K Number
K122490

Validate with FDA (Live)

Device Name
KULAVAC
Manufacturer
Date Cleared
2013-11-13

(455 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KulaVAC Foam Dressing Kit is intended to be used in conjunction with suction devices to deliver negative pressure wound therapy to the wound. KulaVAC Foam Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material, and tissue debris. It is indicated for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

Device Description

KulaVAC is wound dressing kit for Negative Pressure Wound Therapy (NPWT). The kit consists of individually reviewed medical components; polyurethane foam wound dressing, suction head, tubing, connectors, clamps and acrylic adhesive transparent film. Foam dressing is available in three sizes; small, medium and large, and the other accessories are used to connect to a suction device. The kit is supplied sterile and is single-use.

AI/ML Overview

The KulaVAC Foam Dressing Kit, a negative pressure wound therapy (NPWT) device, underwent non-clinical testing to demonstrate its safety and effectiveness.

1. Acceptance Criteria and Reported Device Performance

Test TypeAcceptance Criteria (Implicit)Reported Device Performance
Bench Testing (Mock Wound Model)Verify substantial equivalence to the predicate device in performance. This would imply meeting performance characteristics such as: - Ability to deliver negative pressure: The system should consistently and effectively apply and maintain specified negative pressure levels to the wound bed. - Fluid removal efficiency: The dressing and suction system should effectively remove exudates and infectious material from the wound. - Seal integrity: The adhesive film should maintain a sufficient seal around the foam dressing to prevent air leaks and maintain negative pressure. - Durability/Integrity of components: The foam, tubing, and connectors should withstand the pressures and conditions of NPWT without degradation or failure. - Compatibility with suction devices: The kit should be compatible with standard suction devices used for NPWT.All tests conducted for KulaVAC successfully met the acceptance criteria, demonstrating safety and effectiveness. Specifically, the bench testing using a mock wound model verified substantial equivalence to the predicate device in performance. The compatibility of the device with vacuum sources was also demonstrated.
Wound Dressing Test (EN 13726-1)Conformity to relevant sections of EN 13726-1, which typically covers: - Fluid handling capacity: The ability of the dressing to absorb and retain wound exudate. - Moisture vapor transmission rate (MVTR): The rate at which water vapor passes through the dressing. - Adhesiveness: The ability of the dressing to adhere to the skin. - Dimensional changes: Stability of the dressing's dimensions when wet or stretched.All tests conducted for KulaVAC successfully met the acceptance criteria, demonstrating safety and effectiveness. Wound dressing tests were conducted in accordance with EN 13726-1.
Biocompatibility Testing (ISO 10993-1)Conformity to ISO 10993-1, indicating absence of unacceptable biological risks. This typically includes tests for: - Cytotoxicity: No toxic effects on cells. - Sensitization: No allergic reactions. - Irritation: No skin or tissue irritation. - Hemocompatibility: No adverse effects on blood components (if applicable).All tests conducted for KulaVAC successfully met the acceptance criteria, demonstrating safety and effectiveness. Biocompatibility testing was conducted in accordance with ISO 10993-1.

2. Sample Size and Data Provenance for Test Set

  • Sample Size: The document does not specify the exact sample size used for each "non-clinical test." It broadly states that "non-clinical testing was performed."
  • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given that CG Bio Co., Ltd. is based in Korea and the tests are described as "non-clinical testing," it is reasonable to infer the testing was conducted in a laboratory setting, likely in Korea.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • This information is not provided in the document. The non-clinical tests described (bench testing, wound dressing test, biocompatibility) are objective measurements and do not typically involve experts establishing a ground truth in the way clinical studies or diagnostic AI algorithms do. Instead, they rely on established industry standards and protocols.

4. Adjudication Method (Test Set)

  • This information is not applicable as the tests are non-clinical and do not involve human interpretation or subjective assessment requiring adjudication.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not conducted. The provided text describes non-clinical bench, wound dressing, and biocompatibility tests, primarily focused on demonstrating substantial equivalence to predicate devices through technical performance. There is no mention of human readers or AI assistance in the context of comparative effectiveness.

6. Standalone Performance Study (Algorithm Only)

  • Yes, in essence, the "non-clinical testing" described can be considered a standalone performance study of the device itself. The tests evaluate the physical and functional aspects of the KulaVAC Foam Dressing Kit (e.g., its ability to deliver negative pressure, remove fluid, biocompatibility) without human intervention in the evaluation process, other than to set up and monitor the tests according to protocols. There is no "algorithm" in the context of an AI device, but rather the performance of the physical medical device.

7. Type of Ground Truth Used

  • The ground truth for the non-clinical tests was established through objective measurements and adherence to recognized industry standards and protocols.
    • For bench testing, the ground truth was the established performance characteristics of the predicate devices or engineering specifications.
    • For wound dressing tests, the ground truth was the performance requirements stipulated by the EN 13726-1 standard.
    • For biocompatibility testing, the ground truth was the safety criteria defined by the ISO 10993-1 standard.

8. Sample Size for the Training Set

  • This information is not applicable. The KulaVAC Foam Dressing Kit is a physical medical device, not an AI or machine learning algorithm that requires a "training set" of data.

9. How Ground Truth for Training Set Was Established

  • This information is not applicable as there is no training set for this type of medical device.

{0}------------------------------------------------

K122490 page 1/3

Image /page/0/Picture/1 description: The image contains a logo with the text "CG Bio" in bold, black letters. Above the letters are three curved lines, and below the letters is the text "Cell & Gene Biotechnology" in a smaller font. The logo appears to be for a biotechnology company specializing in cell and gene therapies.

CG Bio Co., Ltd.

