K Number
K121652

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-10-05

(122 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

Device Description

The CVue Advanced HydraVUE (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lenses for daily wear are fabricated from efrofilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. The non-ionic lens material, (efrofilcon A) is a daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers and optionally incorporates D&C Green 6 as an integrated handling tint. The lenses are lathe-cut and made to order for individual prescriptions. Each lens consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (efrofilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution. The lenses will be manufactured in spherical, toric, multifocal toric and irregular cornea configurations. The lens is supplied sterile in vials containing a buffered saline solution.

AI/ML Overview

This 510(k) summary describes the CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lenses for daily wear. The device is being submitted for an expanded indication to include the management of irregular corneal conditions like keratoconus and post-graft fitting, in addition to the correction of visual acuity in aphakic and non-aphakic persons with myopia or hyperopia.

The submission claims substantial equivalence to three predicate devices: IntelliWave3, Metro Soft, and CVue Advanced Definitive, all of which are efrofilcon A silicone hydrogel contact lenses for daily wear.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or threshold format for clinical efficacy. Instead, it relies on the established performance of the predicate devices and the material properties. The acceptance is implied by demonstrating substantial equivalence in material, manufacturing, and intended use as the predicates that have already met regulatory requirements.

CharacteristicAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (CVue Advanced HydraVUE)
Intended UseSame as predicate devices for irregular cornea management and ametropia correctionIndicated for correction of ametropia and management of irregular corneal conditions (keratoconus, post-graft fitting)
Materialefrofilcon Aefrofilcon A
Manufacturing ProcessLathe-cut, custom manufacturedLathe-cut, made to order for individual prescriptions
Water Content74%74%
Oxygen Permeability60 x 10⁻¹¹ (cm²/sec)(ml O₂/ml x mm Hg@35 °C)60 x 10⁻¹¹ (cm²/sec)(ml O₂/ml x mm Hg@35 °C) (Note: reported as 59.8 x 10⁻¹¹ in description)
Specific Gravity1.130 (predicate table)1.048 (hydrated) (description), 1.130 (predicate table)
SterilitySterile for indicated shelf-lifeLenses supplied in glass vials are sterile for the indicated shelf-life
Physical/Material PropertiesConsistent with marketed lensesLens physical and material properties are consistent with currently marketed lenses
Clinical PerformancePreviously established for efrofilcon A materialClinical performance of efrofilcon A lens material has been previously established
BiocompatibilityPreviously establishedPreclinical toxicology and biocompatibility were previously established
Shelf-life StabilityFollows scientific protocolsShelf-life stability studies followed scientific protocols, data scientifically valid

2. Sample Size Used for the Test Set and the Data Provenance

The document does not describe a specific "test set" or clinical study with patient sample sizes for the CVue Advanced HydraVUE Irregular Cornea lenses itself. Instead, it relies on substantial equivalence to predicate devices, whose clinical performance and material properties were previously established.

Therefore, no new clinical study was conducted for this specific 510(k) submission to demonstrate performance based on a new test set. The data provenance for the previously established clinical performance and material data related to efrofilcon A is not detailed in this document. It references Master File MAF #1708 for the efrofilcon A material.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable, as no new clinical study (test set) was performed for this 510(k) submission. Clinical performance was established through previous submissions for the efrofilcon A material itself.

4. Adjudication Method for the Test Set

Not applicable, as no new clinical study (test set) was performed for this 510(k) submission.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a contact lens, not an AI-powered diagnostic or assistive technology. No MRMC study was conducted.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a contact lens. No algorithm-only or standalone performance study was done.

7. The Type of Ground Truth Used

For the safety and effectiveness of the efrofilcon A material, the "ground truth" was established through:

  • Preclinical toxicology and biocompatibility studies: These tests assess the material's interaction with biological systems.
  • Clinical performance data for the efrofilcon A lens material: This would involve patient studies from previous submissions to demonstrate the material's safety and effectiveness in-vivo. The document references Master File MAF #1708 for this data.
  • Physical and material property testing: Laboratory measurements of water content, oxygen permeability, light transmission, etc., with established benchmarks for safe and effective contact lenses.
  • Sterility and shelf-life stability studies: Laboratory validation that the product remains sterile and effective over its shelf life.

8. The Sample Size for the Training Set

Not applicable. This device is a contact lens, not an AI model.

