K Number
K103089

Validate with FDA (Live)

Date Cleared
2011-09-15

(331 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MIS dental implants are intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

When a one stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the occlusal load is appropriate.

MIS short implants are to be used only with straight abutments.

Device Description

The MIS short implants are self tapping, root-form, two piece screw type dental implants, indicated for use in surgical and restorative applications for placement in the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function.

The MIS short implants are provided in 6.0mm length and 4.2mm, 5.0 mm, and 6.0mm diameters, as follows:
Seven internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm
Biocom internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm
Lance internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm

The implants surface is sand blasted and acid etched.
The MIS short implants are two piece devices whereas the implant is to be used in combination with cover screws, healing caps, abutments and superstructures.
The MIS short implants are made of Ti6AL4V ELI complying with standard ASTM F 136-08- Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant.

AI/ML Overview

This 510(k) summary describes a submission for MIS Short Implants. The document establishes substantial equivalence to predicate devices through a comparison of technological characteristics and performance. No specific acceptance criteria or a dedicated study proving performance against them are explicitly detailed in the provided text. Instead, the focus is on demonstrating equivalence based on existing data and non-clinical tests.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria or a specific study designed to meet them. Instead, it relies on demonstrating substantial equivalence to predicate devices. The "performance" is implicitly tied to the safety and effectiveness demonstrated by the predicate devices and confirmed through non-clinical testing and literature review for the new device.

CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Implicit)
Safety & EffectivenessEquivalent to predicate devices (4.5x6.0mm, 6.0x6.0mm, 5.0x6.0mm from Bicon, Inc., and OsseoSpeed 4.0S-6mm from Astra Tech Inc.).Non-clinical fatigue tests performed, and results "found satisfying." Clinical evaluation based on literature review and case studies (30 months follow-up) found "no additional concerns regarding safety and effectiveness."
MaterialMade of Ti6AL4V ELI complying with ASTM F 136-11 standard.Made of Ti6AL4V ELI (Titanium-6Aluminum-4Vanadium ELI) complying with ASTM F 136-08 standard (as stated under "Description of the device").
SterilitySupplied sterile.Supplied sterile (as per comparative table).
Re-useNot reusable.Not reusable (as per comparative table).
ShapeScrew type.Screw type (as per comparative table).
Length6.0 mm.6.0 mm (as per comparative table).
Thread Diameter4.2, 5.0, and 6.0 mm (for MIS Short Implants).4.2, 5.0, and 6.0 mm (as per comparative table).
Abutment TypeUsed only with straight abutments.Used only with straight abutments (as per "Indications for Use" and comparative table).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Non-clinical/Fatigue Test: Not explicitly stated. The document mentions "Fatigue test was performed on MIS short implants and its results were found satisfying." The number of implants tested is not provided.
    • Clinical Evaluation: "case studies with 30 months follow up" – the number of cases/patients is not explicitly stated. The phrasing "case studies" suggests an unspecified number of individual cases rather than a large cohort.
  • Data Provenance:
    • The fatigue tests were performed by the manufacturer, MIS Implants Technologies Ltd. (Israel).
    • The clinical evaluation was "based on literature review and case studies," suggesting a mix of published research and internal data. The geographic origin of the literature or specific case studies is not specified beyond the company being located in Israel. The statement that "MIS short implants are manufactured to the same Titanium alloy" as the predicate devices implies reliance on the established performance of those devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. The document does not describe a process of establishing ground truth by experts for a specific test set. The clinical evaluation references "literature review and case studies," which would rely on diagnoses and outcomes made by clinicians in those cases, but no specific expert panel or their qualifications for establishing ground truth are mentioned.

