K Number
K113721

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-02-15

(58 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TephaFLEX light mesh is intended to reinforce soft tissue where weakness exists in patients undergoing plastic and reconstructive surgery, or for use in procedures involving soft tissue repair, such as for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Device Description

The TephaFLEX light mesh is a resorbable mesh prepared from poly-4-hydroxybutyrate (P4HB). The mesh is prepared from non-dyed or dyed (D&C Violet #2) monofilament P4HB fiber that is knitted into a surgical mesh. It is provided in single sheets of varying widths and lengths and may be cut to the shape or size desired for a specific application.

AI/ML Overview

The provided document describes the TephaFLEX® Light Mesh, a surgical mesh, and its substantial equivalence to predicate devices. However, the document does not contain details of a clinical study or a study proving the device meets specific acceptance criteria in the manner typically associated with diagnostic AI/imaging devices.

The "Safety and Performance" section states: "Mechanical testing, biocompatibility testing, and in vivo animal testing was performed based on recommendations identified in the FDA surgical mesh guidance document: The Guidance for the Preparation of a Pre-market Notification Application for a Surgical Mesh. Specifically, comparative burst strength, suture pull-out strength, tensile strength and tear resistance strength was characterized. In vivo strength retention was characterized via a subcutaneous implantation study. The mechanical and in vivo data collected determined the mesh to be substantially equivalent to the predicate devices."

This indicates that the "study" conducted was a series of non-clinical performance tests (mechanical and animal studies) to demonstrate substantial equivalence, rather than a clinical trial with human subjects, a test set, or ground truth as understood in the context of an AI/imaging device.

Therefore, many of the requested categories (2-7, 9) are not applicable or cannot be extracted from this document, as they pertain to clinical studies or AI algorithm evaluation.

Here's the information that can be extracted or deduced:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance
Mechanical PropertiesThe TephaFLEX® Light Mesh must demonstrate comparable mechanical properties (burst strength, suture pull-out strength, tensile strength, tear resistance strength) to its predicate devices, as specified by the FDA surgical mesh guidance document. This is implicitly the acceptance criterion for demonstrating safety and effectiveness based on equivalence."Comparative burst strength, suture pull-out strength, tensile strength and tear resistance strength was characterized." The conclusion states: "The mechanical... data collected determined the mesh to be substantially equivalent to the predicate devices." (Specific numerical values for these properties are not provided in the summary.)
BiocompatibilityThe TephaFLEX® Light Mesh must demonstrate biocompatibility comparable to predicate devices, following guidelines for surgical mesh."Biocompatibility testing... was performed based on recommendations identified in the FDA surgical mesh guidance document." The conclusion states: "Based on the indications for use, technological characteristics, and safety and performance testing, the TephaFLEX light mesh has been shown to be substantially equivalent to predicate devices..." (Specific results or details of biocompatibility are not provided in the summary.)
In Vivo Strength RetentionThe TephaFLEX® Light Mesh must exhibit strength retention in vivo comparable to predicate devices over a defined period. This is crucial for its intended function as a resorbable mesh."In vivo strength retention was characterized via a subcutaneous implantation study." The conclusion states: "...in vivo data collected determined the mesh to be substantially equivalent to the predicate devices." (Specific duration or quantitative results are not provided in the summary.)

2. Sample Size for the Test Set and Data Provenance:

  • Not applicable. No clinical test set involving human data for algorithm evaluation is described. The "test set" refers to mechanical and animal studies. Details on the sample sizes for these tests (e.g., number of animals, number of mesh pieces tested for mechanical properties) are not provided in the summary.
  • Data Provenance: The studies were internal (Tepha, Inc.) and based on FDA guidance for surgical mesh. The "in vivo animal testing" implies animal studies, but the origin country for these studies is not specified. They were likely prospective, conducted for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

  • Not applicable. This pertains to clinical studies involving expert review, which is not described.

4. Adjudication Method for the Test Set:

  • Not applicable. This pertains to clinical studies involving expert review.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a surgical mesh, not an AI/imaging device, and thus no MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is not an algorithm.

