K Number
K973683

Validate with FDA (Live)

Device Name
OMEGA 21 SYSTEM
Date Cleared
1997-12-24

(89 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Omega 21™ System is a spinal fixation device for pedicle screw fixation and nonpedicle hook and sacral/iliac screw fixation system of the noncervical spine.

When used as a pedicle screw fixation system, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. The screws of the system are limited to L3- S1 or iliac screw fixation with the fusion only at L5-S1.

When used as an unterior fixation system or as a posterior hook and sacral/iliac screw fixation system, the system is indicated for nonpedicle spinal applications for the reduction, alignment or stabilization of the thoracic, lumbar and sacral segments of the spine, in cases of: Degenerative disc disease which is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, spinal deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or revision of failed fusion attempts.

Device Description

The Omega 21™ System is a rod system spinal fixation device. The components of the system are manufactured from Ti-6Al-4V ELI per ASTM F136.

AI/ML Overview

The provided text describes the Omega 21™ System, a spinal fixation device, and its 510(k) summary for safety and effectiveness. However, it does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the context of an AI/ML algorithm or software.

The document primarily focuses on:

  • Device Description: What the Omega 21™ System is.
  • Intended Use/Indications for Use: The medical conditions and patient types for which the device is intended.
  • Materials: What the device components are made from.
  • Comparison to Predicate Devices: How it is similar to other legally marketed spinal fixation systems.
  • FDA Clearance Details: The FDA's decision regarding substantial equivalence, specific limitations, warnings, and regulatory requirements.

The "Bench testing comparing the system to a predicate system demonstrated that the device complies with applicable standards and meets all of its functional requirements" statement (in section 7) is the closest thing to a "study" mentioned. However, this is a very general statement about mechanical testing for a physical device, not a performance study for an AI/ML system with quantifiable metrics like sensitivity, specificity, or F1-score against a ground truth dataset, which are typical for your requested format.

Therefore, I cannot extract the specific information requested in your bulleted list (acceptance criteria table, sample sizes, expert details, adjudication, MRMC, standalone performance, ground truth types, training set details) because the provided text is about a physical medical device (spinal fixation system) and its regulatory clearance, not an AI/ML-driven diagnostic or prognostic device.

{0}------------------------------------------------

510(k) Summary of Safety & Effectiveness

This 510(k) Summary of Safety and Effectiveness for the Omega 21™ System is provided as required per Section 513(1)(3) of the Food, Drug and Cosmetic Act.

Contact Person: Jon Caparotta 1. Submitter: Electro-Biology, Inc. Telephone: (973) 299-9022 6 Upper Pond Road Parsippany, NJ 07054

Date prepared: December 18, 1997

    1. Proprietary Name:Omega 21TM System
Common Name:Spinal Fixation Device
Classification Names:Spondylolisthesis Spinal Fixation Device SystemSpinal Intervertebral Body Fixation OrthosisSpinal Interlaminal Fixation Orthosis
  • Predicate or legally marketed devices that are substantially equivalent: 3.
    • Webb-Morley Spine System Electro-Biology, Inc. .
    • . Isola® Spinal System - AcroMed Corporation
    • TSRH® Spinal System Sofamor Danek .
    • . Osteonics® Spinal System - Osteonics Corporation
    • The SpineLink™ System Electro-Biology, Inc. .
  • Description of the device: The Omega 21™ System is a rod system spinal 4. fixation device.
    1. Intended Use: The Omega 21™ System is a spinal fixation device for pedicle screw fixation and nonpedicle hook and sacral/iliac screw fixation system of the noncervical spine.

When used as a pedicle screw fixation system, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.

1 c+ 2

{1}------------------------------------------------

The screws of the system are limited to L3- S1 or iliac screw fixation with the fusion only at L5-S1.

When used as an anterior fixation system or as a posterior hook and sacral/iliac screw fixation system, the system is indicated for nonpedicle spinal applications for the reduction, alignment or stabilization of the thoracic, lumbar and sacral segments of the spine, in cases of: Degenerative disc disease which is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, spinal deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or revision of failed fusion attempts.

