K Number
K121352

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-08-15

(103 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 5.0 mm balloon diameters are also indicated for post-dilatation of balloon-expandable stents.

Device Description

The Armada 14 XT PTA Catheter is an over-the-wire co-axial design with a balloon at the distal tip. The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The co-axial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device.

AI/ML Overview

The provided text describes a 510(k) summary for the Armada 14 XT PTA Catheter. This document primarily outlines the device's technical characteristics, its substantial equivalence to predicate devices, and the types of performance testing conducted to support this claim. It does not present a study that establishes acceptance criteria and then proves the device meets those criteria in the way a clinical study or a standalone AI performance study would.

Instead, this is a regulatory submission demonstrating substantial equivalence for a medical device (a PTA Catheter) based on bench testing and biocompatibility testing, comparing it to already cleared predicate devices. The "acceptance criteria" here refer to internal engineering specifications and performance benchmarks derived from the predicate devices and general standards for such catheters, not clinical performance metrics in a patient population or AI model performance metrics.

Therefore, many of the requested sections (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone AI performance, training set details) are not applicable to this type of document because it describes a physical medical device submission, not an AI/ML-driven diagnostic or assistive device that would involve such studies.

Here's a breakdown of what can be extracted and what is not applicable:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategoryReported Device Performance
Mechanical Performance- Balloon Crossing Profile met
- Refolded Balloon Profile met
- Guidewire Lumen ID met
- Total Catheter Length met
- Minimum Balloon Burst Strength (RBP) met
- Balloon Compliance (Diameter vs. Pressure) met
- Balloon Inflation/Deflation Time met
- Balloon Fatigue (Repeat Balloon Inflations) met
- Tensile strength - Proximal Adaptation met
- Tensile strength - Proximal Balloon Seal met
- Tensile strength Distal Outer Member to Proximal Shaft met
- Tensile strength Soft Tip to Inner Member Seal met
- Catheter Flexibility & Kink Test met
- Torque Strength met
- Radiopacity met
- Minimum Balloon Burst Strength (RBP) Within Stent met
- Balloon Fatigue Within Stent (Repeat Balloon Inflations in Stent) met
Durability/Stability- Shelf Life (Accelerated Aging) met
Biocompatibility- Cytotoxicity met
- Sensitization met
- Intracutaneous Reactivity met
- Acute Systemic Toxicity met
- Hemolysis met
- Pyrogen met
- Complement Activation met
Conclusion"The in vitro bench and biocompatibility tests demonstrated that the Armada 14 XT PTA Catheters met all acceptance criteria and performed similarly to the predicate devices."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not explicitly stated as numbers of patients or images. The "test set" here refers to physical catheter samples tested in a laboratory setting. The typical sample sizes for such bench tests are defined by internal quality control procedures and relevant ASTM/ISO standards, not usually disclosed in a 510(k) summary in terms of a "study sample size" for performance metrics like sensitivity/specificity.
  • Data Provenance: Not applicable. This is laboratory bench testing and biocompatibility testing of a manufactured medical device, not a study involving patient data from a specific country or setting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. Ground truth, in the context of this device, would be based on established engineering principles, material science, and regulatory standards for mechanical and biological performance. There are no "experts" establishing ground truth in the sense of medical diagnosis or image interpretation for this type of submission. Performance is measured against physical specifications and validated test methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. Adjudication methods are typically used in clinical studies or studies involving human interpretation of data where consensus is needed. This is a bench study of a physical device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. An MRMC study is not relevant to a physical PTA catheter and was not performed. This type of study is for evaluating diagnostic performance, often with AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This device is a physical medical instrument, not an algorithm. Standalone performance for an algorithm is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for the performance tests listed (e.g., Minimum Balloon Burst Strength, Balloon Compliance) is the pre-defined engineering specifications and regulatory limits derived from predicate devices, recognized standards (e.g., ISO, ASTM), and the device's intended use. For biocompatibility, the ground truth is the absence of cytotoxicity, irritation, etc., as per ISO 10993 standards.

