(46 days)
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric image maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions; multi-planar reformats and maximum intensity projections. The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image. Visia™ Dynamic Review can also be used to provide measurements for diameters, areas and volumes. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
When interpreted by a skilled physician, Visia™ Dynamic Review provides information that may be useful in diagnosis. Patient management decisions should not be made based solely on the results of Visia™ Dynamic Review analysis.
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review integrates within typical clinical workflow patterns through receiving and transferring medical images over a computer network. The software can be loaded on a standard offthe-shelf personal computer (PC) and can operate as a stand-alone workstation or in a distributed server-client configuration across a computer network.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric image maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions; multi-planar reformats and maximum intensity projections.
The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source images can also be displayed based on physician preferences using configurable viewing options or hanging protocols.
Visia™ Dynamic Review provides the clinician with a broad set of viewing and analysis tools to annotate, measure, and output selected image views or user defined reports. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Visia™ Dynamic Review device, structured according to your requested information:
1. A table of acceptance criteria and the reported device performance
Based on the provided K113337 510(k) Summary, specific quantitative acceptance criteria and corresponding reported device performance values are not explicitly detailed. The document focuses on the regulatory submission process and general affirmations of safety and effectiveness through nonclinical testing.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly defined in the provided document. The document states: "The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the software in each operational mode and followed the process documented in the Validation Test Plan." And "Validation testing indicated that as required by the risk analysis, designated individuals performed all verification activities and that the results demonstrated that the predetermined acceptance criteria were met." | No specific quantitative performance metrics are provided. The document states: "The complete system configuration has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria." And "Nonclinical tests demonstrate that the device is safe, effective, and is substantially equivalent to the predicate device." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The provided K113337 510(k) Summary does not specify the sample size used for any test set or the data provenance (country of origin, retrospective/prospective). It only mentions "in-house testing criteria" and "Validation Test Plan."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The provided K113337 510(k) Summary does not provide information on the number or qualifications of experts used to establish ground truth for any test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The provided K113337 510(k) Summary does not describe any adjudication method used for a test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not described or performed in the provided K113337 510(k) Summary. The device is software for viewing and analyzing MRI studies, and the document explicitly states, "Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians." There is no mention of AI assistance or human reader improvement with or without AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The K113337 510(k) Summary does not explicitly describe a standalone performance study in terms of quantitative metrics. It states that the "complete system configuration has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria." However, it clarifies that the software's role is to provide information for a skilled physician's interpretation, rather than performing diagnosis independently: "Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians." and "A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The provided K113337 510(k) Summary does not specify the type of ground truth used for its internal testing or validation studies.
8. The sample size for the training set
The provided K113337 510(k) Summary does not mention a training set sample size. This is consistent with the nature of the device, which is described as "Medical Image Processing Software" that performs functions like motion registration, parametric image mapping, subtractions, and multi-planar reformats. These are generally rule-based or algorithmic image processing tasks rather than machine learning models that require explicit training sets in the typical sense.
9. How the ground truth for the training set was established
As no training set is mentioned or implied for the core functionalities of the device, the document does not describe how ground truth for a training set was established.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MeVis Medical Solutions. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent the company's brand.
DEC 3 0 2011
K113337
p'1/3
510(k) Summary
| Submitted By: | MeVis Medical Solutions AGUniversitaetsallee 2928359 Bremen, Germany | |
|---|---|---|
| EstablishmentName: | MeVis Medical Solutions AG | |
| EstablishmentRegistration Number: | Applied for and awaiting assignment by FDA | |
| Contact Person: | Thomas E. TynesPresident & CEOMeVis Medical Solutions, Inc.N27 W24075 Paul Court - Suite 100Pewaukee, WI 53072USAPhone: +1-262-691-9530Fax: +1-262-691-9531 | |
| Date Prepared: | 11/10/2011 | |
| Trade Name: | Visia™ Dynamic Review | |
| Common Name: | Medical Image Processing Software | |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) | |
| ClassificationRegulation Number: | 892.1000 | |
| Class: | II | |
| Panel: | Radiology | |
| Product Code: | LNH |
{1}------------------------------------------------
Device Description
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review integrates within typical clinical workflow patterns through receiving and transferring medical images over a computer network. The software can be loaded on a standard offthe-shelf personal computer (PC) and can operate as a stand-alone workstation or in a distributed server-client configuration across a computer network.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric image maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions; multi-planar reformats and maximum intensity projections.
The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source images can also be displayed based on physician preferences using configurable viewing options or hanging protocols.
Visia™ Dynamic Review provides the clinician with a broad set of viewing and analysis tools to annotate, measure, and output selected image views or user defined reports. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
Intended Use
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric imaqe maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions: multi-planar reformats and maximum intensity projections. The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image. Visia™ Dynamic Review can also be used to provide measurements for diameters, areas and volumes. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
When interpreted by a skilled physician, Visia™ Dynamic Review provides information that may be useful in diagnosis. Patient management decisions should not be made based solely on the results of Visia™ Dynamic Review analysis.
Visia™ Dynamic Review 510(k) 510(k) Summary -- 11/10/2011 Page 2 of 3
{2}------------------------------------------------
Predicate Device Information
| Product | Predicate Device Name | Predicate 510(k)Submission Reference |
|---|---|---|
| Visia™ Dynamic Review | Syngo BreVis | K090038 |
Safety and Effectiveness
The Visia™ Dynamic Review labeling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via MeVis Medical Solution AG's Risk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via software development and verification & validation testing.
Nonclinical Testing and Performance Information
The complete system configuration has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria. The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the software in each operational mode and followed the process documented in the Validation Test Plan.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification activities and that the results demonstrated that the predetermined acceptance criteria were met.
Technological Characteristics
Visia™ Dynamic Review is a software device that does not contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians.
A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.
Conclusion
The 510(k) Pre-Market Notification for Visia™ Dynamic Review contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. MeVis Medical Solutions has determined that its device, Visia™ Dynamic Review, is substantially equivalent to the identified predicate device listed above. A comparison with the legally marketed predicate device indicates that it is substantially equivalent to this device, and that it does not raise any new safety or efficacy concerns. Nonclinical tests demonstrate that the device is safe, effective, and is substantially equivalent to the predicate device.
Visia™ Dynamic Review 510(k) 510(k) Summary - 11/10/2011 Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. To the left of the eagle is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
MeVis Medical Solutions AG % Mr. Thomas E. Tynes President, CEO MeVis Medical Solutions, Inc. N27 W24075 Paul Court, Suite 100 PEWAUKEE WI 53072
DEC 3 0 2011
Re: K113337
Trade/Device Name: Visia™ Dynamic Review Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 10, 2011 Received: November 14, 2011
Dear Mr. Tynes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MeVis Medical Solutions. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent the company's brand.
Indications for Use Statement
510(k) Number (if known): K113337
Device Name: Visia™ Dynamic Review
Indications for Use:
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic Visia ~ Dynamic Review is a sommer Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient visia Dynamic Koviow automatics (parametric image maps). Furthermore, it motion and viousless andrer on one processing functions such as image subtractions; multi-planar reformats performs other user delinou por processions information can be displayed in a variety of formats, and maximally projections in the source image. Visia™ Dynamic Review can also be used including a paraments for diameters, areas and volumes. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
When interpreted by a skilled physician, Visia™ Dynamic Review provides information that may be useful while merplosed by a chilled by a child not be made based solely on the results of Visia™ Dynamic Review analysis.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
ن پن
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S Postel
Office of In Vitro Diagnostic Device
510(k) K.11337
Visia™ Dynamic Review 510(k) Indications for Use Statement - 11/10/2011 Page 1 of 1
N/A