K Number
K120530

Validate with FDA (Live)

Date Cleared
2012-07-05

(134 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DENTIN® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. DENTIN® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Two Stage Implants: CLASSIC, RAPID, PRESTIGE. One Stage Implants: ONE PIECE DENTIN® ONEPIECE Implants 3.0 mmd are intended for placement at the mandibular central and lateral incisors and maxillary and lateral incisors. Indicated also for denture stabilization using multiple implants.

Device Description

DENTIN® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants system; cover screws and healing caps; abutment systems and superstructures; impression copy system & surgical instruments.

AI/ML Overview

This FDA submission for the DENTIN® Dental Implants System (K120530) does not describe a clinical study in the traditional sense, but rather a demonstration of substantial equivalence to predicate devices through various tests and comparisons. Therefore, the concept of "acceptance criteria" is tied to meeting the requirements for substantial equivalence, and "device performance" is demonstrated through compliance with established standards and comparison to predicate devices.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySource/StandardDevice Performance (DENTIN® Dental Implants System)
Intended Use / Indication for UsePredicate Devices (K063364, K003191, K103089, K080162, K023113, K062566, K081396, K090347)"substantially equivalent in terms of intended use, indication for use" to predicate devices. Indicated for surgical and restorative applications in the upper or lower jaw to support prosthetic devices to restore chewing function, including immediate loading when primary stability is achieved. 3.0 mmd ONEPIECE implants for mandibular/maxillary central/lateral incisors and denture stabilization.
Technological Characteristics / DesignPredicate Devices"substantially equivalent in terms of... technological characteristics, performance and user interface." Shares similar design (e.g., tapered implant body contour, self-tapping, internal hex/one-piece options) with predicate devices.
Material (Implants)ISO 5832-3:1996, Predicate DevicesGR-5 Titanium (Ti-6Al-4V ELI) used, which is the same as most predicate devices. Noted difference from NOBELREPLACE (CP4 Titanium) but stated "raises no new issues of safety or effectiveness."
BiocompatibilityISO 7405:2008Not explicitly reported as "test results" in the same way as mechanical tests, but compliance with ISO 7405 indicates biocompatibility has been evaluated. The overall claim of no new safety/effectiveness issues implies acceptable biocompatibility.
Mechanical Strength (Fatigue - Static & Cycling)ISO 14801:2007"conducted Fatigue - Static & Cycling tests which comply with ISO 14801... The test results have demonstrated the high resistance and high ability with the use of DENTIN Dental Implant System. Therefore, DENTIN® Dental Implants System raises no new issues of safety or effectiveness than the predicate devices."
SterilizationNot explicitly stated standard, but implied industry standards for medical devices"sterilization tests... were conducted in order to ensure safety and effectiveness... Test results have demonstrated that the SAL of 10-5 was achieved and all testing requirements were met."
Shelf LifeNot explicitly stated standard, but implied industry standards for medical devices"shelf life testing were conducted in order to ensure safety and effectiveness... Test results have demonstrated that... all testing requirements were met." (for shelf life, specifically)
Risk AssessmentFDA Guidance Document (Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments)"Risk Assessment was conducted and has demonstrated no new safety and/or effectiveness issues than the predicate devices."

2. Sample Size for the Test Set and Data Provenance

The submission does not describe a test set in the context of clinical data for a performance study. It relies on bench testing (mechanical, sterilization, shelf-life) to demonstrate compliance with standards and comparison to legally marketed predicate devices for clinical equivalence.

  • Bench Testing: The sample sizes for the mechanical, sterilization, and shelf-life tests are not specified in the provided document.
  • Data Provenance: Not applicable as no clinical test set data is provided. The tests mentioned are laboratory/bench tests.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. This submission relies on substantial equivalence to predicate devices and adherence to recognized performance standards, not on expert-adjudicated clinical ground truth from a new study.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set with human data requiring adjudication is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This type of study (comparing human readers with and without AI assistance) is not mentioned or relevant to this 510(k) submission for dental implants.

6. Standalone (Algorithm Only) Performance Study

No. This is not an AI/algorithm-based device. The device is a physical dental implant system.

