K Number
K122520

Validate with FDA (Live)

Date Cleared
2013-02-05

(172 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TD-1001 Digital Thermometer is intended to measure the body temperature orally and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.

Device Description

The Digital Thermometer TD-1001 enables easy and accurate readings over the body temperature range. It must be used in conjunction with a disposable probe cover when taking temperature.

From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Digital Thermometer, model TD-1001, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemStandard CompliedAcceptance Criteria (Implied by standard)Reported Device Performance
Display temperature rangeASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Laboratory accuracyASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Operating rangeASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Storage environment testASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Shock testASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Water resistant testASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Cleaning procedure testASTM E1112-00:2011Met requirements of standardMeasurement error (details not provided in summary)
Software validationISO 14971:2007Met requirements of standardVerification and validation of TD-1001 software (details not provided)
Clinical accuracyprEN12470-3:2000Met requirements of standardClinical bias, Clinical repeatability (details not provided)
SafetyIEC 60601-1Met requirements of standardEvaluated by SGS (details not provided)
Electromagnetic compatibility (EMC)IEC 60601-1-2Met requirements of standardEvaluated by SGS (details not provided)

Note: The provided summary states that the device "met the requirements" of the specified standards. It does not explicitly list numerical acceptance criteria or the precise numerical performance results for each test within this summary document. For a full understanding of the acceptance criteria and detailed performance, one would need to refer to the full study reports and the standards themselves (ASTM E1112-00:2011, prEN12470-3:2000, ISO 14971:2007, IEC 60601-1, IEC 60601-1-2).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified in the provided summary text.
  • Data Provenance: Not specified in the provided summary text. The manufacturer is based in Taiwan, but the location of the testing is not mentioned. It is implicitly a prospective study, as it's a test of a new device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the summary. For a digital thermometer, "ground truth" for clinical accuracy would typically involve comparison against a highly accurate reference thermometer, not necessarily expert consensus.

4. Adjudication Method for the Test Set

  • Adjudication methods (like 2+1, 3+1) are typically used for subjective assessments where human interpretation is involved (e.g., medical imaging). For a digital thermometer, the tests mentioned are objective performance comparisons against established measurement standards. Therefore, an adjudication method in this sense is not applicable and not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation and aims to measure the improvement in human reader performance with AI assistance. A digital thermometer is a direct measurement device and does not involve human interpretation in the same way.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study

  • Yes, a standalone performance study was implicitly done. The listed tests (e.g., laboratory accuracy, clinical accuracy) directly assess the performance of the device itself (the "algorithm only," in a broad sense, meaning the device's measurement capability) against established standards, without requiring a human to interpret the device's output beyond reading the display.

7. Type of Ground Truth Used

  • The ground truth for most performance tests (e.g., laboratory accuracy, display temperature range, operating range) would be measurement against a calibrated reference thermometer as defined by the standards (ASTM E1112-00:2011, prEN12470-3:2000). For clinical accuracy, this would involve comparing the device's readings with a reference method in human subjects.

8. Sample Size for the Training Set

  • Not applicable. The TD-1001 is a traditional electronic thermometer, not an AI/machine learning device that requires a "training set" in the conventional sense. Its "training" is in its design and calibration processes, not through data-driven learning.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. See point 8.

{0}------------------------------------------------

A11. 510(k) Summary

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

K12252 The Assigned 510(k) number is:

    1. Submitter's Identification: TaiDoc Technology Corporation 3F, 5F, No.127, Wugong 2nd Rd., Wugu District, New Taipei City, 24888, Taiwan
      Correspondence: Pinjung Chen Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1176 Fax: +886-2-6625-0288 Email: pinjung.chen@taidoc.com.tw

Date of submission: July 31th, 2012

    1. Device name:
      Proprietary name: Digital Thermometer, model TD-1001 Regulatory information:
  • A. Regulation section: 21 CFR 880.2910

  • Classification: Class II B.

