K Number
K130529

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-10-25

(238 days)

Product Code
Regulation Number
882.1620
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RAUMEDIC® - PTO-Monitoring-System measures intracranial pressure, temperature and oxygen and is intended as an adjunct monitor of trends in these parameters, indicating the perfusion status of cerebral tissue local to sensor placement. RAUMEDIC® - PTO-Monitoring-System values are relative within an individual, and should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia or ischemia is a concem.

Use of the parenchymal intracranial pressure monitoring kit with bolt is indicated for children who are at least one year old.

Device Description

The RAUMEDIC® -PTO-Monitoring-System determines the level and change in intracranial pressure (ICP) by using semi-conductor pressure sensors, in temperature using thermistors, and in partial oxygen pressure ptiO2 using a luminescence optical sensor fibre.

The PTO-Monitoring-System consists of intracranial catheters, a cranial bolt kit, the user interface unit (EASY logO), cables to connect the catheters to the EASY logO and cables for the EASY logO to send the measurements to specific patient monitoring units.

Three intracranial catheters are provided: NEUROVENT®-PTO, NEUROVENT® TO and NEUROVENT® PTO 2L. The NEUROVENT®-PTO measures three physiological parameters - ICP, temperature and partial oxygen pressure. The application has to be made with a single lumen Bolt (BOLT-KIT PTO). The NEUROVENT -PTO 2L also measures these three parameters, but has to be placed with a multi lumen Bolt (BOLT KIT PTO 2L) or via tunneling process. The NEUROVENT® TO measures two physiological parameters – temperature and partial oxygen pressure. The application has to be made with a single lumen Bolt (BOLT-KIT PTO).

The NEUROVENT®-PTO or the NEUROVENT®-TO is implanted in the parenchyma using a BOLT KIT PTO. The NEUROVENT®-PTO 2L is implanted in the parenchyma using a BOLT KIT PTO 2L or using a RAUMEDIC® spliceable tunnelling sleeve CH8 (already cleared for market under 510(k) K112017).

For connecting the catheters to the EASY logO two cables – cable PTO and cable LWL - are needed. The cable PTO transfers the electrical signal for ICP and temperature. The cable LWL transfers the optical signal for oxygen pressure.

The EASY logO is the user interface that displays the measurements and alerts the user to the status of the intracranial catheters. The EASY logO displays: oxygen partial pressure pO2, intracranial pressure ICP, pulse amplitude ICPA and temperature T. ICP is a pulsatile signal that consists of the ICP average value and a pulsatile ICPA component, which is superimposed over the ICP average value.

DATALOGGER cables allows the EASY logO to send the measurements to specific patient monitoring units.

AI/ML Overview

The provided text describes the RAUMEDIC® -PTO-Monitoring-System and its performance testing to demonstrate substantial equivalence to previously cleared devices. It does not contain information about a study proving the device meets explicit acceptance criteria in the format requested. However, it does provide performance specifications and comparison to predicate devices, which serve as a de facto set of acceptance criteria for regulatory clearance.

Here’s a summary of the available information, fitted to the request structure where possible.

Device Name: RAUMEDIC® -PTO-Monitoring-System

Intended Use: Measures intracranial pressure (ICP), temperature, and oxygen (ptiO2) as an adjunct monitor of trends, indicating perfusion status of cerebral tissue local to sensor placement. Not for use as the sole basis for diagnosis or therapy. For use in children at least one year old.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly derived from the performance specifications of the predicate devices and relevant standards (ANSI/AAMI NS 28:1988/(R) 2006 for ICP, ASTM E1112-00 (2006) for Temperature, and comparison to K980380 for ptiO2). The tables provided (Table 5, Table 6, and Table 7) list the performance testing and results, demonstrating that the device meets these established comparables.

