K Number
K113769

Validate with FDA (Live)

Date Cleared
2012-01-20

(29 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GPSCath Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.

Device Description

The GPSCath Balloon Dilatation Catheter is designed for dilation of peripheral vessels in the arterial svstem and native or synthetic arteriovenous dialysis fistulae in the treatment of obstructive lesions. The GPSCath Balloon Dilatation Catheter is a 0.035" guide-wire compatible, PTA balloon catheter with a proprietary proximal valve system which allows injection of fluids. By providing an angioplasty balloon with fluid delivery capability, the user is able to treat obstructive lesions within the arterial system and arteriovenous dialysis fistulae without having to lose guidewire position.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the GPSCath™ Balloon Dilatation Catheter, based on the provided document:

Acceptance Criteria and Device Performance

The document states that the purpose of the Special 510(k) is to revise the Instructions for Use (IFU) to illustrate a new balloon rated burst pressure and compliance table. This implies that the acceptance criteria are related to these physical properties and that the device performance met the new, higher specifications. While the exact numerical values for the original and new rated burst pressures, compliance, and other performance criteria are not explicitly detailed in a table, the document confirms that the device met them.

Table of Acceptance Criteria and Reported Device Performance (Inferred)

Acceptance CriterionReported Device Performance
Balloon Rated Burst PressureMet established (new) specifications. This indicates the balloon can withstand a higher internal pressure before bursting compared to the previous specification.
Balloon ComplianceMet established specifications. This refers to how the balloon expands and conforms to the vessel wall at different pressures.
Catheter Body Burst StrengthMet established specifications. This ensures the catheter body itself can withstand the required pressures during use without bursting.
Balloon FatigueMet established specifications. This indicates the balloon can endure repeated inflation and deflation cycles without structural failure.
Balloon Inflation/Deflation TimeMet established specifications. This confirms the balloon inflates and deflates within acceptable timeframes for clinical use.
Overall Performance"Met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, test protocols, and/or customer inputs." This is a general statement confirming successful performance across all tested parameters.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated in the document. The document mentions "bench top testing" but does not give specific numbers of devices or balloons tested for each criterion.
    • Data Provenance: The studies were "in vitro test data" and "bench top tests." This means the data was generated in a lab setting, not from human or animal subjects. The country of origin of the data is implied to be within the US, as Hotspur Technologies, Inc. is located in Mountain View, CA, and submitted the notification to the FDA.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This section is not applicable to this type of device and study. The testing for the GPSCath™ Balloon Dilatation Catheter involved physical bench-top performance metrics (e.g., burst pressure, compliance) rather than diagnostic interpretations requiring expert ground truth. The "ground truth" for these tests would be the measured physical properties of the device itself.
  3. Adjudication method for the test set:

    • This section is not applicable as the study involved objective physical measurements, not subjective evaluations that would require adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This section is not applicable. The GPSCath™ Balloon Dilatation Catheter is a medical device (a balloon catheter) and not an AI-powered diagnostic system. Therefore, MRMC studies involving human readers and AI assistance are not relevant here.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This section is not applicable. This is a physical medical device, not an algorithm, so "standalone" performance in the context of an algorithm is not relevant.
  6. The type of ground truth used:

    • The "ground truth" for this device's performance assessment was physical measurement against predetermined specifications and applicable standards. For example, a burst pressure test would involve applying pressure until the balloon bursts, and that measured pressure would be compared to the specified rated burst pressure.
  7. The sample size for the training set:

    • This section is not applicable. There is no "training set" in the context of this device. The phrase "training set" typically refers to data used to train machine learning models. This product underwent traditional engineering and quality assurance testing.
  8. How the ground truth for the training set was established:

    • This section is not applicable for the same reason as point 7.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Hotspur Technologies. The logo consists of a circular graphic on the left and the company name on the right. The graphic is made up of curved lines that resemble a fingerprint or a stylized wave. The text "Hotspur" is in a bold, sans-serif font, and "Technologies" is in a smaller, lighter font below it.