9. How the Ground Truth for the Training Set was Established

Not applicable. This device is a contact lens, not an AI model.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:K121652
-----------------------------------------

Applicant information:

Date Prepared:1 June 2012
Name:Unilens Corp., USA
Address:10431 72nd Street NorthLargo, FL 33777
Contact person:Alan J. FrazerDirector of Quality AssuranceOfficial Correspondent
Telephone:727-544-2531

Device information

Device Classification:Class II
Classification Number:LPL
Classification Name:Lenses, Soft Contact, Daily Wear
Trade Name:CVue Advanced HydraVUE Irregular Cornea(efrofilcon A) Soft (hydrophilic) Silicone Hydrogelcontact lenses for daily wear

Equivalent Devices:

The CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lenses for daily wear are substantially equivalent to the following predicate devices:

IntelliWave3, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) By Art Optical Contact Lens, Inc.

510(k) number K100221

Metro Soft Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) By Metro Optics of Austin Inc.

510(k) number K100244

CVue Advanced Definitive Daily Wear Soft Contact Lens (efrofilcon A) By Unilens Corp., USA

510(k) number K100456

Device Description:

The CVue Advanced HydraVUE (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lenses for daily wear are fabricated from efrofilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

5 2012

ОСТ

{1}------------------------------------------------

The non-ionic lens material, (efrofilcon A) is a daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers and optionally incorporates D&C Green 6 as an integrated handling tint. The lenses are lathe-cut and made to order for individual prescriptions. Each lens consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN). The blanks from which the lenses are lathe-cut are manufactured by Contamac, Ltd., Bearwalden Business Park, Saffron Walden, Essex CB11 4JX, United Kingdom, UK.

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (efrofilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

Refractive Index1.38
Light TransmissionGreater than 97%
Surface CharacterHydrophilic
Water Content74%
Specific Gravity1.048 (hydrated)
Oxygen Permeability59.8 x 10-11(cm/sec) (ml O2/ml x hPa @ 35 °C),(revised Fatt method)

The Physical properties of the lens are:

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 74% water by weight. The lenses will be manufactured in spherical, toric, multifocal toric and irregular cornea configurations with the following features and properties.

Chord Diameter12.0 mm to 16.00 mm
Center Thickness0.01 mm to 0.50 mm
Base Curve8.0 mm to 9.5 mm
Power Range-20.00D to +20.00D in 0.25D steps
Cylinder Power (Toric)-0.25D to -10.00D
Add Power (Multifocal)+0.50D to +4.00D

The lens is supplied sterile in vials containing a buffered saline solution. Vial labeling is printed with appropriate lot numbering, expiration dating and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.

Intended Use:

The CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned

{2}------------------------------------------------

replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

Testing:

Non-clinical Testing

Preclinical toxicology and biocompatibility were previously established and are not required for this 510(k). All non-clinical testing was conducted according to valid scientific protocols and was presented in K100456. The pre-clinical data for the contact lens blank material is included in K100456 by reference to the efrofilcon A material in Master File MAF #1708.

Test results of the non-clinical testing on the CVue Advanced HydraVUE (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear demonstrate that:

  • Lenses supplied in glass vials are sterile for the indicated shelf-life, .
  • Lens physical and material properties are consistent with currently marketed lenses. ◆

Clinical Data

The clinical performance of the efrofilcon A lens material has been previously established, and therefore is not required for this 510(k). The clinical data for the irregular cornea indications is included in K100456 by reference to the efrofilcon A material in Master File MAF #1708.

The CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lenses for daily wear are identical to the cleared Intelliwave³ Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) for irregular cornea management, cleared under K100221, and to the Metro Soft Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) for irreqular cornea management cleared under K100244.

The CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lenses for daily wear have the identical manufacturing process (lathe cut versus lathe cut) to the cleared Intelliwave3 Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) for irregular cornea management, cleared under K100221, and to the Metro Soft Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) for irregular cornea management cleared under K100244.

