K Number
K130649

Validate with FDA (Live)

Date Cleared
2013-11-01

(235 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Silk Road™ Access Catheter is indicated for the introduction of interventional devices into the peripheral vasculature.

Device Description

The Silk Road Access Catheter is a sterile, non-pyrogenic, single-use access catheter indicated for introduction of interventional devices into the peripheral vasculature. The Silk Road Access Catheter is a single-lumen, coil-reinforced shaft, variable stiffness catheter in a range of diameters and working lengths to accommodate target anatomy. All sizes contain a radiopaque marker on the distal end and a catheter hub on the proximal end. The catheter shaft has a hydrophilic coating on its distal portion to reduce friction during use. The catheter is offered with a dilator in all but the smallest size (SR-045-AC). The Silk Road Access Catheter is a limited duration (<24 hours), externalcommunicating devices, contacting circulating blood.

AI/ML Overview

The Silk Road™ Access Catheter underwent a series of non-clinical tests to demonstrate its safety and effectiveness. The acceptance criteria and reported device performance are summarized below:

1. Acceptance Criteria and Reported Device Performance

AttributeAcceptance CriteriaReported Device Performance
Visual Inspection and Dimensional VerificationAll samples must pass visual and dimensional inspection specifications.Pass
Simulated Prep and UseAll samples must be able to be prepped and used per the IFU.Pass
Coating ParticulateAll samples must track through the track fixture and meet USP 788 microscopy method.Pass
Coating IntegrityAll samples must easily track through the track fixture.Pass
Guidewire-Microcatheter Advance/Withdrawal ForceAll samples must pass advancement and withdrawal force.Pass
Bonding StiffnessCharacterize Bend Stiffness.All samples are acceptable.
Aspiration RateCharacterize Aspiration Rate.All samples are acceptable.
Flush RateCharacterization Flush Rate.All samples are acceptable.
Tensile TestAll bonds must withstand a set force dictated by the device outer diameter for all samples.Pass
Shaft Kink ResistanceThe distal and proximal shaft must not kink around the bend for all samples.Pass
Torsion TestAll samples shall show no evidence of damage (kinking, flattening, separation) to any joints or bonds.Pass
Liquid LeakageAll samples shall show no water leaks large enough to form a falling drop over the entire device.Pass
Air LeakageAll samples shall show no bubbles are present.Pass
Luer Functional TestingAll samples shall meet the requirements of ISO 594-2 (gauging, separation force).Pass
Biocompatibility
Cytotoxicity (MEM Elution L-929)The test article passes if the mouse fibroblast cells do not display signs of toxicity after examination at 24, 48, and 72 hours.The test article scored '0' at 24, 48, and 72 ± 4 hours, considered non-cytotoxic.
Sensitization (Maximum Sensitization, Guinea Pig)The test article passes if the animals do not have signs of a delayed allergic response after being exposed to the test extract when compared to a control group.No sensitization response greater than '0' was observed in animals challenged with test article extracts, considered non-sensitizing.
Irritation (ISO Intracutaneous Reactivity Test)The test article passes if the difference between the mean scores for the test article extract and control are less than or equal to 1.0.Differences in mean test and control scores of dermal observations were less than 1.0, meeting test requirements.
Systemic Toxicity (ISO Acute Systemic Injection)The test article passes if the animals exposed to the test article extract do not show signs of toxicity greater than the concurrent control groups over a 72 hour period.No clinical signs of toxicity were observed in treated animals, and body weight changes were acceptable.
Systemic Toxicity (Material Mediated Pyrogen)The test article passes if the animals exposed to the test article extract do not show a significant increase in body temperature over a 3 hour period.No temperature rise of 0.5 °C at required observation times, determined to be non-pyrogenic.
Hemocompatibility (Four Hour Thromboresistance Evaluation in Dogs)1) Test animals survive general anesthesia and study observation interval without test article complications; and 2) blood test results, pre/post weight differences, and patency and thrombus scores are not subjectively different between the test and control articles.Implantation in two canines resulted in no adverse effects or significant thrombus formation; blood tests and clinical observations indicated no compromise in clotting abilities. Similar thromboresistance characteristics to control devices.
Hemocompatibility (Complement Activation C3a and SC5b-9)The test article passes if complement activation in human plasma (C3a Assay and SC5b-9) is not induced significantly when compared with the comparison article.All biomaterials have the potential to affect complement components; no established acceptable ranges. Test article and comparison article showed similar activation levels.
Hemocompatibility (Platelet and Leukocyte Count)The test article passes if the counts are not statistically and adversely significant compared to the reference material and comparison article.Test article groups and comparison article showed platelet and leukocyte counts within a reasonable percentage of reference material. All biomaterials have the potential to affect blood components; no established acceptable ranges.
Hemocompatibility (Partial Thromboplastin Time (PTT))The test article passes if the PTT is not significantly shortened following contact with a material under standardized conditions.Both test article groups and the comparison article had an average clotting time of 300.0 seconds (100% of negative control), indicating non-activation of the intrinsic coagulation pathway.
Hemocompatibility (Hemolysis - Direct Contact and Extract (ASTM F 756))The test samples are rated per the hemolytic index above the negative control. If 0 – 2%, classified as non-hemolytic, if 2.1-5%, classified as slightly hemolytic, and if ≥5.1%, classified as hemolytic.Test article returned hemolytic index values equal to or lower than the negative control, falling within the non-hemolytic range for both direct contact and extract tests.
Genotoxicity
Bacterial Mutagenicity Test (Ames Assay)The test article passes if the criteria for positive mutagen are not met.The test article did not induce substantial increases in reversion rates associated with mutagenesis; no substantial toxicity noted. Considered non-mutagenic.
In Vitro Mouse Lymphoma AssayThe test article passes if the mutant frequency is less than 1.8 fold higher than that of the concurrent negative control groups.Mutant frequencies and cloning efficiencies were within limits for a negative response. Considered non-mutagenic (non-genotoxic and non-clastogenic).
In Vivo Mouse Micronucleus AssayThe test article passes if it does not show a significant increase in the number of micronucleated polychromatic erythrocytes.No apparent gross manifestations of toxicity nor biologically significant erythropoietic disturbances. No biologically significant increases in mPCE production. Considered non-mutagenic.
Packaging ValidationPackaging must maintain product sterility and protect the product during shipment and storage.Pass
Sterilization ValidationSilk Road Access Catheter must be validated ISO 11135-1:2007.Pass
Shelf LifeSilk Road Access Catheter must perform to specification and the package must maintain a sterile barrier during the shelf-life period as labeled on the product label.Pass

