K Number
K083070

Validate with FDA (Live)

Device Name
TOGGLELOC SYSTEM
Date Cleared
2008-12-16

(62 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ToggleLoc™ System Devices are intended for soft tissue to bone fixation for the following indications:

Shoulder: Bankart lesion repairs Acromio-clavicular repair, Capsular shift/capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps Tenodesis

Foot and Ankle: Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair, Ankle Syndesmosis fixation (Syndesmosis disruptions) and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures (only for ToggleLoc™ with Tophat)

Elbow: Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment

Knee: ACL/PCL repair / reconstruction, ACL/PCL patellar bone-tendon-bone grafts, Double-Tunnel ACL reconstruction, Extracapsular repair: MCL, LCL, and posterior oblique ligament, Illiotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure

Hand and Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx. Volar plate reconstruction

Hip: Acetabular labral repair

Device Description

The ToggleLoc™ System contains toggle buttons and a Tophat suture button preloaded with sutures and/or fiber constructs. The suture button has multiple eyelets for attachment of fiber constructs and/or sutures. The fiber constructs and/or sutures are preloaded for the convenience of the surgeon for soft tissue attachment.

AI/ML Overview

This 510(k) summary for the ToggleLoc™ System describes a medical device seeking clearance based on substantial equivalence to previously marketed devices. It does not contain information about a study proving device performance against specific acceptance criteria.

The submission states that:

  • Non-Clinical Testing: "Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use."
  • Clinical Testing: "None provided as a basis for substantial equivalence."

Therefore, I cannot populate the requested table and information as a study proving device meets acceptance criteria, an acceptance criteria table with reported performance, sample sizes, expert qualifications, or ground truth details were not provided in this document.

In summary:

  1. Table of acceptance criteria and reported device performance: Not provided. The submission states non-clinical testing was performed, but no specific acceptance criteria or performance metrics are detailed.
  2. Sample size used for the test set and data provenance: Not applicable as no clinical or specific non-clinical test set details are provided.
  3. Number of experts used to establish the ground truth for the test set and qualifications: Not applicable, as no ground truth establishment for a test set is described.
  4. Adjudication method: Not applicable.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: No such study was mentioned.
  6. Standalone (algorithm-only) performance: Not applicable, as this is a physical medical device, not an algorithm.
  7. Type of ground truth used: Not applicable, as no specific ground truth for performance evaluation is described.
  8. Sample size for the training set: Not applicable, as this is not an AI/ML device requiring a training set.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

K083070

Image /page/0/Picture/1 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and connected to each other. There is a registered trademark symbol to the right of the letter "T". The word is in black and the background is white.

510(k) Summary

DEC 1 6 2008

Preparation Date: December 4, 2008

Applicant/Sponsor: Biomet Sports Medicine

Contact Person: Robert Friddle

Proprietary Name: ToggleLoc™ System

Common Name: Soft tissue anchor

Classification Name: Fastener, fixation, nondegradable, soft tissue (888.3040)

Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

TightRope™ Syndesmosis Device - K043248 (Arthrex) Sleeve and Button Soft Tissue Devices - K071704 (Biomet Sports Medicine) Titanium Toggle Buttons - K033838 (Biomet Sports Medicine, formerly known as Arthrotek, Inc.)

Device Description: The ToggleLoc™ System contains toggle buttons and a Tophat suture button preloaded with sutures and/or fiber constructs. The suture button has multiple eyelets for attachment of fiber constructs and/or sutures. The fiber constructs and/or sutures are preloaded for the convenience of the surgeon for soft tissue attachment.

Intended Use: The ToggleLoc™ System Devices are intended for soft tissue to bone fixation for the following indications:

Shoulder: Bankart lesion repairs Acromio-clavicular repair, Capsular shift/capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps Tenodesis

Foot and Ankle: Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair, Ankle Syndesmosis fixation (Syndesmosis disruptions) and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures (only for ToggleLoc™ with Tophat)

Elbow: Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment

Knee: ACL/PCL repair / reconstruction, ACL/PCL patellar bone-tendon-bone grafts, Double-Tunnel ACL reconstruction, Extracapsular repair: MCL, LCL, and posterior oblique ligament, Illiotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure

Hand and Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx. Volar plate reconstruction

Hip: Acetabular labral repair

Mailing Address: P.O. Box 587 Warsaw, IN 46581-0587 Toll Free: 800,348,9500 Office: 574.267.6539 Main Fax: 574.372.1718 www.blomet.com

Shipping Address: 56 East Bell Drive Warsaw. IN 46582

{1}------------------------------------------------

Summary of Technologies: The technological characteristics (materials, design, sizing and indications) of the ToggleLoc™ System devices are similar or identical to the predicate device or other previously cleared devices.

Non-Clinical Testing: Non-dinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use.

Clinical Testing: None provided as a basis for substantial equivalence.

All trademarks are property of Biomet, Inc. except TightRope™ which is a trademark of Arthrex.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 6 2008

Biomet Sports Medicine % Mr. Robert Friddle 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581-0587

Re: K083070

Trade/Device Name: ToggleLocTM System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI, JDR Dated: October 13, 2008 Received: October 15, 2008

Dear Mr. Friddle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Robert Friddle

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events, (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark N. Milherson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): KD83070

Device Name: _ ToggleLoc™ System

Indications for Use:

The ToggleLoc™ System Devices are intended for soft tissue to bone fixation for the following indications:

Shoulder: Bankart lesion repair, SLAP lesion repairs Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps Tenodesis

Foot and Ankle: Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair, Ankle Syndesmosis fixation (Syndesmosis disruptions) and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures (only for ToggleLoc™ with Tophat)

Elbow; Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment

Knee: ACL/PCLrepair / reconstruction, ACL/PCL patellar bone-tendon-bone grafts, Double-Tunnel ACL reconstruction, Extracapsular repair: MCL, LCL, and posterior oblique ligament, Illiotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule dosure

Hand and Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction

Hip: Acetabular labral repair

Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page

Division of General, Restorative, and Neurological Devices

4 - Page 1 of 1
510(k) Number L083070

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.