Feature / ParameterAcceptance Criteria (from Predicate/Standards)Reported Device Performance (RAUMEDIC® -PTO-Monitoring-System)
ICP Performance (Based on ANSI/AAMI NS 28:1988/(R) 2006)
Frequency response> 100 Hz167 Hz (Meets Standard)
Slew rates> 7,500 torr/s10,000 torr/s (Meets Standard)
Time constants< 5 ms3 ms (Meets Standard)
Pressure rangeMin 0 to +100 mmHg-50 to +250 mmHg (Meets Standard)
Accuracy (Catheter alone)< maximum deviation point 3.2.2_NS 28 (-50 to +250 mmHg)< maximum deviation point 3.2.2_NS 28 (Meets Standard)
Accuracy (with NPS-3 display)< maximum deviation point 3.2.2_NS 28 (-40 to +100 mmHg)< maximum deviation point 3.2.2_NS 28 (Meets Standard)
Stability (20 ± 1°C, 39 ± 1°C)dPx < 1.52 torr≤ 1.44 torr (Meets Standard)
Zero-Point Stability (24h)< 1 torr≤ 0.7 torr (Meets Standard)
Zero-Point Stability (168h)< 1.5 torr≤ 1.2 torr (Meets Standard)
Accuracy at 10 torr< 1.5 torr≤ 0.9 torr (Meets Standard)
Accuracy at 20 torr< 1.5 torr≤ 1.2 torr (Meets Standard)
Accuracy at 50 torr< 1.5 torr≤ 1.1 torr (Meets Standard)
Accuracy at 100 torr< 1.5 torr≤ 1.2 torr (Meets Standard)
Risk current< 10 µA at 120 V AC< 0.98 µA at 120 V AC (Meets Standard)
Maximum temperatureNo temperature increase > 40 °C (104 °F)max. 39.5 °C (Meets Standard)
Temperature Performance (Based on ASTM E1112-00 (2006))
CleaningNo visible of technical changesNo visible of technical changes (Meets Standard)
ToxicityToxicity TestToxicity Test (Meets Standard)
AccuracyNo difference between the in vivo and in vitro accuracyNo difference between the in vivo and in vitro accuracy (Meets Standard)
Operating Environment (Temp/RH)< Max. Error Temp. Rang. Table 1 ASTM E1112 (for 40°C/15%RH, 40°C/80%RH, 16°C/40%RH, 16°C/95%RH)< Max. Error Temp.Rang. (Meets Standard for all conditions)
Storage (after 1 month)Test point o.k. after storage at -20 °C, 50 °C/15% RH, 35 °C/95% RHo.k. (Meets Standard)
Oxygen Performance (Compared to DIAMETRIC MEDICAL LTD. NEUROTREND MULTIPARAMETER SENSOR K980380)
Drift over 5.5 days (2.5% O2)< 1.5 mmHg≤ 1.2 mmHg (within predicate's < 11.9 mmHg) (Yes)
Drift over 5.5 days (10% O2)< 1.5 mmHg≤ 1.2 mmHg (within predicate's < 46.2 mmHg) (Yes)
Absolute accuracy (0-120 mmHg)0+/- 2.5 mmHg≤ 0.2 mmHg (within predicate's 0+/- 3.0 mmHg) (Yes)
Absolute accuracy (>120 mmHg)150.00+/-15 mmHg (for predicate: 135.0-165.0 mmHg for values from 120-200 mmHg)146.9 mmHg; 152.3 mmHg (within predicate's range) (Yes)
Response time T90< 200 s for change 150+/-15 mmHg to 0.0 mmHg≤ 51 s (within predicate's ≤ 55 s) (Yes)
ptiO2 sensitive surface> 13 mm² (predicate: ≥ 5 mm²)≥ 15.1 mm² (> 5mm² is o.k. in brain) (Yes)
Length ptiO2 sensitive area> 4mm> 4.9 mm (within predicate's > 4 mm) (Yes)
Outer catheter diameter< 1.35 mm (predicate: ≤ 0.5 mm)≤ 0.99 mm (Yes, noted as "Yes" despite a larger diameter, likely due to similar performance or acceptable clinical difference, or comparison to K120252 predicate which has larger diameters). For Oxygen, the comparison is specifically to K980380. The rationale for "Yes" is not explicitly detailed but stated.

2. Sample Size Used for the Test Set and Data Provenance
The document discusses bench testing rather than clinical studies with human subjects. Therefore, details like "test set," "sample size," "country of origin," and "retrospective or prospective" are not applicable in the context of human data. The testing described is primarily physical and analytical.

  • ICP Performance: The tests were performed on the RAUMEDIC® ICP-Monitoring-System (K103206) and the results were transferred, implying testing of components or integrated systems in a lab setting. The specific number of units tested is not stated.
  • Temperature Performance: Performed on the RAUMEDIC® PTO-Monitoring-System, likely in a lab setting. The specific number of units tested is not stated.
  • Oxygen Performance: Performed internally on the RAUMEDIC® PTO-Monitoring-System, likely in a lab setting. The specific number of units tested is not stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable, as the "test set" refers to bench testing against engineering specifications and predicate device performance, not human-annotated data. Ground truth here is defined by physical measurements and established standards.

4. Adjudication Method for the Test Set
Not applicable, as there were no human experts involved in adjudicating the results of the bench testing. The results were compared directly to numerical specifications or predicate device performance.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device for direct physiological measurement, not an AI-based diagnostic or imaging interpretation system.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
The performance testing described (ICP, Temperature, Oxygen) is inherently "standalone" in the sense that it evaluates the device's physical and measurement capabilities independent of a human interpreter. The device provides raw data (measurements) which are then interpreted by clinicians.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance testing is based on:

  • Established international performance standards (ANSI/AAMI NS 28, ASTM E1112).
  • Comparative performance data from legally marketed predicate devices (K120252, K980380, K002765).
  • Direct physical measurements using calibrated equipment in a controlled laboratory environment.

8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set was Established
Not applicable. This is not an AI/ML device that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a handwritten string of characters, which appears to be an identifier or code. The string reads 'K1305-29'. The characters are written in a cursive style with varying stroke thicknesses, and the overall impression is that of a quickly jotted down note.

Image /page/0/Picture/1 description: The image shows the logo for RAUMEDIC, a company that provides lifeline to health. The logo consists of a circular symbol on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health" in a smaller, lighter font. The overall design is clean and professional, conveying a sense of reliability and trustworthiness.

RAUMEDIC AG · Postfach 5 01 · D - 95205 Münchberg

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - W066-0609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002

Regulatory Affairs

Telefon: (0 92 52) 359-0 Telefax: (0 92 52) 359-10 00 info@RAUMEDIC.com www.RAUMEDIC.de

Direktkontakt: Telefon: (0 92 52) 359-2782

Reiner. Thiem@RAUMEDIC.com

Tag 16th of October 2013

Ihre Nachricht vom Ihre Zeichen

H Thiem Sachbearbeiter Unsere Zeichen thi

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

  1. Name, address, phone and fax number of the applicant

RAUMEDIC AG Hermann-Staudinger-Straße 2 95233 Helmbrechts D - Germany Tel.: 0049/9252/359-0 Fax: 0049/9252/359-1000

2. Contact person

Mr. Reiner Thiem Head of Regulatory Affairs Hermann-Staudinger-Straße 2 95233 Helmbrechts D - Germany Tel.: 0049/9252/359-2782

3. Date of preparation of the summary

October, the 16th 2013

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left and the word "RAUMEDIC" in bold, sans-serif font on the right. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line on either side of the text. The circular graphic appears to be a stylized representation of a globe or sphere.

