K Number
K113215

Validate with FDA (Live)

Date Cleared
2012-05-14

(195 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Model 8004CB (Adhesive Version)

The 8004CB Single-Patient use, Non-Sterile, Disposable Regional Oximetry Sensor is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. The sensor is for spot-checking and continuous monitoring of neonate, infant, and pediatric patients weighing less than 88 pounds (40 kilograms).. The sensor may be repositioned or replaced with another 8004CB sensor without baseline re-establishment. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long term care and mobile environments.

Model 8004CB-NA (Non-Adhesive Version)

The 8004CB-NA Non-Adhesive, Single-Patient use, Non-Sterile, Disposable Regional Oximetry Sensor is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. The sensor is for spot-checking and continuous monitoring of neonate, infant, and pediatric patients weighing less than 88 pounds (40 kilograms). The sensor may be repositioned or replaced with another 8004CB-NA sensor without baseline reestablishment. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long term care and mobile environments. I

Model 7600 Regional Oximeter System

Nonin's non-invasive Model 7600 4- Channel Regional Oximeter System is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. It is intended for spot-checking or continuous monitoring of adult, or neonate, infant and pediatric patients. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long-term care and mobile environments.

Device Description

Nonin's® EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series for use with the Model 7600 4-Channel Regional Oximeter System continuously monitors and records the mixed arterial/venous blood oxygen levels (rSO2) through non-invasive near-infrared spectroscopy of tissue under the sensors.

The 7600 system is comprised of three subsystems; sensor, patient oximetry device (pod) and display unit. The sensor allows light absorption measurements at various wavelengths in the near-infrared spectrum (approximately 700 to 900 nanometers). The sensor is sized appropriately for the patient population.

The sensors plug into the patient oximetry device (pod) which controls the light emitted from the sensor LEDs and measures the light returning via the sensor photodiodes. From these measurements, the 7600PA pod determines specific absorption values and calculates the rSO2 value. The pod then communicates the rSO2 readings and other data to the display unit. Up to 4 pods with attached sensors may be used with a single 7600 display unit.

The 7600 display unit displays real-time regional oximetry data. It is a battery-backed, mains-powered device equipped with audio and visual alarm indicators. Real-time data and playback output is accomplished through a Bluetooth transceiver module or serial RS-232 connection.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Absolute rSO2 accuracy (Arms) for 45% to 95% rSO2±5.9%

Note: The document explicitly states the accuracy specification as "absolute rSO2 accuracy (Arms*): 45% to 95% rSO2 ±5.9%." This serves as the acceptance criterion that the device aims to meet. The reported device performance matches this criterion, as the study was conducted to demonstrate this specific accuracy.

Study Details

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: 44 subjects
  • Data Provenance:
    • Country of Origin: Not specified in the provided text.
    • Study Type: Prospective, controlled, multisite study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • The document does not mention the use of experts to establish ground truth in the traditional sense of image or clinical interpretation.
  • Ground Truth Establishment: The ground truth for rSO2 (regional hemoglobin oxygen saturation) was established by comparing the sensor's measurements to actual SavO2 (arterial/venous hemoglobin oxygen) values. These SavO2 values were "determined from venous and arterial blood samples." This implies direct laboratory analysis of blood samples, which serves as the gold standard for oxygen saturation.
  • Qualifications of Experts: Not applicable, as expert consensus on interpretation wasn't the method for ground truth. The ground truth was derived from direct physiological measurements (blood gas analysis).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable/None. The "ground truth" was established through direct physiological measurement (blood gas analysis of SavO2), not through human interpretation or a scenario requiring adjudication of disparate expert opinions.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This was not an MRMC comparative effectiveness study. The study focused on the standalone accuracy of the device itself against a physiological ground truth, not on how a human reader's performance might be augmented by the device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes. The study assessed the "absolute rSO2 accuracy" of the EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series with the Model 7600 Regional Oximeter System against a physiological ground truth (blood gas analysis). This is a standalone performance assessment of the device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: Physiological measurements from blood samples. Specifically, the "SavO2 (value determined from venous and arterial blood samples)" served as the ground truth. A model of 70% venous and 30% arterial blood was used for brain blood.

