K Number
K122503

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-06-25

(313 days)

Product Code
Regulation Number
888.3027
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeuroTherm Parallax Balloon Inflatable Bone Tamp is intended to be used as a conventional bone tamp for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The Parallax Balloon Inflatable Bone Tamp is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.

Device Description

The NeuroTherm Parallax Balloon Inflatable Bone Tamp with an inflatable component at the distal end. The tamp is designed to compress cancellous bone and/or move cortical bone as it inflates. The catheter shaft contains an outer lumen for the tamp inflation and a central lumen for a stylet to facilitate catheter introduction. At full inflation volume, the balloon design allows for a uniform balloon that is concentric to the catheter shaft. The wire mandrel provides stiffness to the balloon catheter to facilitate insertion through the access cannula. The NeuroTherm Parallax Balloon Inflatable Bone Tamp may be used with FDA cleared balloon inflation syringes.

AI/ML Overview

The NeuroTherm Parallax Balloon Inflatable Bone Tamp's acceptance criteria and performance study details are as follows:

1. Table of Acceptance Criteria and Reported Device Performance

The device's acceptance criteria are primarily demonstrated through substantial equivalence to predicate devices, supported by bench testing and compliance with relevant standards. No specific quantitative performance metrics or thresholds are explicitly stated as acceptance criteria beyond those inherent in establishing substantial equivalence.

Acceptance Criterion (Implicit)Reported Device Performance and Evidence
Technology is similar or same as predicate devices.The device description details a double-lumen catheter shaft, side arm adapter, continuous straight arm port, and radiopaque markers, which are comparable to the characteristics of the predicate KyphX Xpander and Inflatable Bone Tamps.
Intended use/Indication for Use in the spine is similar or same as predicate devices.The intended use for reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine, including percutaneous vertebral augmentation with cleared PMMA bone cement, directly aligns with the indications of the predicate KyphX Xpander and Inflatable Bone Tamps for similar spinal applications.
Technical specifications or range of technical specifications are similar or same as predicate devices.While specific ranges are not detailed, the comparison table shows similarities in general design and features. For instance, the presence of a balloon length indicator (printed band vs. color-coded band) and the use of an inflatable bone tamp design align with predicates.
No new issues of safety or effectiveness are raised by differences.The 510(k) summary explicitly states: "Any differences between the NeuroTherm Parallax Balloon Inflatable Bone Tamp and the predicate devices do not raise new issues of safety or effectiveness." This implies that any variations were assessed and deemed acceptable.
Device performs as expected."Bench - Bench testing supports that the NeuroTherm Parallax Balloon Inflatable Bone Tamp performs as expected." (No specific quantitative results are provided in the summary.)
Compliance with relevant consensus standards.The device complies with a list of ASTM, ANSI/AAMI/ISO, and ISO standards related to implants, sterilization, biological evaluation, and luer fittings, ensuring fundamental safety and performance aspects are met.

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not specify a test set involving human data or a specific number of samples for the bench testing beyond the general statement "Bench testing supports that the NeuroTherm Parallax Balloon Inflatable Bone Tamp performs as expected." There is no indication of country of origin or whether any data was retrospective or prospective in detail. It appears the primary "test set" for substantial equivalence was the comparison to existing predicate devices and the internal bench testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the given 510(k) summary. The summary focuses on engineering and regulatory comparisons rather than clinical performance studies requiring expert ground truth in the context of diagnostic or treatment efficacy.

4. Adjudication Method for the Test Set

This information is not provided in the given 510(k) summary, as there is no mention of a formal adjudication process for a clinical test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The device is a physical medical instrument (an inflatable bone tamp), not an AI-powered diagnostic or decision support system that would typically warrant such a study.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This device is an instrument used directly by a surgeon; it's not an algorithm or software requiring a standalone performance evaluation in the context of AI.

7. Type of Ground Truth Used

The ground truth used for demonstrating substantial equivalence was primarily:

  • Predicate Device Characteristics: The established design, intended use, and specifications of the legally marketed Kyphon Xpander II, KyphX Xpander, and KyphX Inflatable Bone Tamps.
  • Bench Test Results: Internal engineering data demonstrating the device's functional performance against its design requirements (though specific metrics are not detailed).
  • Compliance with Standards: Meeting the requirements of various recognized national and international standards related to medical devices.

8. Sample Size for the Training Set

There is no mention of a "training set" as this device is a physical instrument and not a machine learning or AI algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Traditional 510(k) Summary

JUN 2 5 2013

A) SUBMITTED BY: NeuroTherm, Inc. 30 Upton Drive, Suite 2 Wilmington, MA 01887-1083 Registration # 1226344

CONTACT: F. David Rothkopf MEDIcept Inc. 200 Homer Ave Ashland. MA 01721 617-899-3449 Phone 508-231-8861 Fax

B) DEVICE NAME: NeuroTherm Parallax Balloon Inflatable Bone Tamp

COMMON NAME: Arthroscope, Tamp

DEVICE CLASS/: 21 CFR 888.1100 Class II Arthroscope, product code HRX PRODUCT CODE 21 CFR 888.4540 Class I Orthopedic manual surgical instrument, product code HXG (Tamp) Accessory to 21 CFR 888.3027, Cement/Bone Vertebroplasty, product code NDN

C) PREDICATES:

  • . Kyphon Xpander II Inflatable Bone Tamp K101864
  • KyphX Inflatable Bone Tamp . K010246
  • . KyphX Xpander Inflatable Bone Tamp K041454

D) DEVICE DESCRIPTION:

The NeuroTherm Parallax Balloon Inflatable Bone Tamp with an inflatable component at the distal end. The tamp is designed to compress cancellous bone and/or move cortical bone as it inflates. The catheter shaft contains an outer lumen for the tamp inflation and a central lumen for a stylet to facilitate catheter introduction. At full inflation volume, the balloon design allows for a uniform balloon that is concentric to the catheter shaft. The wire mandrel provides stiffness to the balloon catheter to facilitate insertion through the access cannula. The NeuroTherm Parallax Balloon Inflatable Bone Tamp may be used with FDA cleared balloon inflation syringes.

