K Number
K121428

Validate with FDA (Live)

Date Cleared
2012-07-17

(64 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Non-contact Clinical Thermometer, Model THB0F is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.

Device Description

The RII Non-contact Clinical Thermometer, Model THBOF is electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from the Forehead. Its operation is based on measuring the natural thermal radiation emanating from the temples to hairline upper eyebrows. The thermometer consists mainly of the following parts: an IR sensor with a built-in ambient temperature sensor, a thermo-mass, and the associated circuit. The forehead skin, which is thin and beneath, which is the temporal artery, with blood very near to the core temperature. The forehead skin must cover the whole FOV(Field of View) of the sensor. When measuring, radiation fluxes transfer from the skin to the IR detector. The ambient sensor is mounted in the IR sensor package, and near the IR detector element to monitor the ambient temperature. To measure core temperature, the probe of the thermometer are slide through the skin of the a patient's forehead. A activation button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appear on a LCD display. The total operation takes a few seconds.

AI/ML Overview

The Radiant Innovation Inc. Non-contact Clinical Thermometer, Model THB0F, is an infrared thermometer intended for intermittent measurement of human body temperature in people of all ages.

Here's an analysis of its acceptance criteria and the study that proves its performance:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicAcceptance Criteria (Forehead Mode)Reported Device Performance (Forehead Mode)
Accuracy±0.4°F (0.2°C) within 96.8°F to 102°F (36°C to 39°C)Clinical investigation report states that temperature readings difference between digital thermometer and the subject device (THB0F) are within acceptable range, demonstrating the device meets the ASTM E 1965-98 standard for accuracy.
±0.5°F (0.3°C) for other ranges
Measuring Range93.2°F to 108°F (34°C to 42.2°C)93.2°F to 108°F (34°C to 42.2°C)
Display ResolutionNot explicitly stated as an acceptance criterion, but the predicate has 0.1°F (0.1°C)0.1°F (0.1°C)

Additional Acceptance Criteria (General Safety and Standards Compliance):

  • Compliance to applicable voluntary standard ASTM E1965-98: Met.
  • Compliance to recognized consensus standards for safety of medical electrical equipment: Met (EN 60601-1 (IEC 60601-1) for safety and EN 60601-1-2 (IEC 60601-1-2) for electromagnetic compatibility).
  • Compliance to FDA Guidance On The Content of Premarket Notification 510(k) Submissions for Clinical Electronic Thermometers: Met.

2. Sample Size Used for the Test Set and Data Provenance:

The document states, "The clinical investigation report and data analysis is followed the requirements the ASTM E 1965-98." However, it does not explicitly specify the sample size used for the clinical test set or the country of origin of the data. It also does not explicitly state whether the study was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

The document does not specify the number of experts used to establish the ground truth for the clinical test set, nor does it provide their qualifications. The clinical test compares the THB0F's readings against a "digital thermometer." It implies the "digital thermometer" serves as the reference or "ground truth" for the comparison.

4. Adjudication Method for the Test Set:

The document does not describe an adjudication method for the test set. The clinical test appears to be a direct comparison of temperature readings between the subject device and a digital thermometer.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. This device is a standalone thermometer, not an AI assistance tool for human readers.

6. Standalone (Algorithm Only) Performance:

Yes, the primary clinical test performed was a standalone performance study for the device. The study evaluated the accuracy of the THB0Fometer in measuring human body temperature independently. The document states: "The test report shows the three group's temperature readings difference between digital thermometer and the subject device, THB0F are within acceptable range."

7. Type of Ground Truth Used:

The ground truth used for the clinical test was based on measurements from a "digital thermometer." This implies a comparison against an established, accurate clinical thermometer.

8. Sample Size for the Training Set:

The document does not describe a training set for an algorithm, as this device is a hardware clinical thermometer dependent on an infrared sensor and associated circuitry, not a machine learning model.

9. How the Ground Truth for the Training Set Was Established:

As there is no mention of a training set or an algorithm in the context of machine learning, the question of how its ground truth was established is not applicable. The device's performance is validated through physical and clinical testing against established standards and other measurement devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "RADIANT" in large, bold letters. Above the word "RADIANT" is the word "INNOVATION" in smaller letters. Below the word "RADIANT" are four Chinese characters.

EXHIBIT # 1

510(K) SUMMARY

JUL 1 7 2012

This 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: 12 14 28

1. Submitter's Identification:

Radiant Innovation Inc., 1F, No.3 Industrial E. 9th Rd., Science-Based Industrial Park, HsinChu, Taiwan

Contact:

Ms. Monica Chung Product Certification Engineer Radiant Innovation Inc. TEL: +886 3 6111666 Ext. 8128 FAX:+886 3 5670089 E-mail: monicachung@radiantek.com.tw

Date Summary Prepared: May/11/2012

2. Name of the Device:

Non-contact Clinical Thermometer, Model THBOF. Classification Name: Thermometer, Electronic, Clinical Regulatory Class: II Product Code: FLL Regulation Number: 21 CFR 880.2910

3. Predicate Device :

TECNIMED srl, Thermofocus 01500 series Thermometer.(510(k) #K072108)

4. Device Description:

The RII Non-contact Clinical Thermometer, Model THBOF is electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from the Forehead. Its operation is based on measuring the natural thermal radiation emanating from the temples to hairline upper eyebrows.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Radiant Innovation Inc. The logo is divided into two parts, with the left side featuring the word "RADIANT" in bold, stylized letters and the word "INNOVATION" above it in smaller letters. The right side of the logo contains the company name in Chinese characters, followed by "RADIANT INNOVATION Inc." in English.

