K Number
K103150

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-09-23

(332 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Kimberly-Clark Pediatric/ Child Facemask, is intended to be worn by the patient/child (recommended ages 5-12) to provide protection for the respiratory tract. It is a single use, disposable device that is provided non-sterile. This Face Mask is recommended for use in a healthcare setting with appropriate adult supervision.

Device Description

The Kimberly-Clark Pediatric/ Child Facemask is a three layer mask, constructed of nonwoven polyester blends and polypropylene materials. Bindings are nonwoven polyester and earloops are knitted polyester/lycra. A malleable nosepiece is placed within the bindings for comfort and individualized fit around the wearer's nose. The Pediatric/ Child Facemask is appropriately sized to the smaller faces of children across a diverse population. The Pediatric/ Child Facemask is a single use, disposable device, provided non-sterile.

AI/ML Overview

Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study for the Kimberly-Clark Pediatric/Child Facemask:

This document describes a medical device, specifically a surgical mask, and its submission for FDA clearance. The "studies" referred to are bench tests and consumer evaluations, not clinical trials in the sense of comparing diagnostic accuracy or treatment outcomes. Therefore, many standard questions related to a typical AI/diagnostic device study (like MRMC, expert consensus for ground truth, training set size, etc.) are not applicable here.


Acceptance Criteria and Reported Device Performance

StandardTitleAcceptance Criteria (Implied)Reported Device Performance
Mil-M369454CMilitary Specifications: Surgical Mask, disposable 1992Device meets the performance requirements specified in this military standard for surgical masks."All testing conducted on the Pediatric/Child Face Mask met acceptance criteria."
ASTM F2101-07Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol, Staphylococcus AureusDevice meets the BFE threshold defined by the standard for its intended classification (e.g., specific percentage of bacterial filtration)."All testing conducted on the Pediatric/Child Face Mask met acceptance criteria."
PSC CS-191- 53Flammability Test Method (16 CFR 1610) for Flammability of Clothing TextilesMaterial must meet flammability requirements for clothing textiles (e.g., specific burn rate or no ignition)."All testing conducted on the Pediatric/Child Face Mask met acceptance criteria."
ASTM F2299Standard Test Method for Evaluating the Initial Efficiency of Materials Used in Medical Masks to Penetration of Particulates Using Latex SpheresDevice meets the particulate filtration efficiency (PFE) threshold defined by the standard (ee.g., specific percentage of particulate filtration)."All testing conducted on the Pediatric/Child Face Mask met acceptance criteria."
ISO 10993Standards for evaluating the biocompatibility of a medical deviceDevice materials must demonstrate biocompatibility (e.g., non-cytotoxic, non-irritating, non-sensitizing)."All testing conducted on the Pediatric/Child Face Mask met acceptance criteria."
EN 71-3Safety of Toys- Part 3: Migration of Certain ElementsDevice materials must comply with limits for the migration of certain elements (e.g., heavy metals), ensuring safety for incidental contact/ingestion by children."All testing conducted on the Pediatric/Child Face Mask met acceptance criteria."
N/APotential mechanical and physical hazards (e.g., choking, strangulation, tissue tear/injury)Device design and materials minimize these hazards for the pediatric population."Consumer evaluation of potential mechanical and physical hazards...were conducted on the Pediatric/Child Facemask" and "All testing conducted...met acceptance criteria." (Implied successful mitigation).

Study Information (as applicable to a medical device bench test submission):

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated in the provided text for each specific test. For bench testing of materials, this would typically involve a certain number of material samples or manufactured devices.
    • Data Provenance: The tests are described as conducted by Kimberly-Clark, implying internal testing or contracted lab testing. The country of origin of the data (i.e., where the labs were located) is not specified. The tests are general standards, not tied to a specific population's data. These are bench tests, not data derived from retrospective or prospective patient studies.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. For bench testing of a physical product like a mask, "ground truth" is established by the standardized test methods themselves and the precise measurement of physical properties. Experts are involved in conducting and interpreting these tests, but it's not a consensus interpretation of clinical data.
  3. Adjudication method for the test set:

    • Not Applicable. Adjudication methods like 2+1 or 3+1 are used for human interpretation of data where consensus is needed to establish ground truth (e.g., image reading). This is a physical product undergoing standardized performance tests.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-powered diagnostic or interpretive device. It is a physical product (face mask) undergoing performance testing.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm-based device.
  6. The type of ground truth used:

    • Performance against established industry and regulatory standards. The "ground truth" here is determined by whether the device's measured performance (e.g., BFE, PFE, flammability, biocompatibility) meets the numerical thresholds or qualitative criteria set forth in the referenced standards (Mil-M369454C, ASTM F2101-07, PSC CS-191- 53, ASTM F2299, ISO 10993, EN 71-3) and by the "consumer evaluation of potential mechanical and physical hazards."
  7. The sample size for the training set:

    • Not Applicable. There is no "training set" in the context of this type of medical device submission. The device design and materials are developed, and then the finished product is tested against standards.
  8. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

K103150

SEP 2 3 2011

510(k) Summary

Date of Submission of Pre-IDE information:October 8, 2010
Sponsor:Kimberly-Clark
Address:1400 Holcomb Bridge Rd, Roswell, GA30076
Contact:Ann Waterhouse, RAC
Telephone:(678) 352-6719
Facsimile:(920) 225-3108

Name of Device: Kimberly-Clark Pediatric/ Child Facemask Common Name/Classification Name: Mask, Surgical Classification: Class II Product Code: OXZ Regulation Number: 878.4040 Panel: General & Plastic Surgery

Indication for Use Statement:

The Kimberly-Clark Pediatric/ Child Facemask, is intended to be worn by the patient/child (recommended ages 5-12) to provide protection for the respiratory tract. It is a single use, disposable device that is provided non-sterile. This Face Mask is recommended for use in a healthcare setting with appropriate adult supervision.

Device Description:

The Kimberly-Clark Pediatric/ Child Facemask is a three layer mask, constructed of nonwoven polyester blends and polypropylene materials. Bindings are nonwoven polyester and earloops are knitted polyester/lycra. A malleable nosepiece is placed within the bindings for comfort and individualized fit around the wearer's nose. The Pediatric/ Child Facemask is appropriately sized to the smaller faces of children across a diverse population. The Pediatric/ Child Facemask is a single use, disposable device, provided non-sterile.

{1}------------------------------------------------

510(k) Summary, continued

Consumer evaluation of potential mechanical and physical hazards (e.g., choking, strangulation, tissue tear/injury) as well as the following tests were conducted on the Pediatric/ Child Facemask:

StandardTitle
Mil-M369454CMilitary Specifications: Surgical Mask, disposable 1992
ASTM F2101-07Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE)of Medical Face Mask Materials, Using a Biological Aerosol, StaphylococcusAureus
PSC CS-191- 53Flammability Test Method (16 CFR 1610) for Flammability of Clothing Textiles
ASTM F2299Standard Test Method for Evaluating the Initial Efficiency of Materials Used inMedical Masks to Penetration of Particulates Using Latex Spheres
ISO 10993Standards for evaluating the biocompatibility of a medical device
EN 71-3Safety of Toys- Part 3: Migration of Certain Elements

All testing conducted on the Pediatric/Child Face Mask met acceptance criteria.

Predicate Devices:

K061716Prestige Ameritech Face Mask
K082258CrossTex® Surgical Masks
K060776Hong Ya Surgical Face Mask

The Kimberly-Clark Pediatric/ Child Facemask is comparable to the predicate devices cited in terms of materials of composition and performance.

The indications for use are slightly different from predicate devices due to the pediatric population for which this mask is intended to be worn, the differences are not critical to the intended use of the device, and do not affect the safety and effectiveness of the device when used as labeled.

Conclusions:

The Kimberly-Clark Pediatric/ Child Facemask have the same or similar intended use and technological characteristics as the predicate devices. Bench testing demonstrates the safety and efficacy of the Pediatric/ Child Facemask to standards cited. The Kimberly-Clark Pediatric/ Child Facemask does not raise any new questions concerning safety and effectiveness. The Kimberly-Clark Pediatric/ Child Facemask is substantially equivalent to referenced predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem with three stylized human figures, representing health and well-being. The figures are connected and appear to be in motion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Ann Waterhouse Associate Director, Regulatory Affairs Kimberly-Clark 1400 Holcomb Bridge Road. Suite 1154. Bldg 300 Roswell, Georgia 30076

SEP 23 2011

Re: K103150

Trade/Device Name: Kimberly-Clark Pediatric/Child Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: OXZ Dated: September 12, 2011 Received: September 13, 2011

Dear Ms. Waterhouse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Nitt

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K103150

Device Name:

Kimberly-Clark Pediatric/Child Face Mask

Indications For Use:

The Kimberly-Clark Pediatric/ Child Facemask, is intended to be worn by the patient/child (recommended ages 5-12) to provide protection for the respiratory tract. It is a single use, disposable device that is provided non-sterile. This Face Mask is recommended for use in a healthcare setting with appropriate adult supervision.

Prescription Use AND/OR Over-The-Counter Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINEcontinue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clavins-Williams

Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antrol, Dental Devices

510(k) Number: K103150

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.