Search Results
Found 2 results
510(k) Data Aggregation
(259 days)
The MEDINAUT Kyphoplasty System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine, tibia, radius, and calcaneus. This includes percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
The MEDINAUT Kyphoplasty System is designed to reduce compression fracture and create a void in cancellous bone in the spine, tibia, radius, and calcaneus. Bone cement insertion using cement dispensing plunger is applied for spinal use only.
The MEDINAUT Kyphoplasty System consists of the MEDINAUT-X (Inflatable Bone Expander System) and MEDINAUT-I (Cement Dispenser System).
The MEDINAUT-X is comprised of a Balloon Catheter and a Bone Expander Syringe. The Balloon Catheter's main components are the shaft, Y-hub and the inflatable balloon located at the distal tip. The inflatable balloon is covered with a maximum 0.03g of silicone fluid to enhance lubricity when the balloon catheter is inserted in the cannula. Radiopaque markers located at the distal and proximal end of deflated working surface allow fluoroscopic visualization of the deflated balloon catheter during positioning. The Balloon Catheter and Bone Expander Syringe are supplied sterile and are disposable.
The MEDINAUT-I is a cement dispenser system consisting of a Needle Pipe, Needle Pin, Wire Pin, Cannula, Expander, Spacer, Cannula Expander, Cement Pusher, Cement Filler, and Guide Wire. They are supplied sterile and are disposable.
This document is a 510(k) summary for the MEDINAUT Kyphoplasty System. It is an FDA submission seeking approval to market a medical device, asserting its substantial equivalence to previously cleared devices. It primarily details physical and functional attributes and does not contain detailed "acceptance criteria" or "study that proves the device meets the acceptance criteria" in the way one might expect for a diagnostic AI device.
Instead, this document focuses on demonstrating substantial equivalence to predicate devices. This means that the device is shown to be as safe and effective as a legally marketed device (the predicate device) that does not require premarket approval.
Here's an analysis based on the provided text, addressing your points where possible, and noting where the information is not applicable or unavailable in this type of submission:
1. A table of acceptance criteria and the reported device performance
This information is not present in the document as it would be for a diagnostic AI device. For a device like the MEDINAUT Kyphoplasty System, "acceptance criteria" are related to mechanical performance, biocompatibility, and sterilization standards, and "reported device performance" refers to the results of non-clinical bench tests comparing these attributes to pre-set criteria or predicate devices.
The closest equivalent is the "Non-Clinical Testing" section and the comparison table.
Non-Clinical Testing Summary:
Test Type | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Sterilization Validation | In accordance with ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3 | Test results met pre-set criteria. |
Shelf Life | Validated shelf life | Test results validated 1-year shelf life. |
Biocompatibility | In accordance with ISO 10993-4, 5, 10, 11 (Meets ISO 10993) | Test results supported that devices are biocompatible. |
Bench Tests (Mechanical/Performance) | Met pre-set criteria for: | Test results met pre-set criteria for all listed tests. |
- Balloon deflation time | ||
- Burst pressure constrained | ||
- Burst pressure unconstrained | ||
- Fatigue strength | ||
- Balloon dimension (before/after) | ||
- Insertion and withdrawal force | ||
- Tensile bond strength |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable/provided in this context. This document describes a medical device (kyphoplasty system), not an AI/software device that processes data. The "test set" here refers to physical samples of the device undergoing mechanical, chemical, and biological testing, not a dataset of patient information. No patient data or provenance are mentioned because it's a non-clinical submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable/provided. The "ground truth" for this type of device is established through engineering and laboratory standards (e.g., ISO standards for sterilization, biocompatibility, and mechanical properties). There are no "experts" in the sense of clinicians establishing a diagnostic ground truth for patient data. The "experts" involved would be engineers and laboratory technicians performing standardized tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are used in clinical studies or evaluations of diagnostic performance where human interpretation of medical data (e.g., images) is involved. For non-clinical bench testing of a physical medical device, test results are typically validated against established quantitative standards and protocols, not through expert adjudication of uncertain cases.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable/provided. MRMC studies are specific to evaluating diagnostic systems, especially AI in medical imaging, and its impact on human reader performance. This document concerns a physical surgical device for bone fracture reduction, not a diagnostic system. There is no AI component mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable/provided. As stated, this is a physical medical device (kyphoplasty system), not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's safety and effectiveness is established through adherence to international standards (e.g., ISO 11137, ISO 11737, ISO 10993) and engineering specifications. For example:
- Sterilization: Demonstrated bacterial inactivation based on ISO standards.
- Biocompatibility: Demonstrated material safety based on ISO 10993.
- Mechanical Performance: Quantitative measurements (e.g., burst pressure, tensile strength) meeting design specifications and being comparable to predicate devices.
8. The sample size for the training set
This information is not applicable/provided. This device is not an AI model that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable/provided. As above, no training set for an AI model is involved.