4. Adjudication Method for the Test Set

Not applicable. No explicit adjudication method is described for defining a ground truth or reconciling discrepancies, as there isn't a dedicated "test set" in the sense of a dataset requiring expert labeling for a performance study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This device is a dental implant, not an AI-assisted diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. As noted above, this is a physical medical device (dental implant), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the assessment of the MIS Short Implants relies on:

  • Biocompatibility/Material Standards: Compliance with ASTM F 136-08 for the titanium alloy.
  • Mechanical Performance: Results of fatigue testing.
  • Clinical Outcomes/Experience: "Literature review and case studies with 30 months follow up" to assess safety and effectiveness. This would implicitly rely on clinical diagnoses, radiographic evaluations, and patient-reported outcomes to determine the success or failure of the implants.
  • Substantial Equivalence: The primary "ground truth" or benchmark is the established safety and effectiveness of the legally marketed predicate devices.

8. The Sample Size for the Training Set

Not applicable. There is no mention of a "training set" as this is a physical device, not a machine learning algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for an algorithm, there is no ground truth established for one.

{0}------------------------------------------------

KI03089

Image /page/0/Picture/1 description: The image shows the logo for MIS Implants Technologies Ltd. The logo features the letters "MIS" in a stylized font, with the "S" being a curved shape. To the right of the logo is the text "Implants Technologies Ltd." in a smaller font.

510(k) Summary:

SEP 1 5 2011

MIS Short Implants

  • MIS Implants Technologies Ltd. Company Name: MIS Implants Technology
    P.O.Box 7, Bar Lev Industrial Park, 20156, ISRAEL 20150, 1510-4-9016800 Fax: +972-4-9918623

Establishment Registration Number: 3004203816

US Agent:

Motti Weisman - VP Marketing MIS Implants Technologies Inc. MIS Implants Technology
14-25 Plaza Rd. Suite S-3-5 Fair Lawn New Jersey; 07410 Phone: (201) 797-9144 (201) 797-9145 Fax: Fax: 2 (2 mail: service@misimplants.com

Date prepared: August 21st, 2011

Trade Name: MIS Short Implants

Classification name: Implants, Endosseous, Root Form

Common/usual name: Dental Implant

Product Code: DZE

Regulation No.: 872.3640

Class: II

Panel identification: Dental Devices Panel

510 (k) MIS Short Implants Section E - Page 1 of 5

{1}------------------------------------------------

hnologies Ltd.

Predicate Device:

Predicate Device:
4.5x6.0mm and 6.0x6.0mm dental implants cleared under 510(k) K050712 and 4.5x6.0mm and 6.0x6.0mm dental implants cleared and on the 501Arborway,
5.0x6.0mm cleared under 510(k) K042637, both from Bicon, Inc. 501Arborway, 1 Boston, Massachusetts, 02130;

Boston, Massachusetts, 02130;
OsseoSpeed 4.0S – 6 mm implant cleared under 510(k) K063779 from Astra Tech
OsseoSpeed 4.0S – 6 mm implant cleared under 510 kb care and 1245 OsseoSpeed 4.05 = 0 han = 1.
Inc., 890 Winter Street, Suite 310, Waltham, MA 02451 .
Inc., 890 Winter Street, Suite 310, Waltham, MA 02451 .

Description of the device:

Description of the device:
The MIS short implants are self tapping, root-form, two piece screw type dental The MIS short implants are self tapping, 1001-1011; the personations for placement in
implants, indicated for use in surgical and restorative applications for placement in implants, indicated for use in surgical and restoration of institution such as artificial
the upper or lower jaw to provide support for prosthetic devices such as artificial the upper or fower just to f
eeth, in order to restore the patient chewing for a form length

teeth, in order to restore the patient chewing ranetices.
The MIS short implants are provided in 6.0mm length and 4.2mm, 5.0 mm, and

6.0mm diameters, as follows: 6.0mm diameters, as follows:
Seven internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm
1.0mm internal hexagon 6.0mm the diameter 4.20mm, 5.0 mm and 6.0mm Seven internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm
Biocom internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm Biocom internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm
Lance internal hexagon 6.0mm length: diameter 4.20mm, 5.0 mm and 6.0mm
Internal hexagon 6.0mm length a

