K Number
K121947

Validate with FDA (Live)

Date Cleared
2012-09-10

(69 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder free blue nitrile patient exam glove, meets all the requirements of ASTM standard D6319, except for sterility requirements. Non-sterile, powder free, examination gloves made of nitrile and colored blue. The textured gloves are provided in sizes extra small, small, medium, large and extra large. Non-sterile gloves are provided in dispenser boxes.

AI/ML Overview

The provided text describes a 510(k) summary for the "ETS Blue Powder Free Nitrile Patient Exam Glove," a Class I medical device. The summary focuses on demonstrating substantial equivalence to a predicate device by meeting established performance standards.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard)Device Performance
DimensionASTM D6319Meets
Physical PropertiesASTM D6319Meets
Freedom from PinholesASTM D6319 & ASTM D5151Meets
Powder ResidualASTM D6124Meets (Results generated values below 2mg of residual powder)
BiocompatibilityPrimary Skin Irritation in rabbits (ISO 10993-10)Gloves are non-irritating
BiocompatibilityDermal Sensitization in the guinea pig (ISO 10993-10)Gloves do not display any potential for sensitization

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for the specific tests (e.g., number of gloves tested for dimensions, pinholes, etc.). It only indicates that the device "Meets" the specified ASTM and ISO standards, which inherently include sampling plans within those standards.

  • Sample Size: Not explicitly stated for each test. The standards referenced (ASTM D6319, D5151, D6124, ISO 10993-10) would define the required sample sizes for demonstrating compliance.
  • Data Provenance: Not explicitly stated. This information is typically generated by the manufacturer or a contracted testing lab during product development and quality assurance. It's retrospective testing of the manufactured product.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable to this type of device. The "ground truth" for patient examination gloves is established by objective, verifiable physical and chemical properties measured against recognized industry standards (ASTM, ISO). There are no human "experts" establishing a subjective ground truth for these performance criteria in the way there would be for, for example, image interpretation in a diagnostic AI study.

4. Adjudication Method for the Test Set

This information is not applicable. As mentioned above, the "ground truth" is determined by objective physical and chemical testing against standards, not by expert consensus or adjudication of subjective interpretations.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No. This is a Class I medical device (patient examination glove). MRMC studies are relevant for diagnostic devices, particularly those involving human interpretation of complex data (like medical images), where the AI might assist a human reader. This device does not involve that type of interaction or assessment.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No. This is a physical product (glove), not an algorithm or software device. Standalone performance refers to the accuracy of an AI algorithm on its own.

7. The Type of Ground Truth Used

The ground truth used is based on objective measurements against established national and international standards. Specifically:

  • Physical properties (dimensions, tensile strength, ultimate elongation) are measured against ASTM D6319-10.
  • Freedom from pinholes is measured against ASTM D5151-06 and D6319-10.
  • Residual powder is measured against ASTM D6124-06.
  • Biocompatibility (skin irritation and sensitization) is assessed against ISO 10993-10.

8. The Sample Size for the Training Set

Not applicable. This product is a physical medical device. There is no "training set" in the context of machine learning or AI algorithms for this device. The manufacturing process is validated (which might involve statistical process control and testing of batches), but this is distinct from an AI training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. (See point 8).

In summary: The provided 510(k) summary for the ETS Blue Powder Free Nitrile Patient Exam Glove demonstrates its safety and effectiveness by showing that it meets recognized consensus standards (ASTM and ISO) for physical characteristics and biocompatibility, thereby establishing substantial equivalence to a legally marketed predicate device. The nature of this device means that many of the AI/algorithm-specific questions are not relevant.

