K Number
K113340

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-05-04

(172 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Kimberly-Clark Pediatric/ Child Facemask, is intended to be worn by the patient/child (recommended ages 4-12) to provide protection for the respiratory tract. It is a single use, disposable device that is provided non-sterile. This Face Mask is recommended for use in a healthcare setting with appropriate adult supervision.

Device Description

The Kimberly-Clark Pediatric/ Child Facemask is a three layer mask, constructed of nonwoven polyester blends and polypropylene materials. Bindings are nonwoven polyester and earloops are knitted polyester/lycra. A malleable nosepiece is placed within the bindings for comfort and individualized fit around the wearer's nose. The Pediatric/ Child Facemask is appropriately sized to the smaller faces of children across a diverse population. The Pediatric/ Child Facemask is a single use, disposable device, provided non-sterile.

AI/ML Overview

This K113340 submission is for a Kimberly-Clark Pediatric/Child Facemask, which is a physical device, not an AI/ML powered device. Therefore, many of the requested elements for describing an AI/ML study (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size and ground truth establishment) are not applicable.

However, I can provide information based on the provided text regarding the acceptance criteria and study that proves the device meets the acceptance criteria for this specific medical device submission, focusing on the material characteristics and performance relevant to a face mask.

Here's the breakdown of the available information:

Acceptance Criteria and Reported Device Performance

The device, a Kimberly-Clark Pediatric/Child Facemask, was tested against various standards. The summary states that "All results of testing met acceptance criteria." Specific numerical acceptance criteria are not detailed in the provided text, but the standards themselves imply the criteria.

Acceptance Criteria (Implied by Standard)Reported Device Performance (as stated in the summary)
Mil-M369454C (Military Specifications: Surgical Mask, disposable 1992) - Specific criteria not detailed in the provided text.Met acceptance criteria
PSC CS-191-53 (Flammability Test Method (16 CFR 1610) for Flammability of Clothing Textiles) - Specific criteria not detailed.Met acceptance criteria
ASTM F 2299 (Standard Test Method for Evaluating the Initial Efficiency of Materials Used in Medical Masks to Penetration of Particulates Using Latex Spheres) - Specific efficiency thresholds not detailed.Met acceptance criteria
ASTM 2101-07 (Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials. Using a Biological Aerosol of Staphylococcus aureus) - Specific BFE thresholds not detailed.Met acceptance criteria
ISO 10993 (Standards for evaluating the biocompatibility of a medical device) - Specific biocompatibility assessment results not detailed.Met acceptance criteria

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document does not specify the sample size (e.g., number of masks, number of material swatches) used for each of the tests.
  • Data Provenance: Not explicitly stated, but the tests were performed by Kimberly-Clark as part of their 510(k) submission. This is prospective testing conducted specifically for this regulatory submission. There is no information about country of origin for the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a physical device subject to performance testing against established engineering and medical standards, not diagnostic or interpretive tasks requiring expert ground truth for a test set in the same sense as an AI/ML medical device. The "ground truth" would be the objective results of the validated test methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept applies to human interpretation of data for ground truth establishment, not to objective physical product testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, nor does it involve human readers/interpreters.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm or AI system.

7. The type of ground truth used

The "ground truth" for this device's performance is established by the results of standardized physical, chemical, and biological tests (e.g., flammability, particulate filtration efficiency, bacterial filtration efficiency, biocompatibility) against the pre-defined criteria within the referenced standards (Mil-M369454C, PSC CS-191-53, ASTM F 2299, ASTM 2101-07, ISO 10993).

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

K113340

.


Kimberly-Clark* Corporation rtimberly Olare Octoberly-Clark* Pediatric/Clark* Pediatric/Child Face Mask

510K Summary
Date Summary wasPrepared:November 3, 2011
510(k) Submitter:Ann Waterhouse, RACAssociate Director of Regulatory AffairsRoswell, GA 30076678-352-6719 (telephone)920-225-3108 (facsimile)Ann.waterhouse@kcc.com
Device Trade Name:Kimberly-Clark Pediatric/Child Facemask
Device CommonnamesMask, surgical
Device ProductCodes andClassificationNames:OXZ, Class IIMask, surgical (21 CFR 878.4040)
Predicate DevicesThe Kimberly-Clark Pediatric/Child face mask(s) K103150.
Device Description:The Kimberly-Clark Pediatric/ Child Facemask is a three layermask, constructed of nonwoven polyester blends andpolypropylene materials. Bindings are nonwoven polyester andearloops are knitted polyester/lycra. A malleable nosepiece isplaced within the bindings for comfort and individualized fitaround the wearer's nose. The Pediatric/ Child Facemask isappropriately sized to the smaller faces of children across adiverse population. The Pediatric/ Child Facemask is a singleuse, disposable device, provided non-sterile.
Intended Use:The Kimberly-Clark Pediatric/ Child Facemask, is intended to beworn by the patient/child (recommended ages 4-12) to provideprotection for the respiratory tract. It is a single use, disposabledevice that is provided non-sterile. This Face Mask isrecommended for use in a healthcare setting with appropriate

adult supervision.

MAY - 4 2012 .

{1}------------------------------------------------

Kimberly-Clark* Corporation

510(k) for the expansion of age range: Kimberly-Clark* Pediatric/Child Face Mask

TechnologicalCharacteristics andSubstantialEquivalence:The Kimberly-Clark Pediatric/Child face mask is substantiallyequivalent to the predicate device, Kimberly-ClarkPediatric/Child face mask K103150 in intended use andprinciples of operation.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Summary of Testing:

.

The Kimberly-Clark Pediatric/Children's face mask has been tested under the following standards

Mil-M369454CMilitary Specifications: Surgical Mask, disposable 1992
PSC CS-191- 53Flammability Test Method (16 CFR 1610) forFlammability of Clothing Textiles
ASTM F 2299Standard Test Method for Evaluating the InitialEfficiency of Materials Used in Medical Masksto Penetration of Particulates Using LatexSpheres
ASTM 2101-07Test Method for Evaluating the BacterialFiltration Efficiency (BFE) of Medical FaceMask Materials. Using a Biological Aerosol ofStaphylococcus aureus
ISO 10993Standards for evaluating the biocompatibility ofa medical device

All results of testing met acceptance criteria.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Ann Waterhouse Associate Director of Regulatory Affairs Kimberly-Clark 1400 Holcomb Bridge Road Roswell, Georgia 30076

MAY - 4 2012

Re: K113340

Trade/Device Name: Kimberly-Clark Pediatric/Child Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: OXZ Dated: April 26, 2012 Received: April 27, 2012

Dear Ms. Waterhouse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{3}------------------------------------------------

Page 2 - Ms. Waterhouse

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Th for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Kimberly-Clark* Corporation 510(k) for the expansion of age range: Kimberly-Clark* Pediatric/Child Face Mask

Indications for Use

510(k) Number (if known):

Kimberly-Clark Pediatric/Child Face Mask Device Name:

Indications for Use:

The Kimberly-Clark Pediatric/ Child Facemask, is intended to be worn by the patient/child (recommended ages 4-12) to provide protection for the respiratory tract. It is a single use, disposable device that is provided non-sterile. This Face Mask is recommended for use in a healthcare setting with appropriate adult supervision.

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth B. Clausing-Wells

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.