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each non-clinical test beyond phrases like "all samples" or mentioning "replicates" for certain biocompatibility tests. For the animal study, it mentions "two canines" for the Hemocompatibility (Four Hour Thromboresistance Evaluation) test.

The data provenance is not explicitly stated. These appear to be laboratory test results conducted by the manufacturer or contracted labs for the purpose of regulatory submission. The studies are non-clinical, meaning they were performed in a lab or animal setting, not on human subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable as the studies are non-clinical (laboratory and animal studies) and do not involve human interpretation of medical images or data requiring expert consensus or clinical evaluation of ground truth. The "ground truth" for these tests is based on established scientific and regulatory standards in material science, engineering, and toxicology.

4. Adjudication Method for the Test Set

This is not applicable as the studies performed are non-clinical and do not involve human diagnostic decisions or require adjudication for consensus.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The submission is for a medical device (access catheter) and focuses on non-clinical performance, safety, and biocompatibility, not on diagnostic accuracy or reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is not applicable. The device is a physical medical instrument (an access catheter), not a software algorithm, so "standalone algorithm performance" is not relevant.

7. The Type of Ground Truth Used

The "ground truth" for these non-clinical tests is based on:

  • Established specifications and standards: e.g., "All samples must pass visual and dimensional inspection specifications," "meet USP 788 microscopy method," "meet the requirements of ISO 594-2," "validated ISO 11135-1:2007."
  • Biological responses as defined by assays and protocols: e.g., "mouse fibroblast cells do not display signs of toxicity," "animals do not have signs of a delayed allergic response," "no significant increase in body temperature."
  • Physical and mechanical measurements: e.g., "advancement and withdrawal force," "bend stiffness," "aspiration rate."
  • Pathology and histopathology results: from the GLP animal study.

