K Number
K120858

Validate with FDA (Live)

Device Name
IMPLANT BAR
Date Cleared
2012-12-06

(260 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Inclusive® Implant Bridge Framework is indicated for use as a bridge framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.

The Inclusive® Implant Bridge Framework is compatible with the following implant system: Nobel Biocare's Replace 3.5mm, 4.3mm, 5.0mm, and 6.0mm; Nobel Biocare's Branemark System 4.1mm; Nobel Biocare's Active 3.5mm and 4.3mm, Zimmer Dental Screw-Vent 3.5mm, 4.5mm, and 5.7mm; Biomet 3i Certain 3.4mm, 4.1mm, 5.0mm, and 6.0mm; Straumann Bone Level 4.1mm; Astra Tech OsseoSpeed 3.5mm and 4.5mm.

Device Description

Inclusive® Implant Bridge Framework attaches to implants. The Implant Bridge Framework is intended to be finished into a dental prosthesis using standard laboratory materials. The Inclusive® Implant Bridge Framework is customized by following instructions and models specific to each patient.

The Inclusive® Implant Bridge Framework is made of titanium and shipped non-sterile. The Inclusive® Implant Bridge Framework will be attached with titanium screws.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Inclusive® Implant Bridge Framework, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (Implicitly based on predicate device and guidance)Reported Device Performance
MaterialTi-6Al-4V ELI Alloy conforming to ASTM F 136.Ti-6Al-4V ELI Alloy conforming to ASTM F 136.
Indications for UseFor use as a bridge framework in the treatment of partially or totally edentulous jaws for restoring chewing function.The device meets this indication as stated.
DesignPrecision-milled bars made of titanium; comparable to predicate devices.Precision-milled bars made of titanium.
Mechanical PerformanceSufficient mechanical strength for intended use (comparable to predicate devices and ISO 14801:2007).Titanium materials have sufficient mechanical strength for their indicated use. All testing conducted met the acceptance criteria and evaluated the worst-case scenario. Performance testing data indicated the compatibility, and the safety and the effectiveness of the proposed device which meets the mechanical properties.
SterilizationSterilization validation according to ANSI-AAMI ST79-2006.Sterilization tests were validated by following the ANSI-AAMI ST79-2006.
Implant CompatibilityCompatibility with specified implant systems (Nobel Biocare, Zimmer Dental, Biomet 3i, Straumann, Astra Tech).Performance testing data indicated compatibility with listed implant systems.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state a sample size in terms of number of manufactured devices or clinical cases used for the performance testing. It refers to "various static and fatigue tests."
  • Data Provenance: The document does not specify the country of origin for the data or whether the tests were retrospective or prospective. The tests were laboratory-based (material and mechanical testing).

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • This information is not applicable as the study described is a laboratory-based mechanical and material performance study, not a clinical study involving expert interpretation of data or images. Ground truth for material and mechanical properties is established through adherence to international standards (e.g., ASTM F136, ISO 14801:2007).

4. Adjudication Method for the Test Set

  • This information is not applicable as the study described is a laboratory-based mechanical and material performance study. Adjudication methods are typically used in clinical studies or studies involving human assessment.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. The document describes laboratory performance testing only. Therefore, there is no effect size of human readers improving with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • Not applicable. The device is a physical dental implant bridge framework, not an algorithm or AI software. Therefore, there is no "standalone algorithm" performance to report.

7. Type of Ground Truth Used

  • The ground truth used for this device's performance assessment is based on:
    • International Standards: Adherence to material specifications (ASTM F 136 for Ti-6Al-4V ELI Alloy) and mechanical testing standards (ISO 14801:2007 for dynamic fatigue testing).
    • Guidance Documents: Following the FDA's "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" for implant to abutment compatibility.
    • Validated Sterilization Protocols: Adherence to ANSI-AAMI ST79-2006 for sterilization.