The thermometer consists mainly of the following parts: an IR sensor with a built-in ambient temperature sensor, a thermo-mass, and the associated circuit. The forehead skin, which is thin and beneath, which is the temporal artery, with blood very near to the core temperature. The forehead skin must cover the whole FOV(Field of View) of the sensor. When measuring, radiation fluxes transfer from the skin to the IR detector. The ambient sensor is mounted in the IR sensor package, and near the IR detector element to monitor the ambient temperature.

To measure core temperature, the probe of the thermometer are slide through the skin of the a patient's forehead. A activation button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appear on a LCD display. The total operation takes a few seconds.

క. Intended Use:

The Non-contact Clinical Thermometer, Model THB0F is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.

Technological Characteristics and Substantial Equivalence: 6.

The subject device is substantially equivalent to the predicate devices. Thermofocus 01500 series Thermometer (K072108) marketed by TECNIMED srl. The substantial equivalence chart is provided as follows:

CharacteristicsSubject device(THB0F)Predicate device (THP series)
510(k)#KK072108
Measurement MethodInfrared radiation detection
Measuring RangeForehead mode:93.2108°F(3442.2℃)Surface mode:-7.6176°F(-2280℃)Forehead mode:93.2108.5°F(3442.5℃)Surface mode:33.8131°F (155℃ )
AccuracyForehead mode:±0.4°F(0.2°C) within 96.8102°F(3639℃), ±0.5°F(0.3°C) forother rangeSurface mode:±4% of reading or ±4°F(2°C)whichever is greaterForehead mode:±0.4°F(0.2°C) within 96.8102°F(3639°C),±0.5°F(0.3°C ) for other rangeSurface mode:±0.4°F(0.2°C) within 96.9102°F(3639°C),± 0.5°F(0.3℃) within 6896.7°F(2035.9℃)and 102.3108.5°F(39.142.5°C ),± 1.8°F(1°C) within 33.867.9°F(119.9℃)and 108.1131°F(42.655°C )
Display Resolution0.1°F(0.1°C)
Measurement Distance2~3cm2.5cm
Scale Selection°F/°C
Display TypeLCD
Key4 button(Mode/ Memory, On/Off,Light/Set, Scan)3 button(Face, Memory, Home)
Memory60 sets9 sets
Sensor TypeThermopile
CaseABS

Page 2

1F, No.3, Industrial E. 9th Rd., Science-Based TEL: +886 3 5644185 Industrial Park, HsinChu, Taiwan 300, R.O.C. FAX: +886 3 5644170

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains two logos side by side. The logo on the left says "RADIANT INNOVATION" in a stylized font. The logo on the right contains Chinese characters and the text "RADIANT INNOVATION Inc.,".

Weight104.7g90g
Dimension (LxWxH)180.3x47.5x29.2mm165x40x21.9 mm
Power SourceAAA(1.5V)*2AAA(1.5V)*4
Operating Condition50104°F(1040°C)60.8104°F(1640°C)

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: .

  • · Compliance to applicable voluntary standard ASTM E1965-98.
  • · The recognized consensus standards for safety of medical electrical equipment: EN 60601-1 (IEC 60601-1) for safety and EN 60601-1-2 (IEC 60601-1-2) for electromagnetic compatibility are complied.
  • · Guidance Documents included the "FDA Guidance On The Content of Premarket Notification 510(k) Submissions for Clinical Electronic Thermometers".

8. Discussion of Clinical Tests Performed:

The clinical investigation report and data analysis is followed the requirements the ASTM E 1965-98. The test report shows the three group's temperature readings difference between digital thermometer and the subject device, THBOF are within acceptable range. It can conclude that the Non-contact Clinical Thermometer, Model THBOF is acceptable to measure human body's temperature.

9. Conclusions:

The RII Non-contact Clinical Thermometer, Model THB0F, has the same intended use and similar characteristics as the predicate device. Moreover, the subject device demonstrates product safety by successful completion of testing to the EN 60601-1(IEC 60601-1) standard and electromagnetic standard EN 60601-1-2(IEC 60601-1-2). The performance test demonstrates the THB0F meets the ASTM E1965-98 standard and concludes that any differences in their characteristics do not raise any safety and effectiveness issues. Thus, the RII Non-contact Clinical Thermometer, Model THBOF is substantially equivalent to the predicate devic.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three wavy lines extending from its body, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL 1 7 2012

Mr. James Huang Quality Assurance Radiant Innovation, Incorporated 1F, No.3 Industrial East 9th Road Science-Based, Industrial Park HsinChu, Taiwan 30075

Re: K121428

Trade/Device Name: Non-Contact Clinical Thermometer, Model THB0F Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: May, 14, 2012 Received: May 14, 2012

Dear Mr. Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Huang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Fr. A. Betti, S.J.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

EXHIBIT # B

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Radiant Innovation Inc. Non-contact Clinical Thermometer, Model THB0F

Indications For Use:

:

The Non-contact Clinical Thermometer, Model THBOF is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.

Prescription Use (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use ਕਿ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off)7/9/12
Division of Anesthesiology, General HospitalInfection Control, Dental DevicesPage 1 of 1
510(k) Number:K121428

Page 12300 新竹科學園區工業東九路 3 號 1 樓 http://www.radiantek.com.tw

1F, No.3, Industrial E. 9" Rd., Science-Based TEL: +886 3 5644185 Industrial Park, HsinChu, Taiwan 300, R.O.C. FAX: +886 3 5644170

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.