Lance internal nexagon on The implants surface is sand blasted and actues whereas the implant is to be used in
The MIS short implants are two piece devices whereas the implant is and The MIS short implants are two piece academents and superstructures.
combination with cover screws, bealing caps, abutments and supers combination with cover screws, healing caps, abanieman and ASTM
The MIS short implants are made of Ti6ALAV ELI complying with standard ASTM
The MIS short implants are made of The MIS short implants are made of TibALAV LEI compose of the MIS short in the MIS short imm-4Vanadium ELI
F 136-08- Standard Specification for Wrought Titanium-6Aluminum-4Va F 136-08- Standard - 136-06- Surgical Implant.
(Extra Low Interstitial) Alloy for Surgical Implant.

Indications for Use:

Indications for Use:
MIS dental implants are intended to be surgically placed in the bone of the upper or MIS dental implants are intended to be surgically places and as artificial teeth, in
lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in

order jaw around a patient's chewing function. order to restore a patient's chewing function.
When a one stage surgical procedure is applied, the implant may be immediately When a one stage surgical procedure is applical list and the occlusal load is appropriate.
Ioaded when good primary stability is achieved and the occlusal load is appropriate

loaded when good primary stability is
MIS short implants are to be used only with straight abutments.
The short implants are to be used only with straight abutments.

510 (k) MIS Short Implants 310 (K) E - Page 2 of 5

{2}------------------------------------------------

Substantial Equivalence:

Substantial Equivalence:
The MIS short implants have the same intended use as the 4.5x6.0mm dental implans The MIS short implants have the same intentional as as as a manufactured to the mass cleared
6.0x6.0mm cleared under 510(k) KD50712 and 5.0x6.0mm 6.0x6.0mm cleared under 510(x) Ro371 ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... . . . . . . . . . . . . . . . . . . . . . . . . . . . under 510(k) K042637, both from Bicoll, the Street, Speed implant cleared under
Massachusetts, 02130, and the OsseoSpeed 4.0S – 6 mm implant cleared under Massachusetts, 02130, and the OsseoSpeca 4.05 = 6 mm. ........................................................................................................................ 510(k) K063779 from Astra Tech Inc., 690 w hise excepts to these products are
02451, and have equivalent performance characteristics. All these products are 02451, and have equivalent performance characteristics of the missteristics.
manufactured from the same Titanium alloy. All other technological characteristics manufactured from the same Titanum andy. The MIS short implants
are similar and show equivalent performance capabilities. The MIS short implants
are similar and show equival are similar and show equivalent posses.
are therefore substantially equivalent to their predicate devices.
Therefore substantially equivalent to their predicate devices.
Ther

cal characteristics – comparative table:

Technological characteristics - compar...
IndicationsFor UseMIS Short Implantsfrom MIS ImplantsTechnologies Ltd.MIS dentalimplants areintended to besurgically placed inthe bone of the upperor lower jaw archesto provide supportfor prostheticdevices, such asartificial teeth, inorder to restore apatient's chewingfunction.When a one stagesurgical procedureis applied, theimplant may beimmediatelyloaded when goodprimary stability isachieved and theocclusal load is4.5x6.0mm6.0x6.0mmImplants fromBicon, Inc.The 4.5x6.0 mm andthe 6.0mmx6.0mmimplants are designedfor use in edentuloussites in the mandibleor maxilla for supportof a complete dentureprosthesis, a terminalor intermediateabutment for fixedbridgework, partialdentures, or a singletooth replacement.5.0x6.0 mm Implantfrom Bicon, Inc.The 5.0x6.0 mmimplant is intended tobe surgically placedin the maxilla ormandible to providesupport for prostheticdevices to restore thepatient's chewingfunction.OsseoSpeed from AstraTech Inc.The OsseoSpeed™ 4.0S - 6mm is intended to be usedto replace missingmasticatory functional units(teeth) in single or multipleunit applications within themandible or maxilla whereimmediate implant stabilitymay be obtained. Thedevice may be used equallywell in a single-stage ortwo-stage surgicalprocedure. It is indicatedfor immediate implantationin extraction sited orimplantation in partiallyhealed or completelyhealed alveolar ridgesituations. When a one-stage surgical approach is applied, the implant may beimmediately loaded afterimplantation whereimmediate implant stability
MIS Short Implantsfrom MIS ImplantsTechnologies Ltd.4.5x6.0mm6.0x6.0mmImplants fromBicon, Inc.5.0x6.0 mm Implantfrom Bicon, Inc.OsseoSpeed from AstraTech Inc.
appropriate.MIS short implantsare to be used onlywith straightabutments.may be obtained.The fluoride-modifiedsurface, though having afluoride ion level far belowthat needed for cariesprevention in teeth,provides a favorablesubstrate for boneattachment andosseointegration.OsseoSpeed 4.0S – 6 mm isespecially indicated for usein soft bone applicationswhere implants with otherimplant surface treatmentsmay be less effective.Because initial stabilitymay be difficult to obtain inType IV bone, immediateloading of single toothrestorations may not beappropriate in suchsituations.
SuppliedSterileYesYesYesYes
Re-useNoNoNoNo
Material ofConstructionTitanium AlloyTitanium AlloyTitanium AlloyTitanium Alloy
ShapeScrew typeScrew typeScrew typeScrew type
Length6.0 mm6.0 mm6.0 mm4.0
ThreadDiameter4.2, 5.0 and 6.0 mm4.5 and 6.0 mm5.0 mmStraight and up to 20°
AbutmentMaterial ofConstructionofAbutmentsStraightTitanium AlloyStraight and up to 25°Titanium AlloyStraight and up to 25°Titanium AlloyStraight and up to 20°Titanium Alloy
SurfaceNoneNoneNoneNone

510 (k) MIS Short Implants 310 (K) Mis Page 3 of 5

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "nis" in a stylized font. The letters are bold and connected, with the "s" resembling a backwards "u". To the right of the word "nis" is the word "Implan", which is likely the beginning of the word "Implant". The image appears to be a logo or part of a title.

510 (k) MIS Short Implants
Section E - Page 4 of 5

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MIS Implants Technologies Ltd. The logo features the letters "MIS" in a stylized font, with the "S" being a curved line. To the right of the logo is the text "Implants Technologies Ltd." in a smaller font. The text is slightly angled upwards.

Non – clinical tests:

Non – clinical tests:
Fatigue test was performed on MIS short implants and its results were found
ed the state was performed on Marts devices Fatigue test was F

Clinical tests:
A clinical evaluation, based on literature review and case studies with 30 months
. follow up, has been performed.

Conclusion:

Conclusion:
The evaluation of the MIS short implants does not raise any additional concerns
and the MIS short implants of the MIS short implants may therefore be The evaluation of the MIS short implants does not the may therefore be
regarding safety and effectiveness and the MIS short implants may therefore be regarding safety and effectiveness .
considered substantially equivalent to their predicate device

510 (k) MIS Short Implants Section E - Page 5 of 5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Ms. Iman Khorshid V P. QA & RA MIS Implants Technologies, Limited P.O. Box 7 Bar Lev Industrial Park 20156. ISRAEL

Re: Ki03089

Trade/Device Name: MIS Short Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: September 11, 2011 Received: September 13, 2011

Dear Ms. Khorshid:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Paris 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

SEP 1 5 2011

{6}------------------------------------------------

Page 2 - Ms. Khorshid

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reportal/roblem/default.htm for the CDRH 's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.hum.

Sincerely vours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Eyaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Implants Technologies Ltd.

INDICATIONS FOR USE

510(k) Number:K103089
-------------------------

Device Name: MIS Short Implants

Indications for Use: MIS dental implants are intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

When a one stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the occlusal load is appropriate.

MIS short implants are to be used only with straight abutments.

Prescription Use x OR (Part 21 CFR 801 Subpart D)

Over the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

KID30599 510(k) Number:

510 (k) MIS Short Implants Section D- Page 1 of 1

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.