8. The sample size for the training set

  • The document does not explicitly mention a separate "training set" or sample size for training. This typically indicates that the device's algorithms or parameters were either pre-developed, or the accuracy study itself demonstrated the device's performance without explicitly detailing an iterative training/validation process in this summary. For medical devices of this type, initial algorithm development often relies on internal data and testing, with the submission focusing on performance in a clinical validation set.

9. How the ground truth for the training set was established

  • Not applicable, as a distinct training set with established ground truth is not described in this document.

{0}------------------------------------------------

K113215

MAY 1 4 2012

Attachment 3 510(k) Summary

Submitter:

Contact Person:

Phone: Fax: Date Prepared:

Trade Name:

Classification Name and Number:

Product Code:

Predicate Device(s):

Indications for Use:

Nonin Medical, Inc.

Brodie Pedersen Senior Regulatory Engineer Nonin Medical, Inc. 13700 1st Ave. North Plymouth, MN 55441-5443 763-553-9968 763-553-7807 October 31, 2011

EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series for use with Model 7600 Regional Oximeter System with EQUANOX Technology

Oximeter Class II, 21 CFR 870.2700 MUD

Nonin's Model 8004CB Series Oximeter Sensor used with the Model 7600 Equanox Advance Regional Oximeter System is substantially equivalent to the Model 8004CA Oximetry Sensor cleared with the Model 7600 Regional Oximeter System cleared by the FDA in K102715 on 12/17/2010, and the FORE-SIGHT® Oximeter MC-2000 Series Cerebral Oximeter manufactured by CAS Medical Systems Inc. that was cleared by the FDA under K094030 on 12/23/10.

Model 8004CB (Adhesive Version)

The 8004CB Single-Patient use, Non-Sterile, Disposable Regional Oximetry Sensor is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. The sensor is for spot-checking and continuous monitoring of neonate, infant, and pediatric patients weighing less than 88 pounds (40 kilograms). The sensor may be repositioned or replaced with another 8004CB sensor without baseline re-establishment. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long term care and mobile environments.

{1}------------------------------------------------

Indications for Use continued:

Model 8004CB-NA (Non-Adhesive Version)

The 8004CB-NA Non-Adhesive, Single-Patient use, Non-Sterile, Disposable Regional Oximetry Sensor is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. The sensor is for spot-checking and continuous monitoring of neonate, infant, and pediatric patients weighing less than 88 pounds (40 kilograms). The sensor may be repositioned or replaced with another 8004CB-NA sensor without baseline re-establishment. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long term care and mobile environments.

Model 7600 Regional Oximeter System

Nonin's non-invasive Model 7600 4- Channel Regional Oximeter System is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. It is intended for spot-checking or continuous monitoring of adult, or neonate, infant and pediatric patients. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long-term care and mobile environments.

Device Description:

Nonin's® EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series for use with the Model 7600 4-Channel Regional Oximeter System continuously monitors and records the mixed arterial/venous blood oxygen levels (rSO2) through non-invasive near-infrared spectroscopy of tissue under the sensors.

The 7600 system is comprised of three subsystems; sensor, patient oximetry device (pod) and display unit. The sensor allows light absorption measurements at various wavelengths in the near-infrared spectrum (approximately 700 to 900 nanometers). The sensor is sized appropriately for the patient population.

The sensors plug into the patient oximetry device (pod) which controls the light emitted from the sensor LEDs and measures the light returning via the sensor photodiodes. From these measurements, the 7600PA pod determines specific absorption values and calculates the rSO2 value. The pod then communicates the rSO2 readings and other data to the display unit. Up to 4 pods with attached sensors may be used with a single 7600 display unit

{2}------------------------------------------------

The 7600 display unit displays real-time regional oximetry data. It is a battery-backed, mains-powered device equipped with audio and visual alarm indicators. Real-time data and playback output is accomplished through a Bluetooth transceiver module or serial RS-232 connection.

Technological Characteristics and Performance:

EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series is equivalent in construction to Nonin's previously cleared 8004CA regional oximeter sensor. The performance of the EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series, for use on patient's weighing less than 88pounds (40 kg), accuracy specifications are absolute rSO2 accuracy (Arms*): 45% to 95% rSO2 ±5.9%. Based on the measured regional hemoglobin saturation value (rSO₂) of cerebral sensors calibrated to arterial / venous hemoglobin oxygen (SavO2) value, determined from venous and arterial blood samples from 44 subjects ranging in age from 4 days to 10 years. The model used for blood in the brain is 70% venous and 30% arterial. The venous blood was drawn from the right jugular bulb. The accuracy of the sensors in comparison to the blood gas analyzer samples measured over the rSO2 range of 45 -95%. Accuracy data is calculated using the root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment -Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.