7. Type of Ground Truth Used

The "ground truth" for this submission is established through:

  • Compliance with recognized standards: ISO 14801 (mechanical fatigue), ISO 7405 (biocompatibility), ISO 5832-3 (material).
  • Demonstration of equivalence to predicate devices: The established safety and effectiveness of the legally marketed predicate dental implant systems (Alpha-Bio Tec, MIS Implant Technologies, NOBELREPLACE, Implant Direct SwissPlant, CAMLOG) serve as the clinical "ground truth" for the new DENTIN® system. The submission argues that DENTIN® presents no new safety or effectiveness issues compared to these devices.
  • Laboratory test results: Achieving a SAL of 10-5 for sterilization and meeting shelf-life requirements.

8. Sample Size for the Training Set

Not applicable. There is no "training set" as this is not an AI/machine learning device.

9. How the Ground Truth for the Training Set Was Established

Not applicable.

{0}------------------------------------------------

K120530

DICAL REGISTRATION

ס״ר

5 2012 JUL

Page 1 of

New Device

. ..

. :

SECTION 5 ... - . . 510(K) Summary (ST CER 807.92)

510(k) Number K 1 Submission Owner DENTIN® Implants Technologies Ltd. Shai Hagbi - CEO Migdal Tefen Industrial Park, P.O. Box 10, ZIP 24959 ISRAEL Phone: 972-77-4408470 972-4-9941011 Fax: Sterling Medical Registration Official Correspondent 2 Daniela Levy - Regulatory Consultant Contact Person 22817 Ventura blvd. #161 Woodland Hills, CA 91364 Phone: 1-213-787-3026 1-818-456-4222 Fax: Web www.sterlingmedicalregistration.com 3 Submission Date February 2012 Device Trade Name 4 DENTIN® Dental Implants System Regulation Description 5 Root-form Endosseous Dental Implants and Abutments 6 Classification Device Name : Implant, endosseous, rootform Product Code DZE . Subsequent product code: NHA Regulation No 872.3640 ll Class Panel Dental

1

Reason for the Premarket Notification Submission Identification of Legally Marketed Predicate Devices : 8

7

{1}------------------------------------------------

DICAL REGISTRATION

DENTIN® Dental Implants System is substantially equivalent to Alpha-Bio Tec K063364: MIS Implant Technologies Ltd K003191, K103089, K080162; NOBELREPLACE K023113, K062566; Implant Direct SwissPlant K081396 in terms of intended use, indication for use, technological characteristics, performance and user interface.

DENTIN® PEEK Abutments are substantially equivalent to; CAMLOG® Abutments PS K090347 in terms of intended use, indication for use, technological characteristics, performance and user interface.

The predicate devices are a Class II medical device.

  • Device Description: 9
    DENTIN® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants system; cover screws and healing caps; abutment systems and superstructures; impression copy system & surgical instruments.

Two stage, Internal hex implants:-

CLASSIC implants are provided in diameters: 3.3, 3.75, 4.2, 5 & 6 with lengths, 7 (only to 5&6 dm), 8, 10, 11.5, 13, & 16 (only to 3.3, 3.75&4.2 dm)

RAPID implants are provided in diameters: 3.3, 3.75, 4.2, 5 & 6 with lengths 7 (only to 5&6 dm) 8, 10, 11.5, 13, & 16 (only to 3. .3, 3.75&4.2 dm)

PRESTIGE implants are provided in diameters: 3.75, 4.2, 5 & 6, with lengths 7 (only to 5&6 dm), 8, 10, 11.5, 13, & 16 (only to 3.75&4.2 dm)

One stage, one piece implants:-

ONEPIECE implants are provided in diameters 3.0, 3.3 with lengths 10, 11.5, 13, & 16

Healing caps are available in 3 sizes: Standard, Narrow and Wide with Heights: 1, 2, 3, 4, 5, 6 and 7 mm.

