K Number
K100221

Validate with FDA (Live)

Device Name
INTELLIWAVE3
Date Cleared
2010-09-14

(231 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IntelliWave3, sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

The IntelliWave", toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated I he correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.

The IntelliWave", multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The IntelliWave", multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The IntelliWave', irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

Device Description

The IntelliWave3 Silicone Hydrogel Soft Contact Lenses are fabricated from efrofilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The non-ionic lens material, (efrofilcon A) is a group 2, daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers. The lenses are made by lathe-cut for custom RX. It consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN).

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (efrofilcon A) soft birghilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

AI/ML Overview

The provided text describes the IntelliWave3 Silicone Hydrogel Daily Wear Soft Contact Lens and its substantial equivalence determination by the FDA. However, it does not include a table of acceptance criteria or a detailed study proving the device meets specific performance criteria in the format requested. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for a new AI/medical device.

Based on the information provided, here's what can be extracted and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

This information is not explicitly provided in the document in the format of specific, quantifiable acceptance criteria. The document states that "Clinical evaluation demonstrated similar overall performance to the predicate control in the clinically relevant areas of vision, health, and comfort and fit when worn for daily wear." This is a general statement of comparative performance rather than a list of specific acceptance metrics with target values and the device's measured performance against them.

The document does list physical properties of the lens:

Physical PropertyReported Performance
Refractive Index1.38
Light Transmissiongreater than 97%
Surface CharacterHydrophilic
Water Content74 %
Specific Gravity1.048 (hydrated)
Oxygen Permeability59.8 x 10^-11 (cm^2/sec)(ml O2/ml x hPa @ 35°C), (revised Fatt method)
Chord Diameter Range12.0 mm to 16.00 mm
Center Thickness Range0.01 mm to 0.50 mm
Base Curve Range8.0 mm to 9.5 mm
Power Range-20.00D to +20.00D in 0.25 step
Cylinder Power (Toric)-0.25D to -12.00D
Cylinder Power (Multifocal Toric)-0.25D to -4.00D
Add Power (Multifocal)+0.50D to +4.00D

The "acceptance criteria" for these physical properties are implicitly that they fall within acceptable ranges for contact lenses and are comparable to predicate devices, but explicit criteria are not given. For clinical performance, the criterion seems to be "similar overall performance" to the predicate in key areas.


2. Sample Size Used for the Test Set and Data Provenance

The document states:

  • Sample Size: "The study evaluated 63 patients over a three (3) months period."
  • Data Provenance: Not explicitly stated (e.g., country of origin). It also does not specify if the data was retrospective or prospective, although the description of a "bilateral, open-label, parallel group, and randomized daily wear study" strongly suggests it was prospective clinical data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. The "ground truth" for clinical performance likely relies on standardized ophthalmic examinations and patient-reported outcomes, but the adjudication process by experts is not detailed.


4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is not provided.


5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

This is not applicable as the device described is a physical medical device (contact lens), not an AI algorithm intended to assist human readers.


6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

This is not applicable as the device is a physical contact lens, not an AI algorithm. Its performance is inherent in its physical and clinical characteristics when worn by a human.


7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical testing, the "ground truth" was established through standard scientific protocols and toxicology studies according to GLP regulations, indicating laboratory measurements and biological response assessments (e.g., sterility, toxicity, irritation).

For the clinical study, the "ground truth" on performance likely involved:

  • Vision: Visual acuity measurements.
  • Health: Ophthalmic examinations by eyecare practitioners (e.g., assessing corneal health, ocular surface).
  • Comfort and Fit: Patient-reported outcomes and clinical assessment of lens fit by eyecare practitioners.

The comparison was to a "predicate control," meaning direct comparison of these clinical outcomes between the new lens and an existing, legally marketed lens.


8. The Sample Size for the Training Set

This is not applicable as the device is a physical medical device and not an AI algorithm that requires a training set.


9. How the Ground Truth for the Training Set Was Established

This is not applicable for the same reason as above.