4. Name of the device

The RAUMEDIC® -PTO-Monitoring-System is composed of the following components:

  • NEUROVENT -PTO .
  • NEUROVENT®-TO .
  • NEUROVENT®-PTO 2L .
  • EASY logO .
  • BOLT KIT PTO .
  • BOLT KIT PTO 2L .
  • Cable LWL .
  • Cable PTO .
  • Cable DATALOGGER GE/MARQUETTE .
  • Cable DATALOGGER Philips/HP .
  • Cable DATALOGGER Siemens/Dräger Infinity .
  • Cable DATALOGGER Datex-Ohmeda .
  • Cable DATALOGGER Hellige .
  • Cable DATALOGGER SpaceLabs �
  • Cable DATALOGGER Nihon Kohden .

Device Classification Name:

Classification Panel: CFR Section: Device Class: Product Code:

Device, Monitoring, Intracranial Pressure, Temperature and Oxygen Neurology 21 CFR §882.1620 Class II GWM

5. Device Description

The RAUMEDIC® -PTO-Monitoring-System determines the level and change in intracranial pressure (ICP) by using semi-conductor pressure sensors, in temperature using thermistors, and in partial oxygen pressure ptiO2 using a luminescence optical sensor fibre.

The PTO-Monitoring-System consists of intracranial catheters, a cranial bolt kit, the user interface unit (EASY logO), cables to connect the catheters to the EASY logO and cables for the EASY logO to send the measurements to specific patient monitoring units.

Three intracranial catheters are provided: NEUROVENT®-PTO, NEUROVENT® TO and NEUROVENT® PTO 2L. The NEUROVENT®-PTO measures three physiological parameters - ICP, temperature and partial oxygen pressure. The application has to be made with a single lumen Bolt (BOLT-KIT PTO). The NEUROVENT -PTO 2L also measures these three parameters, but has to be placed with a multi lumen Bolt (BOLT KIT PTO 2L) or via tunneling process. The NEUROVENT® TO measures two physiological parameters – temperature and partial oxygen pressure. The application has to be made with a single lumen Bolt (BOLT-KIT PTO).

The NEUROVENT®-PTO or the NEUROVENT®-TO is implanted in the parenchyma using a BOLT KIT PTO. The NEUROVENT®-PTO 2L is implanted in the parenchyma using a BOLT KIT PTO 2L or using a RAUMEDIC® spliceable tunnelling sleeve CH8 (already cleared for market under 510(k) K112017).

Device #:

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word is the tagline "Lifeline to Health" in a smaller, italicized font, flanked by horizontal lines.

For connecting the catheters to the EASY logO two cables – cable PTO and cable LWL - are needed. The cable PTO transfers the electrical signal for ICP and temperature. The cable LWL transfers the optical signal for oxygen pressure.

The EASY logO is the user interface that displays the measurements and alerts the user to the status of the intracranial catheters. The EASY logO displays: oxygen partial pressure pO2, intracranial pressure ICP, pulse amplitude ICPA and temperature T. ICP is a pulsatile signal that consists of the ICP average value and a pulsatile ICPA component, which is superimposed over the ICP average value.

DATALOGGER cables allows the EASY logO to send the measurements to specific patient monitoring units.

6. Indications of Use

The RAUMEDIC® - PTO-Monitoring-System measures intracranial pressure, temperature and oxygen and is intended as an adjunct monitor of trends in these parameters, indicating the perfusion status of cerebral tissue local to sensor placement. RAUMEDIC® - PTO-Monitoring-System values are relative within an individual, and should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia or ischemia is a concern.

Use of the parenchymal intracranial pressure monitoring kit with bolt is indicated for children who are at least one year old.

7. Substantial Equivalence Summary

The RAUMEDIC® - PTO-Monitoring-System is substantially equivalent to the DIAMETRIC MEDICAL LTD. NEUROTREND MULTIPARAMETER SENSOR which was cleared for market under 510(k) K980380 on 14" of July 1999, the LICOX Brain Oxygen Monitoring System which was cleared for market under 510(k) K002765 on 24th of November 2000, the RAUMEDIC® ICP-TEMP- MONITORING-SYSTEM which was cleared for market under 510(k) K120252 on 110 of April 2012 and the RAUMEDIC® ICP-MONITORING-SYSTEM which was cleared for market under 510(k) K103206 on 04* of March 2011.

For further information see device comparison tables below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for RAUMEDIC, a company that provides medical products. The logo consists of a circular symbol on the left and the word "RAUMEDIC" in bold letters on the right. Below the word "RAUMEDIC" is the tagline "Lifeline to Health".