  • Materials: The components of the system are manufactured from Ti-6Al-4V ELI per 6. ASTM F136.
    1. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between EBI Omega 21™ System and other currently marketed spinal systems. It is substantially equivalent* to the predicate devices in regards to intended use, materials and function. Bench testing comparing the system to a predicate system demonstrated that the device complies with applicable standards and meets all of its functional requirements.

*Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully market approval or reclassification and is not intended to be interpreted as an admission or any other in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.))

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jon Caparotta Manager, Regulatory Affairs Electro-Biology, Inc. 6 Upper Pond Road Parsippany, New Jersey 07054-1079

DEC 2 4 1997

K973683 Re: Omega 21™ System Requlatory Class: II Product Codes: MNH, KWP, and KWQ Dated: September 25, 1997 Received: September 26, 1997

Dear Mr. Caparotta:

We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal, Food, Drug, and Cosmetic Act This decision is based on your device system being (Act). found equivalent only to similar device systems labeled and intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. You may, therefore, market your device system subject to the general controls provisions of the Act and the limitations identified below.

The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment for indications other than severe spondylolisthesis, as described above, would cause the device system to be adulterated under 501(f) (1) of the Act.

This device system, when intended for pedicular screw fixation/attachment to the spine for indications other than severe spondylolisthesis, as described above, is a class III

{3}------------------------------------------------

device under Section 513(f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:

  • All labeling for this device, including the package 1. label, must state that there are labeling limitations. The package insert must prominently state that the device system using pedicle screws is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
  • You may not label or in anyway promote this device system 2. for pedicular, screw fixation/attachment to the cervical, thoracic or lumbar vertebral column for intended uses other than severe spondylolisthesis, as described above. The package insert must include the following statements:

WARNINGS :

  • When used as a pedicle screw system, this device system is intended only for grade 3 or 4 spondylolisthesis at the fifth lumbar - first sacral (L5-S1) vertebral joint.
  • The screws of this device system are not intended for insertion into the pedicles to facilitate spinal fusions above the L5-S1 vertebral joint.
  • Benefit of spinal fusions utilizing any pedicle screw fixation system has not been adequately established in patients with stable spines.
  • Potential risks identified with the use of this device system, which may require additional surgery, include :

device component fracture, loss of fixation, non-union, fracture of the vertebra, neurological injury, and vascular or visceral injury.

See Warnings, Precautions, and Potential Adverse Events sections of the package insert for a complete list of potential risks.

{4}------------------------------------------------

  • Any pedicular screw fixation/attachment for intended uses 3 . other than severe spondylolisthesis, as described by item 1, for this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) requlations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment for intended uses other than severe spondylolisthesis, as described above, must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investiqation.
  • Any previous warning statements identified as part of 4 . previous 510(k) clearances or required by OC/Labeling and Promotion which stated that a component/system was not approved for screw fixation into the pedicles of the spine must be replaced by the warnings of items 1 and 2 above .

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

FDA advises that the use of your device system with any other device components but those identified in this 510(k) would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether yours or those of other manufacturers, may also be required.

This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.

{5}------------------------------------------------

Page 4 - Mr. Jon Caparotta

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{6}------------------------------------------------

Statement of Indications for Use:

The Omega 21™ System is a spinal fixation device for pedicle screw fixation and nonpedicle hook and sacral/iliac screw fixation system of the noncervical spine.

When used as a pedicle screw fixation system, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. The screws of the system are limited to L3- S1 or iliac screw fixation with the fusion only at L5-S1.

When used as an unterior fixation system or as a posterior hook and sacral/iliac screw fixation system, the system is indicated for nonpedicle spinal applications for the reduction, alignment or stabilization of the thoracic, lumbar and sacral segments of the spine, in cases of: Degenerative disc disease which is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, spinal deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or revision of failed fusion attempts.

letha with

Prescription Use (Per 21 CFR 801.109)

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.