K Number
K130655

Validate with FDA (Live)

Date Cleared
2013-05-08

(57 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Response Spine System is intended for posterior, non-cervical fixation (pedicle screw fixation T1-S2/ilium and hook fixation (T1-L5)) in pediatric patients is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The device is intended to be used with autograft and/or allograft.

Device Description

The Response Spine System consists of longitudinal members (rods), anchors (hooks and screws), interconnection components (rod-to-rod and anchor-to-rod connectors) and fasteners in a variety of sizes, including fixed angled screws, to accommodate differing anatomic requirements.

AI/ML Overview

The OrthoPediatrics Response Spine System (K130655) relies on substantial equivalence to predicate devices rather than a standalone clinical study to demonstrate its safety and effectiveness. The acceptance criteria and the "study" proving it meets those criteria are articulated through mechanical testing designed to show equivalence.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Set by ASTM F1717 Standard and Equivalence to Predicates)Reported Device Performance (as stated in the 510(k) Summary)
Mechanical Performance: - Static Compression Bending - Dynamic Compression Bending - Static Torsion (Demonstrates that the device's mechanical strength and durability are comparable to predicate devices and meet established standards for spinal implant constructs.)"Static and dynamic compression bending tests and static torsion tests, performed according to ASTM F1717, demonstrated the device to be substantially equivalent to the predicate devices."

Note: The specific numerical acceptance values for these tests are not provided in the summary but would be detailed in the full test report, confirming that the Response Spine System's results fell within acceptable ranges or were comparable to the predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as the evaluation focuses on mechanical testing of devices rather than patient data. The "sample" here refers to the number of devices (rods, screws, hooks, connectors) tested under various mechanical loads.
  • Data Provenance: Not applicable in the traditional sense of patient data. The testing would have been conducted in a laboratory setting, likely in the US, by the manufacturer or a contracted testing facility. This is a retrospective comparison to existing predicate device data and established ASTM standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable in the context of mechanical testing. "Ground truth" for mechanical testing is established by the ASTM F1717 standard itself, which is developed by a consensus of engineering and medical experts in the field of spinal implants. The results are physical measurements.
  • Qualifications of Experts: The "experts" in this context are the engineers and material scientists who designed and conducted the tests, and those who developed the ASTM F1717 standard.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. Mechanical testing results are objective measurements against a standard. There is no human adjudication of "truth" in the way there is for clinical interpretations. The results either meet the standard/equivalence or they don't.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No. This device is a surgical implant undergoing mechanical performance testing, not an imaging or diagnostic device requiring human interpretation of results. Therefore, an MRMC study is not relevant.

6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

  • Standalone Performance: Not applicable. This is a physical medical device, not a software algorithm. Its "performance" is its mechanical integrity under specified loads, which is evaluated directly in a lab.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for this evaluation is based on:
    • Established ASTM F1717 Standard: This standard dictates the methodology and acceptable criteria for static and dynamic testing of spinal implant constructs.
    • Performance of Predicate Devices: The measured mechanical properties (e.g., stiffness, fatigue life) of the Response Spine System are compared to those of the legally marketed predicate devices (OrthoPediatrics' OP Spine System and Medtronic CD Horizon) to establish substantial equivalence.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This is a physical device being evaluated for mechanical equivalence, not a machine learning model requiring a training set.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no training set for this type of device evaluation.

{0}------------------------------------------------

510(k) -- Section 1: OrthoPediatrics Response Spine System

K130655 Page 1 of 2

MAY 0 8 2013

NAME OF FIRM:OrthoPediatrics, Corp.2850 Frontier DriveWarsaw, IN 46582
DATE PREPARED:March 7, 2013
510(K) CONTACT:Mark FoxVice President, Regulatory AffairsTel: (574) 268-6379
PROPOSED TRADENAME:Response Spine System
DEVICECLASSIFICATION:Class II; 21 CFR 888.3070 and 21 CFR 888.3050
CLASSIFICATIONNAME:Pedicle screw spinal system , adolescent idiopathic scoliosisorthosis; spinal interlaminal fixation orthosis
PRODUCT CODE:OSH; KWP
DEVICEDESCRIPTION:The Response Spine System consists of longitudinalmembers (rods), anchors (hooks and screws),interconnection components (rod-to-rod and anchor-to-rodconnectors) and fasteners in a variety of sizes, including fixedangled screws, to accommodate differinganatomicrequirements.
INDICATIONS FORUSE:The Response Spine System is intended for posterior, non-cervical fixation (pedicle screw fixation T1-S2/ilium and hookfixation (T1-L5)) in pediatric patients is indicated as anadjunct to fusion to treat adolescent idiopathic scoliosis. Thedevice is intended to be used with autograft and/or allograft.
MATERIALS:Medical grade titanium alloy - ASTM F136-08e1Medical grade cobalt-chromium-molybdenum - ASTMF1537-08
PREDICATE DEVICES:OrthoPediatrics' OP Spine System (K120291 and K121250)Medtronic CD Horizon (K091445)
TECHNOLOGICCHARACTERISTICS:The fundamental scientific principles and technologicalcharacteristics, including the material, and general design,and sizes of the device are substantially the same as thepredicate devices. Fixed angle (non-swivel) screws havebeen added to the system while the smaller diameterlongitudinal rod has been eliminated from the system.
PERFORMANCEDATA:Static and dynamic compression bending tests and statictorsion tests, performed according to ASTM F1717,demonstrated the device to be substantially equivalent to thepredicate devices.
CONCLUSION:The Response Spine System has the same intended use andtechnological characteristics as the predicate devices.Therefore the Response Spine System is substantiallyequivalent for its intended use.

510(k) Summary

·

.

{1}------------------------------------------------

י

: .

.

.

'

.

.

:

:

.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 8, 2013

OrthoPediatrics, Corporation % Mr. Mark Fox Vice President, Regulatory Affairs 2850 Frontier Drive Warsaw, Indiana 46582

Re: K130655

Trade/Device Name: Response Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: OSH, KWP Dated: March 20, 2013 Received: March 21, 2013

Dear Mr. Fox:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Mark Fox

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

ring theith

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K130655

Device Name: OrthoPediatrics' Response Spine System

Indications for Use:

The Response Spine System is intended for posterior, non-cervical fixation (pedicle screw fixation T1-S2/ilium and hook fixation (T1-L5)) in pediatric patients is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The device is intended to be used with autograft and/or allograft.

Prescription Use ___________ X or Over-The-Counter Use

(21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P. Jean -S

(Division Sign-Off)

Division of Orthopedic Devices

510(k) Number: K130655

N/A