Table 1: Comparison to Predicate RAUMEDIC® ICP-TEMP- MONITORING- SYSTEM (K120252)

FeatureRAUMEDIC® PTO-MONITORING- SYSTEMRAUMEDIC® ICP-TEMP.MONITORING- SYSTEM(K120252)SE?
Trade NameNEUROVENT -PTO,NEUROVENT -PTO 2L.NEUROVENT®-TONEUROVENT P-TEMP.NEUROVENT - TEMP-IFD-S.NEUROVENT - TEMP-IFD-RN/A
Indication ForUseThe RAUMEDIC® -PTO-Monitoring-System is indicated foruse by a qualified neurosurgeon fordirect measurement of intracranialpressure and temperature in theparenchyma as well as todetermine the partial oxygenpressure ptiO2 of the interstitialfluid.The RAUMEDIC® -ICP-TEMP-Monitoring-System is indicated foruse by a qualified neurosurgeon fordirect measurement of intracranialpressure and temperature in theparenchyma and in the ventricleand cerebrospinal fluid drainageapplications.YES - formeasurementof intracranialpressure andtemperature; foroxygenmeasurementsee SE-List withpredicatedevice:K880380
Anatomical SiteBrain parenchymaBrain parenchyma / ventricleYES
CatheterSensors LocationCatheter TipCatheter TipYES
Tip DiameterSFSFYES
Single-use-YESYES
CatheterYES
SterilizationProcessWith Ethylene OxideWith Ethylene OxideYES
BoltBolt with compression capBolt with compression capYES
SpliceableTunnelling SleeveSpliceable Tunnelling Sleeve withTrocarSplicable Tunnelling Sleeve withTrocarYES
Size Of AccessHolefor Bolt:4.2 mm >> NEUROVENT®-PTO,NEUROVENT®-TO6.3 mm → NEUROVENT®-PTO 2Lfor Tunnelling:> 6 mm → NEUROVENT®-PTO 2Lfor Bolt:4.2 mm > NEUROVENT®-P-TEMP.6.3 mm -> NEUROVENT -TEMP-IFD-S,NEUROVENT®-TEMP-IFD-Rfor Tunnelling:≥ 6 mm → NEUROVENT®-TEMP-IFD-S,NEUROVENT®-TEMP-IFD-RYES
Fixation OfCatheterFixation wing (sutured to skin)Fixation wing (sutured to skin)YES
Non-fluidCoupling CatheterYESYESYES
ICP - SensorDesign CatheterPiezoresistive pressure sensor(Semi-conductor / WheatstoneBridge)Piezoresistive pressure sensor( Semi-conductor / WheatstoneBridge)YES
ICP - PressureRange-50 to +250 mmHg-50 to +250 mmHgYES
ICP - MaximumPressure1,250 mmHg1,250 mmHgYES
ICP - Accuracy± 1 mmHg± 1 mmHgYES
ICP - Resolution0.1 mmHg0.1 mmHg
ICP - Zero PointStabilityLess than 1 mmHg during first 24oursLess than 2 mmHg during the first7 daysLess than 1mmHg during first 24hoursLess than 2mmHg during the first 7daysYES
ICP - DynamicResponse Time< 5 ms< 5 msYES
ICP - SensitivitySystem5 µV/V/mmHg on the monitor side5 µV/V/mmHg on the monitor sideYES
FeatureRAUMEDIC® PTO-MONITORING- SYSTEMRAUMEDIC® ICP-TEMP-MONITORING- SYSTEM(K120252)SE?
TEMP - SensorDesignThermistorThermistorYES
TEMP -TemperatureRange25 °C - 45 °C25 °C - 45 °CYES
TEMP - Accuracy± 0.1 °C± 0.1 °CYES
TEMP -Resolution0.01 °C0.01 °CYES
TEMP - ZeroPoint Stability± 0.2 °C± 0.2 °CYES
TEMP - DynamicResponse Time< 150 s< 150 sYES
TEMP - AdapterCable PTOICP-TEMP-AdapterICP-TEMP-Adapter HP / PhilipsYES
p₁O₂ - SensorDesign CatheterFiber Optic / Quenching→ measuring amount of reflectedlight from sensor tip in comparisonto pre-calibrated values withreference solutions (see section"12. Substantial EquivalenceDiscussion")---NO- see SE-Listwith predicatedevice:K980380
p₁O₂ - OxygenPressure Range0 to 150 mmHg---NO- see SE-Listwith predicatedevice:K980380
p₁O₂ - Accuracy± 3 % or ± 2.5 mmHg of themeasured value → the highervalue is applicable for p₁O₂ < 120mmHg:< 10 % of the measured value forp₁O₂ from 120 mmHg to 150 mmHg---NO- see SE-Listwith predicatedevice:K980380
p₁O₂ - Resolution0.1 mmHg---NO- see SE-Listwith predicatedevice:K980380
p₁O₂ - Zero PointStabilitymax. 1.5 mmHg in 5 days(at a ptiO2 of ≤ 10 mmHg)---NO- see SE-Listwith predicatedevice:K980380
p₁O₂ - DynamicResponse Time< 200 s for p₁O₂-decrease from150 ± 15 mmHg to 0.0 mmHg---NO- see SE-Listwith predicatedevice:K980380
p₁O₂ - AdapterCable LWL---NO- see SE-Listwith predicatedevice:K980380
Product CodeGWMGWMYES
Registration #PendingK120252N/A
ApplicantRAUMEDIC AGHermann-Staudinger-Str. 295233 HelmbrechtsRAUMEDIC AGHermann-Staudinger-Str. 295233 HelmbrechtsN/A

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "life.to health" in a smaller font size.

  • Lifeline to Health -
  • ICP measurement only with NEUROVENT®-PTO and NEUROVENT®-PTO 2L

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for RAUMEDIC, a company that provides a lifeline to health. The logo consists of a circular graphic on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health".