DENTIN Abutments system provides: Ball abutment (angulated), Titanium abutment (slim, straight, angulated), Anatomic titanium abutment (straight, angled), Leaf titanium abutment, Plastic abutment (wide, direct), Titanium castable abutment, Esthetic connection abutment, Angular curve (narrow three, regular three, wide), Anqular smooth (minor, thin, wide), Ball minor (angular, straight), Aesthetic connection abutment (minor regular, multi, wide), Angular multi unit, Gauge angle, Immediate temporary conical, Straight curve (narrow three, regular

{2}------------------------------------------------

ק״ד

FREI MEDICAL REGISTRATION

three, wide), Straight (level minor, slim minor), Wise click (angular multi, connection, minor).

Impression copy system consists of: Transfers, Analogs & Accessories and Transfers.

10 Intended use / Indication for Use:

DENTIN® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. DENTIN® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Two Stage Implants: CLASSIC, RAPID, PRESTIGE.

One Stage Implants: ONE PIECE

DENTIN® ONEPIECE Implants 3.0 mmd are intended for placement at the mandibular central and lateral incisors and maxillary and lateral incisors. Indicated also for denture stabilization using multiple implants.

11 Performance Standards or Special Controls

  • ISO 7405 Second edition 2008-12-15 Dentistry Evaluation of biocompatibility of . medical devices used in dentistry.
  • ISO 5832-3:1996 Implants for surgery -- Metallic materials -- Part 3: Wrought . titanium 6-aluminium 4-vanadium alloy.
  • ISO 14801 Second edition 2007-11-15 Dentistry-Implants-Dynamic fatigue test . for endosseous dental implants.
  • FDA guidance document: Class II Special Controls Guidance Document: Root-. form Endosseous Dental Implants and Endosseous Dental Abutments -Guidance for Industry and FDA Staff.

12 Substantial Equivalence

SubstantialEquivalentTableDENTINImplants:CLASSIC,RAPID,PRESTIGEUno NarrowImplantNobelReplaceMIS System-SEVEN InternalHex ImplantsDualfit InternalHex Implants DFI
510k(K080162)(K023113,K062566 )( K003191,K103089K080162)(K063364)
Indicationfor UseDENTIN® DentalImplants SystemThe UNO NarrowImplant isThe NobelBiocare ReplaceThe MIS implantsystem isThe Alpha-BioDental Implant

ਤੇ

{3}------------------------------------------------

K120530

ס״ד

STERLING MEDICAL REGISTRATION

is indicated for use in surgical and restorative applications for placement in the bone ાં the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. DENTIN® Dental Implants System is indicated also for immediate when loading good primary stability ાંડ achieved and with appropriate occlusal loading. Two Stage Implants: CLASSIC. RAPID, PRESTIGE. One Stage lmplants: ONE PIECE DENTIN® ONEPIECE 3.0 Implants mmd are intended for placement at the mandibular central and lateral incisors maxillary and and lateral incisors. Indicated also for denture stabilization multiple using implants.

indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by adjacent teeth and roots, to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function. Mandibular central and lateral incisors must be splinted if using two or more 03.0 mm implants adjacent to one another. The UNO Narrow lmplant is indicated for immediate implantation in extraction sites or implantation in partially healed or completely healed alveolar ridge situations. When a one stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate.

Tiunite Endosseous Implant is intended to be placed in the upper or lower jaw to support prosthetic devices such as artificial teeth, and to restore patient's chewing function. This may be accomplished using a two stage surgical procedure or a single stage surgical procedure. If the single stage surgical procedure is used, these implants may be loaded immediately following insertion provided - at least four implants are placed and splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foraminae) where good initial stability of the implants with or without bi-cortical anchorage, can most often be obtained.

indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.

SystemB is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices. such as artificial teeth, in order to restore the patient's chewing function. Two stage: ATID, DFI,SPI,SFB,ATIE OF, ITO, SPR One stage: ITO, SPR One stage and One Piece: ARRP, ARPB, ARRC 3mmd diameter are intended only for placement at the mandibular central and lateral incisors and maxillary and incisors. lateral Indicated also for denture stabilization using multiple implants stage and One One Piece for temporary use: ARR, ARB, ARS, ARSB permit immediate splint stability for crown, bridge and prosthesis, protect graft sites. Alpha-Bio The Dental Implant System® is indicated also for immediate loading when qood primary stability is achieved and with appropriate occlusal loading. DF1, SPI, ARRP, ARPB. The Alpha-Bio Dental Implant System® is indicated also for