{1}------------------------------------------------

E) INTENDED USE/INDICATIONS FOR USE:

The NeuroTherm Parallax Balloon Inflatable Bone Tamp is intended to be used as a conventional bone tamp for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The Parallax Balloon Inflatable Bone Tamp is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.

{2}------------------------------------------------

K122503 page 3/4

SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION E)

. . . . . .

Product codesNeuroTherm Parallax Balloon InflatableBone TampKyphX XpanderBone TampKyphX InflatableBone Tamp
HRX, HXGAccessory to NDNK101864/K041454HRX, HXGK010246HRX
Intended UseIndication for UseThe NeuroTherm Parallax BalloonInflatable Bone Tamp is intended to be usedas a conventional bone tamp for thereduction and fixation of fractures and/orcreation of a void in cancellous bone in thespine. This includes use duringpercutaneous vertebral augmentation. TheParallax Balloon Inflatable Bone Tamp is tobe used with cleared spinalpolymethylmethacrylate (PMMA) bonecement indicated for use duringpercutaneous vertebral augmentation, suchas kyphoplasty.The KYPHON Xpander isintended to be used as aconventional bone tamp for thereduction of fractures and/orcreation of a void in cancellousbone in the spine (including useduring balloon kyphoplasty with aPMMA-based bone cement that iscleared for use in kyphoplastyprocedures) hand, tibia, radius andcalcaneus.KyphX Inflatable Bone Tampsare intended to be used asconventional bone tamps forthe reduction of fracturesand/or creation of a void incancellous bone in the spine(including use during balloonkyphoplasty with KyphX HV-R Bone Cement), hand, tibia,radius and calcaneus.
Description
TampInflatable bone tamp consisting of a double lumen catheter shaft constructed from two coaxially aligned tubings
Side adaptorA side arm adapter provides access to the catheter lumens allowing inflation and deflation
Straight arm portA continuous straight arm port in inner catheter lumen allows placement of a non-removable stiffening stylet whichattaches to the Luer fitting of the straight arm
MarkersRadiopaque markers for balloon visualizationExit marker band
Balloon lengthindicatorYesPrinted bandYesColor coded bandUnknown
SterilizationEtOGamma irradiation

.

4-3

{3}------------------------------------------------

(F) CONCLUSION

The NeuroTherm Parallax Balloon Inflatable Bone Tamp is similar to or the same as the predicate devices for use as follows:

  • Technology .
  • . Intended use/Indication for Use in the spine
  • Technical specifications or range of technical specifications .

Any differences between the NeuroTherm Parallax Balloon Inflatable Bone Tamp and the predicate devices do not raise new issues of safety or effectiveness. Therefore, the NeuroTherm Parallax Balloon Inflatable Bone Tamp is substantially equivalent to the predicate devices.

G) PERFORMANCE TESTING

There are no applicable performance Consensus Standards or Guidance documents associated with this device.

Bench - Bench testing supports that the NeuroTherm Parallax Balloon Inflatable Bone Tamp performs as expected.

H) COMPLIANCE WITH STANDARDS

  • ASTM F 565-04 (Reapproved 2009)e1 Standard Practice for Care and Handling of . Orthopedic Implants and Instruments (pub date 10/04/2010)
  • ANSI/AAMI/ISO 11135:2007 Sterilization of health care products Ethylene oxide -. Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical device sterilization for EtO sterilized devices i
  • ISO 10993-7:2008 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals
  • ISO 594-1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other ● medical equipment - Part 1: General requirements
  • ISO 594-2 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other . medical equipment - Part 2: Lock fittings
  • ISO 10555-1 First edition 1995-06-5 AMENDMENT 2 2004-05-15 Sterile, single-use . intravascular catheters - Part 1: General requirements
  • ISO 10555-2: 1996 Sterile, single-use intravascular catheters -- Part 2: Angiographic . catheters
  • ISO 10555-3:1996/ Corrigendum 1:2002 Sterile, single-use intravascular catheters Part . 3: Central venous catheters
  • . ISO 10555-4: 1996 Sterile, single-use intravascular catheters -- Part 4: Balloon dilatation catheters

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 25, 2013

NeuroTherm, Incorporated % MEDIcept, Incorporated Mr. F. David Rothkopf 200 Homer Avenue Ashland, Massachusetts 01721

Re: K122503

Trade/Device Name: NeuroTherm Parallax Balloon Inflatable Bone Tamp Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN. HRX Dated: May 13, 2013 Received: May 15, 2013

Dear Mr. Rothkopf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{5}------------------------------------------------

Page 2 - Mr. F. David Rothkopf

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erin J. Keith

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K122503

Device Name: NeuroTherm Parallax Balloon Inflatable Bone Tamp

Indication for Use:

The NeuroTherm Parallax Balloon Inflatable Bone Tamp is intended to be used as a conventional bone tamp for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The Parallax Balloon Inflatable Bone Tamp is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.

Prescription Use X 21CFR 801, Subpart D OR Over-the-Counter Use 21CFR 801.109

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Laurence D. Coyne -A

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K122503

§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.

(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”