Table 2: Comparison to predicate DIAMETRIC MEDICAL LTD.NEUROTREND MULTIPARAMETER SENSOR (K980380)

FeatureRAUMEDIC® PTO-MONITORING- SYSTEMDIAMETRIC MEDICAL LTD.NEUROTRENDMULTIPARAMETER SENSOR(K980380)SE?
Trade NameNEUROVENT®-PTO,NEUROVENT®-PTO 2L,NEUROVENT®-TONeurotrend™ MultiparameterSensor(C7004S)N/A
Indication ForUseThe RAUMEDIC® -PTO-Monitoring-System is indicated for use by aqualified neurosurgeon for directmeasurement of intracranialpressure and temperature in theparenchyma as well as to determinethe partial oxygen pressure ptiO2 ofthe interstitial fluid.The Neurotrend MultiparameterSensor is a disposable, sterile,single-use, device for thecontinuous measurement ofintracranial oxygen, carbondioxide, pH and temperature thatis used in conjunction with asuitable intracranial accessdevice.YES - formeasurementof oxygen; forintracranialpressure andtemperaturemeasurementsee SE-List withpredicatedevice:K120252
Anatomical SiteCatheterBrain parenchymaBrain ventricleNO- see SE-Listwith predicatedevice:K120252
Sensors LocationCatheter TipCatheter TipYES
Tip Diameter5 F2 FNO- see SE-Listwith predicatedevice:K120252
Single-use-CatheterYESYESYES
SterilizationProcessWith Ethylene OxideWith Gamma - RadiationNO- see SE-Listwith predicatedevice:K120252
BoltBolt with compression capA suitable intracranial accessdeviceNO- see SE-Listwith predicatedevice:K120252
SpliceableTunnelling SleeveSpliceable Tunnelling Sleeve withTrocarA suitable intracranial accessdeviceNO- see SE-Listwith predicatedevice:K120252
Size Of AccessHolefor BOLT usage:4.2 mm → NEUROVENT®-PTO,NEUROVENT®-TO6.3 mm → NEUROVENT®-PTO 2Lfor Tunnelling Sleeve usage:≥ 6 mm → NEUROVENT®-PTO 2LNO- see SE-Listwith predicatedevice:K120252
Fixation OfCatheterFixation wing (sutured to skin)YES
Non-fluidYESYESYES
FeatureRAUMEDIC® PTO-MONITORING- SYSTEMDIAMETRIC MEDICAL LTD.NEUROTRENDMULTIPARAMETER SENSOR(Ka80380)SE?
ICP - SensorDesign CatheterPiezoresistive pressure sensor(Semi-conductor / WheatstoneBridge)---NO- see SE-Listwith predicatedevice:K120252
ICP - PressureRange-50 to +250 mmHg---NO- see SE-Listwith predicatedevice:K120252
ICP - MaximumPressure1,250 mmHg---NO- see SE-Listwith predicatedevice:K120252
ICP - Accuracy± 1 mmHg---NO- see SE-Listwith predicatedevice:K120252
ICP - Resolution0.1 mmHg------------------------------------------------------------------------------------------------------------------------------------------------------------------------------NO- see SE-Listwith predicatedevice:K120252
ICP - Zero PointStabilityLess than 1 mmHg during first 24oursLess than 2 mmHg during the first 7days----NO- see SE-Listwith predicatedevice:K120252
ICP - DynamicResponse Time< 5 ms---NO- see SE-Listwith predicatedevice:K120252
ICP - SensitivitySystem5 µV/V/mmHg on the monitor side-NO- see SE-Listwith predicatedevice:K120252
TEMP - SensorDesignThermistorThermocoupleNO- see SE-Listwith predicatedevice:K120252
TEMP -TemperatureRange25 °C - 45 °C10 °C - 42 °CNO- see SE-Listwith predicatedevice:K120252
TEMP - Accuracy+ 0.1 °C+0.1 °CYES
TEMP -Resolution0.01 °C0.01 °CYES
TEMP - ZeroPoint Stability+ 0.2 °C+ 0.2 °CYES
TEMP - DynamicResponse Time< 150 s< 150 sYES
TEMP - AdapterCable PTOPDM CableYES
FeatureRAUMEDIC® PTO-MONITORING-SYSTEMDIAMETRIC MEDICAL LTD.NEUROTRENDMULTIPARAMETER SENSOR(K980380)SE?
ptO2 - SensorDesign CatheterFiber Optic / Quenching→ measuring amount of reflectedlight from sensor tip in comparisonto pre-calibrated values withreference solutions (see section "12.Substantial EquivalenceDiscussion")Fiber Optic / Quenching→ measuring amount of reflectedlight from sensor tip in comparisonto pre-calibrated values withreference solutionsYES
ptO2 - OxygenPressure Range0 to 150 mmHg10 to 430 mmHgNO; the lowerlimit isdemonstratedbyperformancetesting
ptO2 - Accuracy± 3 % or ± 2.5 mmHg of themeasured value → the higher valueis applicable for ptO2 < 120 mmHg;< 10 % of the measured value forptO2 from 120 mmHg to 150 mmHg± 3 % or ± 2.5 mmHg of themeasured value → the highervalue is applicable for ptO2 < 120mmHg:< 10 % of the measured value forptO2 from 120 mmHg to 200mmHgYES
ptO2 - Resolution0.1 mmHg0.1 mmHgYES
ptO2 - Zero PointStabilitymax. 1.5 mmHg in 5 days(at a ptiO2 of ≤ 10 mmHg)< 0.5 % per hourYESwithin thestability rangeof theNEUROTREND
ptO2 - DynamicResponse Time< 200 s for ptO2-decrease from 150± 15 mmHg to 0.0 mmHg< 200 s for ptO2-decrease from150±15 mmHg to 0.0 mmHgYES
ptO2-AdapterCable LWLPDM CableYES
Product CodeGWMGWMYES
Registration #PendingK980380N/A
ApplicantRAUMEDIC AGHermann-Staudinger-Str. 295233 HelmbrechtsGermanyDiametrics Medical, Ltd.Short StreetHigh WycombeBuckinghamshire, UK HP11 2QHN/A