{4}------------------------------------------------

120530

Image /page/4/Picture/1 description: The image contains a single digit, specifically the number 5. The number is black and appears to be slightly tilted to the right. The background is plain white.

immediate loadingon single toothwhen goodprimary stability isachieved and withappropriateocclusal loading.SPI, SFB.All implants withdiameter 3.3mmdshould not useangled abutment.
MaterialGR-5 TitaniumTi-6Al-4V ELIGR-5 TitaniumTi-6Al-4V ELICP4 TitaniumGR-5 Titanium Ti-6Al-4V ELIGR-5 Titanium Ti-6Al-4V ELI
ImplantBodyContourTaperedTaperedTaperedTaperedTapered
AnatomicalSiteOral CavityOral CavityOral CavityOral CavityOral Cavity
Principle ofoperationConventionalprocedureConventionalprocedureConventionalprocedureConventionalprocedureConventionalprocedure
Self tapping
SterilizationGamma RayGamma RayGamma RayGamma RayGamma Ray
PackagingDoublepackagingDoublepackagingDoublepackagingDouble packagingDouble packaging

Summary of Equivalence:

DENTIN® Dental Implants System shares similarity to Alpha-Bio Tec K063364; MIS Implant Technologies Ltd K003191, K103089, K080162; NOBELREPLACE K023113, K062566; in terms of intended use, indication for use, design, technological characteristics, performance and user interface. DENTIN® Dental Implants System shares the same raw material as its predicated devices, the only difference whereas NOBELREPLACE uses pure titanium commercial DENTIN® and the other predicate devices uses titanium alloy, the differences raise no new issues of safety or effectiveness than the predicate devices.

Mechanical Testing - DENTIN® Implants Technologies has conducted Fatigue - Static & Cycling tests which comply with ISO 14801 Second edition 2007-11-15 Dentistry-Implants-Dynamic fatigue test for endosseous dental implants. The test results have demonstrated the high resistance and high ability with the use of DENTIN Dental Implant System. Therefore, DENTIN® Dental Implants System raises no new issues of safety or effectiveness than the predicate devices.

Safety & Effectiveness testing - sterilization tests, shelf life testing were conducted in order to ensure safety and effectiveness related to DENTIN® Dental Implants & Abutments system. Test results have demonstrated that the SAL of 10-5 was

{5}------------------------------------------------

1209 20

TERLING MEDICAL REGISTRATION

achieved and all testing requirements were met. Thus, DENTIN® Dental Implants System raises no new issues of safety or effectiveness than the predicate devices.

б

Risk Assessment was conducted and has demonstrated no new safety and/or effectiveness issues than the predicate devices.

Conclusion:

As verified by clinical and non clinical data, bench testing, mechanical testing, risk assessment and substantial equivalence, DENTIN® Dental Implant System shares similarity with its predicated devices by term of intended use, raw material and technical design. The fundamental scientific technology of the device is identical or very similar to the referenced predicate devices, thus DENTIN® Dental Implant System is considered to be substantially equivalent to its predicate devices and raises no new safety and/or effectiveness issues than the predicate devices.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Dentin Implants Technologies, Limited C/O Ms. Daniela Levy Regulatory Consultant Sterling Medical Registration 22817 Ventura Boulevard #161 Woodland Hills, California 91364

Re: K120530

Trade/Device Name: DENTIN® Dental Implants System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: June 10, 2012 Received: June 12, 2012

Dear Ms. Levy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

5 2012 JUL

{7}------------------------------------------------

Page 2 - Ms. Levy

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson

nthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

סע

- 10 -

EDICAL REGISTRATION

SECTION 4 - Indication for Use Statement

Indications for Use

Indications for Use

510(k) Number (if known): _

Device Name:

DENTIN® Dental Implants System .

Indications for Use:

DENTIN® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. DENTIN® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Two Stage Implants: CLASSIC, RAPID, PRESTIGE.

One Stage Implants: ONE PIECE

DENTIN® ONEPIECE Implants 3.0 mmd are intended for placement at the mandibular central and lateral incisors and maxillary and lateral incisors. Indicated also for denture stabilization using multiple implants.

Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

(Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dental Devices

510(k) Number: K120530

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.