8. The Sample Size for the Training Set

This is not applicable. The device is a physical medical instrument, not an AI or machine learning model, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Traditional 510(k) Silk Road™ Access Catheter

510(k) Summary

NOV 0 1 2013

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

A. Name and Address of Applicant Silk Road Medical, Inc. 735 North Pastoria Avenue Sunnyvale, CA 94085 Phone: (408) 720-9002 . Fax: (408) 720-9013

B. Contact Person

Richard M. Ruedy Vice President, RA/CA/QA (408) 828-7281 ric@silkroadmed.com

Alternate Contact: Diana DeGregorio Lincé Consulting Regulatory Affairs Consultant (925) 980-8047 dianadegregorio@comcast.net

C. Date Prepared

October 23, 2013

D. Device Name

Trade Name:Silk Road™ Access Catheter
Common Name:Percutaneous Catheter
Classification Name:Percutaneous Catheter

E. Device Classification

Classification: Product Code: Device Class:

21 CFR§870.1250 DQY, KRA Class II

{1}------------------------------------------------

F. Predicate Device

Silk Road Medical, Inc. submits that the subject Silk Road™ Access Catheter is substantially equivalent to the predicate, Penumbra Neuron™ Intracranial Access System (K070970, K082290).

G. Device Description

The Silk Road Access Catheter is a sterile, non-pyrogenic, single-use access catheter indicated for introduction of interventional devices into the peripheral vasculature. The Silk Road Access Catheter is a single-lumen, coil-reinforced shaft, variable stiffness catheter in a range of diameters and working lengths to accommodate target anatomy. All sizes contain a radiopaque marker on the distal end and a catheter hub on the proximal end. The catheter shaft has a hydrophilic coating on its distal portion to reduce friction during use. The catheter is offered with a dilator in all but the smallest size (SR-045-AC). The Silk Road Access Catheter is a limited duration (<24 hours), externalcommunicating devices, contacting circulating blood.

H. Indications for Use

The Silk Road Access Catheter is indicated for the introduction of interventional devices into the peripheral vasculature.

l. Technological Comparison

The Silk Road Access Catheter has similar features as compared to the predicate device as shown in the table below:

ManufacturerModel Name510(k) NumberPenumbraNeuron™ Intracranial Access SystemK070970, K082290Silk Road Medical Inc.Silk Road Access CatheterK130649
Indications forUseThe Neuron™ Intracranial Access System isindicated for the introduction of interventionaldevices into the peripheral, coronary, andneuro vasculature.The Silk Road™ Access Catheter isindicated for the introduction ofinterventional devices into the peripheralvasculature.
AnatomicalLocationsPeripheral, Coronary, and Neuro VasculaturePeripheral Vasculature
Materials-Catheter shaftPTFE Lined, nylon co-polymer catheter, withhydrophilic coatingSame
Materials-Catheter ShaftSupportStainless SteelSame
Catheter ShaftConstructionProximal - braid reinforcedDistal - coil reinforcedFull length coil reinforced
RadiopacityRadiopaque marker at distal tipSame
SterilizationMethodEOSame
TipConfigurationStraightSame

