device Name:Abbott m2000 system
Regulation Number:21CFR §862.2570
Regulation Name:Instrumentation for clinical multiplex test systems
Regulatory Class:Class II
Product Code:OOI, JJH
Dated:May 21, 2010
Received:May 24, 2010

Dear Ms. Martin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the

{13}------------------------------------------------

Page 2 – Paula Martin

  1. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Valy a Ayms

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K092705

Device Name: Abbott m2000 System

Indications for Use:

The Abbott m2000 system is intended for in vitro diagnostic use in performing FDA cleared and approved nucleic acid testing in clinical laboratories. It comprises the Abbott m2000sp and the Abbott m2000rt instruments.

The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.

The Abbott m2000rt is an automated system for performing fluorescence-based PCR to provide quantitative and qualitative detection of nucleic acid sequences.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Ube Sch

Division Sign-Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K092705

Page 1 of 1

§ 862.2570 Instrumentation for clinical multiplex test systems.

(a)
Identification. Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.