K Number
K123485

Validate with FDA (Live)

Date Cleared
2013-03-18

(125 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The JuqqerKnotless™ Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows:

Shoulder
Acromio-clavicular Separation, Anterior Shoulder Instability Repair, Bankart lesion repair, Biceps tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs

Foot and Ankle
Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs

Elbow
Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral/Medial Repairs, Lateral Epicondy/itis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction

Knee
Iliotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Liqament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment/Repair, Posterior Oblique Ligament Repair, VMO Advancement

Hand and Wrist
Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Volar Plate Reconstruction

Hip
Labral

Device Description

The JuggerKnotless™ Soft Anchors consist of a coreless sleeve structure and a knotless construct incorporating ZipLoop™ Technology. The anchors are intended for use in soft tissue fixation by bunching against bone when deployed.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the JuggerKnotless™ Soft Anchors:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Not explicitly stated as such, but inferred from testing)Reported Device Performance
Equivalent fixation strength to predicate devices"The test results indicate that the Biomet Sports Medicine JuggerKnotless™ Soft Anchors provide equivalent fixation strength to the predicate devices and would be functional within their intended use."

Explanation of "Acceptance Criteria" for this device:
The document does not explicitly list numerical "acceptance criteria" in the way one might see for diagnostic software (e.g., target specificity/sensitivity). Instead, for this type of medical implant, the primary "acceptance criterion" for non-clinical testing is typically demonstrated equivalence in performance (specifically, mechanical fixation strength) to an already legally marketed and accepted predicate device. This is a common approach for 510(k) submissions where substantial equivalence is claimed.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The document mentions "mechanical pullout testing," but the number of anchors or tests performed is not specified.
  • Data Provenance: The testing was "non-clinical laboratory testing," implying it was conducted in a controlled lab setting, likely in the USA where Biomet Sports Medicine is located. It is by definition retrospective in the context of device development, as it assesses the device's properties before market release.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This question is not applicable to the provided document. The "ground truth" for this type of device (medical implant fixation) is established through objective mechanical testing (fixation strength measurements) against a control, not through expert consensus or interpretation of images.

4. Adjudication Method for the Test Set

This question is not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in studies involving human interpretation (e.g., radiologists reviewing images), not for objective mechanical testing of device performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • Was it done? No. The document explicitly states: "Clinical Testing: None provided as a basis for substantial equivalence." MRMC studies typically involve human readers interpreting cases, often to compare AI-assisted performance with unassisted performance. This was a non-clinical, mechanical performance study.

  • Effect size of human readers improving with AI vs without AI assistance: Not applicable, as no MRMC study or clinical study involving human-AI interaction was performed or reported.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Was it done? Yes, in a sense. The mechanical testing performed on the JuggerKnotless™ Soft Anchors is a form of "standalone" evaluation of the device's physical performance, independent of human interaction during its assessment. The device itself is an implant, not an algorithm. The testing evaluated the device's intrinsic mechanical property (fixation strength) without human-in-the-loop performance assessment of that specific property.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth used was objective mechanical measurements of "fixation strength." This is determined through physical testing protocols that measure parameters like pullout force, displacement, or ultimate load to failure, compared against a predicate device.

8. The Sample Size for the Training Set

This question is not applicable. This device is a physical medical implant, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The device's design and manufacturing are based on engineering principles and material science, not data training.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as there is no training set for this device.

{0}------------------------------------------------

K123485 (1/2)

Image /page/0/Picture/1 description: The image shows the logo for Biomet Sports Medicine. The word "BIOMET" is in a bold, outlined font on the top line. Below that, the words "SPORTS MEDICINE" are in a smaller, regular font.

