K Number
K123019

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-10-22

(24 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The system is indicated for primary or revision surgery in skeletally mature individuals for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

The system is intended for use either with or without bone cement in total hip arthroplasty.

Device Description

The Longevity IT Highly Crosslinked Polyethylene Liners are intended to be used with either Continuum or Trilogy IT Acetabular components in Total Hip Arthroplasty. The liners are available with neutral, elevated, offset, and oblique liner faces in 28, 32, 36, and 40mm articulation diameters.

AI/ML Overview

The provided document describes the modification of Longevity IT Highly Crosslinked Polyethylene Liners and relies on non-clinical performance data for substantial equivalence. Therefore, there is no acceptance criteria or study that proves device meets acceptance criteria in the context of clinical performance as this was not required.

However, the document does detail the non-clinical performance assessment:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
Design and GeometryMet performance requirements and demonstrated substantial equivalence to predicates.
Wear TestingPerformed to demonstrate substantial equivalence to predicates.
Range of Motion AnalysisPerformed to demonstrate substantial equivalence to predicates.
Materials and ProcessesEquivalent to predicates.

2. Sample Size Used for the Test Set and Data Provenance

This information is not applicable as no clinical test set was used. The assessment was based on non-clinical studies. The provenance for the non-clinical testing is implied to be Zimmer, Inc.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not applicable as no clinical test set or human expert review was performed.

4. Adjudication Method for the Test Set

This information is not applicable as no clinical test set was used.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A MRMC comparative effectiveness study was not conducted as no clinical study involving human readers or patient cases was performed.

6. Standalone Performance Study

No standalone clinical performance study was conducted. The assessment was based solely on non-clinical testing and comparison to predicate devices.

7. Type of Ground Truth Used

The ground truth for the non-clinical assessment was based on:

  • Engineering analysis
  • Risk analysis
  • Results from wear testing
  • Results from range of motion analysis
  • Design and geometry evaluation against established performance requirements for hip prosthesis liners.

8. Sample Size for the Training Set

This information is not applicable as there was no training set in the context of a machine learning or AI algorithm. The "training" for this device's assessment was based on the performance history and characteristics of the predicate devices.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable. The "ground truth" for showing equivalence was based on the previously cleared predicate devices' performance and characteristics, which themselves would have been established through a combination of non-clinical and, where applicable, clinical data. For the current submission, the focus was on demonstrating that the modified device's non-clinical performance matched that of the predicates.

{0}------------------------------------------------

K123019(1/3)

Image /page/0/Picture/1 description: The image shows a logo with the letter Z inside of a circle. The letter Z is stylized with sharp angles and lines. Below the circle is the word "zimmer" in a sans-serif font. The logo appears to be for a company or brand named Zimmer.

P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131

OCT 22 2012

رار Summary of Safety and Effectiveness

Sponsor:

Contact Person:

Date:

Trade Name:

Common Name:

Classification Name and Reference:

Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708

Rebecca M. Brooks Sr. Specialist, Regulatory Affairs Telephone: (574) 371-8033 Fax: (574) 372-4605

September 27, 2012

Longevity IT Highly Crosslinked Polyethylene Neutral Liners

Longevity IT Highly Crosslinked Polyethylene Elevated Liners

Longevity IT Highly Crosslinked Polyethylene Offset Liners

Longevity IT Highly Crosslinked Polyethylene Oblique Liners

Total Hip Prosthesis

LPH - Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented 21 CFR § 888.3358

JDI - Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented 21 CFR § 888.3350

LZO - Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented 21 CFR § 888.3353

ਹੋਰੇ

{1}------------------------------------------------

Page 2 September 27, 2012

Predicate Device:

Device Description:

Intended Use:

Comparison to Predicate Device:

Continuum and Trilogy IT Acetabular Systems, manufactured by Zimmer, K091508, cleared September 11, 2009

Longevity IT Highly Crosslinked Polyethylene Elevated Liners, manufactured by Zimmer, K093846, cleared February 4, 2010

Continuum and Trilogy IT Acetabular Systems & Longevity IT Highly Crosslinked Polyethylene Elevated Liners, manufactured by Zimmer, K101229, cleared December 3, 2010

Continuum and Trilogy IT Acetabular Systems & Longevity IT Highly Crosslinked Polyethylene Elevated, Offset, & Oblique Liners, manufactured by Zimmer, K103662, cleared April 15, 2011

The Longevity IT Highly Crosslinked Polyethylene Liners are intended to be used with either Continuum or Trilogy IT Acetabular components in Total Hip Arthroplasty. The liners are available with neutral, elevated, offset, and oblique liner faces in 28, 32, 36, and 40mm articulation diameters.

The system is indicated for primary or revision surgery in skeletally mature individuals for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip. fracture of the pelvis, and diastrophic variant.

The system is intended for use either with or without bone cement in total hip arthroplasty.

No changes are being made to the designs of the subject Longevity IT Liners. The proposed modification is limited to expanding the scope of compatible femoral heads. The Longevity IT Highly Crosslinked Polyethylene Neutral, Elevated,

30

{2}------------------------------------------------

Page 3 September 27, 2012

Performance Data (Nonclinical and/or Clinical):

Offset, and Oblique Liners are manufactured, packaged, and sterilized using equivalent materials and processes as their predicates. The subject devices also have the same intended use and performance characteristics as their predicates.

Non-Clinical Performance and Conclusions:

Non-clinical testing as well as engineering and risk analyses were performed to demonstrate substantial equivalence of the subject liners to the predicate devices. The specific non-clinical testing and analyses completed include wear testing and range of motion analyses.

The liners were also evaluated in terms of design and geometry to demonstrate that the devices meet performance requirements and are substantially equivalent to their predicates. This information and testing results formed the basis for a determination of substantial equivalence.

Clinical Performance and Conclusions:

Clinical data and conclusions were not needed for this device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes forming its body and wing. The eagle faces right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Zimmer, Incorporated % Ms. Rebecca M. Brooks Senior Specialist, Regulatory Affairs 1800 West Center Street Warsaw, Indiana 46580

OCT 22 2012

Re: K123019

Trade/Device Name: Longevity® IT Highly Crosslinked Polyethylene neutral, Offset, and Oblique Liners

Regulation Number: 21 CFR 888.3358

Regulation Name: Hip joint metal/ceramic/polymer semi-constrained porous-coat-ed uncemented prosthesis

Regulatory Class: Class II Product Code: LPH, JDI, LZO Dated: September 27, 2012 Received: September 28, 2012

Dear Ms. Brooks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 19 we any nor to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Ms. Rebecca M. Brooks

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Mellecker

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Longevity IT Highly Crosslinked Polyethylene Neutral, Elevated, Offset, and Oblique Liners

Indications for Use:

The system is indicated for primary or revision surgery in skeletally mature individuals for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

The system is intended for use either with or without bone cement in total hip arthroplasty.

Asb

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

8123019 510(k) Number

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.