{0}------------------------------------------------

! ・

510 (k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

K100221

SEP 1 4 2010

Applicant information:

Device

Date Prepared:August 16, 2010
Name:AddressArt Optical Contact Lens, Inc.3175 3 Mile Road NWWalker, Michigan 49534
Contact Person:Mike Johnson, FCLSADirector of Consultative Services
Phone number:(616) 559-5167
Consultant:Martin DalsingMedvice Consulting, Inc.806 Kimball AvenueGrand Junction, CO 81501
Phone number(970) 243.5490
Information:
Device Classification:Class II
Classification Number:LPL
Classification Name:Lenses, Soft Contact. Daily Wear
Trade Name:IntelliWave³, Silicone Hydrogel Daily Wear Soft ContactLens (ofrofilcon A)

{1}------------------------------------------------

K100 221
PAGE 2 OF 6

Equivalent Devices:

The IntelliWave', Silicone Hydrogel Daily Wear Soft Contact Lenses (efrofilcon A) are substantially equivalent to the following predicate devices:

Predicate devices:

"IntelliWavel and Intelliwave2 (acofilcon B, hioxifilcon B)" by Art Optical Contact Lens. Inc.

510(k) number: K073621

"Biofinity (comfilcon A)" by Coopervision, Inc. 510(k) number; K052560

"ActiFresh 400 (lidofilcon A)" By Hydron Ltd. 510(k) number: K983637

Device Description:

The IntelliWave3 Silicone Hydrogel Soft Contact Lenses are fabricated from efrofilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The non-ionic lens material, (efrofilcon A) is a group 2, daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers. The lenses are made by lathe-cut for custom RX. It consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN).

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (efrofilcon A) soft birghilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

{2}------------------------------------------------

K10022
PAGE 3 of 6

The Physical properties of the lens are:

Refractive Index1.38
Light Transmissiongreater than 97%
Surface Characterhydrophilic
Water Content74 %
Specific Gravity1.048 (hydrated)
Oxygen Permeability59.8 x 10-11 (cm2/sec) (ml O2/ml x hPa @ 35°C), (revised Fattmethod).

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 74% water by weight. The lenses will be manufactured in spherical, toric, multifocal, multifocal toric and irregular Cornea configurations with the following features and properties.

• Chord Diameter12.0 mm to 16.00 mm
• Center Thickness0.01 mm to 0.50 mm
• Base Curve8.0 mm to 9.5 mm
• Power Range-20.00D to +20.00D in 0.25 step
• Cylinder Power (Toric)-0.25D to -12.00D
• Cylinder Power (Multifocal Toric)-0.25D to -4.00D
• Add Power (Multifocal)+0.50D to +4.00D

The lens is supplied sterile in vials containing a buffered saline solution. Vial labeling is printed with appropriate lot numbering, expiration dating and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.

Intended Use:

The IntelliWave3, sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

The IntelliWave', toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated I he correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.

{3}------------------------------------------------

The IntelliWave , multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters.

IGE 4 of 6

The IntelliWave", multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The IntelliWave , irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

Testing:

Non-clinical Testing A series of in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the IntelliWave (efrofilcon A) Silicone Hydrogel Soft Contact Lenses packaged in glass vials. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols.

Test results of the non-clinical testing on the IntelliWave3 (efrofilcon A) Silicone Hydrogel Soft Contact Lenses demonstrate that:

  • Lenses supplied in glass vials are sterile for the indicated shelf-life, .
  • The packaging material and extracts are not toxic and not irritating, and .
  • Lens physical and material properties are consistent with currently marketed . lenses.

A bilateral, open-label, parallel group, and randomized daily wear study was Clinical Testing A onlineral, open have, the safety and efficacy of the Intelli Wave³ (efrofilcon A) Silicone Hydrogel Soft Contact Lens. The study evaluated 63 patients over a three (3) months period. Clinical evaluation demonstrated similar overall performance to (2) months por readicate control in the clinically relevant areas of vision, health, and comfort and fit when worn for daily wear.

{4}------------------------------------------------

K100221
PAGE 5 of 6

Substantial Equivalence:

The following matrix illustrates the production method, lens function and material characteristics of the IntelliWave", (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens, as well as the predicate devices.

