K Number
K122336

Validate with FDA (Live)

Date Cleared
2013-01-09

(160 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Eleven Blade Solutions, Inc. Q-Fix™ Suture Anchor System is intended for soft tissue to bone fixation for:
Shoulder
Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff repair, Biceps tenodesis
Foot and Ankle
Medial/lateral repair and reconstruction, Mid and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair
Elbow
Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment
Knee
Extra-capsular repair: MCL, LCL, and posterior oblique ligament, Illiotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure
Hand and Wrist
Collateral ligament repair, Sacpholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction
Hip
Acetabular labral repair

Device Description

The Q-Fix™ Suture Anchor is an all-suture anchor device designed for soft tissue to bone fixation by expanding against bone when deployed. Made of a polyester sleeve-type structure with one or more strands of UHMWPE suture threaded through it, the anchor is provided preloaded on a disposable inserter. The inserter is used to deploy the anchor into the bone, with the suture remaining slideable within the anchor to facilitate attachment of the soft tissue. The one-time use anchor is provided sterile, preloaded on the inserter. Additional accessory instruments, including a drill bit and drill guide with obturator are single use devices provided sterile for use during the orthopedic procedure.

AI/ML Overview

The Q-Fix™ Suture Anchor System is a medical device designed for soft tissue to bone fixation. The provided document is a 510(k) summary, which indicates that the device is seeking substantial equivalence to already marketed predicate devices. This type of submission focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than proving absolute safety and efficacy through extensive clinical trials. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the demonstration of substantial equivalence through bench testing compared to predicate devices.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative "acceptance criteria" in a table format with corresponding "reported device performance" values for the Q-Fix™ Suture Anchor System. Instead, it states that bench studies were conducted to evaluate performance characteristics and demonstrate equivalence to predicate devices. The implicit acceptance criterion is that the Q-Fix™ Suture Anchor System's performance characteristics are "equivalent to the predicate devices and thus provide equivalent fixation strength within their intended use."

The performance characteristics evaluated were:

Performance CharacteristicReported Device Performance (Implicit Acceptance)
Anchor Pull-out StrengthEquivalent to predicate devices
Cyclic DisplacementEquivalent to predicate devices
Device Insertion TestingEquivalent to predicate devices

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "bench studies" for Anchor Pull-out Strength, Cyclic Displacement, and Device Insertion Testing. However, it does not specify the sample sizes used for these tests. It also does not provide information on the country of origin of the data or whether the studies were retrospective or prospective, as these are bench tests and not human or animal studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable in the context of this 510(k) summary. The "ground truth" for bench testing of a suture anchor system is typically based on established engineering and biomechanical standards, not on expert clinical consensus. The assessments are objective measurements of physical properties.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable. Adjudication methods are typically used in clinical studies or image interpretation studies where expert consensus is needed. For bench testing of physical properties, measurements are objective and do not require expert adjudication in the same way.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. An MRMC comparative effectiveness study is used for evaluating diagnostic aids or imaging devices where human interpretation is involved. The Q-Fix™ Suture Anchor System is a surgical implant, and its performance is evaluated through biomechanical bench testing.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable. The Q-Fix™ Suture Anchor System is a physical medical device, not an algorithm or AI system. Its performance is inherent to its physical properties and design.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For the bench studies, the "ground truth" for assessing device performance is based on objective biomechanical measurements against established engineering and material science principles, and comparison to the performance characteristics of the predicate devices. It is not based on expert consensus, pathology, or outcomes data, as these are typically relevant for clinical studies.

8. The sample size for the training set

This question is not applicable. This device is a physical medical implant, not an AI or machine learning algorithm. Therefore, there is no "training set."

9. How the ground truth for the training set was established

This question is not applicable, as there is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "ELEVEN BLADE SOLUTIONS". The word "ELEVEN" is in large, bold, sans-serif font, with the "V" stylized to look like two blades. Below "ELEVEN" is the phrase "BLADE SOLUTIONS" in a smaller, bold, sans-serif font.