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Underneath the word "RAUMEDIC" is the tagline "Lifeline to Health".

  • Lifeline to Health ----

1

Device #:_____________________________________________________________________________________________________________________________________________________________________

,

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "life. united. health."

  • Lifeline to Health ----
  • ICP measurement only with NEUROVENT®-PTO and NEUROVENT®-PTO 2L

Device #:_

:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Raumedic logo. The logo consists of a circular symbol on the left and the word "RAUMEDIC" in bold, sans-serif font on the right. Below the word is the tagline "Lifeline to Health" with a line above and below the text.

Table 3: Comparison to Predicate LICOX Brain Oxygen Monitoring System (K002765)

FeatureRAUMEDIC® PTO-MONITORING-SYSTEMRAUMEDIC® ICP-MONITORING-SYSTEMLICOX Brain OxygenMonitoring SystemSE?
Trade NameEASY logoNPS3CMP® MonitorN/A
Indication for UseThe RAUMEDIC® -PTO-Monitoring-System is indicatedfor use by a qualifiedneurosurgeon fordirect measurementof intracranialpressure andtemperature in theparenchyma as wellto determine thepartial oxygenpressure ptiO2 of theinterstitial fluid.The RAUMEDIC® -ICP-Monitoring-System is indicatedfor use by a qualifiedneurosurgeon fordirect measurementof intracranialpressure in theparenchyma.The LICOX BrainOxygen PressureMonitoring Systemdirectly measuresPartial Pressure ofOxygen in the brain.The LICOX system canalso utilize braintemperature informationfor temperaturecompensation of thepO2.YES
ICP MeasurementYESYESNOYES
TemperatureMeasurementYESNOYESYES
pO2MeasurementYESNOYESYES
Pressure Range-40 to +100 mmHg-40 to +100 mmHg---YES
TemperatureRange15 °C - 45 °C---0°C - 46 °CNO; limits aredemonstratedbyperformancetesting inaccordance toASTM E1112
pO2 Range0 to 150 mmHg---0 to 150 mmHgYES; limits aredemonstratedbyperformancetesting
ScreenYESYESYESYES
MonitoringContinuousContinuousContinuousYES
Power SourceA/C wall outletD/C batteryA/C wall outletYES
Data OutputAnalog / Serial---Analog / SerialYES
CalibrationAuto ZeroAuto ZeroAuto ZeroYES
AlarmsTechnical AlarmsTechnical AlarmsTechnical AlarmsYES
Case MaterialPlasticPlasticPlasticYES
Product CodeGWMGWMGWMYES
Registration #PendingK103206K002765N/A
ApplicantRAUMEDIC AGHermann-Staudinger-Str. 295233 HelmbrechtsGermanyRAUMEDIC AGHermann-Staudinger-Str. 295233 HelmbrechtsGermanyIntegra NeuroSciences5955 Pacific CenterBlvd.San Diego, CA 92121USAN/A

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Raumedic logo. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word is the tagline "Lifeline to Health" in a smaller, italicized font. The logo appears to be for a healthcare or medical-related company.

8. Device Testing

The Biocompatibility studies for the material/components of the medical devices NEUROVENT® PTO, NEUROVENT®-TO, NEUROVENT®-PTO 2L, BOLT KIT PTO and the BOLT KIT PTO 2L were conducted per ISO 10993 standard.

In the toxicological evaluation reports in accordance to ISO 10993/1 for the tested devices NEUROVENT®-PTO (100801-40-B rev 01), BOLT KIT PTO (101164-40-A rev 01) and BOLT KIT PTO 2L (122471-40) it is stated, that the available test results, because of testing the EQ-sterilized devices for cytotoxicity according to ISO 10993/5 and performing additional chemical analysis in accordance to ISO 10993/18, reveal no toxicologically relevant information, considering biological effects like cytotoxicity, tissue irritation (implant reactions), sensitization, systemic, subchronic toxicity and genotoxicity based on the intended use of the devices, that would affect the biological safety of the patient. For test matrix see table below.

The additional component of the BOLT KIT PTO 2L mandrel was also tested for cytotoxicity in accordance to ISO 10993/5. The test report 112078-20 indicate that the mandrels do not release substances in cytotoxic concentration during a permanent 24 h contact of 4.5 cm² surface area to 1 ml physiological fluid.

For all tests the material insolubilities are in compliance with the requirements of ISO 10993/1 for the intended use. There is no evidence that any effects hazardous to the patient will arise by leachable ingredients and/or residues.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for RAUMEDIC, a company that provides lifeline to health. The logo consists of a circular symbol on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health" with a line on either side.