Conclusions Drawn from Studies

Validity of Scientific Data

Several laboratories under Good Laboratory Practice regulations conducted toxicology studies, Microbiology, chemistry, shelf-life stability studies and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7

Substantial Equivalence

Information presented in this Premarket Notification establishes that the IntelliWave 3, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens is as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the requested indication.

Risks and Benefits

The risks of the subject device are the same as those normally attributed to the wearing of Silicone Hydrogel, Daily Wear Soft Contact Lens. The benefits to the patient are the same as those for other Silicone Hydrogel contact lenses.

{5}------------------------------------------------

IntelliWave3New DeviceArt Optical IntelliWave1 2(acofilcon B)(hioxifilcon B)predicate deviceCooperVision Biofinity(comfilcon A)predicate deviceHydron Ltd. ActiFresh 400(lidofilcon A)predicate device
Intended UseIndicated for daily wear for thecorrection of visual acuity in aphakicand not aphakic persons with non-diseased eyes with myopia orhyperopia. The lens may also beprescribed for management ofirregular corneal conditions such askeratoconus and post graft fitting.Indicated for daily wear for thecorrection of visual acuity in aphakicand not aphakic persons with non-diseased eyes with myopia orhyperopia.Indicated for daily wear for thecorrection of visual acuity in aphakicand not aphakic persons with non-diseased eyes with myopia orhyperopia.Indicated for daily wear for thecorrection of visual acuity in aphakicand not aphakic persons with non-diseased eyes with myopia orhyperopia.
Functionalitysame aspredicate deviceThe contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retina.The contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retina.The contact lenses act as a refractivemedium that focus light rays from nearand distant objects on the retina.
IndicationsDaily wear, Silicone Hydrogel Soft(hydrophilic) contact lensDaily wear, Soft (hydrophilic) contactlensDaily wear, Silicone Hydrogel Soft(hydrophilic) contact lensDaily wear, Soft (hydrophilic) contactlens
ProductionMethodLathe-Cut, custom manufacturedLathe-Cut, custom manufacturedCast Molded, mass producedLathe-Cut, custom manufactured
FDA Group #Group # 11 (Silicone Hydrogel) > 50%Water, Nonionic PolymersGroup # 1 < 50% Water, NonionicPolymersGroup # 1 (Silicone Hydrogel) < 50%Water, Nonionic PolymersGroup # 11 > 50% Water, NonionicPolymers
USAN nameefrofilcon Ahioxifilcon B, acofilcon Bcomfilcon Alidofilcon A
Water Content74%48%48.0%74.0%
OxygenPermeability60 x 10-11 (cm2/sec)(ml O2/ml x mm Hg @ 35 degrees C),(revised Fatt method).15.89 x 10-11 (cm2/sec)(ml O2/ml x mm Hg @ 35 degrees C),(revised Fatt method).128 x 10-11 (cm2/sec)(ml O2/ml x mm Hg @ 35 degrees C),(revised Fatt method).28 x 10-11 (cm2/sec)(ml O2/ml x mm Hg @ 35 degrees C),(revised Fatt method).
Specific Gravity1.1391.1421.1421.060

ــ

_. _. _. ... .................................................................................................................................................................

GE 6 of

  • K10022-1
    PAGE 6 OF 6

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Art Optical Contact Lens, Inc. c/o Medvice Consulting, Inc. Mr. Martin Dalsing Official Correspondent 806 Kimball Avenue Grand Junction, CO 81501

Re: K100221

Trade/Device Name: IntelliWave Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A)

Regulation Number: 21 CFR 886,5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: August 16, 2010 "Received: August 19, 2010

Dear Mr. Dalsing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

.

{7}------------------------------------------------

Page 2 - Mr. Martin Dalsing

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

SEP 1 4 2010

INDICATIONS FOR USE STATEMENT

IntelliWave3, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) Device Name:

INDICATIONS FOR USE:

The IntelliWave', sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

The IntelliWave", toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.

The IntelliWave", multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The IntelliWave ; multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The IntelliWave', irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)
------------------------------------------------------------------------------------------

or
(Optional Format 1-2-96)

Over-The-Counter Use____________________
--------------------------------------------

Marc Robboy
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,Nose and Throat Devices

510(k) Number K100221

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.