8. Sample Size for the Training Set

  • This information is not applicable. As the device is a physical medical device and the study is for performance testing, there is no "training set" in the context of machine learning or AI models. The "training" for the manufacturing process would be established through quality control and process validation, which is not detailed in terms of sample size in this summary.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable for the same reasons as #8. Ground truth for manufacturing processes is established through engineering specifications, validated manufacturing procedures, and quality control checks, not through a "training set" in the AI sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for PRISMATIK DENTALCRAFT, INC. The logo consists of a black triangle on the left and the company name on the right. The company name is written in bold, black letters.

K120858

005-510 (k) Summary-807.92(c)

DEC 0 6 2012

This 510 (K) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.

SUMITTER INFORMATION A.

Company Name:

Company Address:

Prismatik Dentalcraft, Inc.

2212 Dupont Dr., Suite IJK Irvine, CA 92612

Company Phone:

Company FAX:

Primary Contact Person:

Secondary Contact Person:

Date Summary Prepared:

949-399-1940

949-553-0924

Armin Zehtabchi. (949) 225-1234

Marilyn Pourazar, (949) 225-1269

November 28, 2012

B. DEVICE IDENTIFICATION

Trade/Proprietary Name:

21 CFR Common Name:

21 CFR Reference:

Classification:

Product Code:

Panel:

Inclusive® Implant Bridge Framework

21 CFR 872.3630

Endosseous Dental Implant Abutment

Class II

NHA

Dental

C. IDENTIFICATION OF PREDICATE DEVICE

Trade/Proprietary Name:.

Nobel /Procera Implant Bridge-K091848, Biomet 3i-CAM StructSure Overdenture Bars-K101582

DEVICE DESCRIPTION D.

Inclusive® Implant Bridge Framework attaches to implants. The Implant Bridge Framework is intended to be finished into a dental prosthesis using standard laboratory materials. The Inclusive® Implant Bridge Framework is customized by

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a black triangle on the left and the company name on the right. The company name is written in a bold, sans-serif font.

following instructions and models specific to each patient.

The Inclusive® Implant Bridge Framework is made of titanium and shipped non-sterile. The Inclusive® Implant Bridge Framework will be attached with titanium screws.

E. INDICATION FOR USE

Indications for Use: The Inclusive® Implant Bridge Framework is indicated for use as a bridge framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.

The Inclusive® Implant Bridge Framework is compatible with the following implant system: Nobel Biocare's Replace 3.5mm, 4.3mm, 5.0mm, and 6.0mm; Nobel Biocare's Branemark System 4.1mm; Nobel Biocare's Active 3.5mm and 4.3mm, Zimmer Dental Screw-Vent 3.5mm, 4.5mm, and 5.7mm; Biomet 3i Certain 3.4mm, 4.1mm, 5.0mm, and 6.0mm; Straumann Bone Level 4.1mm; Astra Tech OsseoSpeed 3.5mm and 4.5mm.

SUBSTANTIAL EQUIVALENCE F.

The Inclusive® Implant Bridge Framework is substantially equivalent to the Nobel /Procera Implant Bridge and Biomet 3i-CAM StructSure Overdenture Bars. The Inclusive® Implant Bridge Framework is substantially equivalent in indications for use, material, design and performance.

Page 13 of 27

{2}------------------------------------------------

PRISMATIK DENTALCRAFT, INC.
Callery Company of September 1998 to the states of the states of the states of the states of the states of the states of the states of the states of the states of the states1

Comparison of Predicate Devices

. .

.