Page 1/2

K122336
Q-Fix™ Suture Anchor System
January 7, 2013

JAN 9 2013

510(k) Summary

General Company Information
Name:Eleven Blade Solutions, Inc.
Contact:Michael KolberRegulatory Affairs
Address:147 Hillbrook Dr.Los Gatos, CA 95032
Telephone:408-505-6626
Fax:408-402-8361
Date Prepared:January 7, 2013
General Device Information
Product Name:Q-Fix™ Suture Anchor System
Common Name:Suture Anchor
Classification:21CFR888.3040; Smooth or threaded metallicbone fixation fastener
Device Class:Class II
Product Code:MBI
Predicate Devices
ManufacturerDevice Name510(k) Number
Biomet Sports MedicineJuggerknot Soft AnchorK110145
Teleflex MedicalForceFiberK063778
Description
The Q-Fix™ Suture Anchor is an all-suture anchor device designed for soft tissue tobone fixation by expanding against bone when deployed. Made of a polyester sleeve-type structure with one or more strands of UHMWPE suture threaded through it, theanchor is provided preloaded on a disposable inserter. The inserter is used to deploythe anchor into the bone, with the suture remaining slideable within the anchor tofacilitate attachment of the soft tissue.The one-time use anchor is provided sterile, preloaded on the inserter. Additionalaccessory instruments, including a drill bit and drill guide with obturator are singleuse devices provided sterile for use during the orthopedic procedure.
Intended Use (Indications)
The Eleven Blade Solutions, Inc. Q-Fix™ Suture Anchor System is intended for softtissue to bone fixation for:
Shoulder
Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsularshift/capsulolabral reconstruction, Deltoid repair, Rotator cuff repair, Bicepstenodesis
Foot and Ankle
Medial/lateral repair and reconstruction, Mid and forefoot repair, Hallux valgus

.

{1}------------------------------------------------

122336

Image /page/1/Picture/1 description: The image shows the logo for "ELEVEN BLADE SOLUTIONS". The word "ELEVEN" is in large, bold, sans-serif font, with a stylized image of two blades forming a "V" in the middle. Below "ELEVEN", in a smaller, bold, sans-serif font, is the text "BLADE SOLUTIONS".

Page 2/2

K122336 Q-Fix™ Suture Anchor System January 7, 2013

reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair

Elbow

Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment

Knee

Extra-capsular repair: MCL, LCL, and posterior oblique ligament, Illiotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure

Hand and Wrist

Collateral ligament repair, Sacpholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction

Hip

Acetabular labral repair

Substantial Equivalence

This submission supports the position that the Eleven Blade Solutions, Inc. O-Fix™ Suture Anchor System is substantially equivalent to the Biomet Juggerknot (K110145) and the Teleflex ForceFiber (K063778). The O-Fix™ Suture Anchor System has the following similarities to the previously cleared predicate device: the same intended use, same operating principle, similar technologies, and similar manufacturing process. Design verification activities were performed as a result of the risk analysis. The 510(k) notice contains summaries of bench studies, which were conducted to evaluate the performance characteristics of the O-Fix™ Suture Anchor System. Anchor Pull-out Strength, Cyclic Displacement, and Device Insertion Testing were performed. The data presented demonstrate that the performance characteristics of the Q-Fix™ Suture Anchor System are equivalent to the predicate devices and thus provide equivalent fixation strength within their intended use.

Conclusions

Eleven Blade Solutions, Inc. believes that the information provided demonstrates that the proposed device is substantially equivalent to the predicate devices and does not raise any new issues of safety or efficacy. Based on the indications for use, technological characteristics, and comparison to predicated devices, the Q-Fix™ Suture Anchor System has been shown to be substantially equivalent to predicate devices as described under the Federal Food, Drug and Cosmetic Act.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the graphic.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Eleven Blade Solutions, Incorporated % Mr. Michael Kolber Regulatory Affairs 147 Hillbrook Drive Los Gatos, California 95032

Letter dated: January 9, 2013

Re: K122336

Trade/Device Name: Eleven Blade Solutions, Incorporated. Q-Fix™ Suture Anchor System Regulation Number: 21 CFR 888.3040

Regulation Name: Smooth or threaded metallic bone fixation fastener.

Regulatory Class: Class II

Product Code: MBI

Dated: November 23, 2012

Received: November 26, 2012

Dear Mr. Kolber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Mr. Michael Kolber

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "ELEVEN BLADE SOLUTIONS" in a bold, sans-serif font. The word "ELEVEN" is stacked on top of "BLADE SOLUTIONS". The text is black and the background is white. The image appears to be a logo or heading.

Indications for Use

510(k) Number (if known): K122336

Device Name: Eleven Blade Solutions, Inc. Q-Fix™ Suture Anchor System

Indications for Use: The Eleven Blade Solutions, Inc., Q-Fix™ Suture Anchor System is intended to be used for soft tissue to bone fixation for:

Shoulder

Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction. Deltoid repair, Rotator cuff tear repair, Biceps tenodesis

Foot and Ankle

Medial/lateral repair and reconstruction, Mid and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction. Achilles tendon repair

Elbow

Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment

Knee

Extra-capsular repair: MCL, LCL, and posterior oblique ligament, Illiotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure

Hand and Wrist

Collateral ligament repair, Sacpholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction

Hip

Acetabular labral repair

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton E. Dmitriev, PhD 2013.01.09 14:50:59 -05'00' Division of Orthopedic Devices

N/A