K Number
K113300

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-12-09

(30 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter is also intended for post dilatation of balloon-expandable peripheral vascular stents.

Device Description

The ATB Advance® PTA Dilatation Catheter is an over-the-wire catheter indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter is also intended for post dilatation of balloon-expandable peripheral vascular stents. The 6 Fr balloon catheter will be compatible with a 0.035 inch wire guide and an 8 Fr sheath. It will be supplied sterile, intended for one-time use.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the ATB Advance® PTA Dilatation Catheter:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Balloon Minimum Burst Strength: Burst at or above the minimum rated burst pressure.Testing shows the balloons will burst at or above the minimum rated burst pressure, with all failure modes being linear tears. Predetermined acceptance criteria met.
Balloon Compliance: Meet labeled diameter within tolerance at the nominal pressure under simulated body temperature conditions.Testing shows that, under simulated body temperature conditions, each balloon will meet its labeled diameter within tolerance at the nominal pressure. Predetermined acceptance criteria met.
Balloon Profile: Compatible with an 8 Fr sheath (< 0.113 inch profile).Measurement of the diameter of catheter shaft, bonds, and folded balloon shows that the device is compatible with an 8 Fr sheath (< 0.113 inch profile). Predetermined acceptance criteria met.
Balloon Fatigue: Free from leakage and damage on inflation after 10 cycles of inflation/deflation (inflating to rated burst pressure, holding for 30 seconds and deflating).Testing shows that balloons are free from leakage and damage on inflation, withstanding 10 cycles of inflation/deflation (inflating to rated burst pressure, holding for 30 seconds and deflating). Predetermined acceptance criteria met.
Sheath Compatibility: Compatible with an 8 Fr sheath.Qualitative and quantitative evaluations show that the balloons are compatible with an 8 Fr sheath. Predetermined acceptance criteria met.
Balloon Bond Strength: Tensile force during proper clinical use should not fracture or rupture the balloon catheter bond (in conformance with ISO 10555-1: 1995).Testing shows the tensile force during proper clinical use should not fracture or rupture the balloon catheter bond. In conformance with the applicable sections of ISO 10555-1: 1995, predetermined acceptance criteria were met.

2. Sample Size Used for the Test Set and Data Provenance

The provided text describes bench testing only. There is no mention of a "test set" in the context of patient data. The testing involves physical properties of the device (balloon, catheter).

  • Sample Size: Not explicitly stated for each test (e.g., "how many balloons were tested for burst strength"). It only states "Testing shows..." which implies multiple units were tested to demonstrate conformity.
  • Data Provenance: N/A, as this is laboratory bench testing, not clinical data.
  • Retrospective/Prospective: N/A

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

N/A. This 510(k) summary focuses on device design and performance through engineering and laboratory testing, not on clinical evaluations requiring expert interpretation of patient data or ground truth establishment.


4. Adjudication Method for the Test Set

N/A. Adjudication methods are typically used for clinical studies with subjective interpretations. The tests described are objective, measurable physical properties.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document does not mention any MRMC study, human readers, AI, or comparative effectiveness with or without AI assistance. This device is a medical instrument (balloon catheter), not an AI-powered diagnostic tool.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No. This concept is irrelevant to a physical medical device like a balloon catheter.


7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

N/A. For physical device testing, the "ground truth" is typically defined by engineering specifications, industry standards (like ISO 10555-1: 1995), and predefined acceptance thresholds derived from those specifications. For example, the ground truth for "Balloon Minimum Burst Strength" is the "minimum rated burst pressure."


8. The Sample Size for the Training Set

No "training set" is applicable here. The device is a physical product, not a machine learning model.


9. How the Ground Truth for the Training Set Was Established

No "training set" and therefore no ground truth establishment for it.

