The Valved Tearaway Introducers are intended to obtain central venous access to facilitate catheter insertion into the central venous system. The Valved Tearaway Introducer Generation II is designed to reduce blood loss and the risk of air intake but it is not a hemostasis valve. It is not intended to create a complete two way seal nor is it intended for arterial use.

Device Description

The Medcomp Valved Tearaway Introducer is a two-part single use device used to obtain vascular access and facilitate intravascular catheter insertion. The Medcomp Valved Tearaway Introducer consists of a peel-able introducer sheath which contains a "duckbill" seal to substantially minimize air/blood from entering or escaping when the dilator is removed. The dilator is comprised of a cylindrical tube with a locking hub; the sheath is also a cylindrical tube with a hub and valve. When the dilator is fully seated it snaps into the hub of the introducer sheath to prevent independent movement of either piece when in use. The dilator extends beyond the sheath to provide a zero tolerance clearance between sheath and dilator. The sheath and dilator when used in conjunction with an introducer needle and quidewire provide a means to obtain a percutaneous opening to the vascular system to facilitate the insertion of a catheter. After removing the dilator a catheter can then be placed through the sheath. Breaking the sheath's hub and peeling the sheath away from the catheter then allows the sheath to be removed. The dilator is composed of Nylon-based Triax with Barium Sulfate for visibility under fluoroscopy by the attending physician during insertion. The sheath is composed of PTFE to provide a smooth, consistent peel.

AI/ML Overview

The provided text describes a medical device submission (K124046) for a "Valved Tearaway Introducer Generation II." This submission is a Traditional 510(k), which is a premarket submission made to the FDA to demonstrate that the device to be marketed is at least as safe and effective, that is, substantially equivalent, to a legally marketed predicate device.

The acceptance criteria and the study proving the device meets them are based on bench testing and biocompatibility testing, and a comparison to a predicate device. This is typical for a 510(k) submission for this type of device, which is not an AI/ML powered device. Therefore, many of the requested categories (like sample size for test sets, data provenance, number/qualification of experts for ground truth, adjudication method, MRMC studies, standalone algorithm performance, or training set details) are not applicable to this type of medical device submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that "the proposed device exceeds all the testing criteria the predicate device was tested against" and "has comparable or improved results for many of the performance tests." However, explicit numerical acceptance criteria and specific performance results are not provided in the summary. The summary lists the types of tests conducted.

Test NameAcceptance Criteria (Not explicitly stated in summary)Reported Device Performance (Summary statement)
Air LeakageNot explicitly stated"exceeds all the testing criteria the predicate device was tested against" and "comparable or improved results when compared to the predicate device." (Results available in Section 18 of full submission)
Dilator Insertion/ExtractionNot explicitly stated"exceeds all the testing criteria the predicate device was tested against" and "comparable or improved results when compared to the predicate device." (Results available in Section 18 of full submission)
Force at BreakNot explicitly stated"exceeds all the testing criteria the predicate device was tested against" and "comparable or improved results when compared to the predicate device." (Results available in Section 18 of full submission)
Force at Break: Dilator HubNot explicitly stated"exceeds all the testing criteria the predicate device was tested against" and "comparable or improved results when compared to the predicate device." (Results available in Section 18 of full submission)
Liquid LeakageNot explicitly stated"exceeds all the testing criteria the predicate device was tested against" and "comparable or improved results when compared to the predicate device." (Results available in Section 18 of full submission)
BiocompatibilityISO 10993 requirements"materials used meet the requirements of ISO 10993."

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified in the summary. For bench testing of physical devices, the "sample size" would refer to the number of devices or components tested, which is not detailed here.
  • Data Provenance: The studies are described as "Bench Testing" and "Biocompatibility Testing," implying laboratory-controlled experiments. There is no indication of human subject data, retrospective, or prospective studies involving patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This is a physical device, and the "ground truth" for its performance is established through standardized engineering and materials testing protocols, not expert clinical consensus on images or patient data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not Applicable. As per point 3, there's no expert adjudication involved in bench testing of physical device characteristics.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a physical medical instrument, not an AI/ML-powered diagnostic or assistive tool. Therefore, MRMC studies are not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This device is a physical medical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Standardized Bench Test Measurements / ISO 10993 Compliance: For the performance tests (Air Leakage, Dilator Insertion/Extraction, Force at Break, Liquid Leakage), the "ground truth" is defined by the objective measurements obtained from the specified test protocols and compared against the performance of the predicate device (and implicit or explicit engineering specifications). For biocompatibility, the ground truth refers to adherence to the requirements of the ISO 10993 standard.

8. The sample size for the training set:

  • Not Applicable. This is not an AI/ML device; there is no "training set."

9. How the ground truth for the training set was established:

  • Not Applicable. As per point 8, there is no training set.

