K Number
K122465

Validate with FDA (Live)

Date Cleared
2012-10-16

(64 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free examination glove is a disposable device intended for medical purposes, that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Midas Touch™ - Nitrile Examination Gloves Powder free, Pink

AI/ML Overview

The provided text is a 510(k) premarket notification letter from the FDA for medical gloves, specifically "Midas Touch 101 - Nitrile Examination Gloves-Powder Free, Pink."

This document does not contain information about acceptance criteria and a study proving a device's performance in the sense of an AI/ML or diagnostic device. Instead, it concerns the FDA's determination of substantial equivalence for a patient examination glove to predicate devices.

Therefore, I cannot provide the requested information about acceptance criteria and a study that proves the device meets those criteria, as the document does not pertain to such a study or device.

The document primarily focuses on:

  • Regulatory Clearance: K122465, indicating the FDA has reviewed the submission and determined substantial equivalence.
  • Device Type: Nitrile Examination Gloves, Powder Free, Pink.
  • Intended Use: "A powder-free examination glove is a disposable device intended for medical purposes, that is worn on the examiner's hand or finger to prevent contamination between patient and examiner."
  • Regulatory Classification: Class I, Product Code LZA (Patient Examination Glove).
  • Manufacturing Information: Industrial Clothings, Limited, Prime Polymers Division, from Sri Lanka.

For a medical glove, acceptance criteria would typically relate to physical properties (e.g., tensile strength, elongation), barrier integrity (e.g., freedom from pinholes, viral penetration), sterility (if applicable), and biocompatibility. The "study" proving these would be laboratory testing against recognized standards (e.g., ASTM standards for medical gloves). This specific FDA letter, however, only states that substantial equivalence has been determined, implying that such testing was submitted and found acceptable by the FDA, but it does not detail the tests, methodology, or results within the letter itself.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Industrial Clothings, Limited
Mr. S.N. Hari
Deputy General Manager
Prime Polymers Division
D17, Seethawaka Export Processing Zone
Avissawella, Sri Lanka 10700

OCT
16 2012

Re: K122465

Trade/Device Name: Midas Touch 101 - Nitrile Examination Gloves-Powder Free, Pink Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: August 9, 2012 Received: August 20, 2012

Dear Mr. Hari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2- Mr. Hari

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Anton D. Naser

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for "MIDAS SAFETY". The logo consists of a stylized hand on the left, followed by the word "MIDAS" in large, bold letters. Below "MIDAS" is the word "SAFETY" in smaller letters. Underneath the words is a line, and below the line are the words "INNOVATIVE SAFETY SOLUTIONS".

Industrial Clothings Limited (Prime Polymers Division)

510(k) Submission for Pink Nitrile Examination Gloves – Powder free

: K122465

INDICATIONS FOR USE

Applicant

: Industrial Clothings Ltd., Prime Polymers Division

510(k) number (if known)

Device name

Indications for use

: Midas Touch™ - Nitrile Examination Gloves Powder free, Pink

: A powder-free examination glove is a disposable device intended for medical purposes, that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

PNPF

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth P. Clamie-Williams

ston Sign-Off)
ion of Anesthesiology, General Hospital
Hon of Anesthesiology, General Devices

510(k) Number: K122465

4-1 | P a g e

Amendment 2

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.