K Number
K120472

Validate with FDA (Live)

Device Name
QUANTRA
Manufacturer
Date Cleared
2012-06-22

(127 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Quantra™ is a software application intended for use with images acquired using digital breast X-ray systems. Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates; and area breast density as a ratio of fibroglandular tissue area and total breast area estimates. It segregates breast density into BI-RADS-like breast composition categories, which may be useful in the reporting of consistent breast composition values as mandated by certain state regulations. Quantra provides these numerical values for each image, breast, and subject, to aid radiologists in the assessment of breast tissue composition. Quantra produces adjunctive information; it is not an interpretive or diagnostic aid. Quantra runs on a Windows platform.

Device Description

Quantra is a software application that estimates breast tissue volume and area density. The estimations are made from images acquired using digital breastray systems.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Hologic Quantra™ device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text describes validation methods and comparisons rather than explicit, stand-alone acceptance criteria with quantitative thresholds. However, it implicitly demonstrates that the device's performance met the expectation of correlating with existing methods and providing similar distributions across diverse equipment.

Metric/MeasureAcceptance Criteria (Implicit)Reported Device Performance
Volumetric Breast Density (Vbd)Correlation with predicate device (K082483, Quantra Volumetric Assessment) and MRI cases; comparison to mode BI-RADS density rating from 15 radiologists.Demonstrated correlation with predicate device and MRI cases.
Area Breast Density (Abd)Correlation with an expert's hand-drawn dense areas using predicate device (Sectra IDS5 Workstation); comparison to mode BI-RADS density rating from 15 radiologists.Demonstrated correlation with expert hand-drawn dense areas on Sectra IDS5 Workstation.
BI-RADS-like Breast Composition (q_abd)Comparison of continuous q_abd score to the mean BI-RADS values assigned by 15 radiologists; comparison across different FFDMs to ensure similar distributions.Demonstrated consistency with mean BI-RADS values from 15 radiologists.
Consistency of Density MeasuresStatistical evaluation of density measures between CC and MLO views of the same breast, and left/right breasts of the same women.All Quantra density measures were evaluated statistically and found to be consistent.
Distribution SimilarityComparison of Vbd, Abd, and q_abd values across large populations of cases from Hologic (Selenia, Dimensions), GE (Senographe, Essential), and Siemens (Mammomat Novation) FFDMs.Distributions were similar across different FFDMs.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Density Measures:
    • Vbd/Abd/q_abd compared to BI-RADS: "a large database of cases" (exact number not specified).
    • Vbd compared to MRI: "MRI cases of the same patients" (exact number not specified).
    • Vbd/Abd/q_abd compared across FFDMs: "a large population of cases" from Hologic, GE, and Siemens (exact number not specified).
    • Consistency (CC/MLO, Left/Right): "substantially large number of images" from Hologic, GE, and Siemens (exact number not specified).
  • Data Provenance: Not explicitly stated (e.g., country of origin). The data included cases from Hologic (Selenia and Dimensions), GE (Senographe and Essential), and Siemens (Mammomat Novation) FFDMs, implying diversity in equipment. The data was "retrospective" as it was used for validation of an already developed software.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: 15 radiologists.
  • Qualifications: "radiologists" - no further specific qualifications (e.g., years of experience, subspecialty) are provided in the document.

4. Adjudication Method for the Test Set

  • The text describes comparing Quantra values to the "mode (most common) BI-RADS density rating from 15 radiologists" and the "mean value of the 15 radiologists." This indicates a form of consensus-based ground truth derived from multiple readers, but it doesn't specify a formal adjudication method like "2+1" or "3+1" (where differing opinions are resolved by a tie-breaker). It appears to use statistical aggregations (mode or mean) of expert ratings.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

  • No, a formal MRMC comparative effectiveness study "with AI vs without AI assistance" to measure the effect size of human readers improving with AI assistance was not explicitly described for this 510(k) submission. The studies detailed focus on validating the standalone performance of the Quantra software against existing clinical practices and predicate devices.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, the primary focus of the performance/bench testing described is on the standalone performance of the Quantra algorithm. It generates volumetric and area breast density estimates and segregates breast density into BI-RADS-like categories without human input during the calculation process. Its output is "adjunctive information" for radiologists, implying it functions independently before a radiologist's assessment.

7. The Type of Ground Truth Used

  • Expert Consensus: The "mode" and "mean" BI-RADS density ratings from 15 radiologists served as a form of expert consensus ground truth for comparison.
  • Predicate Device Output: Results from the predicate device (Quantra Volumetric Assessment for Vbd, Sectra IDS5 Workstation with expert hand-drawn annotations for Abd) were used as ground truth for correlation.
  • Other Imaging Modalities: MRI cases were used as ground truth for Vbd validation.

8. The Sample Size for the Training Set

  • The sample size for the training set is not specified. The document mentions a "reference database" of "approximately 1,000 patients" used to create Vbd-score and Vfg-score (reflecting standard deviations from the mean in this database), but it's unclear if this was the training set or a separate reference population.

9. How the Ground Truth for the Training Set was Established

  • The document does not explicitly describe how the ground truth for any potential training set was established. The focus of the "Performance/Bench Testing" section is on validation against established references and expert opinions, not on the training process itself. If the "reference database" of 1,000 patients was part of a training set, the method for establishing its ground truth is not detailed in the provided information.