8. The sample size for the training set

  • Not Applicable. This is a physical device, not an AI model. There is no training set in the machine learning sense. The device "training" would be the engineering design, prototyping, and manufacturing process based on established medical device development principles.

9. How the ground truth for the training set was established

  • Not Applicable. As there is no AI training set, this question is not relevant.

{0}------------------------------------------------

510(k) Summary

. The 510(k) Summary is submitted in accordance with 21 CFR 8807,92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.

AUG 1 5 2012

Submitter's Name Submitter's Address Telephone Fax Contact Person Date Prepared Device Trade Name Device Common Name Device Classification Name Predicate Device Names

Abbott Vascular 3200 Lakeside Drive, Santa Clara, CA 95054 (408) 845-0682 (408) 845-3743 Ivalee Cohen May 2, 2012 Armada 14 XT PTA Catheter PTA Catheter Catheter, angioplasty, peripheral, transluminal, LIT • Abbott Vascular TREK OTW & MINI TREK OTW Coronary Dilatation Catheter (K103011, cleared 1/10/11 and K110617, cleared 6/2/11)

· Abbott Vascular Armada 14 PTA Catheter (K102705, cleared 12/7/10)

Device Description

The Armada 14 XT PTA Catheter is an over-the-wire co-axial design with a balloon at the distal tip. The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The co-axial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device.

Indication for Use

The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 5.0 mm balloon diameters are also indicated for postdilatation of balloon-expandable stents.

{1}------------------------------------------------

Technological Characteristics

The subject Armada 14 XT PTA Catheter has the same fundamental scientific technology (operating principle) and indication for use as the Armada 14 predicate device and shares the same materials, design, labeling, packaging materials, shelf life and sterilization process as the TREK and MINI-TREK predicate devices.

Performance Data

Performance testing was successfully completed on the Armada 14 XT PTA Catheter.

The following tests were conducted:

  • . Balloon Crossing Profile
  • Refolded Balloon Profile .
  • Guide wire Lumen ID .
  • . Total Catheter Length
  • Minimum Balloon Burst Strength (Rated Burst Pressure, RBP) .
  • Balloon Compliance (Diameter vs. Pressure) .
  • . Balloon Inflation/ Deflation Time
  • Balloon Fatigue (Repeat Balloon Inflations) ●
  • . Tensile strength - Proximal Adaptation
  • . Tensile strength - Proximal Balloon Seal
  • Tensile strength Distal Outer Member to Proximal Shaft .
  • Tensile strength Soft Tip to Inner Member Seal ●
  • Catheter Flexibility & Kink Test
  • Torque Strength .
  • . Radiopacity
  • Minimum Balloon Burst Strength (RBP) Within Stent �
  • Balloon Fatigue Within Stent (Repeat Balloon Inflations in Stent) ●
  • Shelf Life (Accelerated Aging)

Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation.

Conclusion

The in vitro bench and biocompatibility tests demonstrated that the Armada 14 XT PTA Catheters met all acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing program and, therefore, these devices may be considered substantially equivalent to the predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 15 2012

Abbott Vascular c/o Ivalee Cohen 3200 Lakeside Drive Santa Clara, CA 95054

Re: K121352

Trade/Device Name: Armada 14 XT PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 02, 2012 Received: May 04, 2012

Dear Ms. Cohen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a provisions of the Federal rood, Drily, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate informations against missianuly and warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Ivalee Cohen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of modical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electionic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118001,htmfor for for for for for for for for for for for for for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, Jease note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

.s.kiel

Go Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

K121352 510(k) Number (if known):

Device Names:

Armada 14 XT PTA Catheter

Indications for Use:

The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 5.0 mm balloon diameters are also indicated for post-dilatation of balloon-expandable stents.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Cuañillele

(Division Sign-Off) Division of Cardiovascular Devices

Page 1 of

化1252 510(k) Number_

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).