Summary of Device and Equivalence Claim:
The document states that the MEDINAUT Kyphoplasty System (K153296) is substantially equivalent to its primary predicate device (MEDINAUT Kyphoplasty System, K133669) because it is identical in raw materials, design, and manufacturing processes. The purpose of the current 510(k) is to add more anatomical sites (tibia, radius, and calcaneus) to the indications for use. This expanded indication is supported by referencing an additional predicate device, the KyphX Inflatable Bone Tamp (K010246, K041454, K981251), which already has these broader claims and shares similar design, function, size, shape, materials, and physical specifications.
The "study that proves the device meets the acceptance criteria" (in this context, substantial equivalence to predicates for the expanded indications) is the non-clinical testing enumerated in section 8, which includes sterilization, shelf life, biocompatibility, and various bench tests. These tests confirmed that the device performs as intended and is safe when compared to established standards and the characteristics of the predicate devices.
Ask a specific question about this device
(313 days)
The NeuroTherm Parallax Balloon Inflatable Bone Tamp is intended to be used as a conventional bone tamp for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The Parallax Balloon Inflatable Bone Tamp is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
The NeuroTherm Parallax Balloon Inflatable Bone Tamp with an inflatable component at the distal end. The tamp is designed to compress cancellous bone and/or move cortical bone as it inflates. The catheter shaft contains an outer lumen for the tamp inflation and a central lumen for a stylet to facilitate catheter introduction. At full inflation volume, the balloon design allows for a uniform balloon that is concentric to the catheter shaft. The wire mandrel provides stiffness to the balloon catheter to facilitate insertion through the access cannula. The NeuroTherm Parallax Balloon Inflatable Bone Tamp may be used with FDA cleared balloon inflation syringes.
The NeuroTherm Parallax Balloon Inflatable Bone Tamp's acceptance criteria and performance study details are as follows:
1. Table of Acceptance Criteria and Reported Device Performance
The device's acceptance criteria are primarily demonstrated through substantial equivalence to predicate devices, supported by bench testing and compliance with relevant standards. No specific quantitative performance metrics or thresholds are explicitly stated as acceptance criteria beyond those inherent in establishing substantial equivalence.
Acceptance Criterion (Implicit) | Reported Device Performance and Evidence |
---|---|
Technology is similar or same as predicate devices. | The device description details a double-lumen catheter shaft, side arm adapter, continuous straight arm port, and radiopaque markers, which are comparable to the characteristics of the predicate KyphX Xpander and Inflatable Bone Tamps. |
Intended use/Indication for Use in the spine is similar or same as predicate devices. | The intended use for reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine, including percutaneous vertebral augmentation with cleared PMMA bone cement, directly aligns with the indications of the predicate KyphX Xpander and Inflatable Bone Tamps for similar spinal applications. |
Technical specifications or range of technical specifications are similar or same as predicate devices. | While specific ranges are not detailed, the comparison table shows similarities in general design and features. For instance, the presence of a balloon length indicator (printed band vs. color-coded band) and the use of an inflatable bone tamp design align with predicates. |
No new issues of safety or effectiveness are raised by differences. | The 510(k) summary explicitly states: "Any differences between the NeuroTherm Parallax Balloon Inflatable Bone Tamp and the predicate devices do not raise new issues of safety or effectiveness." This implies that any variations were assessed and deemed acceptable. |
Device performs as expected. | "Bench - Bench testing supports that the NeuroTherm Parallax Balloon Inflatable Bone Tamp performs as expected." (No specific quantitative results are provided in the summary.) |
Compliance with relevant consensus standards. | The device complies with a list of ASTM, ANSI/AAMI/ISO, and ISO standards related to implants, sterilization, biological evaluation, and luer fittings, ensuring fundamental safety and performance aspects are met. |
2. Sample Size Used for the Test Set and Data Provenance
The provided 510(k) summary does not specify a test set involving human data or a specific number of samples for the bench testing beyond the general statement "Bench testing supports that the NeuroTherm Parallax Balloon Inflatable Bone Tamp performs as expected." There is no indication of country of origin or whether any data was retrospective or prospective in detail. It appears the primary "test set" for substantial equivalence was the comparison to existing predicate devices and the internal bench testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not provided in the given 510(k) summary. The summary focuses on engineering and regulatory comparisons rather than clinical performance studies requiring expert ground truth in the context of diagnostic or treatment efficacy.
4. Adjudication Method for the Test Set
This information is not provided in the given 510(k) summary, as there is no mention of a formal adjudication process for a clinical test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. The device is a physical medical instrument (an inflatable bone tamp), not an AI-powered diagnostic or decision support system that would typically warrant such a study.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This device is an instrument used directly by a surgeon; it's not an algorithm or software requiring a standalone performance evaluation in the context of AI.
7. Type of Ground Truth Used
The ground truth used for demonstrating substantial equivalence was primarily:
- Predicate Device Characteristics: The established design, intended use, and specifications of the legally marketed Kyphon Xpander II, KyphX Xpander, and KyphX Inflatable Bone Tamps.
- Bench Test Results: Internal engineering data demonstrating the device's functional performance against its design requirements (though specific metrics are not detailed).
- Compliance with Standards: Meeting the requirements of various recognized national and international standards related to medical devices.
8. Sample Size for the Training Set
There is no mention of a "training set" as this device is a physical instrument and not a machine learning or AI algorithm.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1