510(k) Summary

MAR 1 8 2013

October 31, 2012 Preparation Date:

Applicant/Sponsor: Biomet Sports Medicine

Contact Person: Elizabeth Wray / Global RA Project Manager (574)-267-6673

JuggerKnotless™ Soft Anchors Proprietary Name:

Common Name: Soft Tissue Fixation Device

Classification Name: Fastener, fixation, nondegradable, soft tissue (21CFR &88.3040) MBI Staple, fixation, bone (21CFR §888.3030) JDR

Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

K080088 Biomet Sports Medicine Anchor Devices and ZipLoop™ Constructs K110145 JuggerKnot™ Soft Anchors

Device Description:

The JuggerKnotless™ Soft Anchors consist of a coreless sleeve structure and a knotless construct incorporating ZipLoop™ Technology. The anchors are intended for use in soft tissue fixation by bunching against bone when deployed.

Intended Use / Indications for Use:

The JuggerKnotless™ Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows:

Shoulder

Acromio-clavicular Separation, Anterior Shoulder Instability Repair, Bankart lesion repair, Biceps tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs

Foot and Ankle

Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs

Elbow

Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral Medial Repairs, Lateral Epicondylitis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction

Mailing Address

Shipping Address:
56 East Bell Drive
Warsaw, IN 46582

{1}------------------------------------------------

Knee

Iliotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Ligament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment/Repair, Posterior Oblique Ligament Repair, VMO Advancement

Hand and Wrist

Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Volar Plate Reconstruction

  • Hip
    Labral

Summary of Technologies:

The technological characteristics (materials, design, sizing and indications) of the JuggerKnotless™ Soft Anchors are similar or identical to the predicate devices or other previously cleared devices.

Non-Clinical Testing:

Non-clinical laboratory testing was performed to verify the fixation strength of the JuqgerKnotless™ Soft Anchors in mechanical pullout testing as compared to the predicate devices for specific indications for use. The efficacy of the JuggerKnotless™ Soft Anchors was compared to that of the Biomet Sports Medicine 2.4mm PEEK Hitch Anchor. The test results indicate that the Biomet Sports Medicine JuggerKnotless™ Soft Anchors provide equivalent fixation strength to the predicate devices and would be functional within their intended use.

Clinical Testing:

None provided as a basis for substantial equivalence.

All trademarks are the property of Biomet, Inc.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 18, 2013

Biomet Sports Medicine % Ms. Elizabeth Wray Senior Regulatory Affairs Specialist P.O. Box 587 Warsaw, Indiana 46581

Re: K123485

Trade/Device Name: JuggerKnotless™ Soft Anchors Regulation Number: 21 CFR 888.3040, 21 CFR 888.3030 Regulation Name: Smooth or threaded metallic bone fixation fastener, Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MBI, JDR Dated: February 25, 2013 Received: February 26, 2013

Dear Ms. Wray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Elizabeth Wray

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Erin DKeith

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): __K123485

Device Name: JuggerKnotless™ Soft Anchors

Indications For Use:

The JuqqerKnotless™ Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows:

Shoulder

Acromio-clavicular Separation, Anterior Shoulder Instability Repair, Bankart lesion repair, Biceps tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P Lesion Repairs

Foot and Ankle

Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Lateral/Medial Stabilization Repairs/Reconstruction, Midfoot/Forefoot Reconstruction/Repairs

Elbow

Biceps Tendon Reattachment, Biceps Tendon Reconstruction, Lateral/Medial Repairs, Lateral Epicondy/itis Repair (Tennis Elbow Repair), Ulnar & Radial Collateral Ligament Reconstruction

Knee

Iliotibial Band Tenodesis, Joint Capsule Closure, Lateral/Medial Collateral Liqament Repair, Patellar Ligament Repair, Patellar Tendon Repair, Patellar Realignment/Repair, Posterior Oblique Ligament Repair, VMO Advancement

Hand and Wrist

Collateral Ligament Repair (Gamekeeper's Thumb), Scapholunate Ligament Reconstruction, Tendon Transfers in Phalanx, Volar Plate Reconstruction

Hip

Labral

Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley Ph.D.
Division of Orthopaedic Devices

Page 1 of 1

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.