Special 510(k) Notification: GPSCath™ Balloon Dilatation Catheter Increased Rated Burst Pressure

CONFIDENTIAL

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

A. Submitter Information

Submitter's Name: Address:

Telephone: Fax: Contact Person: Hotspur Technologies, Inc. 880 Maude Ave., Suite A Mountain View, CA 94043 650-969-3150 408-608-1597 Eric Ankerud, Executive Vice President Clinical, Regulatory, Quality

JAN 2 0 2012

B. Subject Device

Trade Name: Common/Usual Name: Classification Name:

GPSCath Balloon Dilatation Catheter Balloon Catheter Catheter, Angioplasty, Peripheral, Transluminal/Percutaneous Catheter (21 CFR 870.1250, Product Code LIT)

C. Predicate Device Name(s) Trade Name(s):

GPSCath Balloon Dilatation Catheter (previously cleared under the name of PTA-Duo PTA Balloon Catheter, 510(k) #K101047)

D. Device Description:

The GPSCath Balloon Dilatation Catheter is designed for dilation of peripheral vessels in the arterial svstem and native or synthetic arteriovenous dialysis fistulae in the treatment of obstructive lesions. The GPSCath Balloon Dilatation Catheter is a 0.035" guide-wire compatible, PTA balloon catheter with a proprietary proximal valve system which allows injection of fluids. By providing an angioplasty balloon with fluid delivery capability, the user is able to treat obstructive lesions within the arterial system and arteriovenous dialysis fistulae without having to lose guidewire position.

E. Intended Use:

The GPSCath Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lessions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.

F. Reason for Modification:

The purpose of this Special 510(k) is to revise the GPSCath Balloon Dilatation Catheter Instructions for Use (IFU) to illustrate a new balloon rated burst pressure and compliance table based on the following bench top testing: Balloon Compliance, Catheter Body Burst Strength, Balloon Rated Burst Pressure, Balloon Fatigue and Balloon Inflation/ Deflation Time Testing. Hotspur also wishes to notify FDA of a new commercial trade name for the PTA-Duo PTA Balloon Catheter which was originally cleared under 510(k) #K101047. Hotspur has replaced the PTA-Duo PTA Balloon Catheter name with the new commercial trade name of GPSCath Balloon Dilatation Catheter. Additionally, this Special 510(k) includes minor revisions to the IFU for consistency and clarification.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image contains the logo for Hotspur Technologies. The logo consists of a stylized, circular graphic on the left, resembling a series of concentric arcs or waves. To the right of the graphic, the word "Hotspur" is written in a slightly bold, sans-serif font. Below "Hotspur," the word "Technologies" is written in a smaller, lighter font, also sans-serif. The overall design is simple and modern.

G. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:

The modified GPSCath Balloon Dilatation Catheter and the predicate PTA-Duo PTA Balloon Catheter are both indicated for treatment of obstructive lesions by high pressure dilation in the arterial system and of native or synthetic arteriovenous dialysis fistulae.

The GPSCath Balloon Dilatation Catheter contains an inflatable balloon for dilation of obstructive lesions. The usable length of both devices is 55 cm. The modified GPSCath Balloon Dilatation Catheter and predicate device are substantially equivalent in terms of intended use, fundamental scientific technology, target population, operating principles, and method of sterilization.

H. Performance Data:

The modified GPSCath Balloon Dilatation Catheter was evaluated using in vitro test data to confirm the performance characteristics as compared to the cleared device, the GPSCath Balloon Dilatation Catheter which was originally cleared as the PTA-Duo PTA Balloon Catheter under 510(k) #K101047 device.

Bench top tests, specifically Balloon Compliance, Catheter Body Burst Strength, Balloon Rated Burst Pressure, Balloon Fatigue and Balloon Inflation/ Deflation Time Testing, were completed to demonstrate that the modified GPSCath Balloon Dilatation Catheter met the established specifications necessary for consistent performance during its intended use.

I. Conclusions:

The modified GPSCath Balloon Dilatation Catheter met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, test protocols, and/or customer inputs. The modified GPSCath Balloon Dilatation is substantially equivalent to the legally marketed predicate device and does not raise any new safety or effectiveness questions. Device modifications are documented in Hotspur's Design History File in accordance with design controls, 21 CFR 820.30, and Hotspur's internal quality system procedures.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN 2 0 2012

Hotspur Technologies, Inc. c/o Eric Ankerud Executive VP Clinical, Regulatory, and Quality 880 Maude Ave., Suite A Mountain View, CA 94043

Re: K113769

Trade/Device Name: GPSCath Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: December 20, 2011 Received: December 22, 2011

Dear Mr. Ankerud:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Eric Ankerud

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

M.A. Hilbebun

a Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Hotspur Technologies. The logo consists of a stylized symbol on the left and the company name on the right. The symbol is made up of curved lines that resemble a stylized letter C. The company name is written in a simple, sans-serif font, with the word "Hotspur" in a larger font size than the word "Technologies".

CONFIDENTIAL

STATEMENT OF INDICATIONS FOR USE

510(k) Number: Device Name: Indication For Use:

K113769

GPSCath™ Balloon Dilatation Catheter The GPSCath Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. L. Kleham

Division Sign-Off) Division of Cardiovascular Devices

12113769 510(k) Number_

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).