{0}------------------------------------------------

120472

Hologic, Inc. 35 Crosby Drive, Bedford, MA 01730 USA Main: +1.781.999.7300 Fax: +1.781.280.0669 JUN: 2 2 2012

B. Administrative Information

B.1 510(k) Summary of Safety & Effectiveness (as required by 21 CFR §807.92c)

Date Prepared: February 15, 2012 Submitted by:

Hologic, Inc. 35 Crosby Drive Bedford, MA 01730 USA

Name, Title and Phone Number of Contact:

Catherine A. Williams Director, Regulatory Affairs Phone: (408) 352-0201 FAX: (408) 352-0101 Email: catherine.williams@hologic.com

Trade Name and Common Name:

Trade Name:Quantra™
Software Version:2.0
Common Name:Picture Archiving and Communications System

Device Classification:

Regulatory Class:II
Classification Panel:Radiology
Image Processing System21 CFR §892.2050
Product Code90-LLZ

Predicate Devices:

The predicate devices for Quantra software are certain software functions contained in the following devices:

K082483, September 12, 2008 K050196, February 24, 2005 K102556, October 7, 2010 Limited]

Quantra (Volumetric Assessment) [Hologic, Inc.] Sectra IDS5 Workstation [Sectra Imtec AB] Volpara Imaging Software [Matakina Technology

{1}------------------------------------------------

Device Description:

Quantra is a software application that estimates breast tissue volume and area density. The estimations are made from images acquired using digital breastray systems.

Quantra has been designed and will be manufactured in accordance with the following standards:

  • · ISO 14971 Medical Devices - Application of Risk Management to Medical Devices
  • ISO 62304: Medical Device Software - Software Life Cycle Processes

The performance of the software is also tested in accordance with Hologic's SOPs and testing procedures to demonstrate adequate performance.

Intended Use:

· Quantra™ is a software application intended for use with images acquired using digital breast X-ray systems. Quantra calculates volumetric breast demsity as a minori fibroglandular tissue and total breast volume estimates; and area breast density as of fibroglandular tissue area and total breast area estimates. It segregates breast de into BI-RADS-like breast composition categories, which may be useful in the report of consistent breast composition values as mandated by certain state regul · provides these numerical values for each image "breast" and subject, to ai the assessment of breast tissue composition. Quantra produces adjunctive information is not an interpretive or diagnostic aid. Quantra runs on a Windows platform.

Technological Characteristics:

Quantra is a software application that processes digital mammography images. The device does not contact the patient, nor does it control any life-sustaining devices.

Performance/Bench Testing:

The volumetric breast density measures were validated by demonstrating correlation with the Quantra predicate device and with MRI cases of the same patients. The me also were compared to the mode (most common) BI-RADS density rating from 15 radiologists on a large database of cases. Finally, the Vbd values were compared a large population of cases from Hologic (Selenia and Dimensions), GE (Seniographe and similar. 11 - 1 - 1 - 1 - 1 - 1 - 1 - 1

The area breast density measure was validated using a correlation-based assessme compare Quantra values with dense area measurement based on hand-drawn dense areas annotated by an expert using the predicate device Sectra IDS5 workstation. mode (most common) BI-RADS density rating from 15 radiologists on a large database of cases. Finally, the Abd values were compared across. large population of cases from Hologic (Selenia and Dimensions), GE (Senographe and

{2}------------------------------------------------

Essential), and Siemens (Mammomat Novation) FFDMs to ensure the distributions were similar.

The volumetric breast density (Vbd) and volume of fibroglandular tissue (Vfg) values. were used to create two scores (Vbd-score and Vfg-score), which reflect the number of standard deviations between a subject's V.bd or Vig value and the corresponding mean value of approximately 1,000 patients in a reference database.

The BI-RADS-like breast composition measure (O abd), similar to the density grade measure in the Matakina predicate device, was evaluated based on a comparison of Q abd with BI-RADS values assigned by 15 radiologists on a large set of digital mammography cases. The continuous q. abd score was compared to the mean value of the 15 radiologists on the same large database of cases described above. Finally, the q abd values were compared across a large population of cases from Hologic (Selenia and Dimensions), GE (Senographe and Essential), and Siemens (Mammomat Novation) FFDMs to ensure the distributions were similar.

All Quantra density measures were evaluated statistically between CC and MLO views of the same breast and left and right breasts of the same women, using a substantially large number of images from Hologic (Selenia and Selenia Dimensions), GE (Senographe and Senographe Essential), and Siemens (Mammomat Novation) digital breast X-rav systems).

General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are, controlled via software development, verification and validation testing.

Conclusion:

The 510(k) Premarket Notification for Quantra contains adequate information and data to enable FDA/CDRH to determine substantial equivalence to the predicate devices:

The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "Moderate".

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration · 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

JUN 2 2 2012

Ms. Catherine A. Williams Director, Regulatory Affairs Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730

Re: K120472

Trade/Device Name: QuantraTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications svstem Regulatory Class: II Product Code: LLZ Dated: May 18, 2012 Received: May 21, 2012

Dear Ms. Williams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket while and to finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 July 11 Species and the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket s 150. First Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

B.2 Indication(s) for Use Statement

510(k) Number (if known): K120472

Device Name:

Quantra™M

Indications for Use:

Quantra™ is a software application intended for use with images acquired using digital breast X-ray systems. Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates; and area breast density as a ratio of fibroglandular tissue area and total breast area estimates. It segregates breast density into BI-RADS-like breast composition categories, which may be useful in the reporting of consistent breast composition values as mandated by certain state regulations. Quantra provides these numerical values for each image, breast, and subject, to aid radiologists in the assessment of breast tissue composition. Quantra produces adjunctive information; it is not an interpretive or diagnostic aid. Quantra runs on a Windows platform.

Prescription Use X (Part 21 CFR§801 Subpart D) Over-The-Counter - Use (21 CFR §807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR .

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Division Sign-
Divisiókf Radiologica: Devices
Office of In Vitro Diagnostic Device Evaluation and Safety.
K120472

N/A