{0}------------------------------------------------

K113300

Special 510(K) Premarket Notification PTA Balloon Catheter: ATB Advance® PTA Dilatation Catheter Cook Incorporated 4 November 2011

DEC - 9 2011

510(k) Summary

Submitted By:

Elysia Poor Regulatory Affairs Specialist Cook Incorporated 750 Daniels Wav, P.O. Box 489 Bloomington, IN 47402 812-339-2235

Device:

ATB Advance® PTA Dilatation Catheter Trade Name: Proposed Classification: Catheter, Percutaneous (74 DQY)

Indications for Use:

For percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter is also intended for post dilatation of balloon-expandable peripheral vascular stents.

Predicate Devices:

ATB Advance® PTA Dilatation Catheter cleared under the following 510(k) Premarket Notifications:

Device Description:

The ATB Advance® PTA Dilatation Catheter is an over-the-wire catheter indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter is also intended for post dilatation of balloon-expandable peripheral vascular stents. The 6 Fr balloon catheter will be compatible with a 0.035 inch wire guide and an 8 Fr sheath. It will be supplied sterile, intended for one-time use.

Substantial Equivalence:

Cook currently markets the PTA Balloon Catheter which is considered substantially equivalent to the ATB Advance PTA Dilatation Catheter. The identical indications for use and technological characteristics of the ATB Advance® PTA Dilatation Catheter as compared to the predicate device support a determination of substantial equivalency. The ATB Advance® PTA Dilatation Catheter has

{1}------------------------------------------------

Special 510(K) Premarket Notification PTA Balloon Catheter: ATB Advance® PTA Dilatation Catheter Cook Incorporated 4 November 2011

been modified from the predicate ATB Advance® PTA Dilatation Catheter to include a balloon size combination of 14 mm diameter x 2 cm length.

Test Data:

The ATB Advance® PTA Dilatation Catheter was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:

  • 1 . Balloon Minimum Burst Strength - Testing shows the balloons will burst at or above the minimum rated burst pressure, with all failure modes being linear tears. Predetermined acceptance criteria met.
    1. Balloon Compliance - Testing shows that, under simulated body temperature conditions. each balloon will meet its labeled diameter within tolerance at the nominal pressure. Predetermined acceptance criteria met.
    1. Balloon Profile - Measurement of the diameter of catheter shaft, bonds, and folded balloon shows that the device is compatible with an 8 Fr sheath (< 0.113 inch profile). Predetermined acceptance criteria met.
    1. Balloon Fatigue - Testing shows that balloons are free from leakage and damage on inflation, withstanding 10 cycles of inflation/deflation (inflating to rated burst pressure, holding for 30 seconds and deflating). Predetermined acceptance criteria met.
    1. Sheath Compatibility - Qualitative and quantitative evaluations show that the balloons are compatible with an 8 Fr sheath. Predetermined acceptance criteria met.
    1. Balloon Bond Strength - Testing shows the tensile force during proper clinical use should not fracture or rupture the balloon catheter bond. In conformance with the applicable sections of ISO 10555-1: 1995, predetermined acceptance criteria were met.

In conclusion, the results of these tests provide reasonable assurance that the device is as safe and effective as the predicate device and support a determination of substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and the USA. The eagle is positioned in the center of a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" encircling it.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC - 9 2011

Cook, Inc. c/o Ms. Elysia Poor Regulatory Affairs Specialist 750 Daniels Way Bloomington. IN 47404

Re: K113300

Trade/Device Name: ATB Advance PTA Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: LIT, DOY Dated: November 4, 2011 Received: November 9, 2011

Dear Ms. Poor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Elysia Poor

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

M. G. Killebrew

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Special 510(K) Premarket Notification PTA Balloon Catheter: AT13 Advance® I'TA Dilatation Catheter Cook Incorporated 4 November 2011

510(k) Number (if known):

Device Name: ATB Advance® PTA Dilatation Catheter

Indications for Use:

For percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter (ATB) is also intended for post dilatation of balloon-expandable peripheral vascular stents.

K113300

Prescription Use X (Per 21 CFR 801 Subpart D)

OR

Over-the-Counter Use (21 CFR 807 Subpart C)

l

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. A. Keller

ivision Sian-Off) ivision of Cardiovascular Devices

510(k) Numbe

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).