  • C. Product Code: FLL, Clinical electronic thermometer

  • D. Panel: General Hospital (80)

    1. Intended Use:

TD-1001 Digital Thermometer is intended to measure the body temperature orally and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.

{1}------------------------------------------------

4. Device Description:

The Digital Thermometer TD-1001 enables easy and accurate readings over the body temperature range. It must be used in conjunction with a disposable probe cover when taking temperature.

From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts.

Substantial Equivalence Information:

  • A. Predicate device name:
    Omron electronic thermometer Models MC-246

  • Predicate K number: K091676 B.
    Comparison with predicate: C.

The Digital Thermometer, model TD-1001, has the following similarities to the predicate device:

  • Same intended use. D
  • Same operating principle and technological characteristics. I
  • Same measurement area (oral) D
  • Same functions and accuracy specification. D
  • 同 Same battery type.

The differences:

  • Temperature displayed range 国
  • D Operating temperature range
  • Storage temperature range
  • 门 Battery life
    1. Test Principle:

The Digital thermometer TD-1001 is the electronic thermometer operated by a thermistor as the temperature sensor and an ASIC (Application Specified IC) for

{2}------------------------------------------------

signal processing. The basic operation principle is that a change of thermistor, caused by changes of temperature, provide signal to ASIC. ASIC gets the sensor's signal then processes the signal and calculates the result, after that displays the temperature result by a LCD.

    1. Performance Characteristics:
      The Digital thermometer TD-1001 was validated by the tests according to ASTM E1112-00:2011 standard and met the requirements of prEN12470-3:2000 standard.

A brief description for each test was given in this section. Table 1 lists items of tests, related standard complied and format of data presentation.

ItemStandard compliedData presentationAttachment
Displaytemperature rangeASTM E1112-00:2011Measurement error4.1.1
LaboratoryaccuracyASTM E1112-00:2011Measurement error4.1.2
Operating rangeASTM E1112-00:2011Measurement error4.1.3
Storageenvironment testASTM E1112-00:2011Measurement error4.1.4
Shock testASTM E1112-00:2011Measurement error4.1.5
Water resistant testASTM E1112-00:2011Measurement error4.1.6
Cleaning proceduretestASTM E1112-00:2011Measurement error4.1.7
Software validationISO 14971:2007Verification andvalidation of TD-1001software4.2
Clinical accuracyprEN12470-3:2000Clinical biasClinical repeatability4.3
SafetyIEC 60601-1Evaluated by SGS4.4
Electromagneticcompatibility(EMC)IEC 60601-1-2Evaluated by SGS4.5

Table 1. Summary of test

{3}------------------------------------------------

7. Conclusion:

Based on the information provided in this submission, TD-1001 Digital thermometer is shown to be substantially equivalent to the predicate Omron electronic thermometer Models MC-246.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a stylized representation of human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 5, 2013

Ms. Pinjung Chen Regulatory Affairs Specialist Taidoc Technology Corporation 6F, No. 127, Wugong 2nd Road Wugu District New Taipei City, Taiwan 24888

Re: K122520

Trade/Device Name: Digital Thermometer, model TD-1001 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: January 15, 2013 Received: January 22, 2013

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Chen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

hor

Keith
Harshfield.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

A0. Indications for Use

Indications for Use

510(k) Number: K/22520

Device Name: Digital Thermometer, model TD-1001

Indications for Use:

.

TD-1001 Digital Thermometer is intended to measure the body temperature orally and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.

Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use __ X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off Office of Device Evaluation (ODE) 510(k)_k122520

Page 1 of

Sajjad H. Syed
Digitally signed by Sajjad H. Syed
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Sajjad H. Syed
0.9.2342.19200300.100.1.1=2000601742
Date: 2013.02.05 11:08:03 -05'00'

n Sign-Off)
of Anesthestology, General Hospital
of Anesthestology, General Devices on of Anestholorological

510(k) Number: K122520

A0- Page 1 of 1

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.