Elements of ComparisonPrismatik's Inclusive® Implant Bridge FrameworkNobel /Procera Implant Bridge-K091848Biomet 3i's CAM StructSure Overdenture Bars-K101582Prismatik's Inclusive Mini Implant-K100932Biomet 3i's CAM StructSure Precision Milled Bars-K080864
MaterialTi-6Al-4V ELI Alloy conforming to ASTM F 136.Ti-6Al-4V ELI Alloy conforming to ASTM F 136.Ti-6Al-4V ELI Alloy conforming to ASTM F 136.Ti-6Al-4V ELI Alloy conforming to ASTM F 136.Ti-6Al-4V ELI Alloy conforming to ASTM F 136.
IndicationsThe Inclusive® Implant Bridge Framework is indicated for use as a bridge framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.The Nobel Procera Implant Bridge is indicated for use as a bridge framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.The 3i Patient-Specific CAM StructSure® Precision Milled Bars are intended for use as an accessory to endosseous dental implant to support a prosthetic device in a partially or edentulous patient.Inclusive® Mini Implants are self-tapping threaded titanium screws indicated for long-term applications. Inclusive® Mini Implants may also be used for provisional applications. These devices will allow immediate loading and long-term stabilization of dentures and provisional stabilization of dentures while standard implants heal.Jo be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.The CAM StructSURE® Precision Milled Bars are intended for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient. It is intended for use to support multiple tooth prostheses in the mandible or maxilla. The prostheses can be screw retained.
DesignPrecision-milled bars made of titanium.Precision-milled bars made of titanium. Design features and technological characteristics have been previously cleared for market (K091848).Precision-milled bars made of titanium. Design features and technological characteristics have been previously cleared for market (K101582).Precision-milled bars made of titanium.Precision-milled bars made of titanium.
PerformanceThe titanium materials have sufficient mechanical strength for their indicated use.Performance and technological characteristics have been previously cleared for market (K091848).Performance and technological characteristics have been previously cleared for market (K101582).The titanium materials have sufficient mechanical strength for their indicated use.Performance and technological characteristics have been previously cleared for market (K0808864).

Page 14 of 27

:

/

{3}------------------------------------------------

PERFORMANCE DATA G.

The following FDA's Guidance Document "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments was used for the purpose of Implant to Abutment Compatibility. Various static and fatigue tests were performed by following the ISO 14801: 2007- Dentistry - Implants -- Dynamic fatigue test for endosseous dental implants. All testing conducted met the acceptance criteria and evaluated the worst case scenario. Performance testing data indicated the compatibility, and the safety and the effectiveness of the proposed device which meets the mechanical properties. In addition, the sterilization tests were validated by following the ANSI-AAMI ST79-2006: Comprehensive guide to steam sterilization and sterility assurance in health care facilities. Furthermore, the use of Titanium alloy as a material with acceptable performance for Inclusive® Implant Bridge Framework is well documented in the dental literature.

COMPARISON OF TECHNOLOGICAL DIFFERENCES H.

There are no known technological differences between the Inclusive® Implant Bridge Framework and those of the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles connected by flowing lines, representing the department's focus on health and human well-being.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

December 6, 2012

Ms. Kathleen Dragovich Manager, Regulatory Affairs / Quality Assurance Prismatik Dentalcraft, Incorporated 2212 Dupont Drive, Suite P Irvine, California 92612

Re: K120858

Trade/Device Name: Inclusive® Implant Bridge Framework Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: November 16, 2012 Received: November 19, 2012

Dear Ms. Dragovich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kwame O. Ulmer for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

004-Indications for Use Statement

510 (K) Number (if known): To be determined

Inclusive® Implant Bridge Framework Device Name:

Indications for Use: The Inclusive® Implant Bridge Framework is indicated for use as a bridge framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.

The Inclusive® Implant Bridge Framework is compatible with the following implant system: Nobel Biocare's Replace 3.5mm, 4.3mm, and 6.0mm; Nobel Biocare's Branemark System 4.1mm; Nobel Biocare's Active 3.5mm and 4.3mm, Zimmer Dental Screw-Vent 3.5mm, 4.5mm, and 5.7mm; Biomet 3i Certain 3.4mm, 4.1mm, 5.0mm, and 6.0mm; Straumann Bone Level 4.1mm; Astra Tech OsseoSpeed 3.5mm and 4.5mm.

Prescription Use: Yes X No

Over-the-Counter Use: Yes |_ No X

(Part 21 CFR 801 Subpart D)

(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

2012.12.06 13:40:31 Susan Runner DDS, MA 2012.1

(Division Sign-Off)
Division of Anesthesiology, General Hospita,
Infection Control, Dental Devices

510(k) Number:K120 858
--------------------------

Page 6 of 12

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)