{0}------------------------------------------------

510(k) SUMMARY

DATE:

OWNER:

February 17, 2012

SEP 10 2012

Northstar Healthcare Holdings 70 Sir John Rogerson's Quay Dublin 2, Ireland

OFFICIAL

CORRESPONDENT:

Alex Nagy Manager, Quality Systems and Regulatory Affairs Cypress Medical Products LLC 1200 South Route 31 McHenry, IL 60050 Telephone: 815-385-0100 Fax: 815-385-0114

DEVICE NAME:

Trade Name:ETS Blue Powder Free Nitrile Patient Exam Glove
Common Name:Patient Examination Gloves
Classification:Patient Examination Gloves
Class:Class I
Product Code:LZA

PREDICATE DEVICE(S):

Predicate510(k)Device NameIndicationClearanceDateCompany
K062764Nitrile Powder FreeExamination Gloves(White, Blue)The patient examination glove is adisposable device intended for medicalpurposes that is worn on the examiner's handto prevent contamination between patient andexaminer.27 Feb 07KossanLatexIndustriesSDN BHN.SelangorDarulEhsan,Malaysia

DEVICE DESCRIPTION: Powder free blue nitrile patient exam glove, meets all the requirements of ASTM standard D6319, except for sterility requirements.

Northstar Healthcare Holdings

{1}------------------------------------------------

STATEMENT OF INTENDED USE:

The nitrile examination glove is a disposable device intended for medical purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.

TECHNOLOGICAL CHARACTERISTICS: The ETS Blue Powder Free Nitrile Patient Exam Glove is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with nitrile using similar manufacturing processes.

FeatureNitrile Powder Free Examination Gloves Testedfor Use with Chemotherapy Drugs (Blue,White)K062764PredicateETS Blue Powder Free NitrilePatient Exam GlovesProposed
Intended UseIntended for medical purposes that is worn onthe examiner's hand to prevent contaminationbetween patient and examiner. Tested for Usewith Chemotherapy Drugs.Intended for medical purposes thatis worn on the examiner's hand toprevent contamination betweenpatient and examiner.
Indications for UseStatementThe patient examination glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand to preventcontamination between patient and examiner.This device is not intended to be used as achemical barrier.The examination glove is adisposable device intended formedical purposes that is worn on theexaminer's hand or finger to preventcontamination between patient andexaminer.
DescriptionNon-sterile, powder free, examination glovesmade of nitrile and colored white. The texturedgloves are provided in sizes extra small, small,medium, large and extra large. The smoothgloves are provided in sizes small, medium,large and extra large.Same
PresentationNon-sterile gloves are provided in dispenserboxes.Same
MaterialNitrileSame
SterilizationNon-sterileSame
Single UseYesSame
AmbidextrousYesSame
DimensionsMeets ASTM D6319-10Same
Tensile StrengthMeets ASTM D6319-10Same
Ultimate ElongationMeets ASTM D6319-10Same
Freedom fromPinholesMeets ASTM D5151-06 and ASTM D6319-10Same
Residual PowderMeets ASTM D6124-06Same
BiocompatibilityPasses Primary Skin Irritation in RabbitsSame
TestsPasses Closed Patch Sensitization in Guinea PigsSame

Northstar Healthcare Holdings

{2}------------------------------------------------

ASSESSMENT OF NONCLINICAL DATA:

CharacteristicStandardDevice Performance
DimensionASTM D6319Meets
Physical PropertiesASTM D6319Meets
Freedom from PinholesASTM D6319 & ASTM D5151Meets
Powder ResidualASTM D6124Meets
Results generated values
below 2mg of residual
powder
BiocompatibilityPrimary Skin Irritation in rabbitsGloves are non-irritating
(ISO 10993-10)
Dermal Sensitization in the guineaGloves do not display any
pig (ISO 10993-10)potential for sensitization

CONCLUSIONS:

The ETS Blue Powder Free Nitrile Patient Exam Gloves meet the requirements of established standards ASTM D6319-10, ASTM D6124-06, ASTM D5151-06, and ISO 10993-10.

Based on the comparison of intended use, design, materials and performance, the ETS Blue Powder Free Nitrile Patient Exam Gloves are substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Northstar Healthcare Holdings C/O Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062

SEP 10 2012

Re: K121947

Trade/Device Name: ETS Blue Powder Free Nitrile Patient Exam Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LZA Dated: August 28, 2012 Received: August 28, 2012

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

For Q. Shem.D.

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): _ K 12 1947

ETS Blue Powder Free Nitrile Patient Exam Glove Device Name:

Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of _510(k) Number:

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.