CONFIDENTIAL

{2}------------------------------------------------

Silk Road Medical

ManufacturerModel Name510(k) NumberPenumbraNeuron™ Intracranial Access SystemK070970, K082290Silk Road Medical Inc.Silk Road Access CatheterK130649
CatheterDimensionsNeuron 6F Delivery CatheterInner Diameter: 0.053"-0.070" (1.3-1.8 mm)Access CatheterInner Diameter 0.045"-0.078" (1.1-2.0 mm)
DimensionsOuter Diameter: 0.067"-0.079" (1.7-2.0 mm)Working Length: 95-115 cm (37.4" - 45.3")Outer Diameter 0.057" 0.090" (1.4-2.3mm)Working Length 47-64 cm (18.5" - 25.2")
Delivery AidComponent andDimensionsComponent: Neuron 5F Select CatheterComponent: Dilator
DimensionsInner Diameter 0.040" (1.0 mm)Inner Diameter 0.021"-0.042" (0.5-1.1 mm)
DimensionsOuter Diameter 0.065" (1.6 mm)Working Length 120-130 cm (47.2 - 51.2")Outer Diameter 0.052"-0.072" (1.3-1.8mm)Working Length 63 - 80 cm (24.8" - 31.5")
SuppliedAccessoryDevicesRotating Hemostasis Valve (RHV)Peel away guidewire introducer(Note: The Neuron Delivery Catheter and theNeuron Select Catheter are both part of theNeuron Intracranial Access System but aresupplied separately)Dilator (dimensions noted above)

The technological characteristics and principals of operation of the Silk Road Access Catheter is substantially equivalent to the named predicate device.

J. Non-Clinical Performance Data

The following table contains the non-clinical testing which was conducted to support a determination of substantial equivalence to the predicate device.

AttributeAcceptance CriteriaResult
Visual Inspection and DimensionalVerificationAll samples must pass visual anddimensional inspectionspecifications.Pass
Simulated Prep and UseAll samples must be able to beprepped and used per the IFU.Pass
Coating ParticulateAll samples must track through thetrack fixture and meet USP 788microscopy method.Pass
Coating IntegrityAll samples must easily trackthrough the track fixturePass
Guidewire-MicrocatheterAdvance/Withdrawal ForceAll samples must passadvancement and withdrawal forcePass
Bonding StiffnessCharacterize Bend StiffnessAll samples are acceptable
Aspiration RateCharacterize Aspiration RateAll samples are acceptable
Flush RateCharacterization Flush RateAll samples are acceptable
Tonsile TestAll bonds must withstand a set forcedictated by the device outerdiameter for all samplesPass
Shaft Kink ResistanceThe distal and proximal shaft mustnot kink around the bend for allsamples.Pass
Torsion TestAll samples shall show no evidenceof damage (kinking, flattening,separation) to any joints or bondsPass
Liquid Leakage.All samples shall show no waterleaks large enough to form a fallingdrop over the entire devicePass
Air LeakageAll samples shall show no bubblesare presentPass
Luer Functional TestingAll samples shall meet therequirements of ISO 594-2(gauging, separation force)Pass