{0}------------------------------------------------

K124046 1 of 3

JUN 0 3 2013

Section 5

510(k) SUMMARY

Traditional 510K

Submitter Information:

Submitter:MEDCOMP®1499 Delp DriveHarleysville, PA 19438(215) 256-4201 Telephone(215) 256-9191 Fax
Contact:Timothy Holwick, Regulatory Associate
Date Prepared:December 28, 2012
Device Name:Valved Tearaway Introducer Generation II
Common Name:Introducer, Catheter
Classification Name:Blood Access device and accessories (a)(2)(b)(2)
C.F.R. Section:870.1340
Classification Panel:Cardiovascular
Class:II, DYB

Predicate Devices:

Primary:

K090394, Valved Tearaway Introducer Generation I, concurrence date September 1, 2009. Class II CFR 870.1340

Device Description:

The Medcomp Valved Tearaway Introducer is a two-part single use device used to obtain vascular access and facilitate intravascular catheter insertion. The Medcomp Valved Tearaway Introducer consists of a peel-able introducer sheath which contains a "duckbill" seal to substantially minimize air/blood from entering or escaping when the dilator is removed. The dilator is comprised of a cylindrical tube with a locking hub; the sheath is also a cylindrical tube with a hub and valve. When the dilator is fully seated it snaps into the hub of the introducer sheath to prevent independent movement of either piece when in use. The dilator extends beyond the sheath to provide a zero tolerance clearance between sheath and dilator. The sheath and dilator when used in conjunction with an introducer needle and quidewire provide a means to obtain a percutaneous opening to the vascular system to facilitate the insertion of a catheter. After removing the dilator a catheter can then be placed through the sheath. Breaking the sheath's hub and peeling the sheath away from the catheter then allows the sheath to be removed. The dilator is composed of Nylon-based Triax with Barium Sulfate for visibility under fluoroscopy by the attending physician during insertion. The sheath is composed of PTFE to provide a smooth, consistent peel.

Intended Used:

The Valved Tearaway Introducers are intended to obtain central venous access to facilitate catheter insertion into the central venous system.

Medcomp Valved Tearaway Introducer Generation II

{1}------------------------------------------------

Indications for Use:

The Valved Tearaway Introducers are intended to obtain central venous access to facilitate catheter insertion into the central venous system. The Valved Tearaway Introducer Generation II is designed to reduce blood loss and the risk of air intake but it is not a hemostasis valve. It is not intended to create a complete two way seal nor is it intended for arterial use.

Comparison to Predicate Devices:

The Valved Tearaway Introducer Generation II is substantially equivalent to the predicate devices in terms of intended use, materials, anatomical location, basic design, performance, labeling, manufacturing process and method of sterilization.

Performance Standards:

Performance standards have not been established by the FDA under section 514 of the Federal Food, Drug and Cosmetic Act.

Summary of Non-Clinical Performance Testing:

The following Performance - Bench Testing results and protocols are provided in this submission:

  • Air Leakage .
  • Dilator Insertion/Extraction .
  • . Force at Break
  • . Force at Break: Dilator Hub
  • Liquid Leakage �

All results can be found in Section 18. As illustrated with side-by-side testing results against the predicate device in the Section 12 Substantial Equivalence Matrix, the performance testing demonstrates that the proposed device exceeds all the testing criteria the predicate device was tested against. In addition, the matrix shows that the proposed device has comparable or improved results for many of the performance tests when compared to the predicate device.

Biocompatibility:

Results for all biocompatibility testing demonstrate the materials used meet the requirements of ISO 10993.

In Section 15 of this submission, a biocompatibility summary and test reports are provided. The following materials are identical to the predicate device (K090394) and therefore the biocompatibility testing for those components can be found in that submission:

  • Sheath: Gray Teflon Tubing w/ 9% BiO3 ●
  • Valve: Silicone, C6-570 .
  • Dilator Hub: HDPE .
  • . Dilator Tube: Gray HDPE - Soltex K44-24-122 w/ BaSO4 and UV Stabilizer

{2}------------------------------------------------

K124046

3 of 3

The following remaining components were tested and the reports have been included in Section 15:

• Valved Peelable Snap Hub:Nylon/ABS (Triax)
• Snap Cap:Nylon/ABS (Triax)
  • . Colorants: Gray, Green, Orange and Lt. Blue
    The testing provided, when read with the testing for relevant components of K090394, demonstrate that the materials used meet ISO 10993 requirements for limited contact external communicating device with indirect blood contact.

Technological Characteristics:

The principles of operation are the same as the predicate devices. There are no new questions raised regarding the safety or effectiveness of the device.

Summary of Substantial Equivalence:

The proposed device meets the performance criteria of design verification as specified by ISO standards, guidance documents and test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed device is substantially equivalent to the legally marketed predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or flowing lines, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 3. 2013

Medcomp c/o Timothy Holwick Regulatory Associate 1499 Delp Drive Harleysville, PA 19438

Re: K124046

Trade/Device Name: Valved Tearaway Introducer Generation II Regulation Number: 21 CFR 870.1340 Regulation Name: Introducer, Catheter Regulatory Class: Class II Product Code: DYB Dated: May 2, 2013 Received: May 3. 2013

Dear Mr. Holwick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Timothy Holwick

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely vours.

Matthew G퀘咖ebrenner

for Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use
K124046510(k) Number (if known):
medcomp.Device Name: Valved Tearaway Introducer Generation II
1499 Delp DriveIndications for Use:
Harleysville, PA 19438Tel: 215-256-4201The Valved Tearaway Introducers are intended to obtain central venous access tofacilitate catheter insertion into the central venous system.
Fax: 215-256-1787
www.medcompnet.com
Prescription Use XOver-The-Counter UseAND/OR(21 CFR 801 Subpart C)(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGEOF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Matthew Grillillebrenner
Page 1 of 1
4-1

Comments of the country

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).