(259 days)
JXB182 Non-contact Infrared Thermometer is an infrared thermometer intended to measure forehead temperature of infants and adults without contacting human body. It can be used by consumers in household environment and doctor in clinic as reference.
All objects, solid, liquid or gas, emit energy by radiation. The intensity of this energy depends on the temperature of the object. The JXB-182 infrared thermometer is therefore able to measure the temperature of a person by the energy the person emits. This measurement can be taken thanks to an external temperature probe on the device which permanently analyses and registers the ambient temperature. Therefore, as soon as the operator holds the thermometer near the forehead and activates the radiation sensor, the measurement is taken instantly by detection of the infrared heat generated by the arterial blood flow. Body heat can therefore be measured without any interference from the heat of the surrounding environment. The JXB-182 thermometer without contact has been developed using latest infrared technology . This technology allows temporal artery (TA) temperature to be taken at a distance of about 5cm away from the forehead. Then the result will be displayed on LCD. Besides, it owns Led backlight which can show three different colors in different temperature ranges and three-language prompt function.
The provided text describes the JXB-182 Non-contact Infrared Thermometer and its substantial equivalence to a predicate device.
Here's the breakdown of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Item / Criteria | Acceptance Criteria (Predicate) | Reported Device Performance (JXB-182) |
|---|---|---|
| Measurement Method | Infrared Radiation Detection | Infrared Radiation Detection |
| Range (Body Mode) | 32-42.9° C | 32-42.9° C |
| Accuracy | ± 0.3°C | ± 0.3°C |
| Precision | ± 0.3°C (34-35.9° C), ±0.2°C (36-39°C), ±0.3°C (39-42.5°C) | ± 0.3°C (34-35.9° C), ±0.2°C (36-39°C), ±0.3°C (39-42.5°C) |
| Power Supply | DC 3V (2AA Batteries) | DC 3V (2AA Batteries) |
| Material | ABS | ABS |
| Operating Condition (Temperature) | 10° C to 40° C | 10° C to 40° C |
| Display | LCD | LCD |
| Display Resolution | 0.1° C | 0.1° C |
| Measurement Distance | 5cm-8cm | 3cm~5cm |
| Operating Condition (Humidity rate) | ≤ 85% | ≤ 95% |
2. Sample Size for the Test Set and Data Provenance
The document states: "Subject device JXB 182 is compliant to the ASTM E 1965-98(Reapproved 2009), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met."
However, it does not specify the sample size used for the clinical tests nor the data provenance (e.g., country of origin, retrospective or prospective nature of the study).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical tests. It only mentions that activities were performed by "designate individual(s)."
4. Adjudication Method for the Test Set
The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. This device is a standalone infrared thermometer, not an AI-assisted diagnostic tool that would involve human readers. Therefore, there's no information on humans improving with AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance evaluation was conducted. The clinical tests were performed to ensure the device itself met the performance criteria outlined in ASTM E 1965-98. The device's performance characteristics (accuracy, precision, range) are reported for the device operating independently.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The type of ground truth used for the clinical tests is implied to be reference temperature measurements, as per the industry standard ASTM E 1965-98, which sets specifications for infrared thermometers. This standard typically involves comparing the device's measurements to a highly accurate reference thermometer.
8. The sample size for the training set
The document does not provide information on a training set size. As this is a medical device, and not explicitly an AI/ML product developed with distinct training and test sets in the typical AI sense, the concept of a "training set" for the device's core functionality (temperature measurement) is not outlined. Performance is validated against established standards.
9. How the ground truth for the training set was established
Since no specific "training set" is mentioned in the context of AI/ML development, how its ground truth would be established is not applicable or described in this document. The device's design and engineering would have been developed against known thermodynamic principles and calibrated using established standards.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font. The letter B has a design on the upper right side of the letter. There is a horizontal line under the letters.
Jinxinbao Electronic Co., Ltd
Section 5 510(k) Summary
(As required by 21 CFR 807.92(a))
5.1 Submitter Information
- Company: Jinxinbao Electronic Co., Ltd
- · Address: No.38, Huanzhenxi Road, Dagang Town, Nansha, Guangzhou, China
- · Phone: 86-20-34938449
- · Fax: 86-20-22865301
- · Email: xhoucc@jxb-htb.net
· Contact: Jiacheng Guo, General Manager
· Date: November 25, 2012
5.2 Device Information
· Trade/Proprietary Name: Non-contact Infrared Thermometer-JXB 182
· Common Name: Infrared Thermometer
· Classification Information: Classification: 2
Review Panel: General Hospital
Name: Thermometer, Electronic, Clinical
Regulation Number: 21 CFR 880.2910
Product Code: FLL
• Predicate Device: Remote Infrared Thermometer, Model RC002, K090361
- · Device Description:
All objects, solid, liquid or gas, emit energy by radiation. The intensity of this energy depends on the temperature of the object. The JXB-182 infrared thermometer is therefore able to measure the temperature of a person by the energy the person emits. This measurement can be taken thanks to an external temperature probe on the device which permanently analyses and registers the ambient temperature. Therefore, as soon as the operator holds the thermometer near the forehead and activates the radiation sensor, the measurement is taken instantly by detection of the infrared heat generated by the arterial blood flow. Body heat can therefore be measured without any interference
OCT 15 2013
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters J, X, and B in a bold, sans-serif font. The letters are black and appear to be slightly distressed or textured. The letter B is partially obscured by a pattern of small dots or speckles, which adds a visual element to the design.
from the heat of the surrounding environment.