{3}------------------------------------------------

AttributeAcceptance CriteriaResult
BiocompatibilityCytotoxicity:MEM Elution L-929ISO/USPThe test article passes if the mouse fibroblast cells do not display signs of toxicity after examination at 24, 48 and 72 hours.Result - The test article scored '0' at 24, 48, and 72 ± 4 hours.Conclusion - The test article is considered non-cytotoxic under the conditions of this test.
Sensitization:MaximumSensitization(Guinea Pig)The test article passes if the animals do not have signs of a delayed allergic response after being exposed to the test extract when compared to a control group.Result - None of the negative control animals challenged with the control vehicle were observed with a sensitization response greater than '0'. None of the animals challenged with the test article extracts were observed with a sensitization response greater than '0'. The normal saline extract of the test material had a sensitization response of '0' under valid test conditions. The cottonseed oil extract of the test material had a sensitization response of '0' under valid test conditions.Conclusion - The test article did not elicit a sensitization response.
Irritation:ISO IntracutaneousReactivity TestThe test article passes if the difference between the mean scores for the test article extract and control are less than or equal to 1.0.Result - The differences in the mean test and control scores of the extract dermal observations were less than 1.0.Conclusion - The ISO Intracutaneous Reactivity Test have been met by the test article.
Systemic Toxicity:ISO AcuteSystemic InjectionThe test article passes if the animals exposed to the test article extract do not show signs of toxicity greater than the concurrent control groups over a 72 hour period.Result - None of the test article extract treated animals were observed with clinical signs consistent with toxicity at any of the observation periods. Body weight changes were within acceptable parameters over the course of the study.Conclusion - These findings indicate that the requirements of the ISO Acute Systemic Injection Test have been met by the test article.
Systemic Toxicity:Material MediatedPyrogenThe test article passes if the animals exposed to the test article extract do not show significant increase in body temperature over a 3 hour period.Result - The baseline temperatures for all the rabbits in this study were within 1 °C at the start of the test and no animals had a baseline temperature above 39.8 °C or less than 38.5 °C. During the 3 hour observation period, none of the rabbits administered with the test article extract had a temperature rise 0.5 °C at the required observation time points.Conclusion - The test article was determined to be non-pyrogenic.
Hemocompatibility:Four HourThromboresistanceEvaluation in Dogsinterval without test articlecomplications; andThe test articles are considered thromboresistant if1) the test animals survive general anesthesia and a study observationsigns. There was minimalResult - Implantation of the test and control devices in the jugular veins of two canines resulted in no adverse effects or clinical
2) the blood test results, pre/postthrombus formation associatedwith one test device that was
AttributeAcceptance CriteriaResult
weight differences, and patencyand thrombus scores are notsubjectively different between thetest and control articlesunremarkable given the absenceof anticoagulants during theimplant period. Additionally, thecombined analysis of APTT,platelet counts, device weights,and clinical observationsindicated that neither animal'sclotting abilities werecompromised after implantation ofthe devices.Conclusion - The test devicesappear to have similarthromboresistance characteristicsas the control devices.
Hemocompatibility:ComplementActivation C3a andSC5b-9The test article passes ifcomplement activation in humanplasma (C3a Assay and SC5b-9) isnot induced significantly whencompared with the comparisonarticle.Results - The test article and thecomparison article wereevaluated for their capacity toactivate the complement system.Under the conditions of the C3aassay, test article group one,group two, and the comparisonarticle exhibited activation at 6017ng/mL, 8239 ng/mL, and 6048ng/mL. This was 3.0%, 8.3%, and3.0%, respectively, of thenormalized C3a concentrationproduced by CVF. Under theconditions of the SC5b-9 assay,test article group one, group two,and the comparison articleexhibited activation at 4212ng/mL, 13020 ng/mL, and 4047ng/mL. This was 0.1 %, 2.3%,and 0.0%, respectively, of thenormalized SC5b-9 concentrationproduced by CVF.Conclusion - All biomaterialshave the potential to affect themake-up of the complementcomponents of the blood.However, at this time there are noranges or levels established asacceptable.
Hemocompatibility:Platelet andLeukocyte CountThe test article passes if the countsare statistically and adverselysignificant compared to thereference material and comparisonarticle.Result - Test article group oneresults for the leukocyte andplatelet counts showed 96% and103%, respectively, of thereference material. Test articlegroup two results for theleukocyte and platelet countsshowed 100% and 105%,respectively. of the referencematerial. The comparison articleresults for the leukocyte andplatelet counts showed 98% and114%, respectively of thereference material.Conclusion - While allbiomaterials have the potentialto affect the make-up of thevarious components of the blood,at this time there are no ranges orlevels that have been establishedas acceptable.
Hemocompatibility:PartialThromboplastinThe test article passes if the PTT isnot significantly shortened followingcontact with a material underResult - Both test article groupsand the comparison article had anaverage clotting time of 300.0

