K Number
K112017

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-10-11

(89 days)

Product Code
Regulation Number
882.1620
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RAUMEDIC® -ICP-Monitoring-System ventricular is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. Use of the ventricular intracranial pressure monitoring kit with bolt is contra-indicated in children under one year old. The RAUMEDIC® catheters are MR Unsafe.

Device Description

The RAUMEDIC® -ICP-Monitoring-System ventricular is composed of NEUROVENT® components, spliceable tunneling sleeves, and bolt and drill kits. The device is used for intracranial pressure monitoring and cerebrospinal fluid drainage applications, and is implanted in the ventricle using various methods.

AI/ML Overview

Despite the user's request, the provided document does not contain the detailed information necessary to describe acceptance criteria, device performance, or study details such as sample sizes, ground truth establishment, or multi-reader multi-case studies for a diagnostic AI device.

The document is a 510(k) summary for a medical device called the "RAUMEDIC® -ICP-Monitoring-System ventricular." This device is an intracranial pressure monitoring system, not an AI-powered diagnostic tool. The information provided is typical for a traditional medical device submission, focusing on:

  • Device Identification: Name, applicant, contact, date.
  • Device Description: Components (NEUROVENT® components, tunneling sleeves, bolt and drill kits).
  • Intended Use/Indications for Use: Ventricular pressure monitoring and cerebrospinal fluid drainage, contraindications (children under one year old for the bolt kit), and MR safety status (MR unsafe).
  • Substantial Equivalence: Comparison to predicate devices (K103206 and K062584).
  • Testing: Mentions biocompatibility studies and performance testing to demonstrate safety and compliance with standards, but does not provide details on specific performance metrics, acceptance criteria, or study methodologies that would be relevant for an AI diagnostic device.
  • Regulatory Classification: Class II, product code GWM.

Therefore, I cannot fulfill the request for information on acceptance criteria and study details as they pertain to an AI diagnostic device, because the provided text describes a physical medical device for monitoring and fluid drainage, not an AI software.

{0}------------------------------------------------

)<112017

Image /page/0/Picture/1 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word is the tagline "Lifeline to Health" between two horizontal lines. The graphic on the left appears to be a stylized representation of an eye.

OCT 1 1 2011

i

RAUMEDIC

Lifeline to Health

RAUMEDIC AG Postfach 5 01 D-95205 MünchbergForschung & Entwicklung
Food and Drug AdministrationTelefon: (0 92 52) 359-0
Center for Devices and Radiological HealthTelefax: (0 92 52) 359-10 00
Document Mail Center - W066-0609info@RAUMEDIC.com
10903 New Hampshire Avenuewww.RAUMEDIC.de
Silver Spring, MD 20993-0002Direktkontakt:
Telefon: (0 92 52) 359-2782
Ihre Nachricht vomSachbearbeiter H Thiem
Ihre ZeichenUnsere Zeichen thi
Reiner.Thiem@RAUMEDIC.com
Tag
03.06.2011

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

  1. Name, address, phone and fax number of the applicant
RAUMEDIC AG
Hermann-Staudinger-Straße 2
95233 Helmbrechts
D - Germany
Tel.: 0049/9252/359-0
Fax: 0049/9252/359-1000
  1. Contact person
Mr. Reiner Thiem
Head of Regulatory Affairs
Hermann-Staudinger-Straße 2
95233 Helmbrechts
D- Germany
Tel.: 0049/9252/359-2782
  1. Date of preparation of the summary

June, the 03rd 2011

Device #: ____________________Page 1 of 3
---------------------------------------------

{1}------------------------------------------------

K112017

Image /page/1/Picture/1 description: The image shows the Raumedic logo. The logo consists of a circular symbol on the left, followed by the word "RAUMEDIC" in bold, uppercase letters. Below the word is the tagline "Lifeline to Health" with a line on either side. The logo appears to be for a healthcare or medical company.

Image /page/1/Picture/2 description: This document describes the RAUMEDIC® -ICP-Monitoring-System ventricular device. The device is composed of NEUROVENT® components, spliceable tunneling sleeves, and bolt and drill kits. The device is used for intracranial pressure monitoring and cerebrospinal fluid drainage applications, and is implanted in the ventricle using various methods. The document also includes the device classification name, panel, CFR section, class, and product code.

{2}------------------------------------------------

K11 2017

Image /page/2/Picture/1 description: The image shows the Raumedic logo. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line on either side of the text. The circular graphic appears to have a stylized "C" shape within it.

Image /page/2/Picture/2 description: This document describes the intended use, equivalence, and testing of the RAUMEDIC® -ICP-Monitoring-System ventricular device. The device is indicated for ventricular pressure monitoring and cerebrospinal fluid drainage applications. It is substantially equivalent to predicate devices cleared under 510(k) K103206 and K062584. Biocompatibility studies and performance testing have demonstrated the device's safety and compliance with relevant standards.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

TUV SUD America, Inc. c/o Mr. Olaf Teichert Foreign Affairs 1775 Old Highway 8 NW Suite 104 New Brighton, MN 55112-1891

OCT 1 1 2011

Re: K112017

Trade/Device Name: Raumedic ICP Monitoring System Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM Dated: July 11, 2011 Received: July 14, 2011

Dear Mr. Teichert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

4 J-

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Olaf Teichert

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melvin R. Fields, DMD

Malvina B. Eydelman M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K11201+

Image /page/5/Picture/1 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line on either side of the text. The logo is simple and professional, and the tagline suggests that the company is involved in the healthcare industry.

Indications for Use

510(k) Number (if known): K
Device Name:Device, Monitoring, Intracranial Pressure
Indications for Use:
The RAUMEDIC® -ICP-Monitoring-System ventricular is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications.
Use of the ventricular intracranial pressure monitoring kit with bolt is contra-indicated in children under one year old.
The RAUMEDIC® catheters are MR Unsafe.
Prescription Use (Part 21 CFR 801 Subpart D)✓ AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Prescription Use (Per 21 CFR 801.109)X
------------------------------------------

(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) NumberK112017
------------------------

§ 882.1620 Intracranial pressure monitoring device.

(a)
Identification. An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.(b)
Classification. Class II (performance standards).