The JXB-182 thermometer without contact has been developed using latest infrared technology . This technology allows temporal artery (TA) temperature to be taken at a distance of about 5cm away from the forehead. Then the result will be displayed on LCD. Besides, it owns Led backlight which can show three different colors in different temperature ranges and three-language prompt function. • Intended Use:
JXB-182 Non-contact Infrared Thermometer is an infrared thermometer intended to measure forehead temperature of infants and adults without contacting human body. It can be used by consumers in household environment and doctor in clinic as reference.
| Item | Subject Device | Predicate Device |
|---|---|---|
| Device Name | Non-contact Infrared Thermometer | Remote Infrared Thermometer |
| Model JXB-182 | Model RC002 | |
| Measurement | Infrared Radiation Detection | Infrared Radiation Detection |
| Method | ||
| Range | In body mode: 32-42.9° C | In body mode: 32-42.9° C |
| Accuracy | ± 0.3°C | ± 0.3°C |
| Precision | ± 0.3°C(34-35.9° C) | ± 0.3°C(34-35.9° C) |
| ±0.2°C(36-39°C) | ±0.2°C(36-39°C) | |
| ±0.3°C(39-42.5°C) | ±0.3°C(39-42.5°C) | |
| Measurement | 3cm~5cm | 5cm-8cm |
| Distance | ||
| Power Supply | DC 3V (2AA Batteries) | DC 3V (2AA Batteries) |
| Material | ABS | ABS |
| Operating | Temperature: 10° C to 40° C | Temperature: 10° C to 40° C |
| Condition | Humidity rate: ≤ 95% | Humidity rate: ≤ 85% |
| Display | LCD | LCD |
| Display Resolution | 0.1° C | 0.1° C |
5.3 Comparison of Required Technology Characteristics
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font and are arranged horizontally. The letter B has a pattern of small circles on it. The logo is black and white.
Jinxinbao Electronic Co., Ltd
| Dimension | 175×50×52mm(L×W×H) | 196×150×50mm(L×W×H) |
|---|---|---|
| Weight | 148g | 220g |
| With Led Backlight | Yes, with three colors | Yes |
| Automatic Stop | 5 seconds | 5 seconds |
| LanguagePrompt | Yes, three languages | No |
| Function | (French, English, Spanish) |
Brief Summary:
The subject device - JXB 182 Non-contact Infrared Thermometer enjoys almost the same design and technological characteristics with the predicate device, such as the same measurement method, accuracy, precision, power supply, display, material and measurement range, and the similar operating condition and measurement distance.
Though they differ in dimension and weight, such trivial difference in appearance will not affect the core function of the device, let alone affecting the comparison of substantial equivalence. A lso they are different in Led backlight and language prompt function, that is, the subject device has two more functions - three-color backlight and three-language prompt function than the predicate device. But such addition will not affect their core usage of the device. Besides, the tests mentioned below have demonstrated the subject device is as safe and effective as the predicate device.
5.4 Discussion of Tests Performed
Clinical Tests:
Subject device JXB 182 is compliant to the ASTM E 1965-98(Reapproved 2009), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
Non-clinical Tests:
The subject device was tested to evaluate its safety and effectiveness according to the following applicable standards:
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font and are arranged horizontally. The letter B has some decorative elements on the right side. The logo is black and white.
· Electrical Safety:
IEC 60601-1 Medical electrical equipment Part 1: General requirement for basic safety and essential performance
· Electromagnetic Compatibility:
IEC 60601-1-2: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
· Performance Effectiveness:
ASTM E 1965-98(Reapproved 2009), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature.
5.5 Conclusion
The subject device, JXB-182 Non-contact Infrared Thermometer has the same intended use and similar characteristics with the predicate device. Besides, the safety of the subject has been demonstrated by the following standards: IEC 60601-1(Electrical Safety), IEC 60601-1-2(Electromagnetic Compatibility). And since the subject is composed of the same material as the predicate device, they enjoy the same biocompatibility. Moreover, the subject device meets the specific requirements of ASTM E 1965, which further embodies that the subject device is both safe and effective.
So from the above information, it is reasonable to conclude that the subject device, JXB-182 Non-contact Infrared Thermometer, is substantially equivalent to the predicate device, RC002 Remote Infrared Thermometer.
4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 0903 New Hampshire Avenue cument Control Center - WO66-G60 Silver Spring, MD 20993-0002
October 15, 2013
Jinxinbao Electronic Company, Limited C/O Ms. Helen Nan, General Manager Wenzhous Cytech Information Service Company, Limited Room 404, Building 7, Jinhuichang Homeland, Liuhongqiao Road WENZHOUS CITY, 325000, Zhejiang Province CHINA
Re: K130231
Trade/Device Name: JXB182 Non-contact Infrared Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: II Product Code: FLL Dated: November 25, 2012 Received: August 7, 2013
Dear Ms. Nan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Nan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mary S. Brunner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters J, X, B, and D. The letters are in a bold, sans-serif font. The letters are arranged horizontally, with the J on the left and the D on the right. The letters are all the same size. The logo is black and white.
Section 4 Indications for Use Statement
510(k) Number (if known): K 13023 !
Device Name: JXB182 Non-contact Infrared Thermometer
Indications for Use:
JXB182 Non-contact Infrared Thermometer is an infrared thermometer intended to measure forehead temperature of infants and adults without contacting human body. It can be used by consumers in household environment and doctor in clinic as reference.
Prescription Use
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 80) Subpart C) ·
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
l
AND/OR
Image /page/6/Picture/13 description: The image shows the logo of the Food and Drug Administration (FDA). The logo is in black and white and appears to be a stylized version of the letters "FDA". The letters are blocky and have a slightly pixelated appearance. The logo is relatively small and centered in the image.
Richard C. Chapman 2013.10.15 10:16:38 -04'00'
Page 1 of 1
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.