7. The type of ground truth used:

  • Not applicable to clinical ground truth for AI algorithms. For the non-clinical studies:
    • Mechanical testing: The "ground truth" would be the established performance characteristics of the predicate devices and accepted industry standards for surgical mesh properties.
    • Biocompatibility: The "ground truth" would be established safety profiles for materials and predicate devices, assessed through standardized biocompatibility tests.
    • In vivo strength retention: The "ground truth" would be the expected or observed strength retention profiles of the predicate devices in animal models.

8. The sample size for the training set:

  • Not applicable. This refers to training data for an AI algorithm. No AI algorithm is part of this device.

9. How the ground truth for the training set was established:

  • Not applicable. This refers to training data for an AI algorithm.

{0}------------------------------------------------

K11372)

Image /page/0/Picture/2 description: The image shows the word "tepha" in a bold, sans-serif font. A curved line starts above the "t", wraps around the word, and ends below the "a", creating a visual element that emphasizes the word. The text is black against a white background.

840 Memorial Drive Cambridge, MA 02139 tel: (617) 995-5400 fax: (617) 995-5401

Section IX Summary of Safety and Effectiveness (Prepared on December 16, 2011)

Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Tepha, Inc. is submitting the following summary of information respecting safety and effectiveness:

Trade Name:TephaFLEX® Light Mesh
Sponsor:Tepha, Inc.99 Hayden Avenue, Suite 360Lexington, MA 02421
Contact Person:Mary P. LeGraw, V.P., Regulatory Affairs
Device Classification Name:CFR §878.3300Surgical Mesh
Classification:According to Section 13 of the Federal Food, Drug and Cosmetic Act, thedevice classification is Class II, Performance Standards.
Predicate Devices:TephaFLEX® Surgical Mesh, TephaFLEX® Composite Mesh, SupraMesh
Device Description:The TephaFLEX light mesh is a resorbable mesh prepared from poly-4-hydroxybutyrate (P4HB). The mesh is prepared from non-dyed or dyed(D&C Violet #2) monofilament P4HB fiber that is knitted into a surgicalmesh. It is provided in single sheets of varying widths and lengths andmay be cut to the shape or size desired for a specific application.
Indications for Use:TephaFLEX light mesh is intended to reinforce soft tissue whereweakness exists in patients undergoing plastic and reconstructivesurgery, or for use in procedures involving soft tissue repair, such as forthe repair of hernia or other fascial defects that require the addition of areinforcing or bridging material to obtain the desired surgical result.
Safety and Performance:Mechanical testing, biocompatibility testing, and in vivo animal testingwas performed based on recommendations identified in the FDA surgicalmesh guidance document: The Guidance for the Preparation of a Pre-market Notification Application for a Surgical Mesh. Specifically,comparative burst strength, suture pull-out strength, tensile strength andtear resistance strength was characterized. In vivo strength retention wascharacterized via a subcutaneous implantation study. The mechanicaland in vivo data collected determined the mesh to be substantiallyequivalent to the predicate devices.
Conclusion:Based on the indications for use, technological characteristics, and safetyand performance testing, the TephaFLEX light mesh has been shown tobe substantially equivalent to predicate devices under the Federal Food,Drug and Cosmetic Act.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three overlapping, abstract shapes that resemble birds in flight or flowing water.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

FEB 1 5 2012

Tepha, Inc. % Ms. Mary P. LeGraw V.P., Regulatory Affairs 99 Hayden Avenue, Suite 360 Lexington, Massachusetts 02421

Re: K113721

Trade/Device Name: TephaFLEX® Light Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OOD Dated: January 27, 2012 Received: January 30, 2012

Dear Ms. LeGraw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

Page 2 - Ms. Mary P. LeGraw

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Eunice Keith

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K 11 3721

Indications for Use

510(k) Number (if known): Unknown

Device Name: TephaFLEX® Light Mesh

Indications for Use:

TephaFLEX light mesh is intended to reinforce soft tissue where weakness exists in patients undergoing plastic and reconstructive surgery, or for use in procedures involving soft tissue repair, such as for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Prescription Use: X AND/OR (21 CFR 801 Subpart D)

Over-The-Counter (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Daniel Kucera MXIM

(Division Sign-Off) Division of Surgical. Othonedic. and Restorative Devices

510(k) Number K113721

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.