Table 4: Toxicological tests performed on the devices
-----------------------------------------------------------
MaterialTestCommentsResult
NEUROVENT-PTOCytotoxicity4.5 cm2/ml DMEM-FBS, 24 h, 37 °CModerately cytotoxic
ISO 10993-5L 929 cell cultures, quantitative(the moderately
determination of cell proliferationcytotoxic effects
observed are within the
typical value range
obtained with aliphatic
polyurethanes)
Chemical analysis3 cm²/ml ethanol/water (1:20), 24 h, 3732.3 µg/cm²/24 h
150 10993-18°C, GC-FID/GC-MS. quantitative and(cyclohexanone)
qualitative determination of organic(Cyclohexanone is
leachableswidely used as solvent,
bonding and processing
agent for medicaldevices. The
toxicological profile is
defined. It has not been
classified by EU as
carcinogenic, mutagenic
or reprotoxic. As well,
there is no indication of
sensitizing, irritating
and/or other toxic
properties during
application of device)
3 cm²/ml isopropanol,characterized (mainly
24 h, 37 °C, GC-MS characterization ofcyclohexanone)
organic extractables4.5 cm2/ml DMEM-FBS, 24 h, 37 °C
Intracranial screw(milled) with fixingCytotoxicityISO 10993-5L 929 cell cultures, quantitativen.n.
cap and siliconedetermination of cell proliferation
seal of BOLT KIT
Chemical analysis3 cm²/ml ethanol/water (1:20), 24 h, 37n.u.
ISO 10993-18°C, GC-FID/GC-MS. quantitative and(< 0.3 µg/cm²/24 h)
qualitative determination of organic
leachables
3 cm²/ml isopropanol,characterized
24 h, 37 °C, GC-MS characterization of(silicone residues of the
organic extractablesseal)
Intracranial screwCytotoxicity4.5 cm2/ml DMEM-FBS, 24 h, 37 °Cn.n.
(injection mouled)of BOLT KITISO 10993-5L 929 cell cultures, quantitativedetermination of cell proliferation
Chemical analysis3 cm2/ml ethanol/water (1:20), 24 h, 37n.n.
ISO 10993-18°C, GC-FID/GC-MS. quantitative and(< 0.3 µg/cm²/24 h)
qualitative determination of organic
leachables
3 cm2/ml isopropanol,n.n.
24 h, 37 °C, GC-MS characterization of
organic extractables
BOLT KIT PTO 2LCytotoxicity4.5 cm2/ml DMEM-FBS, 24 h, 37 °Cn.n.
Patient-contactingISO 10993-5L 929 cell cultures, quantitative
surfaces EOdetermination of cell proliferation
sterilizedBOLT KIT PTO 2L*Chemical analysis3 cm2/ml ethanol/water (1:20), 24 h, 37
Patient-contactingISO 10993-18°C, GC-FID/GC-MS. quantitative andப.ப.(3 µg/cm²/24 h)
sufacesqualitative determination of organic
leachables
3 cm²/ml isopropanol,characterized (mainly
24 h, 37 °C, GC-MS characterization ofadhesive residues
organic extractables
  • the inner surfaces of BOLT CH9 were coextracted for technical reasons

n.n. no toxicologically relevant effects observed in comparison to the controls

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for RAUMEDIC, a company that provides a lifeline to health. The logo consists of a stylized circle on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health".

In addition, Performance testing in accordance to the following standard ANSI/AAMI NS 28:1988/(R) 2006 was performed on the RAUMEDIC® ICP-Monitoring-System (K103206). Table 5 provides the testing performed to evaluate the ICP-Monitoring System.

Since the catheter NEUROVENT®-PTO or NEUROVENT®-PTO 2L is equal in dimension, made from the same materials and used the same ICP measurement properties to the tested catheters, the results of performance testing for the ICP-Monitoring System can be transferred to the NEUROVENT®-PTO or NEUROVENT®-PTO 2L.

PerformanceTestingANSI/AAMI NS28:1988/(R) 2006Test conditions;requirementsResultsMeetStandard?
FrequencyresponsePoint 1.2. PB_27.11.09;> 100 Hz167 Hzyes
Slew ratesPoint 1.2. PB_27.11.09;> 7,500 torr/s10.000 torr/syes
Time constantsPoint 3.2. PB_27.11.09;< 5 ms3 msyes
Pressure rangePoint 4.1. PB_27.11.09;Min 0 to +100 mmHg-50 to + 250 mmHgyes
AccuracyPoint 5.2. PB_27.11.09;-50 to +250 mmHg (catheter alone)< maximum deviation point3.2.2_NS 28yes
-40 to + 100 mmHg in combinationwith NPS-3 display device< maximum deviation point3.2.2_NS 28yes
Stability ofpressuremeasurementsPoint 6.2. PB_27.11.09;Acc. to guidelines 4.1.2.1 (8),temperature 20 ± 1°C (68 ± 2°F) and39 ± 1 °C (102 ± 2°F)dPx < 1.52 torr≤ 1.44 torryes
Specific zero-point testingPoint 7.2. PB_27.11.09;drift ona 24-hour basis, < 1 torr; drift andon a 168-hour basis, < 1.5 torr≤ 0.7 torrand≤ 1.2 torryes
Accuracy at 10,20, 50 and 100torrPoint 8.2. PB_27.11.09;After least 10 days on10 torr: < 1.5 torr,≤ 0.9 torryes
20 torr. < 1,5 torr≤ 1.2 torryes
50 torr: < 1.5 torrand100 torr: < 1.5 torr≤ 1.1 torrand≤ 1.2 torryes
Risk currentPoint 9.2. PB_27.11.09;< 10 µA at 120 V AC< 0.98 µA at 120 V ACyes
MaximumtemperaturePoint 10.2. PB_27.11.09;No temperature increase> 40 °C (104 °F)max. 39.5 °Cyes

Table 5: ICP Performance testing-bench

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular symbol on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line above and below the text.