{4}------------------------------------------------

Traditional 510(k) Silk Road™ Access Catheter

CONFIDENTIAL

{5}------------------------------------------------

Traditional 510(k) Silk Road™ Access Catheter

AttributeAcceptance CriteriaResult
Time (PTT)standardized conditions.seconds. This value was found to be 100% of the negative control.Conclusion - Both test article groups and the comparison article are considered to be non-activators of the intrinsic coagulation pathway and pass the test.
Hemocompatibility:Hemolysis - Direct Contact andExtract (ASTM F 756)The test samples are rated per the hemolytic index above the negative control. If 0 – 2%, classified as non-hemolytic, if 2.1-5%, classified as slightly hemolytic, and if ≥5.1%, classified as hemolytic.Results -Direct: The negative control replicates returned a hemolytic index of 0.1 %. The positive control replicates returned a hemolytic index of 11.8%. The test article returned a hemolytic index value equal to that found in the negative control and fell within the non-hemolytic range.Extract: The negative control replicates returned a hemolytic index of 0.1 %. The positive control replicates returned a hemolytic index of 14.7%. The test article returned a hemolytic index value lower than that found in the negative control and clearly fell within the non-hemolytic range.Conclusion - The test article is considered non-hemolytic under the test conditions employed and passes this test.
Genotoxicity:BacterialMutagenicity Test-Ames AssayThe test article passes if the criteria for positive mutagen are not metResults - The test article did not induce substantial increases in reversion rates of the type that are associated with mutagenesis. Furthermore, no substantial test article toxicity was noted that may have interfered with the ability of the test system to detect mutagens. As none of the tester strains showed an increase in reversion rates when treated with the test article, the test article is determined not to have caused an increase in point mutations, exchanges or deletions.Conclusions - The test article is considered non-mutagenic.
Genotoxicity:In Vitro MouseLymphoma AssayThe test article passes if the mutant frequency is less than 1.8 fold higher than that of the concurrent negative control groups.Results - The controls for the assay performed as required; qualifying both the assay and cell culture system as valid. The mutant frequencies and cloning efficiencies of preparations treated with test article were within the limits defined for a negative response.Conclusion - The test article is considered to be non-mutagenic (non-genotoxic and non-clastogenic) in this test system.
Genotoxicity:In Vivo MouseMicronucleusAssayThe test article passes if it does not show a significant increase in the number of micronucleated polychromatic erythrocytes.Result - In general there were no apparent gross manifestations of toxicity nor biologically significant erythropoietic disturbances
AttributeAcceptance CriteriaResult
Furthermore, there were nobiologically significant increasesin mPCE production in the testarticle treated groups ascompared to the concurrentnegative controls.Conclusion - The test article isconsidered non-mutagenic in thistest system.
Packaging ValidationPackaging must maintain productsterility and protect the productduring shipment and storage.Pass
Sterilization ValidationSilk Road Access Catheter must bevalidated ISO 11135-1:2007Pass
Shelf LifeSilk Road Access Catheter mustperform to specification and thepackage must maintain a sterilebarrier during the shelf-life period aslabeled on the product label.Pass

{6}------------------------------------------------

Safety and Performance Testing - Animal

A GLP animal study was performed to evaluate the safety, performance and handling of the Silk Road Access Catheter as compared to a control device. Based on pathology and histopathology results, the safety acceptance criteria for the study were met. Performance and handling observations were made based on detailed characteristics of the device. No untoward observations were found by the clinician.

The physical, mechanical and in vitro and in vivo performance testing of the subject Silk Road Access Catheter demonstrate that the product is safe and effective for its labeled indications and is Substantially Equivalent to the currently marketed predicate.

K. Conclusions

The Silk Road Access Catheter has been carefully compared to legally marketed devices with respect to intended use and technological characteristics. In addition, nonclinical testing was conducted to validate the performance of the device and ensure the Silk Road Access Catheter functions as intended and meets design specifications. The comparison and non-clinical results demonstrate that the device is substantially equivalent to the predicate device for its intended use.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms reaching upwards, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 1, 2013

Silk Road Medical. Inc. % Mr. Richard M. Ruedy Vice President, RA/CA/QA 735 North Pastoria Avenue Sunnyvale, CA 94085

Re: K130649

Trade/Device Name: Silk Road™ Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, KRA Dated: September 20, 2013 Received: September 23, 2013

Dear Mr. Ruedy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements; including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{8}------------------------------------------------

Page 2 - Mr. Richard M. Ruedy

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Victor Krauthamer -A

Victor Krauthamer, Ph.D. Acting Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K130649

Device Name: Silk Road™ Access Catheter

Indications For Use:

The Silk Road™ Access Catheter is indicated for the introduction of interventional devices into the peripheral vasculature.

Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Victor Krau 2013.11.01

Page 1 of 1

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).