510(k) SUMMARY

1. Submitter Information

Sponsor

Name:CG Bio Co., Ltd.
Address:223-23 Sangdaewon-dong, Jungwon-gu, Seongnam,Gyeonggi-do, 462-120 Koreawww.cgbio.co.kr
Tel.:+82-31-732-0840
Fax.:+82-31-732-0842

US Agent

NOV 1 3 2013

Name:Mr. Shawn Park
Address:Daewoong America Inc.155 Gibbs Street, Suite #431 Rockville, MD 20850
Tel.:301-309-1957, 301-956-0588
Fax.:301-309-1959
E-Mail:yann@daewoong.co.kr

Official Correspondent

Name:Ms. Jung-Eun ImRegulatory Affairs
Address:163-3 Samsung-dong, Kangnam-gu, Seoul, 135-715 Korea
Tel.:+82-2-550-8053
Fax.:+82-2-550-8660
E-Mail:mvrdv@daewoong.co.kr

2. Subject Device Information

Type of 510(k) submission:Traditional
Trade Name:KulaVAC
Common/Usual Name:Foam Dressing
Classification Name:Accessory to Powered Suction Pump
Regulation Number:21 CFR 878.4780
Product Code:OMP
Regulation Class:Class II
Panel:General & Plastic Surgery

{1}------------------------------------------------

K122490 Page 2/3

Image /page/1/Picture/1 description: The image shows a logo for CG Bio. The logo has the letters CG in large font on the left, with the word Bio in a smaller font on the right. Below the letters is the text "Cell & Gene Biotechnology" in a small font.

CG Bio Co., Ltd.

SponsorK.C.ISmith & Nephew, Inc.
Device NameV.A.C.® GranuFoam® DressingRenasys™-F NPWT Foam Dressing
510(k) NumberK032310K082211
Product CodeOMPOMP
Regulation No.21 CFR 878.478021 CFR 878.4780
Regulation ClassIIII

3. Predicate Device Information

4. Device Description

KulaVAC is wound dressing kit for Negative Pressure Wound Therapy (NPWT). The kit consists of individually reviewed medical components; polyurethane foam wound dressing, suction head, tubing, connectors, clamps and acrylic adhesive transparent film.

Foam dressing is available in three sizes; small, medium and large, and the other accessories are used to connect to a suction device.

The kit is supplied sterile and is single-use.

5. Indications for Use

KulaVAC Foam Dressing Kit is intended to be used in conjunction with suction devices to deliver negative pressure wound therapy to the wound. KulaVAC Foam Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material, and tissue debris. It is indicated for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

6. Non-Clinical Testing

Non-clinical testing was performed to assess the safety and effectiveness of KulaVAC. The non-clinical tests included bench testing using a mock wound model to verify the substantial equivalence to the predicate device in performance, wound dressing test in accordance with EN 13726-1 and biocompatibility testing in accordance with ISO 10993-1, respectably. And the compatibility of the device with the vacuum sources was demonstrated. All tests conducted for KulaVAC had successfully met the acceptance criteria, which demonstrate the safety and effectiveness of the device.

{2}------------------------------------------------

K122490 Page 3/3

Image /page/2/Picture/1 description: The image shows the logo for CG Bio. The logo features the letters "CG" in a large, bold font, with the word "Bio" in a smaller font to the right of the "G". Above the letters is a curved line. Below the letters is the text "Cell & Gene Technology" in a small font.

CG Bio Co., Ltd.

7. Comparison to Predicate Device

KulaVAC Foam Dressing Kit is substantially equivalent in design, materials, technology, function and intended use to its predicate devices. The minor technological differences between KulaVAC and its predicate devices raise no new issues of safety or effectiveness.

Elements ofComparisonSubject DevicePredicate DevicesVerdict
Product NameKulaVACV.A.C.®GranuFoam®DressingRenasys™-F NPWTFoam Dressing
510(k) NumberApplyingK032310K032310-
Product CodeOMPOMPOMPSE
Regulation No.21 CFR 878.478021 CFR 878.478021 CFR 878.4780SE
Materials- Foam Dressing : Polyurethane- Film Dressing : Acrylic adhesive polyurethane filmSE
ComponentsComposed of foam, tubing and adhesive filmSE
Intended UseCombination with a suction device to deliver negativepressure to the wound to promote wound healing by drainageof fluids and infectious materialsSE
IndicationsChronic, acute, traumatic, subacute and dehisced wounds,partial-thickness burns, ulcers (such as diabetic or pressure),flaps and grafts.SE

8. Conclusion

KulaVAC Foam Dressing Kit is substantially equivalent in design, materials, technology, function and intended use to its predicate devices. The minor technological differences between KulaVAC and its predicate devices raise no new issues of safety or effectiveness.

Performance data demonstrate that KulaVAC is as safe and effective as its predicate device, and comparable to the various controlled suction devices. Thus, KulaVAC is substantially equivalent.

9. Summary Prepared Date

18 October 2013

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 13, 2013

CGBio Co., Ltd. % JungEun Im 223-23 Sangdaewon-dong, Jungwon-gu, Seongnam, Gyeonggi-do. 462-120. Korea

Re: K122490

Trade/Device Name: KulaVAC Foam Dressing Kit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 18, 2013 Received: October 29, 2013

Dear JungEun Im:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 -- JungEun Im

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for vour device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You mav obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

for Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K122490

Device Name: KulaVAC

Indications for Use:

KulaVAC Foam Dressing Kit is intended to be used in conjunction with suction devices to deliver negative pressure wound therapy to the wound. KulaVAC Foam Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material, and tissue debris. It is indicated for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jiyoung Dang -S 2013.11.08 16:01:57 -05'00'

Page 1 of 1

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.