{3}------------------------------------------------

Substantial Equivalence

The following matrix illustrates the production and material characeristics of the CVue Advanced HydraVUE
(efrofileon A) Soft (hydrophilic) Silicone Hydrogel contact lenses f

UseCVue Advanced HydraVue(efrofilcon A) Irregular CorneaNew indicationArt OpticalIntelliWave³ Irregular Cornea(efrofilcon A)Predicate deviceMetro Soft irregular cornea(efrofilcon A)Predicate deviceCVue Advanced HydraVueToric Multifocal (efrofilcon A)Predicate device
Indicated for daily wear for thecorrection of ametropia (myopia andhyperopia) in aphakic and non-aphakicpersons and may be prescribed inotherwise non-diseased eyes thatrequire a Soft Contact Lens for themanagement of irregular cornealconditions such as keratoconus andpost graft fitting.Indicated for daily wear for thecorrection of ametropia (myopia andhyperopia) in aphakic and non-aphakicpersons and may be prescribed inotherwise non-diseased eyes thatrequire a Soft Contact Lens for themanagement of irregular cornealconditions such as keratoconus andpost graft fitting.Indicated for daily wear for thecorrection of ametropia (myopia andhyperopia) in aphakic and non-aphakicpersons and may be prescribed inotherwise non-diseased eyes thatrequire a Soft Contact Lens for themanagement of irregular cornealconditions such as keratoconus andpost graft fittingIndicated for daily wear for thecorrection of visual acuity in aphakic andnot aphakic persons with non-diseasedeyes with myopia or hyperopia, possesrefractive astigmatism and/or arepresbyopic.
abilitySame as predicate device: The contactlenses act as a refractive medium thatfocus light rays from near and distantobjects on the retinaThe contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retinaThe contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retinaThe contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retina
Daily wear, Silicone Hydrogel Soft(hydrophilic) Contact LensDaily wear, Silicone Hydrogel Soft(hydrophilic) Contact LensDaily wear, Silicone Hydrogel Soft(hydrophilic) Contact LensDaily wear, Silicone Hydrogel Soft(hydrophilic) Contact Lens
On MethodLathe-cut, made to order for individualprescriptionsLathe-cut, custom manufacturedLathe-cut, custom manufacturedLathe-cut, custom manufactured
SAN nameefrofilcon Aefrofilcon Aefrofilcon Aefrofilcon A
ntent74%74%74%74%
OxygenPermeability$60 x 10^{-11}$ (cm²/sec(ml O₂/ml x mmHg@35 °C) (revised Fatt method)$60 x 10^{-11}$ (cm²/sec(ml O₂/ml x mmHg@35 °C) (revised Fatt method)$60 x 10^{-11}$ (cm²/sec(ml O₂/ml x mmHg@35 °C) (revised Fatt method)$60 x 10^{-11}$ (cm²/sec(ml O₂/ml x mmHg@35 °C) (revised Fatt method)
SpecificGravity1.1301.1301.1301.139

{4}------------------------------------------------

Conclusions Drawn from the Studies

Validity of the Scientific Data

As previously presented in predicate 510(k)s, the toxicology studies, microbiology and chemistry studies all followed scientific protocols. The shelf-life stability studies followed scientific protocols, and the data were determined to be scientifically valid under 21 CFR 860.7.

Substantial Equivalence

Information presented in this Premarket Notification establishes that the CVue Advanced HydraVUE (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel contact lens for daily wear is as safe and effective as the predicate devices when used in accordance with the labeled directions for use and for the requested indication.

Risks and Benefits

The risks of the subject device are the same as those normally attributed to the wearing of silicone hydrogel, daily wear soft contact lenses. The benefits to the patient are the same as those for other silicone hydrogel contact lenses when compared to the predicates.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing segments, representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 5 2012

Unilens Corporation % Mr. Alan J. Frazer Director of Quality Assurance 10431 72nd Street North Largo, FL 33777

Re: K121652

Trade/Device Name: CVue Advanced HydraVUE (efrofilcon A) soft (hydrophilic) silicone hydrogel contact lens for dailywear Regulation Number: 21 CFR 886.5925

Regulation Name: Lenses, Soft Contact, Daily Wear

Regulatory Class: Class II Product Code: LPL Dated: September 14, 2012

Received: September 18, 2012

Dear Mr. Frazer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K121652

Device Name: CVue® Advanced™ HydraVUE (efrofilcon A) soft (hydrophilic) contact lenses

Indications for Use:

The CVue Advanced HydraVUE Irregular Cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

Eye care practitioners may prescribe the lens for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection system.

Prescription Use _ X Over-The-Counter Use _ AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

KI21652 510(k) Number_

Page | of |

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.