Nonin's EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series used with the Model 7600 Regional Oximeter System was developed following the risk management processes defined by ISO 14971:2007 Medical Devices - Application of Risk Management to Medical Devices. The EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series has successfully undergone laboratory safety and performance testing in accordance with ISO 9919:2005, Medical Electrical Equipment -Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use, IEC 60601-1:1995 Medical electrical equipment - part 1: General requirements for safety and IEC 60601-1-2:2007 General requirements for basic safety and essential performance - Collateral standards: Electromagnetic compatibility - requirements and tests.

Evaluation of materials was performed in accordance with ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.

Functional and Safety Testing:

{3}------------------------------------------------

A prospective controlled multisite study involving a diverse ethnic pediatric population ranging in age from 4 days to 10 years used SavO2 values to simultaneously validate the rSO2 values measured by the Model 7600 4-Channel Regional Oximeter System with the EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series.

Performance testing and clinical testing were successfully completed in order to ensure that it has appropriate performance, functional features and is substantially equivalent to the predicate device.

Conclusion:

Nonin's EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series used with the Model 7600 Regional Oximeter System is substantially equivalent to the Model 7600 4-Channel Regional Oximeter with Equanox Technology and Bluetooth wireless Technology K102715 cleared 12/17/2010 including regional sensors 8004CA and 8000CA. It is equivalent to the FORE-SIGHT® Oximeter MC-2000 Series Cerebral Oximeter manufactured by CAS Medical Systems Inc. that was cleared by the FDA under K094030 on 12/23/10.

The conclusion drawn is that the revised indications for use and labeling are substantially equivalent to the predicate device and do not raise new questions of safety and effectiveness.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 1 4 2012

Nonin Medical, Inc. c/o Mr. Brodie C. Pedersen Senior Regulatory Engineer 13700 15t Aivenue North Plymouth, MN 55441

Re: K113215

Trade/Device Name: EQUANOX Advance Neonatal/Pediatric Sensor 8004CB Series for use with Model 7600 Regional Oximeter System with EQUANOX

Technology

Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: MUD Dated: April 10, 2012 Received: April 11, 2012

Dear Mr. Pedersen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{5}------------------------------------------------

Page 2 - Mr. Brodie C. Pedersen

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

1. Indications for Use Statement

510(K) Number:

Device Name:

Nonin Medical, Inc. EQUANOX Advance ™ Neonatal/Pediatric Sensor 8004CB Series for use with the Model 7600 4 Channel Regional Oximeter System with Equanox™ Technology Indications for Use:

Model 8004CB (Adhesive Version)

The 8004CB Single-Patient use, Non-Sterile, Disposable Regional Oximetry Sensor is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. The sensor is for spot-checking and continuous monitoring of neonate, infant, and pediatric patients weighing less than 88 pounds (40 kilograms).. The sensor may be repositioned or replaced with another 8004CB sensor without baseline re-establishment. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long term care and mobile environments.

Model 8004CB-NA (Non-Adhesive Version)

The 8004CB-NA Non-Adhesive, Single-Patient use, Non-Sterile, Disposable Regional Oximetry Sensor is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. The sensor is for spot-checking and continuous monitoring of neonate, infant, and pediatric patients weighing less than 88 pounds (40 kilograms). The sensor may be repositioned or replaced with another 8004CB-NA sensor without baseline reestablishment. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long term care and mobile environments. I

Model 7600 Regional Oximeter System

Nonin's non-invasive Model 7600 4- Channel Regional Oximeter System is intended for use as an absolute real-time adjunct monitor of hemoglobin oxygen saturation of blood underneath the sensor at cerebral and somatic sites. It is intended for spot-checking or continuous monitoring of adult, or neonate, infant and pediatric patients. It is intended for use in environments including the operating room, surgical recovery, critical care, emergency room, long-term care and mobile environments.

Prescription Use _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

K113215

510(k) Number

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).