(15 days)
Quantra™ is a software application intended for use with Hologic digital mammography systems. Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra provides these numerical values for each breast to aid radiologists in the assessment of breast tissue composition. Quantra produces adjunctive information; it is not an interpretive or diagnostic aid.
Quantra is a software application which runs on the Hologic Cenova DICOM server (Class I exempt per 21 CFR § 892.2010 and 21 CFR § 892.2020).
Quantra is a software application that estimates breast tissue volumes. The estimations are made from mammography images produced by full-field digital mammography (FFDM) systems.
Here's an analysis of the provided text regarding the Quantra™ device's acceptance criteria and studies:
1. Table of Acceptance Criteria and Reported Device Performance:
The provided document does not explicitly state specific numerical acceptance criteria for the Quantra™ device's performance. It primarily focuses on the device's intended use, regulatory classification, and substantial equivalence to predicate devices. The "performance of the software" is mentioned as being tested "in accordance with Hologic's SOPs and testing procedures to demonstrate adequate performance," but no specific metrics or thresholds for "adequate performance" are detailed in this summary.
Therefore, a table of acceptance criteria and reported device performance cannot be created from the given information.
2. Sample Size Used for the Test Set and Data Provenance:
The document does not provide any information regarding the sample size used for a test set or the data provenance (e.g., country of origin, retrospective or prospective nature) of any study data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
The document does not provide any information regarding the number of experts used to establish ground truth or their qualifications.
4. Adjudication Method for the Test Set:
The document does not provide any information about an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it provide any information on the effect size of human readers improving with or without AI assistance.
6. Standalone (Algorithm Only) Performance Study:
The document does not explicitly detail a standalone performance study. While it describes the software's functionality in "calculating volumetric breast density," it doesn't present data from a study that solely evaluates the algorithm's performance against a ground truth without human interaction.
7. Type of Ground Truth Used:
The document does not specify the type of ground truth used for any potential testing or validation.
8. Sample Size for the Training Set:
The document does not provide any information regarding the sample size used for a training set.
9. How Ground Truth for the Training Set Was Established:
The document does not provide any information on how ground truth for a training set was established.
Summary of what is present in the document versus what is missing regarding a performance study:
The provided 510(k) summary primarily focuses on the regulatory aspects of the Quantra™ device:
- Intended Use: Calculating volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates, to aid radiologists in assessment (adjunctive, not diagnostic).
- Regulatory Classification: Class II, Picture Archiving and Communications System.
- Predicate Devices: Sectra IDS5 Workstation and WorkstationOne™ Breast Imaging Workstation.
- Safety and Effectiveness Concerns: Covered by instructions for use, cautions, warnings, risk analysis, software development, verification, and validation testing.
- Conclusion: Substantial equivalence to predicate devices.
Critical Missing Information for a Performance Study:
The document is notably lacking specific details about any clinical validation or performance study that would typically include:
- Defined acceptance criteria (e.g., accuracy, precision, correlation with a gold standard).
- Details of a test dataset (size, characteristics, provenance).
- Methodology for establishing ground truth.
- Results of any performance metrics.
- Information on reader studies or standalone algorithm performance.
This suggests that for this 510(k) submission, the primary evidence for substantial equivalence was likely based on the technological characteristics and intended use being similar to existing cleared predicate devices, supported by internal Hologic SOPs and testing procedures for software performance, rather than a full-scale clinical performance study with detailed numerical results presented in the summary.
{0}------------------------------------------------
SEP 1 2 2008
Quantra™M 510(k) Submission
510(k) Summary of Safety & Effectiveness B.1
(as required by 21 CFR § 807.92c)
Date Prepared: August 15, 2008
Submitted by:
Hologic, Inc. 35 Crosby Drive Bedford, MA 01730, USA
Name, Title and Phone Number of Contact:
Catherine A. Williams Director, Regulatory Affairs Phone: (408) 745-5128 FAX: (408) 744-1905 Email: catherine.williams(@hologic.com
Trade Name and Common Name:
| Trade Name: | Quantra™ |
|---|---|
| Software Version: | 1.1 |
| Common Name: | Picture Archiving and Communications System |
Classification:
| Regulatory Class: | II |
|---|---|
| Classification Panel: | Radiology |
| Image Processing System | 21 CFR § 892.2050 Product Code 90-LLZ |
Predicate Devices:
The predicate devices for Quantra software are certain software functions contained in the following devices:
Sectra IDS5 Workstation [Sectra Imtec AB] K050196. Feb 24, 2005 WorkstationOne™ Breast Imaging Workstation [Three Palm Software, LLC] K073272, Dec 11, 2007
Device Description:
Quantra is a software application that estimates breast tissue volumes. The estimations are made from mammography images produced by full-field digital mammography (FFDM) systems.
{1}------------------------------------------------
Quantra has been designed and will be manufactured in accordance with the following standards:
- ISO 13485 Medical Devices - Quality Management Systems .
- . ISO 14971 Medical Devices - Application of Risk Management to Medical Devices
- Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology) NEMA DICOM PS 3 . (ISO 12052)
- Medical Device Software Software Life Cycle Processes . ANSI/AAMI SW68:2001
- . 21 CFR & 820 US Food and Drug Administration, Quality System Regulation General Principles of Software Validation; Final Guidance for Industry and N/A . FDA
The performance of the software is also tested in accordance with Hologic's SOPs and testing procedures to demonstrate adequate performance.
Intended Use:
Quantra™ is a software application intended for use with Hologic digital mammography systems. Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra provides these numerical values for each breast to aid radiologists in the assessment of breast tissue composition. Quantra produces adjunctive information; it is not an interpretive or diagnostic aid.
Quantra is a software application which runs on the Hologic Cenova DICOM server (Class I exempt per 21 CFR & 892.2010 and 21 CFR & 892.2020).
Technological Characteristics:
Quantra is a software application that processes digital mammography images. The device does not contact the patient, nor does it control any life-sustaining devices.
General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing.
Conclusion:
The 510(k) Pre-Market Notification for Quantra contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate devices.
The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "Minor".
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird image. The logo is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 2 2008
Hologic, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K082483
Trade/Device Name: QuantraTM Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 27, 2008 Received: August 28, 2008
Dear Mr. Job;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
forque M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
B.2 Indication(s) for Use Statement
510(k) Number (if known):
Device Name:
Quantra™M
KO82483
Indications for Use:
Quantra™ is a software application intended for use with Hologic digital mammography systems. Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra provides these numerical values for each breast to aid radiologists in the assessment of breast tissue composition. Quantra produces adjunctive information; it is not an interpretive or diagnostic aid.
Quantra is a software application which runs on the Hologic Cenova DICOM server (Class I exempt per 21 CFR § 892.2010 and 21 CFR § 892.2020).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal Radiological Devices 510(k) Number
Section B
N/A