Performance testing in accordance to the following standard ASTM E1112-00 (2006) was
performed on the RAUMEDIC® PTO-Monitoring-System. Table 6 provides the testing performed to evaluate the PTO-Monitoring System.

Table 6: Temperature Performance testing-bench

PerformanceTestingASTM E1112-00(2006)Test conditions;requirementsResultsMeetStandard?
CleaningPoint 5.2 PB_08.12.11No visible of technical changesNo visible of technical changesyes
ToxicityPoint 5.3 PB_08.12.11Toxicity TestToxicity Testyes
AccuracyPoint 5.4 PB_08.12.11No difference between the in vivo andin vitro accuracyNo difference between the in vivo and in vitro accuracyyes
OperatingEnvironmentPoint 5.5 PB_08.12.1140 °C, 15 % RH< Max. Error Temp. Rang. Table 1ASTM E111240 °C, 80 % RH< Max. Error Temp. Rang. Table 1ASTM E111216 °C, 40 % RH< Max. Error Temp. Rang. Table 1ASTM E111216 °C, 95 % RH< Max. Error Temp. Rang. Table 1ASTM E1112< Max. Error Temp.Rang.< Max. Error Temp.Rang.< Max. Error Temp.Rang.< Max. Error Temp.Rang.yesyesyesyes
StoragePoint 5.5.2 PB_08.12.11above Test point o.k.after storage period of one month on-20 °C, RH not applicableandafter storage period of one month on50 °C, 15% RHandafter storage period of one month on35 °C, 95 % RHo.k.yes

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for RAUMEDIC, a company that provides lifeline to health. The logo consists of a circular symbol on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health" with a line on either side.

Internal Performance testing was performed on the RAUMEDIC® PTO-Monitoring-System. Table 7 provides the testing performed to evaluate the PTO-Monitoring System.

Table 7: Oxygen Performance testing-bench

PerformanceTestingTest conditions;requirementsResultsDIAMETRICMEDICALLTD.NEUROTRENDMULTIPARAMETERSENSOR(K980380)SE?
Drift over 5.5 dayson 2.5 % oxygen saturation< 1.5 mmHgon 10 % oxygen≤ 1.2 mmHg< 11.9 mmHgyes
saturation< 1.5 mmHg< 46.2 mmHgyes
Absolutemeasurementaccuracy of the ptiO2Range 0....120mmHg0+/- 2.5 mmHg≤ 0.2 mmHg0+/- 3.0 mmHgyes
Absolutemeasurementaccuracy of the ptiO2Range>120 mmHg150.00+/-15 mmHg146.9 mmHg152.3 mmHg135.0 mmHg165.0 mmHgyes
Response time T90< 200 s for change150+/-15 mmHg to0,0 mmHg≤ 51 s≤ 55 syes
ptiO2 sensitivesurface> 13 mm²≥ 15.1 mm²≥ 5 mm²yes> 5mm²is o.k. in brain
Length of the ptiO2sensitive area> 4mm> 4.9 mm> 4 mmyes
Outer catheterdiameter< 1.35 mm≤ 0.99 mm≤ 0.5 mmyes

The oxygen function of NEUROVENT® -PTO is comparable to DIAMETRIC MEDICAL LTD.NEUROTREND MULTIPARAMETER SENSOR (which is cleared to market under 510(k) (K980380) for the test criteria shown in the table above.

9. Conclusion:

The RAUMEDIC® -PTO-Monitoring-System is equivalent to the predicate devices in that:

  • The devices have the same intended use and indication for use. ●
  • . The devices are made of the same materials or substantially equivalent materials.
  • The devices have equivalent form, function, procedures and features. .
  • The devices demonstrate equivalent performance. .

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 25, 2013

Raumedic AG c/o TUV SUD America Mr. Alexander Schapovalov 1775 Old Hwy 8 NW Suite #104 New Brighton, MN 55112

Rc: K130529

Trade/Device Name: Raumedic-PTO-Monitoring System Regulation Number: 21 CFR §882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM Dated: August 22, 2013 Received: August 26, 2013

Dear Mr. Schapovalov:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRII does not evaluate information related to contract liability warrantics. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{15}------------------------------------------------

Page 2 - Mr. Alexander Schapovalov

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (2) CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joyce M. Whang -S

for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

Indications for Use

510(k) Number (if known): _ K130529

Device Name: RAUMEDIC® -PTO-Monitoring-System

Indications For Use:

The RAUMEDIC® - PTO-Monitoring-System measures intracranial pressure, temperature and oxygen and is intended as an adjunct monitor of trends in these parameters, indicating the perfusion status of cerebral tissue local to sensor placement. RAUMEDIC® - PTO-Monitoring-System values are relative within an individual, and should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia or ischemia is a concem.

Use of the parenchymal intracranial pressure monitoring kit with bolt is indicated for children who are at least one year old.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

Page 1 of 1__1

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Joyce M. Whang -S

§ 882.1620 Intracranial pressure monitoring device.

(a)
Identification. An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.(b)
Classification. Class II (performance standards).